510(k) Data Aggregation

    K Number
    K182231

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-01

    (257 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brush Biopsy Sets are intended to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula or calyces under direct vision.

    Device Description

    The Brush Biopsy Set and Deflectable Brush Biopsy Set are designed to be used through rigid or flexible endoscopes in order to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula, or calyces under direct vision. The Brush Biopsy Set consists of a stainless steel shaft, a catheter, and a nylon brush at the distal tip. The Brush Biopsy Set is available in two sizes: 3.2 French with a length of 115 centimeters, or 5 French with a length of 67 centimeters. The 3.2 French version has a stainless steel wire shaft, a catheter constructed of radiopaque polytetrafluoroethylene, and a connector cap and pin vise located at the proximal end. The 5 French version has a stainless steel coil shaft, a catheter constructed of vinyl radiopaque tubing, and a side-arm adapter attached to the proximal end.

    The Deflectable Brush Biopsy Set consists of a stainless steel shaft, a radiopaque vinyl catheter, and a nylon brush at the distal tip. The Deflectable Brush Biopsy Set is 5 French with a length of 67 centimeters. The Deflectable Brush Biopsy Set has a 3-ring handle at the proximal end that allows for deflection of the brush tip to within 15° from the shaft.

    The stainless steel shaft of the brush assembly is straight for the length of the catheter, and protrudes 7 centimeters from the distal tip of the catheter on the 5 French Deflectable Brush Biopsy Set to allow for deflection.

    The sets will be supplied sterile and are intended for one-time use. The sets are packaged in a peel-open pouch with a three-year shelf life.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Brush Biopsy Set and Deflectable Brush Biopsy Set, as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text describes performance testing rather than specific, quantified acceptance criteria with corresponding numerical results. Instead, it lists the types of tests performed to demonstrate that the device "met applicable design and performance requirements." It concludes that the device "will perform as intended" based on these tests.

    Therefore, I cannot create a table with specific numerical acceptance criteria and reported device performance. The document only lists the categories of tests conducted.

    Acceptance Criteria (Category)Reported Device Performance (Summary)
    BiocompatibilityMet applicable design and performance requirements
    Dimensional Verification of the CatheterMet applicable design and performance requirements
    Simulated Use and Bristle IntegrityMet applicable design and performance requirements
    Corrosion Following Exposure to Artificial UrineMet applicable design and performance requirements
    Tensile Strength (Catheter Shaft, Wire Shaft to Distal Tip, Proximal Wire Shaft)Met applicable design and performance requirements
    SterilizationMet applicable design and performance requirements
    Packaging/DistributionMet applicable design and performance requirements
    Shelf Life/StabilityMet applicable design and performance requirements
    Acute Performance in a Porcine Urinary ModelPerformed as intended

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for any of the tests (e.g., number of devices tested for tensile strength, number of porcine models used, etc.). It also does not explicitly state the provenance of the data (country of origin, whether retrospective or prospective). The testing appears to be conducted by the manufacturer, Cook Incorporated.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to the type of device and testing described. The Brush Biopsy Set and Deflectable Brush Biopsy Set are physical medical devices used for specimen collection. The "ground truth" in this context would relate to the physical properties and functionality of the device itself (e.g., did it break, did it corrode, did it successfully collect a specimen in a model). Expert interpretation of images or clinical data (which is where expert ground truth and adjudication are typically relevant) is not mentioned as part of the performance data.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated in point 3. Adjudication methods like 2+1 or 3+1 are used for establishing a consensus ground truth among multiple experts, particularly in image interpretation or clinical diagnosis studies. The performance data here pertains to engineering and material testing of a physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study assesses how AI assistance impacts human readers' performance in tasks like diagnostics. The device described is a physical tool for biopsy, not an AI-powered diagnostic system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone algorithm performance study was not done. The device is a physical medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" in this context would be based on:

    • Engineering specifications and standards: For dimensional verification, tensile strength, and corrosion.
    • Direct observation and measurement: For simulated use, bristle integrity, and acute performance in the porcine model (e.g., successful specimen collection, device integrity during use).
    • Laboratory analysis: For biocompatibility and sterilization efficacy.
    • Environmental testing: For packaging/distribution and shelf life/stability.

    There is no mention of expert consensus, pathology (beyond the specimens collected by the device), or outcomes data used to establish the "ground truth" for the device's technical performance.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument; it does not involve a training set as would be used for machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.