510(k) Data Aggregation

    K Number
    K250047

    Validate with FDA (Live)

    Device Name
    bite away two
    Manufacturer
    Date Cleared
    2025-09-12

    (245 days)

    Product Code
    Regulation Number
    890.5740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The bite away® two is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes.

    Device Description

    The mibeTec bite away® two device is a light weight, portable, handheld, battery powered, user-operated device that produces mild heat for direct contact with the affected areas of the skin. The heat is initiated by the user through the activation of the unit by depressing a non-locking button. The device is provided with two (2) side-by-side buttons. The user has the choice of a short 3 second heat treatment or a 5 second heat treatment depending on which button is depressed. The activation of the unit is signified with the illumination of an LED light and an audible chirp through the use of an electronic buzzer. An audible error message is activated when there is a user error. The device is not connected to the user as the user is in complete control of the heat treatment and as such, self-delivers the heat treatment to themselves by contacting the device to their own anatomy / treatment site. As in any user-controlled heat therapy, the user determines the tolerance to such delivered heat and applies and removes the heat treatment per their own tolerance; thereby virtually alleviating the risks of inadvertent overheating of the skin. The short duration of treatment time of 3 and 5 seconds and the small surface area of the heated ceramic plate that contacts the patient further alleviates risk of overheating the skin as the device automatically stops heating the element after 3 or 5 seconds; limiting the maximum amount of heat to be delivered to the site.

    The mibeTec bite away® two device utilizes two AAA batteries (1.5 volts each) that power a MOSFET that heats a 7mm ceramic disc to approx. 51C +/- 1% when activated by the user through the use of one of the two electro-mechanical NOC (normal open contact) buttons (3 seconds or 5 seconds of heating). The device consists of a few major components: a hard plastic outer case, a printed circuit board, two activation buttons, two AAA batteries, and a heated ceramic plate that contacts the patient.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K220514

    Validate with FDA (Live)

    Device Name
    bite away neo
    Manufacturer
    Date Cleared
    2022-06-23

    (121 days)

    Product Code
    Regulation Number
    890.5740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The bite away® neo is indicated for use to provide temporary relief of the pain and itching from insect stings and bites such as bees, wasps and mosquitoes.

    Device Description

    The bite away® neo device is a light weight, portable, hand-held, battery powered, user- operated device that produces mild heat for direct contact with the affected areas of the skin. The heat is initiated by the user through the activation of the unit by depressing a non-locking button. The device is provided with two (2) side-by-side buttons. The user has the choice of a short 3 second heat treatment or a 5 second heat treatment depending on which button is depressed. The activation of the unit is signified with the illumination of an LED light and an audible chirp through the use of an electronic buzzer. An audible alarm is activated when there is a user error. The device is not connected to the user as the user is in complete control of the heat treatment and as such, self-delivers the heat treatment to themselves by contacting the device to their own anatomy / treatment site. The short duration of 3 and 5 seconds of heating the device's heated ceramic plate that contacts the patient further alleviates risk of overheating the skin as the device automatically stops heating the element after 3 or 5 seconds; limiting the maximum amount of heat to be delivered to the site.

    The bite away® neo device utilizes two AA batteries (1.5 volts each) that power a resistor that heats a 7mm ceramic disc to approximately 50°C when activated by the user through the use of one of the two electro-mechanical NOC (normal open contact) buttons (3 seconds of heating). The device consists of a few major components: a hard plastic outer case, a printed circuit board, an activation button. AA batteries, and a heated ceramic plate that contacts the printed circuit board consists of the following components: electro-mechanical NOC buttons, microcontroller with programmed firmware, MOSFET transistor heating element, ceramic pad, resistors, coils, capacitor, thermistor temperature sensor and buzzer. The bite away® neo device is approximately 7 inches long and weighs approximately 40 grams.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from MibeTec, GmbH to the FDA for their device, the bite away® neo. The device is a powered heating pad intended for the temporary relief of pain and itching from insect stings and bites.

    Since this document is a 510(k) submission, the study performed is a non-clinical performance study to demonstrate substantial equivalence to a predicate device, not a comparative effectiveness study with human readers (MRMC study) or a standalone algorithm performance study. The core of this submission is comparison to a predicate device and demonstration that the new device is as safe and effective.

    Here's an breakdown of the requested information based on the provided document:


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical thresholds for clinical efficacy in this document, as this is a 510(k) submission demonstrating substantial equivalence based primarily on non-clinical performance and design characteristics, rather than de novo clinical efficacy trials. The implicit acceptance criterion is that the device is as safe and effective as the predicate device and meets relevant safety and performance standards.

    The document demonstrates this by:

    • Identical Indications for Use: The bite away® neo and the predicate device (Riemser Bite Away, K160943) are indicated for the same purpose.
    • Similar Design Principles: Both are user-operated, hand-held, battery-powered, reusable devices that deliver mild heat to the skin via a ceramic disc.
    • Compliance with Safety Standards: The device underwent and passed electrical safety, EMC, software verification/validation, and usability testing according to international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62366, IEC 60601-1-6) and FDA guidance for biocompatibility (ISO 10993-1).

    Table of Acceptance Criteria (Implicit) and Reported Device Performance:

    Acceptance Criterion (Implicitly Met)Reported Device Performance (vs. Predicate K160943)
    Indications for Use EquivalenceIdentical: Both devices are indicated for temporary relief of pain and itching from insect stings and bites (bees, wasps, mosquitoes). The difference in claim of "increasing localized blood flow" for the predicate is stated to not affect safety or efficacy.
    Intended Use EquivalenceIdentical: Delivery of mild heat to the skin/dermis.
    Temperature Profile EquivalenceSimilar Temperature Range: Subject device: 50 - 53°C (device surface); Predicate: 50 - 53°C (device surface). Max skin surface temp for subject: 48.6°C. Rationale: Identical temperature range, no additional risks.
    Energy Transfer Mechanism EquivalenceIdentical: Both use a heated ceramic disc.
    Power Source EquivalenceIdentical: Both use AA Batteries (3 volts DC).
    Operator Control / Application Method EquivalenceIdentical: Hand-held, user-operated, applied directly to the insect bite, user in complete control of heat treatment.
    BiocompatibilityCompliant: Patient contact polymers and ceramics evaluated against ISO 10993-1, including Cytotoxicity, Sensitization, and Irritation.
    Electrical Safety & Electromagnetic Compatibility (EMC)Compliant: Tested against IEC 60601-1 (MOD), IEC 60601-1-2, and IEC 60601-1-11.
    Software Verification and ValidationCompliant: Tested following FDA Guidance for "moderate" level of concern software; complies with IEC 62366 (implied).
    UsabilityCompliant: Tested against IEC 60601-1-6.
    Risk Mitigation (Differences from Predicate)Reduced Heating Time: Subject device has 3 and 5 seconds vs. predicate's 3 and 6 seconds. Rationale: "Greater safety with less heating time." Smaller Heated Area (7mm vs 9mm): Rationale: "Small disc size delivers heat to less surface area; reducing risk. User can apply to multiple areas for more surface area treatments." These differences are argued to enhance safety.
    Material & Component SimilaritiesHard plastic outer case, LED light, audible signal, microprocessor.
    Regulatory Classification EquivalenceIdentical: Powered Heating Pad, Class II, Product Code IRT, CFR Section 890.5740, OTC Use.

    2. Sample Size for the Test Set and Data Provenance

    The document explicitly states: "No clinical studies were performed to support safety or effectiveness of the subject device."

    Therefore, there is no "test set" in the traditional sense of a clinical data set used to prove efficacy. The "testing" refers to non-clinical bench testing (biocompatibility, electrical safety, software validation, usability) against established standards. The data provenance is from laboratory testing and engineering analysis conducted by the manufacturer, MibeTec, GmbH (Germany), as part of their premarket notification submission.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. As no clinical studies were performed, there was no "ground truth" derived from expert consensus on medical images or patient outcomes. The "ground truth" for the non-clinical tests is adherence to engineering and safety standards, which are defined by international bodies and regulatory agencies like the FDA. The experts involved would be the engineers, quality assurance personnel, and regulatory affairs specialists at the manufacturer and the testing laboratories.


    4. Adjudication Method for the Test Set

    N/A. There was no clinical test set requiring adjudication. Adjudication methods (e.g., 2+1, 3+1) are typically used for establishing ground truth in clinical image interpretation studies or clinical trials, which were not conducted here.


    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "No clinical studies were performed to support safety or effectiveness of the subject device." Therefore, no MRMC study was conducted, and no effect size for human readers improving with or without AI assistance is provided. (Note: This device is a heating pad, not an AI-powered diagnostic tool, so an MRMC study would generally not be applicable in this context).


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    N/A. This device is a powered heating pad, not an AI algorithm. There is no software or algorithm whose performance is being evaluated in a standalone manner. The software mentioned is for the device's internal control (e.g., heating duration, temperature regulation), verified as "moderate" level of concern.


    7. The Type of Ground Truth Used

    The "ground truth" for this 510(k) submission is established through:

    • Engineering and Safety Standards: Compliance with international standards (IEC 60601 series, ISO 10993, IEC 62366).
    • Predicate Device Equivalence: The legally marketed predicate device (K160943) serves as the primary benchmark for demonstrating safety and effectiveness. The premise of a 510(k) is that if the new device is substantially equivalent to a legally marketed predicate, it is considered safe and effective.

    8. The Sample Size for the Training Set

    N/A. As this is not an AI/machine learning device, there is no "training set" in the conventional sense. The device's design (e.g., microcontroller firmware) would be developed and refined through engineering iterations, but not trained on a distinct dataset like an AI model.


    9. How the Ground Truth for the Training Set Was Established

    N/A. See #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160943

    Validate with FDA (Live)

    Device Name
    Bite Away
    Manufacturer
    Date Cleared
    2016-08-26

    (143 days)

    Product Code
    Regulation Number
    890.5740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bite Away is indicated for use to provide temporary relief of the pain and itching from insect stings and bites such as bees, wasps and mosquitoes by temporarily increasing localized blood flow.

    Device Description

    The Riemser Pharma Bite Away device is a light weight, portable, hand held, battery powered, user-operated device that produces mild heat for direct contact with the affected areas of the skin. The heat is initiated by the user through the activation of the unit by depressing a non-locking button. The device is provided with two (2) side-by-side buttons. The user has the choice of a short 3 second heat treatment or a 6 second heat treatment depending on which button is depressed. The activation of the unit is signified with the illumination of an LED light and an audible chirp through the use of an electronic buzzer. An audible alarm is activated when there is a user error. The device is not connected to the user as the user is in complete control of the heat treatment and as such, self-delivers the heat treatment to themselves by contacting the device to their own anatomy / treatment site. The short duration of 3 and 6 seconds of heating the device's heated ceramic plate that contacts the patient further alleviates risk of overheating the skin as the device automatically stops heating the element after 3 or 6 seconds; limiting the maximum amount of heat to be delivered to the site.

    The Riemser Pharma Bite Away device utilizes two AA batteries (1.5 volts each) that power a resistor that heats a 9mm ceramic disc to 51.5℃ +/- 1% when activated by the user through the use of one of the two electro-mechanical NOC (normal open contact) buttons (3 seconds or 6 seconds of heating). The device consists of a few major components: a hard plastic outer case, a printed circuit board, an activation button, AA batteries, and a heated ceramic plate that contacts the patient. The printed circuit board consists of the following components: electro-mechanical NOC buttons, microcontroller with programmed firmware, MOSFET transistor heating element, ceramic pad, resistors, coils, capacitor, thermistor temperature sensor and buzzer.

    The Riemser Pharma Bite Away device is approximately 6 inches long and weighs approximately 25 grams. The Bite Away device is provided in 2 shapes, both of which are identical in every way except for the shape. One product is an elongated configuration called "Cobra" and the other configuration is an oblong configuration called "Mouse". Although these products have different shapes and names, the products are identical in every way to include raw materials, subcomponents, electronic circuits, electronic components, and imbedded software code, fabrication processes, etc., and only differs in the shape of the outer case (elongated Cobra model versus the oblong Mouse model).

    AI/ML Overview

    This document describes the marketing authorization for the "Bite Away" device, a powered heating pad intended for temporary relief of pain and itching from insect stings and bites. The study provided focuses on clinical data to support the device's indications for use.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The concept of "acceptance criteria" is not explicitly defined as a numeric threshold in the document for the clinical study's primary endpoint. Rather, the study aims to demonstrate a statistically significant reduction in pain, itching (pruritus), and swelling as measured by a Visual Analog Scale (VAS) after using the Bite Away device. The reported performance shows significant improvement across these metrics.

    Acceptance Criteria (Implicit)Reported Device Performance
    Primary Endpoint: Statistically significant reduction in VAS scores for swelling, pain, and pruritus after treatment.Swelling: Mean VAS score decreased from 4 (prior) to 2 (2-5 min post-treatment) and 1 (10 min post-treatment) (p < 0.0001).
    Pain: Mean VAS score decreased from 6 (prior) to 2 (2 min post-treatment), 1 (5 min post-treatment), and 0 (10 min post-treatment) (p < 0.0001).
    Pruritus: Mean VAS score decreased from 5 (prior) to 2 (2 min post-treatment) and 0 (5 and 10 min post-treatment) (p < 0.0001).
    Safety: No mention of specific safety acceptance criteria, but generally implied to be safe for its intended use.Not explicitly quantified in this section directly, but device safety is typically assessed through adverse event reporting which is not detailed here. The clinical study is a "single arm, open label" which might not be designed to extensively assess comparative safety but rather efficacy in a real-world setting.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: 146 subjects
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study data. It is a prospective, single-arm, open-label clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The study utilized a Visual Analog Scale (VAS) for pain, swelling, and pruritus. This is a subjective patient-reported outcome. Therefore, there were no external "experts" used to establish a ground truth in the traditional sense (e.g., radiologist opinions). The "ground truth" was the subjects' own reported experience via the VAS scores.

    4. Adjudication Method for the Test Set

    Since the "ground truth" was based on patient-reported VAS scores, there was no adjudication method described for external review of these subjective reports.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was NOT done. The device (Bite Away) is a direct treatment device, not an AI-powered diagnostic or assistive tool for human readers. The study evaluates the device's direct therapeutic effect on patients.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The "Bite Away" is a physical device, not an algorithm. Therefore, the concept of a "standalone" algorithmic performance study does not apply to this device. The clinical study evaluates the device's performance when used by a human (user-operated).

    7. The Type of Ground Truth Used

    The ground truth used was patient-reported outcomes (PROs) via a Visual Analog Scale (VAS) for pain, swelling, and pruritus.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve an AI algorithm that requires a training set. The clinical study evaluates the device's performance in a direct therapeutic context.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI algorithm being developed or evaluated, there is no training set or ground truth establishment for a training set.