Found 2 results

510(k) Data Aggregation

    K Number
    K243606

    Validate with FDA (Live)

    Device Name
    Babyroo TN300
    Date Cleared
    2025-01-17

    (57 days)

    Product Code
    Regulation Number
    880.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Babyroo TN300 is an open care radiant warmer that provides a controlled source of heat and regulation of skin temperature for neonates and infants. The optional integrated resuscitation module provides emergency respiratory support administered by clinicians and includes the functionality of suction. Additionally, the device provides weighing (optional) and pulse oximetry (optional) of neonates and infants. The device is designed for use with a body weight up to 10 kg (22 lb).

    The device is indicated for thermoregulation, skin temperature regulation, weighing (optional), pulse oximetry (optional), and resuscitation (optional) of neonates and infants.

    Device Description

    The purpose of this premarket notification is to activate, for the US market, existing optional functionality which enables the display of SpO2 information by the subject Babyroo TN300 via its graphical user interface (GUI). The SpO2 display functionality is facilitated through connection with existing, cleared pulse oximetry accessories which generate the SpO2 information that is displayed on the subject Babyroo TN300 GUI.

    As originally cleared under K230278:
    The Babyroo TN300 is an open care infant radiant warmer that provides controlled source heat and skin temperature display for use with neonates and infants. The device can be configured for either labor and delivery or the newborn intensive care unit (NICU) and can be used for intra hospital transfer. Warming therapy is interrupted during intra hospital transfer and patient is not supplied with heat.

    The Babyroo TN300 device is offered with a fixed height or adjustable height trolley configuration and provides two heat sources for infant warming: a radiant warmer and an optional heating plate with conductive gel mattress. The device's bed can be tilted up to 15° in Trendelenburg and reverse Trendelenburg directions and the design of the device is intended to facilitate uniform heat distribution over the entire mattress surface across the range of bed tilt angulation. An optional removable canopy is available for intra-hospital transfer.

    Infant warming is facilitated via three (3) available thermorequlation modes:

    • Manual mode
    • Skin temperature mode
    • Kangaroo mode

    The Babyroo TN300 is available with an optional Resuscitation module. The optional Resuscitation module is pneumatically powered, can be connected to central gas supplies or gas cylinders, and provides emergency resuscitation and suction to the patient. The optional Resuscitation module includes adjustment for gas flow, peak inspiratory pressure, O2 concentration, and suction functionality and is available in three (3) variants:

    • Resuscitation module with gas mixer and AutoBreath®.
    • Resuscitation module with gas mixer.
    • Resuscitation with O2 only.

    The optional AutoBreath® function facilitates pneumatically driven, automatic respiratory rate and positive end-expiratory pressure control.

    The Babyroo TN300 can be configured to include an optional integrated electronic scale, as well as, optional heated gel mattress, optional integrated single or dual storage drawers, optional gas cylinder holders, and optional x-ray tray.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modification to the Babyroo TN300 infant warmer device. The modification involves activating existing optional software functionality to display SpO2 information generated by existing, cleared pulse oximetry accessories. The core device functions for warming, resuscitation, and weighing remain unchanged from the predicate device (K230278).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Key Takeaway: The submission focuses on demonstrating substantial equivalence for the new functionality (SpO2 display and associated alarms) by relying on existing clearances of the core device and the OEM pulse oximetry accessories, in conjunction with specific non-clinical verification testing of the integrated system. No new clinical studies were performed.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for a modification (enabling SpO2 display), the acceptance criteria primarily revolve around demonstrating that the new functionality performs as expected and does not introduce new safety or effectiveness concerns, especially concerning the integration with existing, cleared pulse oximetry technology. The performance is assessed by confirming conformity to relevant standards and effective data transfer.

    Acceptance Criterion (for SpO2 display functionality)Reported Device Performance (as per document)
    Conformity to ISO 80601-2-61:2017 (Medical electrical equipment -Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment)"Verification testing with respect to conformity to ISO 80601-2-61:2017 is included." and "Verification testing is included to demonstrate conformity of the subject device - in combination with the cleared OEM pulse oximetry accessories - to the FDA-recognized consensus standard relating to pulse oximetry."
    Effective Transfer of Pulse Oximetry Data (from OEM accessories to device display)"Verification testing that confirms the effective transfer of the data generated by the OEM pulse oximetry devices is included in this premarket notification in support of substantial equivalence." and "OEM integration testing demonstrating the integrity of the SpO2 information generated by the cleared accessories and displayed by the subject Babyroo TN300 device is included."
    SpO2 Alarms Functionality (Upper/Lower SpO2, Desaturation, Pulse Rate)The device "Facilitates the setting of SpO2 alarms for: - Upper limit for SpO2 - Lower limit for SpO2 - Lower alarm limit for desaturation - Upper alarm limit for pulse rate - Lower alarm limit for pulse rate." (Performance is implicitly accepted through the successful completion of ISO 80601-2-61:2017 conformity and data transfer testing, as these alarm functions are integral to pulse oximetry equipment.)
    Cybersecurity of External Interfaces (including SpO2 interface)"Cybersecurity assessment of external interfaces, including the SpO2 interface, is included in support of substantial equivalence." and "Software Documentation: With reference to September, 2023, Guidance for Industry and Food and Drug Administration Staff: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."
    Overall Safety and Performance of the Modified Device"The subject Babyroo TN300 was tested in accordance with applicable standards, guidance, and internal design requirements, including performance testing and functional/operation testing. Testing included the cleared OEM accessories that are relevant to the modification that is the subject of this premarket notification. The results of the non-clinical performance testing support substantial equivalence." This broadly covers the device's continued safe operation with the new feature integrated.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No human clinical data are included in support of substantial equivalence." This means there isn't a "test set" in the traditional sense of patient data.

    The testing relied on non-clinical performance testing:

    • Sample Size: Not numerical. The testing involved the Babyroo TN300 device itself, presumably one or more units, integrated with the specific OEM Masimo SET® pulse oximetry accessories. The "sample" here refers to the actual device and integrated components under test in a laboratory/engineering setting.
    • Data Provenance: The data comes from laboratory/engineering testing performed by the manufacturer (Draeger Medical Systems, Inc.). It is implicitly prospective in the sense that the testing was conducted specifically for this 510(k) submission to demonstrate the safety and effectiveness of the new functionality. No geographic origin for the test data is specified, but it would typically be conducted at the manufacturer's R&D facilities.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Since no human clinical data or expert evaluations of images/readings were used for the test set (as this is a device modification for display of existing, cleared parameters), there were:

    • Number of Experts: N/A.
    • Qualifications of Experts: N/A.

    The acceptance criteria were demonstrated through objective non-clinical performance testing against recognized standards (e.g., ISO 80601-2-61) and internal verification activities. The "ground truth" for the SpO2 display functionality would be the accurate and reliable transmission and display of the SpO2 and pulse rate data generated by the cleared OEM pulse oximetry accessories themselves, as validated in the OEM integration testing.

    4. Adjudication Method for the Test Set

    • Adjudication Method: N/A. As no human interpretation of data (e.g., images, vital signs patterns) was involved in a "test set" requiring ground truth establishment through expert consensus or adjudication, this is not applicable. The performance was verified through direct measurements and compliance with technical standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • MRMC Study: No. The document explicitly states: "No human clinical data are included in support of substantial equivalence." This type of study is typically performed to evaluate diagnostic accuracy or human performance with AI assistance, neither of which is the primary focus of this specific device modification (which is display of a physiological parameter).
    • Effect Size of Human Readers Improvement with AI vs Without AI Assistance: N/A, as no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    • Standalone Study: Yes, in a practical sense, the "OEM Integration Validation" is a form of standalone testing. This is described as validating "the effective transfer of pulse oximetry signals generated by a clinical simulator through the cleared OEM Masimo SET® pulse oximetry technology board to the subject Babyroo TN300 device's display." This tests the algorithm's ability (within the Babyroo TN300's software) to receive, process, and display the SpO2 data accurately from the source. The "algorithm" here isn't a complex diagnostic AI but rather the software logic for data handling and display.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for validating the SpO2 display functionality was the known, simulated pulse oximetry signals generated by a "clinical simulator" and then interpreted by the already cleared OEM Masimo SET® pulse oximetry technology board. The test ensured that the Babyroo TN300 accurately displayed what the cleared OEM technology was reporting. Essentially, the ground truth was the output of the "cleared OEM Masimo SET® pulse oximetry technology," and the new device's display was compared directly to that expected output.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: N/A. This device modification does not involve a machine learning or AI algorithm that requires a "training set" in the traditional sense. The software functionality being activated is to display existing, cleared data.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: N/A, as no training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230278

    Validate with FDA (Live)

    Device Name
    Babyroo TN300
    Date Cleared
    2023-05-30

    (118 days)

    Product Code
    Regulation Number
    880.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Babyroo TN300 is an open care radiant warmer that provides a controlled source of heat and regulation of skin temperature for neonates and infants. The optional integrated resuscitation module provides emergency respiratory support administered by clinicians and includes the functionality of suction. Additionally, the device provides weighing (optional) of neonates and infants. The device is designed for use with a body weight up to 10 kg (22 lb).

    The device is indicated for thermoregulation, skin temperature regulation, weighing (optional), and resuscitation (optional) of neonates and infants.

    Device Description

    The Babyroo TN300 is an open care infant radiant warmer that provides controlled source heat and skin temperature display for use with neonates and infants. The device can be configured for either labor and delivery or the newborn intensive care unit (NICU) and can be used for intra hospital transfer. Warming therapy is interrupted during intra hospital transfer and patient is not supplied with heat.

    The Babyroo TN300 device is offered with a fixed height or adjustable height trolley configuration and provides two heat sources for infant warming: a radiant warmer and an optional heating plate with conductive gel mattress. The device's bed can be tilted up to 15° in Trendelenburg and reverse Trendelenburg directions and the design of the device is intended to facilitate uniform heat distribution over the entire mattress surface across the range of bed tilt angulation. An optional removable canopy is available for intra-hospital transfer.

    Infant warming is facilitated via three (3) available thermoregulation modes:

    • Manual mode
    • Skin temperature mode
    • Kangaroo mode

    The Babyroo TN300 is available with an optional Resuscitation module. The optional Resuscitation module is pneumatic powered, can be connected to central gas supplies or gas cylinders and provides emergency resuscitation and suction to the patient. The optional Resuscitation module includes adjustment for gas flow, peak inspiratory pressure, O2 concentration, and suction functionality and is available in three (3) variants:

    • Resuscitation module with gas mixer and AutoBreath®.
    • Resuscitation module with gas mixer.
    • Resuscitation with O2 only.

    The optional AutoBreath® function facilitates pneumatically-driven, automatic respiratory rate and positive end-expiratory pressure control.

    The Babyroo TN300 can be configured to include an optional integrated electronic scale, as well as, optional heated gel mattress, optional integrated single or dual storage drawers, optional gas cylinder holders, and optional x-ray tray.

    The Babyroo TN300 device has an expected service life of 10 years.

    AI/ML Overview

    The provided text is a 510(k) summary for the Dräger Babyroo TN300 infant radiant warmer. It details the device's indications for use, its comparison to a predicate device, and the non-clinical performance testing conducted to demonstrate substantial equivalence.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal table of acceptance criteria with corresponding reported device performance values in a single, clear format as typically expected for specific performance metrics (e.g., a specific temperature range, with a specific accuracy requirement, and the measured accuracy).

    Instead, the "7. Substantial Equivalence Comparison and Discussion" section (pages 6-7) serves as the de-facto presentation of performance characteristics and how they compare to the predicate device. For the characteristics where the proposed device differs, the "Comment" column often implies the acceptance criteria (i.e., that the performance meets safety and effectiveness requirements, even if different from the predicate).

    Here's a partial reconstruction of a table based on the provided comparison, focusing on performance-related aspects and inferring acceptance criteria where possible:

    Performance CharacteristicAcceptance Criteria (Inferred/Stated)Reported Device Performance (Babyroo TN300)Study that Proves Acceptance (Implied from text)
    Body Weight IndicationUp to 10 kg (22 lb)Up to 10 kg (22 lb)Stated as identical to predicate, and established in contraindications.
    Irradiance (30% power)10 mW/cm²10 mW/cm²Implied successful non-clinical performance testing against this specified value.
    Irradiance (60% power)Not specified for predicate, accepted for proposed device.18 mW/cm²Implied successful non-clinical performance testing.
    Irradiance (100% power)Not specified for predicate, accepted for proposed device.32 mW/cm²Implied successful non-clinical performance testing.
    Pre-warm ProcedureSame power sequence (100%, 60%, 30%) as predicate, with specified durations and alarms.100% for 3 min ("Pre"), 60% for 11.5 min ("Pre"), 30% ("30") until clinician sets value. Alarms at 14 min and shut off at 15 min if above 30%.Implied successful non-clinical performance testing.
    Skin Temperature ModeSet value control in 0.1 °C (0.1 °F) steps, range 34°C to 37°C, extended 37.1°C to 38°C.Same as predicate.Implied successful non-clinical performance testing.
    Kangaroo Warming ModeAdequate warming by parent's body heat, device maintains 30% radiant warmer power, continuous patient temperature monitoring.Patient warmed by parent's body heat; device switches to manual mode with 30% radiant warmer power; continuous patient temperature monitoring.Verification of requirements relating to Kangaroo mode included in support of substantial equivalence (i.e., non-clinical performance testing).
    Skin Temperature Measurement RangeInclusive of predicate (18°C to 43°C).13°C to 43°C (55.4°F to 109.4°F)Implied successful non-clinical performance testing.
    Skin Temperature Measurement AccuracyComparable safety and effectiveness to predicate's ± 0.2°C display accuracy.Overall Accuracy: ± 0.3°C (0.54°F)Verification testing included in support of substantial equivalence (i.e., non-clinical performance testing).
    Skin Temperature Display Resolution0.1°0.1°Implied successful non-clinical performance testing.
    Primary Patient Outlet Adjustable Airway Pressure LimitDoes not exceed predicate (50 cmH2O), typical < 30 cm H2O, allows for 30-40 cm H2O for spontaneous ventilation.0 cmH2O to 40 cmH2OImplied successful non-clinical performance testing against safety and effectiveness.
    Primary Patient Outlet Fixed Airway Pressure Limit50 cmH2O ± 10%50 cmH2O ± 10%Implied successful non-clinical performance testing.
    Primary Patient Outlet Flow Control Range0 L/min to 15 L/min0 L/min to 15 L/minImplied successful non-clinical performance testing.
    Auxiliary Supply Pressure LimitDoes not exceed predicate (160 cmH2O).75 cmH2O ± 10%Implied successful non-clinical performance testing against safety and effectiveness.
    Blender Module Adjustable O2 Concentration21% to 100%21% to 100%Implied successful non-clinical performance testing.
    I:E Ratio (AutoBreath®)Non-adjustable, fixed internally at 1:2 nominal (1:1.6 to 1:2.4)Non-adjustable, fixed internally at 1:2 nominal (1:1.6 to 1:2.4)Implied successful non-clinical performance testing.
    Adjustable PEEP (AutoBreath®)< 2 cmH2O (at 5 L/min), ≤ 4 cmH2O (at 10 L/min), > 14 cmH2O (at 15 L/min)Same as predicate.Implied successful non-clinical performance testing.
    Suction Circuit Adjustable Suction Pressure0 kPa to 20 kPa (0 mmHg to 150 mmHg)0 kPa to 20 kPa (0 mmHg to 150 mmHg)Implied successful non-clinical performance testing.
    Suction Circuit Maximum Flow Rate< 20 L/min< 20 L/minImplied successful non-clinical performance testing.
    AutoThermo SW PackageSafety and effectiveness not affected, requirements verified.Allows "Tolerate cooling" and "warm-up" functions.Verification of requirements relating to AutoThermo function included in support of substantial equivalence (non-clinical performance testing).
    Bed-tiltSafety and effectiveness maintained up to 15° for continuous tilt.Continuous ± 15° from horizontal. Tactile detents at 0° and ± 10°.Testing included to verify requirements of the bed-tilt specifications of the proposed device (non-clinical performance testing).

    Study that Proves Acceptance Criteria:

    The primary study proving the device meets the acceptance criteria is non-clinical performance testing conducted according to recognized international consensus standards. The document explicitly states:

    • "The subject Babyroo TN300 was tested in accordance with applicable standards, guidance, and internal design requirements, including performance testing, functional/operation testing, biocompatibility, human factors, risk analysis and verification of risk control measures. Testing included accessories and optional components. The results of the non-clinical performance testing support substantial equivalence." (Page 12)
    • "The performance test data were provided to support substantial equivalence included:" followed by a list of IEC and ISO standards (pages 12-13).

    2. Sample Size Used for the Test Set and the Data Provenance

    The document details non-clinical performance testing. For non-clinical (laboratory/bench) testing of medical devices, the "sample size" typically refers to the number of devices or components tested. The document does not specify the sample size (number of devices/components) used for the non-clinical performance tests.

    Data Provenance: This is non-clinical performance testing, so there is no patient data provenance information like country of origin or retrospective/prospective status. The testing was conducted internally or by testing facilities on physical devices/components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the studies were non-clinical performance tests. Ground truth in this context would be defined by the technical specifications and standards (e.g., a thermometer should read within X tolerance, a ventilator should deliver Y pressure). The "experts" would be the engineers and technicians conducting and verifying the tests against these established standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable as the studies were non-clinical performance tests. Adjudication methods like 2+1 are typically used in clinical studies for endpoint assessment or expert review of clinical images/data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The Babyroo TN300 is an infant radiant warmer with optional resuscitation and weighing modules. It is a hardware device with software, but it does not involve "human readers" or AI assistance in interpreting diagnostic data, which is the context for MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The Babyroo TN300 is a physical medical device, not an image analysis algorithm or AI software for diagnosis that would have standalone performance. Its software controls the device functions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance testing, the "ground truth" or acceptance criteria are defined by:

    • International Consensus Standards: A long list of IEC and ISO standards is provided (pages 5, 12-13), which set specific performance, safety, and essential performance requirements.
    • Internal Design Requirements: The document states testing was against "internal design requirements," meaning the manufacturer's own specifications for the device.
    • Predicate Device Performance: For characteristics where the proposed device is equivalent to the predicate, the predicate's established performance serves as a benchmark for equivalence.

    8. The sample size for the training set

    This information is not applicable. This is a hardware medical device with controlling software, not a machine learning model that requires a distinct "training set" of data. The software development process follows IEC 62304.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no "training set" in the context of this device. Ground truth for the device's functional performance refers to compliance with established engineering specifications and international standards, rather than labels on a dataset for machine learning.