510(k) Data Aggregation

    K Number
    K250223

    Validate with FDA (Live)

    Date Cleared
    2025-02-25

    (29 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BruxZir® Esthetic NOW is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals.

    Device Description

    BruxZir® Esthetic NOW is a chairside zirconia material offering enhanced esthetic performance. These pre-shaded, fully sintered zirconia CAD/CAM blocks are used by dentists in the construction of custom-made dental restorations that require no sintering, staining, glazing or polishing.

    AI/ML Overview

    The provided text describes the regulatory submission for a dental material (BruxZir® Esthetic NOW) and its comparison to a predicate device. It includes performance data, but NOT information typically associated with AI/ML device studies, such as sample sizes for test sets, data provenance, expert qualifications for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance.

    Therefore, the response will focus on the available information regarding acceptance criteria and the study that proves the device meets those criteria, highlighting the lack of AI/ML-specific details where applicable.


    Here's a summary of the acceptance criteria and study findings based on the provided document:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance (BruxZir® Esthetic NOW)
    Flexural Strength: >800 MPa (Type II Class 5 per ISO 6872:2024)Average flexural strength of worst-case shades was above the minimum threshold of 800 MPa (meeting ISO 6872:2024 for Type II Class 5 ceramic).
    Chemical Solubility: <100 µg/cm² (ISO 6872:2024)Measured values for worst cases were <100 µg/cm² (meeting ISO 6872:2024 requirement).
    Coefficient of Thermal Expansion (CTE): 10-11 x 10-6/°C with standard deviation no greater than 0.5 x 10-6/°C (ISO 6872:2024)Measured values for worst cases were between 10-11 x 10-6/°C and a standard deviation of CTE no greater than 0.5 x 10-6/°C (meeting ISO 6872:2024 requirement).
    Shade Evaluation: Meets shade match requirements (Implied by comparison to VITA Classical and reference Bleach shade guides)Evaluations concluded that BruxZir® Esthetic NOW meets shade match requirements and works as intended.
    Biocompatibility: Biocompatible (per ISO 10993-1:2018; no biocompatibility concern)Cytotoxicity testing results determined no biocompatibility concern regarding yttria concentration, colorants, or other elements.
    Radioactivity: Uranium-238 activity concentration no more than 1.0 Bq/g (ISO 6872:2024)Measured values for worst cases were below the Uranium-238 activity threshold of 1.0 Bq/g (meeting ISO 6872:2024 requirement).
    Packaging Validation: Suitable to withstand distribution environment (per ASTM D4169-22)After testing per ASTM D4169-22, it was determined that the respective packaging for BruxZir® Esthetic NOW blocks is suitable for use.

    2. Sample size used for the test set and the data provenance

    The document specifies "worst-case shades" for Flexural Strength, Chemical Solubility, CTE, and Radioactivity testing, and "all shades offered" for Shade Evaluation. However, the exact numerical sample sizes for these tests (e.g., number of blocks tested per shade) are not provided.

    The data provenance is from non-clinical testing conducted by the manufacturer, Prismatik Dentalcraft, Inc. No information regarding country of origin of data or whether it was retrospective/prospective is specified, as these are material property tests rather than clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    For "Shade evaluation," the document states "qualified evaluators." However, the number of experts and their specific qualifications are not provided. For other non-clinical tests (Flexural Strength, Solubility, CTE, Radioactivity, Biocompatibility, Packaging Validation), ground truth is established by standardized testing methods (e.g., ISO, ASTM) and material science principles, not typically by expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This information is relevant for studies involving human interpretation or subjective assessment with consensus mechanisms, typically in clinical trials or image assessment studies. The described tests are objective material property measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental material (zirconia milling block), not an AI/ML-based diagnostic or assistive device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. As this is not an AI/ML device, this type of study is irrelevant.

    7. The type of ground truth used

    The ground truth for the performance criteria is established by:

    • Standardized measurement thresholds: Specific values defined by international standards (ISO 6872:2024, ISO 10993-1:2018, ASTM D4169-22) for material properties (Flexural Strength, Chemical Solubility, CTE, Radioactivity) and biocompatibility.
    • Visual comparison: For shade evaluation, it relied on comparison against VITA Classical shade guides and reference Bleach shade guides by qualified evaluators.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no training set mentioned or implied.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240882

    Validate with FDA (Live)

    Date Cleared
    2024-05-01

    (30 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

    Device Description

    BruxZir® Esthetic blanks are zirconia milling blanks used for the production of highly esthetic zirconia dental restorations. The high esthetics and high strength of BruxZir® Esthetic restorations make them viable for use in all regions of the mouth. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. BruxZir® Esthetic blanks are available in white, bleach, and pre-shaded varieties that can be easily adjusted to match the final shade.

    AI/ML Overview

    The provided text describes the submission for a dental device, BruxZir® Esthetic (K240882), and compares it to a predicate device (K213425). The focus of the submission is to demonstrate substantial equivalence through non-clinical testing. This is not a submission for an AI/ML powered device, and therefore, many of the requested elements related to AI/ML specific studies (e.g., MRMC studies, standalone AI performance, training set details, ground truth establishment for AI models) are not applicable.

    However, I can extract the relevant information regarding acceptance criteria and the study that proves the device meets those criteria, specifically concerning the physical and chemical properties of the dental material.

    Here's a breakdown of the requested information based on the provided text, while noting the inapplicability of AI/ML specific questions:


    Device: BruxZir® Esthetic (K240882)
    Device Type: Zirconia milling blanks for dental restorations. This is a material, not an AI/ML powered device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Flexural Strength>800 MPa (per ISO 6872:2015/Amd 1:2018, Type II Class 5)Met the criteria (>800 MPa)
    Visual Shade EvaluationGrade of "Pass" for Visual Shade Match for all shades.Met the criteria
    Color ConsistencyΔE calculation needs to be less than 2.00 between samples measured.Met the criteria
    TranslucencyWithin 1% of the equivalent BruxZir Anterior-16 shade at 700nm.Met the criteria
    Coefficient of Thermal Expansion (CTE)Between 10-11 x 10^-6 /°C, and standard deviation of CTE no greater than 0.5 x 10^-6 /°C (per ISO 6872:2015/Amd 1:2018).Met the criteria (Reported range: 10-11 x 10^-6 /°C; implies standard deviation also met based on compliance statement)
    Chemical Solubility<100 µg/cm² (per ISO 6872:2015/Amd 1:2018)Met the criteria (<100 µg/cm²)
    Radioactivity (Uranium-238)Not exceed 1.0 Bq/g (per ISO 6872:2015/Amd 1:2018)Met the criteria (No more than 1.0 Bq/g)
    BiocompatibilityBiocompatible, safe, and non-toxic (per ISO 10993-1:2018)Met the criteria (Biocompatible and non-toxic)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes for each non-clinical test. It states that "the test samples were made using the standard manufacturing process and QC process for BruxZir® Esthetic."

    • Data Provenance: The tests were conducted by Prismatik Dentalcraft, Inc. directly, as part of their design verification "according to BruxZir® Esthetic Design Verification Protocol." This indicates the data is prospective, generated specifically for validating the device's properties based on established industry standards (ISO). The country of origin is not explicitly stated for the testing location, but the applicant (Prismatik Dentalcraft, Inc.) is based in Irvine, California, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable as the studies conducted are non-clinical, laboratory-based physical and chemical property tests, not clinical evaluations requiring expert interpretation of medical images or patient outcomes. The "ground truth" for these tests is the quantitative measurement against an established ISO standard.

    4. Adjudication Method for the Test Set

    Not applicable. The tests are quantitative measurements against predefined ISO standards. There is no human subjective adjudication process for these material property tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This is not an AI-powered device, nor is it a diagnostic imaging device that would typically involve human readers. The submission focuses on the material properties of dental zirconia.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This is not an algorithm or software device.

    7. The Type of Ground Truth Used

    The ground truth for these non-clinical tests is established by industry-standardized test methods (e.g., ISO 6872:2015/Amd 1:2018, ISO 10993-1:2018). The "ground truth" is adherence to the specified quantitative values and qualitative interpretations (e.g., biocompatible).

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.