    Intended Use

    The Browne Metrex 1.8% Glutaraldehyde Indicator for Metricide® 28 and Metricide Plus 300 Solutions (Browne Metrex 1.8% Glutaraldehyde Indicator) is a glutaraldehyde concentration monitor for use in glutaraldehyde-containing germicide solutions with a minimum effective concentration of 1.8% glutaraldehyde.

    The Browne Metrex 1.8% Glutaraldehyde Indicator is dedicated for use with Metricide® 28 and Metricide Plus 30® Solutions.

    Device Description

    The Browne Metrex 1.8% Glutaraldehyde Indicator for Metricide 28 and The Drownload Solution and the substantially equivalent device are chemical indicator strips intended to monitor the concentration of glutaraldehyde in glutaraldehyde-containing germicide solutions. The devices indicate, via a color change, if the germicide concentration exceeds the MEC for the solution.

    The Browne Metrex 1.8% Glutaraldehyde Indicator consists of a polypropylene The Drowns 110ward results of pad on one end. The indicator pad is impregnated with an indicator solution that changes color from yellow to purple in liquid chemical germicides with a glutaraldehyde concentration above the MEC of 1.8%.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Browne Metrex 1.8% Glutaraldehyde Indicator. It outlines the device's intended use and performance testing.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the intended use and the specific glutaraldehyde concentration thresholds. The reported performance refers to the device's ability to meet these thresholds.

    Acceptance Criteria (Implied)Reported Device Performance
    Accurately indicate glutaraldehyde concentration < 1.8%Strip appears patchy purple/yellow or yellow (at 60 secs)
    Accurately indicate glutaraldehyde concentration >= 2.1%Strip exhibits a uniform purple color (at 60 secs)
    Accurately indicate glutaraldehyde concentration between 1.8-2.1%Strip may appear yellow, purple/yellow, or purple (at 60 secs)
    No false positives in solutions containing 1.8% glutaraldehyde (when used according to IFU)"No false positives were recorded in solutions containing 1.8% glutaraldehyde, when the testing was performed according to the Instructions for Use."
    Effective monitor for glutaraldehyde component of Metricide 28 and Metricide Plus 30 Solution with a MEC of 1.8%."The data demonstrates that the Browne Metrex 1.8% Glutaraldehyde Indicator is an effective monitor for the glutaraldehyde component of the Metricide 28 and Metricide Plus 30 Solution with a glutaraldehyde MEC of 1.8%."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document states that testing was performed in "solutions containing 1.8% and 2.1% glutaraldehyde." However, it does not specify the exact number of indicators tested or the number of solution samples used for the test set.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. Given the nature of a 510(k) submission, it is highly likely that this was a prospective, in-house laboratory study conducted by the manufacturer (Albert Browne Ltd., located in the United Kingdom).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (a chemical indicator strip) does not typically involve human expert interpretation for "ground truth" establishment in the same way, for example, a medical imaging AI would. The "ground truth" for the test set is the actual, measured glutaraldehyde concentration of the solutions used.

    • Number of Experts: Not applicable in the context of expert interpretation. The ground truth is established by chemical analysis of the solutions.
    • Qualifications of Experts: Not applicable for establishing ground truth for this device type. The accuracy of the glutaraldehyde concentration would rely on standard chemical laboratory practices and equipment.

    4. Adjudication Method for the Test Set

    Not applicable. The color change of the indicator is a direct chemical reaction, not a subjective interpretation requiring adjudication among human readers. The "reading" of the indicator is a comparison to a color chart or a visual assessment based on specific color descriptions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This device is a chemical indicator, not a diagnostic imaging or interpretive device that requires human readers. Therefore, an MRMC comparative effectiveness study is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence. The device itself is a standalone chemical indicator. Its performance is measured by its intrinsic chemical reaction and the resulting color change, without human-in-the-loop performance influencing the primary "reading" (though a human must visually interpret the color change against the provided guidance). The performance described ("No false positives were recorded...") is the standalone performance of the indicator.

    7. The Type of Ground Truth Used

    The ground truth used is the known, measured glutaraldehyde concentration of the germicide solutions. This would be established through a separate, calibrated chemical assay method.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. This is a chemical indicator, not an AI/ML device that requires a training set. The performance is based on the inherent chemical properties of the indicator.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of chemical indicator. The device's mechanism is based on established chemical principles.