Found 2 results

510(k) Data Aggregation

    K Number
    K151325

    Validate with FDA (Live)

    Date Cleared
    2015-08-04

    (78 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bovie J-Plasma Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.

    The Bovie J-Plasma Handpiece is compatible only with Bovie J-Plasma generators.

    Device Description

    The Bovie J-Plasma Handpiece is a sterile, single use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical for the delivery of helium gas plasma for cutting, coagulating, and ablating soft tissue.

    The Bovie J-Plasma Handpiece is available with a retractable cutting blade tip for excising tissue or a needle tip. Both tip configurations serve as electrodes to generate helium plasma. The device is available in 15cm, 33cm, and 45cm lengths. The device can be used in laparascopic or open surgical procedures.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Bovie J-Plasma Handpiece (K151325). The document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with specific acceptance criteria and detailed device performance metrics in a table format.

    Therefore, much of the requested information regarding acceptance criteria, reported device performance, sample sizes for test and training sets, data provenance, expert ground truth establishment, MRMC studies, standalone performance, and ground truth types is not explicitly available in the provided document.

    However, I can extract the information that is present:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria with corresponding reported device performance metrics. Instead, it lists various performance tests and implies their successful completion to demonstrate equivalence.

    Test TitleObjectiveResults/Performance (Implied)
    Mechanical Verification and FunctionalityVerify the mechanical functionality of the handpiece.Met performance requirements. (Implied by conclusion of SE)
    Electrical VerificationVerify the electrical functionality and safety of the handpiece.Met performance requirements. (Implied by conclusion of SE)
    Plasma CharacteristicsMeasure effect of generator settings on the plasma stream, plasma stream characterization.Met performance requirements. (Implied by conclusion of SE)
    Performance EvaluationConfirm device performance on various tissue types.Met performance requirements. (Implied by conclusion of SE)
    Usability EvaluationVerify and validate the usability of the J-Plasma Pistol Grip Handpiece as it relates to safety.Met performance requirements. (Implied by conclusion of SE)
    Shipping StudyVerify the device meets requirements after shipping and handling.Met performance requirements. (Implied by conclusion of SE)

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the document. The document mentions "Performance testing" but does not detail the sample sizes for any of the tests. Data provenance (country of origin, retrospective/prospective) is also not stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document. The tests described are primarily bench tests and engineering assessments, not clinical or diagnostic studies that would typically involve expert ground truth.

    4. Adjudication method for the test set:
    This information is not provided in the document. As there is no mention of expert ground truth, an adjudication method is not applicable in this context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not provided in the document. The device in question is an electrosurgical handpiece, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study with human readers and AI assistance would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This information is not provided in the document. The device is a surgical instrument and does not involve an algorithm for standalone performance evaluation in the context of AI.

    7. The type of ground truth used:
    For the listed performance tests (mechanical, electrical, plasma characteristics, performance evaluation, usability, shipping), the "ground truth" would likely be based on engineering specifications, industry standards (e.g., IEC standards listed), and predefined functional requirements. For "Performance Evaluation on various tissue types," the ground truth would be the observed effect of the device (cutting, coagulation, ablation) on those tissues, assessed against desired outcomes or predicate device performance. However, the specific methodology for establishing this ground truth (e.g., histological analysis, expert observation) is not detailed.

    8. The sample size for the training set:
    This information is not provided in the document. The document describes a medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established:
    This information is not provided in the document. Not applicable as this is not an AI model.

    Summary of what is available:

    The document primarily serves as a 510(k) summary for a medical device and describes various performance tests (bench tests) conducted to ensure the device meets requirements and is substantially equivalent to a predicate. It does not provide detailed results or specific acceptance criteria in a quantitative, tabular format beyond the general statement that the device "meets performance requirements." The "study" here refers to the collection of these performance tests. There is no mention of clinical studies, human readers, AI, or large datasets for algorithm training or testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112233

    Validate with FDA (Live)

    Date Cleared
    2012-01-27

    (176 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bovie® J-Plasma Handpiece with Retractable Cutting Blade is an electrosurgical device that utilizes helium gas and a retractable blade for cutting and coagulation of soft tissue for use during both open and laparoscopic surgical procedures.

    Device Description

    The Bovie® J-Plasma Handpiece with Retractable Cutting Tip are single use devices designed to be compatible only with the ICON GS generator (GS100) and reusable cable (GS-807-RC) cleared under 510k K-090586. There are device models for both open and laparoscopic surgical procedures (GS-018C and GS-270C respectively). These handpieces contain a programmable memory device that requires a unique code for the generator to operate. Each handpiece has a button for finger activation of the generator. Once activated, helium gas flows out the distal end of the device where gas molecules are excited by the high electric field resulting in a plasma stream. The handpiece that is intended for open procedures does not have an extended shaft. The handpiece intended for laparoscopic procedures has a 27 cm extended tip. Both handpieces contain a retractable cutting blade that can be extended manually for excising tissue.

    AI/ML Overview

    This FDA 510(k) summary describes a device that is an electrosurgical tool, and therefore, the acceptance criteria and study information typically associated with AI/ML-driven diagnostics or imaging devices are not provided in this document. This submission focuses on demonstrating substantial equivalence to a predicate device, primarily through comparison of design, materials, and performance characteristics for electrosurgical functionality.

    Here's a breakdown of the information that can be extracted from the document, and what cannot be found due to the nature of the device:

    Device: Bovie® J-Plasma Handpiece with Retractable Cutting Blade

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not present quantitative acceptance criteria or detailed performance metrics in the format typically seen for AI/ML diagnostic tools. Instead, it focuses on demonstrating equivalence to a predicate device through shared characteristics and compliance with electrical safety standards.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Indications for UseEquivalent or broader than predicate device."The Bovie® J-Plasma Handpiece with Retractable Cutting Blade is an electrosurgical device that utilizes helium gas plasma for coagulation of soft tissue during both open and laparoscopic surgical procedures. ... utilizes a retractable blade for cutting of soft tissue during both open and laparoscopic surgical procedures." (Covers both plasma coagulation and blade cutting, for both open and laparoscopic procedures, expanding on the predicate's "coagulation of soft tissues during open soft tissue surgery.")
    Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-2-2."Both named devices and the predicate are designed to comply with the following electrical standards: IEC 60601-1: ... General Requirements for Safety (2005) IEC 60601-2-2: ... Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories (2009)"
    Output Energy (Coagulation)Equivalent to predicate device."40 Watts (100% Setting on ICON GS Generator)" (Same as predicate)
    Delivery SystemEquivalent to predicate device."Monopolar" (Same as predicate)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    Not Applicable. This information is relevant for studies involving data-driven algorithms (like AI/ML models). This document pertains to an electrosurgical hardware device, and thus, this type of data is not generated or tested in the same manner. The "testing" for this device would involve engineering verification and validation, performance testing (e.g., power output, safety features), and biocompatibility, not data set evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not Applicable. As this is not an AI/ML diagnostic or imaging device, there is no "ground truth" in the sense of expert consensus on medical image interpretation. The ground truth for this device would relate to its physical performance and safety (e.g., does it cut tissue as intended, does it coagulate effectively, does it meet electrical safety standards).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not Applicable. Adjudication methods are used to resolve disagreements among human annotators or diagnosticians creating a ground truth for AI model training or evaluation. This is not relevant for an electrosurgical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not Applicable. An MRMC study is specific to diagnostic imaging devices where human readers interpret medical images, often with and without AI assistance. This device is a surgical tool, not an imaging or diagnostic AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not Applicable. This device is a manual surgical tool; it does not operate as an "algorithm only" or in a "standalone" fashion without human interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not Applicable. As mentioned above, there isn't a "ground truth" of this nature for an electrosurgical device. The "truth" is established through engineering and performance testing against predefined specifications and safety standards (e.g., power meter readings, tissue ablation tests, sterility testing).

    8. The sample size for the training set:

    Not Applicable. This device does not use a "training set" as it is not an AI/ML device.

    9. How the ground truth for the training set was established:

    Not Applicable. This device does not have a "training set" or "ground truth" to be established in the context of AI/ML.