Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220870

    Validate with FDA (Live)

    Date Cleared
    2022-10-20

    (209 days)

    Product Code
    Regulation Number
    866.3966
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioFire® Global Fever Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with BioFire® FilmArray® 2.0 and BioFire® FilmArray® Torch Systems. The BioFire Global Fever Panel detects and identifies selected bacterial, viral, and protozoan nucleic acids directly from EDTA whole blood collected from individuals with signs and/or symptoms of acute illness or recent acute febrile illness and known or suspected exposure to the following target pathogens: chikungunya virus (serotypes 1, 2, 3 and 4), Leptospira spp., and Plasmodium spp. (including species differentiation of Plasmodium falciparum and Plasmodium vivax/ovale). Evaluation for more common causes of acute febrile illness (e.g., infections of the upper and lower respiratory tract or gastroenteritis, as well as non-infectious causes) should be considered prior to evaluation with this panel. Results are meant to be used in conjunction with other clinical, epidemiologic, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities.

    Positive results do not rule out co-infections with pathogens not included on the BioFire Global Fever Panel. Not all pathogens that cause acute febrile illness are detected by this test, and negative results do not rule out the presence of other infections. In the United States, patient travel history and consultation of the CDC Yellow Book should be considered prior to use of the BioFire Global Fever Panel as some pathogens are more common in certain geographical locations.

    For In Vitro Diagnostic Use.

    Device Description

    The BioFire Global Fever Panel is a multiplexed nucleic acid-based test for the detection and identification of six pathogens which cause acute febrile illness (AFI) from whole blood specimens on BioFire FilmArray systems. The BioFire Global Fever Panel detects and identifies the following pathogens: chikungunya virus, dengue virus, Leptospira spp., and Plasmodium spp., including species differentiation between P. falciparum and P. vivax/ovale. The BioFire Global Fever Panel was originally described and was granted De Novo classification in DEN200043.

    The BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel is an assayed quality control intended for monitoring the diagnostic performance of the BioFire Global Fever Panel. The Control Kit consists of Positive and Negative External Controls in a FilmArray Control Injection Vial format. The Positive External Control contains external assayed quality control material consisting of a set of non-infectious DNA segments dried on the filter of a FilmArray Control Injection Vial and detected by the Global Fever Panel. The Negative External Control contains no DNA and is also provided in the Control Injection Vial format. Analysis of the controls is carried out by specific pouch modules that are included in the BioFire Global Fever Panel Pouch Module Package. The BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel was fully described and cleared in K202382.

    The purpose of this submission is to add BioFire FilmArray Torch as an additional instrument system for use with the BioFire Global Fever Panel and BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel, which were previously marketed for use with BioFire FilmArray 2.0 Systems. The FilmArray Torch is a modular configuration of the FilmArray 2.0 that minimizes instrument footprint by stacking up to twelve individual FilmArray Torch Modules on top of a single FilmArray Torch System Base. This 510(k) request describes modifications to the BioFire Global Fever Panel Pouch Module Package software and validation efforts to support adding FilmArray Torch Systems to the intended use of both the BioFire Global Fever Panel and associated BIOFIRE SHIELD Control Kit.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state formal "acceptance criteria" in a separate section. Instead, the performance data is presented as a comparison to the predicate device (BioFire FilmArray 2.0 system), implying that non-inferiority or comparable performance to the established predicate is the implicit acceptance criterion.

    MetricAcceptance Criteria (Implied: Comparable to FilmArray 2.0)Reported Device Performance (FilmArray Torch)
    Overall Agreement with Expected Result (BioFire Global Fever Panel for all analytes, all concentrations combined)99.4% [98.7-99.7%] (FilmArray 2.0)98.9% [98.1-99.4%] (FilmArray Torch)
    Detection Rate: Leptospira interrogans (Moderate Positive, 3xLoD)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Leptospira interrogans (Low Positive, 1xLoD)95.6% [89.1-98.3%]94.4% [87.6-97.6%]
    Detection Rate: Leptospira interrogans (Negative)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Dengue virus DENV-2 (Moderate Positive, 3xLoD)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Dengue virus DENV-2 (Low Positive, 1xLoD)100% [95.9-100%]98.9% [94.0-99.8%]
    Detection Rate: Dengue virus DENV-2 (Negative)100% [95.9-100%]98.9% [94.0-99.8%]
    Detection Rate: Plasmodium falciparum (Moderate Positive, 3xLoD)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Plasmodium falciparum (Low Positive, 1xLoD)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Plasmodium falciparum (Negative)100% [95.9-100%]100% [95.9-100%]
    Detection Rate: Plasmodium spp. (BEI / MRA-1238) (Moderate Positive, 3xLoD)100% [95.9-100%]98.9% [94.0-99.8%]
    Detection Rate: Plasmodium spp. (BEI / MRA-1238) (Low Positive, 1xLoD)96.7% [90.7-98.9%]95.6% [89.1-98.3%]
    Detection Rate: Plasmodium spp. (BEI / MRA-1238) (Negative)100% [95.9-100%]100% [95.9-100%]
    Overall Agreement with Expected Result (BIOFIRE SHIELD Control Kit, Positive Control)Not explicitly stated (implied: high agreement)100% [97.2-100%]
    Overall Agreement with Expected Result (BIOFIRE SHIELD Control Kit, Negative Control)Not explicitly stated (implied: high agreement)99.3% [95.9-99.9%]
    Overall Agreement with Expected Result (BIOFIRE SHIELD Control Kit, all controls combined)Not explicitly stated (implied: high agreement)99.6% [97.9-99.9%]

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set for BioFire Global Fever Panel (Table 3):

      • Sample Size: For each analyte (e.g., Leptospira interrogans), there were 3 concentrations tested (Moderate Positive, Low Positive, Negative). Each concentration was tested with 3 FilmArray 2.0 systems and 3 FilmArray Torch systems. On each system, 30 replicates were run.
        • For a single analyte and concentration, 30 replicates/system * 3 systems = 90 replicates for each platform.
        • Total replicates per analyte (3 concentrations): 90 replicates/platform * 3 concentrations = 270 replicates for each platform.
        • Total replicates across all analytes and concentrations for comparison: 1080 replicates on FilmArray 2.0 and 1080 replicates on FilmArray Torch.
        • The exact number of unique "samples" (batches of contrived material) is not specified, but the number of test runs is clear.
      • Data Provenance: The replicates are "contrived samples containing representative pathogens... at concentrations near the limit of detection (LoD)" and "negative samples containing no analyte." The exact country of origin is not stated, but the study was conducted by BioFire Defense, LLC (Salt Lake City, UT, USA). It is a prospective study as new testing was performed to evaluate the FilmArray Torch system.
    • Test Set for BIOFIRE SHIELD Control Kit (Table 4):

      • Sample Size:
        • Positive External Controls: 135 replicates (45 replicates/system * 3 FilmArray Torch systems).
        • Negative External Controls: 135 replicates (45 replicates/system * 3 FilmArray Torch systems).
        • Total: 270 replicates.
      • Data Provenance: Details are for the BioFire FilmArray Torch platform only. The data is prospective, generated specifically for this evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For both studies (BioFire Global Fever Panel and BIOFIRE SHIELD Control Kit), the ground truth was established by the known concentration/presence of the analyte in the contrived samples or controls. While operators performed the tests, there is no mention of "experts" establishing the ground truth of the test material, as it's an analytical performance study using characterized samples.

    4. Adjudication Method for the Test Set

    There is no mention of an adjudication method in the context of expert review. The "expected result" for each test run (Detected/Not Detected) was based on the known composition of the contrived samples or control materials.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance?

    This type of study was not performed, nor is it applicable. The device is an in vitro diagnostic test that provides automated interpretation and report generation. There is no "human reader" component in the interpretation of the results to be improved by AI assistance. The document explicitly states: "Automated test interpretation and report generation; user cannot access raw data."

    6. If a Standalone (i.e. Algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone study was done. The BioFire Global Fever Panel runs on the FilmArray system, which is an automated device. The performance data presented in Table 3 and Table 4 reflects the algorithm's performance (i.e., the device's ability to detect and identify targets) directly on the FilmArray Torch system, without human intervention in the result interpretation. The objective of this submission was to evaluate the performance of the existing panel on a new instrument system (FilmArray Torch), which itself is automated.

    7. The Type of Ground Truth Used

    The ground truth used was known composition of contrived samples/controls. This means:

    • For positive samples, the specific analytes and their concentrations (e.g., 3xLoD, 1xLoD) were pre-determined.
    • For negative samples and controls, the absence of the target analyte was pre-determined.

    8. The Sample Size for the Training Set

    The document does not provide information on a training set. This submission is for adding a new instrument system (BioFire FilmArray Torch) for an already existing and cleared device (BioFire Global Fever Panel). The "BioFire Global Fever Panel was originally described and was granted De Novo classification in DEN200043." Therefore, any initial development and potential "training" (if applicable for the underlying algorithm) would have occurred during the development phase for DEN200043, and those details are not part of this 510(k) summary. The current study is a verification/validation for expanded instrument compatibility.

    9. How the Ground Truth for the Training Set Was Established

    As no training set information is provided in this document, the method for establishing its ground truth is also not available here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1