Found 1 results

510(k) Data Aggregation

    K Number
    K200153

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-28

    (280 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Direct bonding for light-cured composites to tooth surface
    2. Bonding of dual-cured core build up composites to tooth surface
    3. Intraoral repair of composite, PFM and ceramic restoration using cements
    4. Sealing of tooth preparation for indirect restoration
    Device Description

    Bright Bond Universal is a light-cured dental adhesive that combines etching, priming, and bonding functions in a single bottle solution. It provides strong and durable bonds between dentin/enamel and light-curable direct/indirect restorative materials.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for a dental bonding agent, "Bright Bond Universal," in the context of its 510(k) premarket notification to the FDA. The study presented here is a bench testing study, not a clinical study involving human patients or multi-reader studies. Therefore, many of the requested details about human expert involvement, MRMC studies, and patient data provenance are not applicable to the information contained in this document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated against a set of physical and chemical specifications for a dental bonding agent and biocompatibility standards.

    No.TestMethodAcceptance CriteriaReported Device PerformancePass/Fail
    1CytotoxicityISO 10993-5None cytotoxicity-Pass
    2IrritationISO 10993-10None oral irritation-Pass
    3SensitizationISO 10993-10None sensitization-Pass
    4Acute systemic toxicityISO 10993-11None systemic toxicity-Pass
    5GenotoxicityISO 10993-3None genotoxicity-Pass
    6ImplantationISO 10993-6Biocompatible-Pass
    7Chronic toxicityISO 10993-11No chronic toxicity-Pass
    1Visual testBare eyesNo impurities and No specific changesNo impurities and No specific changes-
    2Capacity testWeight differenceCapacity error of; Standard Capacity ± 5 %2.75 %-
    3Package testBare eyesNo damageNo damages, cracks-
    4Film ThicknessISO 4049:2009 7.5≤ 50 µm2 µm-
    5Sensitivity to Ambient LightISO 4049:2009 7.9Must be physically uniformHomogeneous-
    6Depth of CureISO 4049:2009 7.10≥ 1.5 mm3.0 mm-
    7Bonding Strength (Dentin)ISO 29022:2013 7≥ 8 MPa16 MPa-
    8Bonding Strength (Enamel)ISO 29022:2013 7≥ 8 MPa14 MPa-

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact sample sizes for each bench test. However, these tests are typically conducted on a sufficient number of material samples to ensure statistical validity for the specific test (e.g., multiple specimens for bonding strength, multiple replicates for biocompatibility assays).
    • Data Provenance: The tests were conducted by GENOSS Co., Ltd. (South Korea). The data is from laboratory bench testing of the physical properties and biocompatibility of the bonding agent. It is inherently "prospective" in the sense that the tests were specifically performed to demonstrate the device's characteristics for regulatory submission. It does not involve patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The "ground truth" for these tests is established by industry-standardized test methods (e.g., ISO standards) and measurable physical/chemical properties, not by human expert opinion or interpretation of images. The results are quantitative measurements.

    4. Adjudication Method for the Test Set

    This is not applicable. The tests involve objective measurements (e.g., weight difference, film thickness, bonding strength in MPa, observation of cell viability or tissue reaction), rather than subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. This type of study is relevant for AI/radiology devices where human readers interpret medical images. This document describes a dental material, not an imaging device or an AI algorithm for image interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This is not an AI algorithm. The performance described is the inherent performance of the dental bonding material itself.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on:

    • Standardized Test Methods: Specific ISO standards (e.g., ISO 10993 for biocompatibility, ISO 4049 and ISO 29022 for physical properties) define how the tests are performed and how results are interpreted.
    • Quantitative Measurements: Performance is measured against predetermined quantifiable thresholds (e.g., MPa for bonding strength, µm for film thickness, visual assessment of impurities).
    • Biocompatibility Endpoints: For biocompatibility, the ground truth is established by the absence of adverse biological responses as defined by the ISO 10993 series of standards.

    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" as this is not an AI/machine learning model. The device's formulation and manufacturing processes are likely developed through research and development, but there isn't a "training set" of data in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set.