510(k) Data Aggregation

    K Number
    K190815

    Validate with FDA (Live)

    Device Name
    BrainScope TBI
    Date Cleared
    2019-09-11

    (166 days)

    Product Code
    Regulation Number
    882.1450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BrainScope TBI is a multi-modal, multi-parameter assessment indicated for use as an adjunct to standard clinical practice to aid in the evaluation of patients who have sustained a closed head injury, and have a Glasgow Coma Scale (GCS) score of 13-15 (including patients with concussion/mild traumatic brain injury (mTBI)).

    BrainScope TBI provides a multi-parameter measure (CI)) to aid in the evaluation of concussion in patients between the ages of 13-25 years who present with a GCS score of 15 following a head injury within the past 72 hours (3 days), in conjunction with a standard neurological assessment of concussion. The CI is computed from a multivariate algorithm based on the patient's electroencephalogram (EEG), augmented by neurocognitive measures and selected clinical symptoms.

    The BrainScope TBI Structural Injury Classification ("SIC") uses brain electrical activity (EEG) to determine the likelihood of structural brain injury visible on head CT for patients between the ages of 18-85 years (have a GCS score of 13 – 15), have sustained a closed head injury within the past 72 hours (3 days) who are being considered for a head CT. BrainScope TBI should not be used as a substitute for a CT scan. Negative likely corresponds to those with no structural brain injury visible on head CT. Positive likely corresponds to those with a structural brain injury visible on head CT. Equivocal may correspond to structural brain injury visible on head CT or may indicate the need for further observation or evaluation.

    BrainScope TBI provides a measure of brain Function Index, (BFI)) for the statistical evaluation of the human electroencephalogram (EEG), aiding in the evaluation of head injury as part of a multi-modal, multi-parameter assessment, in patients 18-85 years of age (have a GCS score of 13 - 15) who have sustained a closed head injury within the past 72 hours (3 days).

    The BrainScope TBI device is intended to record, measure, analyze, and display brain electrical activity utilizing the calculation of standard quantitative EEG (QEEG) parameters from frontal locations on a patient's forehead. The BrainScope TBI calculates and displays raw measures for the following standard QEEG measures: Absolute and Relative Power, Asymmetry, Coherence and Fractal Dimension. These raw measures are intended to be used for post hoc analysis of EEG signals for interpretation by a qualified user.

    BrainScope TBI also provides clinicians with quantitative measures of cognitive performance in patients 13-85 years of age to aid in the assessment of an individual's level of cognitive function. These measures interact with the CI and can be used stand alone.

    BrainScope TBI also stores and displays electronic versions of standardized clinical assessment tools that should be used in accordance with the assessment tools' general instructions. These tools do not interact with any other device measures, and are stand alone.

    Device Description

    BrainScope TBI (model: Ahead 500) is a portable, non-invasive, non-radiation emitting, point of care device intended to provide results and measures to support clinical assessments and aid in the diagnosis of concussion / mild traumatic brain injury (mTBI). The BrainScope TBI includes a new multivariate classification algorithm that analyzes a patient's electroencephalogram (EEG), augmented by neurocognitive performance and selected clinical symptoms to compute a multi-modal index called the Concussion Index (CI). BrainScope TBI provides the healthcare provider with a multi-parameter measure to aid in the evaluation of concussion following a head injury within the past 72 hours (3 days). The BrainScope TBI (Ahead 500) retains all the capabilities of the predicate (BrainScope TBI, model: Ahead 400) including the Structural Injury Classification (SIC) and the Brain Function Index (BFI). It also contains configurable, selectable computerized cognitive performance tests and digitized standard clinical assessment tools intended to provide a multi-modal panel of measures to support the clinical assessment of concussion / mTBI.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the BrainScope TBI (model: Ahead 500) device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance (95% CI)
    Sensitivity0.690.8599 (0.8050, 0.9041)
    Specificity0.5650.7078 (0.6588, 0.7535)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 580 subjects
      • 229 matched controls
      • 144 healthy volunteers
      • 207 subjects who sustained closed head injury and were removed from play
    • Data Provenance: The study was conducted across 10 US clinical sites, including High Schools, Colleges, and Concussion Clinics. The study design appears to be prospective, given it involved testing subjects at different time points and with specific inclusion/exclusion criteria.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used or their specific qualifications (e.g., "Radiologist with 10 years of experience") for establishing the ground truth.

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, the document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study to assess how much human readers improve with AI vs. without AI assistance. The study focuses on the standalone performance of the Concussion Index (CI) algorithm.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was done for the Concussion Index (CI). The reported sensitivity and specificity values are for the algorithm's performance in classifying concussions.

    7. Type of Ground Truth Used

    The clinical reference standard (ground truth) incorporated elements from guidelines published in the International Conference on Concussion in Sport (McCrory 2017; 2013) as well as the National Collegiate Athletic Association (NCAA) concussion policy. This suggests a clinical diagnosis/consensus-based ground truth, likely established by clinicians based on established guidelines and possibly direct observations or outcomes related to concussion (e.g., "removed from play"). It's not explicitly stated to be solely pathology or patient outcomes data, but rather a combination of clinical criteria.

    8. Sample Size for the Training Set

    The document states that the "cutoff (threshold) CI [was] derived from an algorithm development study that was independent of the validation study," but it does not provide the sample size for this algorithm development (training) study.

    9. How the Ground Truth for the Training Set Was Established

    The document implies that the ground truth for the "algorithm development study" (training set) would have been established using similar clinical criteria as the validation study, i.e., "consistent with similar changes seen in subjects with concussion," incorporating elements from the International Conference on Concussion in Sport guidelines and NCAA concussion policy. However, it does not explicitly detail the process for establishing ground truth for the training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183241

    Validate with FDA (Live)

    Date Cleared
    2019-02-19

    (90 days)

    Product Code
    Regulation Number
    882.1450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BrainScope TBI is a multi-modal, multi-parameter assessment indicated for use as an adjunct to standard clinical practice to aid in the evaluation of patients who have sustained a closed head injury within the past 72 hours (3 days), are between the ages of 18-85 years, have a Glasgow Coma Scale (GCS) score of 13-15 (including patients with concussion / mild traumatic brain injury (mTBI), and are being considered for a head CT. BrainScope TBI should not be used as a substitute for a CT scan.

    The BrainScope TBI Structural Injury Classification ("SIC") uses brain electrical activity to determine the likelihood of structural brain injury visible on head CT. Negative likely corresponds to those with no structural brain injury visible on head CT. Positive likely corresponds to those with a structural brain injury visible on head CT. Equivocal may correspond to structural brain injury visible on head CT or may indicate the need for further observation or evaluation.

    BrainScope TBI provides a measure of brain function Index, (BFI)) for the statistical evaluation of the human electroencephalogram (EEG), aiding in the evaluation of head injury as part of a multi-modal, multi-parameter assessment.

    The BrainScope TBI device is intended to record, measure, analyze, and display brain electrical activity utilizing the calculation of standard quantitative EEG (QEEG) parameters from frontal locations on a patient's forehead. The BrainScope TBI calculates and displays raw measures for the following standard QEEG measures: Absolute and Relative Power, Asymmetry, Coherence and Fractal Dimension. These raw measures are intended to be used for post hoc analysis of EEG signals for interpretation by a qualified user.

    BrainScope TBI also provides clinicians with quantitative measures of cognitive performance to aid in the assessment of an individual's level of cognitive function. These measures do not interact with any other device measures, and are stand alone.

    BrainScope TBI also stores and displays electronic of standardized clinical assessment tools that should be used in accordance with the assessment tools' general instructions. These tools do not interact with any other device measures, and are stand alone.

    Device Description

    BrainScope TBI is a portable, non-invasive, non-radiation emitting, point of care device intended to provide results and measures to support clinical assessments and aid in the diagnosis of mild traumatic brain injury (mTBI). It also contains configurable, selectable computerized cognitive performance tests and digitized standard clinical assessment intended to provide a multi-modal panel of measures to support the clinical assessment of concussion / mTBI. BrainScope TBI provides healthcare professionals with a set of validated and clinically accepted library of concussion / mTBI assessments.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state formal acceptance criteria with numerical thresholds. Instead, it focuses on demonstrating substantial equivalence to predicate devices. The performance data presented primarily confirms that the new modifications (cognitive performance tests, PECARN, wireless connectivity) work as specified and do not negatively impact existing functionalities.

    Table of Acceptance Criteria (Inferred from Substantial Equivalence Claim) and Reported Device Performance:

    Acceptance Criteria (Inferred/Implicit)Reported Device Performance
    For Cognitive Performance Tests:Functionality of new cognitive testsNormative data was collected from 707 healthy individuals (age 13-85) to construct databases for the cognitive tests. Test data demonstrated that the modifications (additional cognitive performance tests and PECARN) were implemented as per specifications. The new tests and Reliable Change Index (RCI) output are "well accepted in clinical practice for assessment of Adult and Adolescent patient population."
    For Standard Clinical Assessments:Integration of PECARNTest data demonstrated that the modifications (additional cognitive performance tests and PECARN) were implemented as per specifications. PECARN was added to the existing library of digitized standard clinical assessments. The expanded availability of clinical assessment tools does not affect safety and effectiveness and increases utility.
    For Wireless Connectivity (OTA):Functionality of wireless connectivityTest data demonstrated that the modifications (wireless connectivity) were implemented as per specifications. The BrainScope TBI has wireless connectivity to accept Over the Air (OTA) software upgrades. This provides additional data transfer capabilities.
    For Existing EEG Algorithms (SIC, BFI, QEEG):No impact on existing functionalityThe new modifications "did not impact existing device functionality including core EEG based algorithms" (e.g., Structural Injury Classification (SIC) and Brain Function Index (BFI)). The device maintains the same technical characteristics as the predicate for EEG parameters (e.g., bandwidth, CMRR, noise floor, ADC resolution, sampling rate, electrode placement, electrode positions, electrode material, real-time EEG display, EEG-based classification algorithm).
    Basic Safety and EMC Standards:Conformity to relevant standardsThe BrainScope TBI device conforms to "all same basic safety and EMC standards as the predicate." It was also tested to the most recent recognized consensus standard for EMC (IEC 60601-1-2 Ed. 4.0 2014) and other listed standards (e.g., IEC 60601-1/A1:2012, IEC 60601-1-6/A1:2013, IEC 60601-2-26:2012, ANSI/AAMI EC12:2000/(R)2010, ANSI/AAMI/ISO 10993-1:2009, ANSI/AAMI/ISO 10993-5:2009, ANSI/AAMI/ISO 10993-10:2010, MIL-STD-810G, IEC 60529 (2004), ASTM D4169 09).
    Overall Safety and Effectiveness:Comparable to predicate devicesPerformance data demonstrated that the BrainScope TBI is "as safe and effective as the predicates."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 707 healthy individuals.
    • Data Provenance: The document states that normative data was collected from "707 healthy individuals" for the cognitive tests. It does not specify the country of origin, but given the context of FDA submission, it is typically expected to be from a US-based population or a population generalizable to the US. It is a prospective collection for establishing normative data for the new cognitive tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test set of the new cognitive performance tests. It refers to the new tests and RCI output being "well accepted in clinical practice," which implies expert consensus in the field, but no specific quantification of experts is provided within this document. For the EEG-based Structural Injury Classification (SIC), the ground truth is based on the visible injury on head CT, which is a clinical standard.

    4. Adjudication method for the test set:

    Not explicitly stated for the cognitive performance tests or the overall assessment of device modifications. Given the nature of normative data collection for cognitive tests, the "ground truth" is often statistical (e.g., population averages, standard deviations) rather than expert adjudication on individual cases for classification. For the SIC, the ground truth is based on head CT findings, which are objective imaging results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    The document does not mention an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device is described as an "adjunct to standard clinical practice" and "should not be used as a substitute for a CT scan," suggesting it provides additional information rather than directly assisting in CT interpretation by a reader.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the Structural Injury Classification (SIC) and Brain Function Index (BFI) are described as "EEG based algorithms" which "use brain electrical activity to determine the likelihood of structural brain injury visible on head CT." This implies a standalone algorithmic assessment of the EEG data, providing "Negative, Equivocal and Positive outputs" for SIC and a "statistical evaluation of the human EEG" for BFI.

    7. The type of ground truth used:

    • For Structural Injury Classification (SIC): Visible structural brain injury on head CT. This is a form of outcomes data or a clinical standard.
    • For Cognitive Performance Tests: Normative data derived from a population of healthy individuals. This establishes a baseline for cognitive function.
    • For QEEG parameters: Standards related to the accurate recording, measuring, analyzing, and displaying of brain electrical activity, as well as comparison to the predicate device's performance characteristics.

    8. The sample size for the training set:

    The document does not specify the sample size for the training set. It only mentions the "normative data collected from 707 healthy individuals" used to construct databases for the cognitive tests, which likely serves as a reference/validation set for these specific components. The EEG-based algorithms (SIC, BFI) were likely developed and trained on separate, larger datasets that are not detailed in this particular summary, as these algorithms are shared with the predicate device (BrainScope One).

    9. How the ground truth for the training set was established:

    The document does not provide details on how the ground truth for the training set was established for the core EEG algorithms (SIC, BFI) as these algorithms were already established and cleared under the predicate device (BrainScope One). For the new cognitive performance tests and PECARN, the ground truth for the "normative data" was established by collecting data from "healthy individuals," implying a healthy control group without known head injury or cognitive impairment.