Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231408

    Validate with FDA (Live)

    Date Cleared
    2023-08-11

    (88 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

    Device Description

    The subject device is a patient examination glove made of synthetic nitrile latex compound. It is Non-Sterile, Powder free and is Blue in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per ISO 10993-11.

    AI/ML Overview

    The provided text describes Shandong Intco Medical Products Co, Ltd.'s Basic Synguard Nitrile Exam Gloves. This is a medical device submission, and the acceptance criteria and study described relate to the physical and biocompatibility performance of these gloves.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsASTM D6319-19
    Length230 mm minPass (240 mm min)
    WidthXS: 70±10, S: 80±10, M: 95±10, L: 110±10, XL: 120±10, XXL: 130±10Pass (XS: avg 78.3mm, S: avg 86.5mm, M: avg 97.6mm, L: avg 108.7mm, XL: avg 115.4mm, XXL: avg 123.7mm)
    ThicknessPalm: Minimum 0.05 mm, Finger: Minimum 0.05 mmPass (Palm 0.06mm min, Finger 0.082mm min)
    Physical PropertiesASTM D6319-19
    Before Aging
    Tensile Strength14 MPa, minPass (22.6 MPa, min)
    Elongation500%, minPass (531%, min)
    After Aging
    Tensile Strength14 MPa, minPass (21.9 MPa, min)
    Elongation400%, minPass (437%, min)
    Freedom from holesASTM D5151-19 (No leakage at sampling level of G-1, AQL 2.5)Pass (No leakage, 312 of 315 gloves passed)
    Residual PowderASTM D6124-06 (Reapproved 2017) (< 2mg per glove)Pass (average 0.17 mg per glove)
    Biocompatibility
    Skin SensitizationISO 10993-10 (Not a sensitizer)Under the conditions of the study, not a sensitizer
    Intracutaneous ReactivityISO 10993-10 (Not an irritant)Under the condition of study, not an irritant
    Acute Systemic ToxicityISO 10993-11 (No evidence of systemic toxicity)Under the conditions of this study, there was no evidence of systemic toxicity. (also stated as "the device extracts do not pose a systemic toxicity" in the predicate comparison table)

    2. Sample size used for the test set and the data provenance

    • Freedom from holes: 315 gloves were tested.
    • Other physical properties (Dimensions, Physical Properties, Residual Powder): The document does not explicitly state the sample size for each test, but it references ASTM D6319-19, ASTM D6124-06, and ASTM D5151-19, which would define the sampling plans. For "Freedom from holes," the sample size of 315 and AQL 2.5 is typical for ASTM D5151.
    • Biocompatibility: The sample sizes for the ISO 10993 tests are not explicitly stated in this document but are defined by the respective ISO standards.
    • Data Provenance: The studies were conducted by Shandong Intco Medical Products Co., Ltd. (China) for their device. The nature of these tests (physical and biocompatibility testing on manufactured products) categorizes them as prospective performance evaluations of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the device is a medical glove, and the "ground truth" is established through standardized laboratory testing against objective criteria (e.g., tensile strength, dimensions, presence of holes, chemical residues, biological reactions in test models) as defined by ASTM and ISO standards, not expert interpretation of cases.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as the device is a medical glove, and the performance is measured directly by physical and chemical tests, not through human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The device is a medical glove, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is a medical glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is established by:

    • Pre-defined physical and chemical specifications from ASTM standards (e.g., minimum tensile strength, maximum residual powder, dimensional tolerances).
    • Biological responses (e.g., no irritation, no sensitization, no systemic toxicity) in animal or in vitro test models, as defined by ISO 10993 standards.

    8. The sample size for the training set

    This section is not applicable. The device is a medical glove, and this is not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    This section is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1