Found 2 results

510(k) Data Aggregation

    K Number
    K103345

    Validate with FDA (Live)

    Date Cleared
    2011-02-11

    (88 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Breathe Technologies™ Ventilator, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients who are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional or home care environment. The device is intended for operation by trained personnel, patients, or caregivers under the direction of a physician.

    Device Description

    The Breathe Technologies Ventilator with Accessories (BT-V2S) is a small, wearable, ventilator that interfaces with proprietary nasal and tracheostomy breathing circuits.

    AI/ML Overview

    The provided text describes the Breathe Technologies Ventilator (BT-V2S) and its accessories, and a 510(k) submission for expanded indication to include use in the Home Care Environment. The information focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed acceptance criteria or a study design in the way you've outlined for a typical AI/software device.

    Specifically, the document discusses general standards and guidelines the device complies with, and "Human Factors Studies" to demonstrate proper operation by intended users. It does not provide quantitative performance metrics, sample sizes, ground truth establishment, or expert-based evaluations in the context of an AI/algorithm-driven device.

    Therefore, many of the requested fields cannot be directly extracted from the provided text. I will fill in the available information and indicate where data is not present.

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria or performance metrics directly from a study in the format typically seen for AI/diagnostic devices. Instead, it states that the device is compliant with various industry standards and guidelines, and that "Human Factors Studies" demonstrated proper operation.

    Acceptance Criteria (Implied from compliance/study goals)Reported Device Performance (Summary from text)
    Compliance with FDA Draft Reviewer Guide for Ventilators (July 1995)Demonstrated compliance
    Compliance with ASTM F1100-90 (1997)Demonstrated compliance
    Compliance with IEC 60601-1 (1988), Amd 1 (1991-11), Amd 2 (1995)Demonstrated compliance
    Compliance with IEC 60601-2-12 (2001-10)Demonstrated compliance
    Compliance with ASTM F1246-91 (1991, Reapproved 2005)Demonstrated compliance
    Proper operation by intended users (Human Factors) for Home Use"Demonstrated that the intended users... can properly operate the device."
    Conformance with FDA Guideline: Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management. July 18, 2000Documented conformance
    Conformance with IEC 62366: 2007Documented conformance
    Conformance with ANSI/AMEE HE75:2009Documented conformance

    2. Sample size used for the test set and the data provenance

    The document mentions "three Human Factors Studies" but does not provide sample sizes for these studies or their data provenance (e.g., country of origin, retrospective/prospective nature), beyond stating "The first study was with clinicians, and the second was with respiratory therapy patients."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a mechanical ventilator, and the studies mentioned are Human Factors studies focused on usability and proper operation, not diagnostic accuracy requiring expert ground truth in the traditional sense of medical image analysis. The "experts" would likely be the clinicians and patients participating in the Human Factors studies.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods are typically relevant for establishing ground truth in diagnostic studies, which is not the type of study described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-driven device, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI-driven device.

    7. The type of ground truth used

    For the Human Factors studies, the "ground truth" was likely defined by observational data and user feedback demonstrating the ability of users (clinicians, patients, caregivers) to safely and effectively operate the device according to its instructions and design, and to identify and address any usability issues. This is not a pathology or outcomes data ground truth in the diagnostic sense.

    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The "training" here refers to the actual training of users on how to operate the device.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102525

    Validate with FDA (Live)

    Date Cleared
    2010-12-02

    (91 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Breathe Technologies Ventilator (BT-V2S), with Accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients, with or without a tracheostomy, that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician.

    Device Description

    This application is being filed to allow a new breathing circuit for the Breathe Technologies Ventilator (BT-V2). The Ventilator is fundamentally unchanged. The only difference between the predicate and application submissions is that the application requests clearance to market the BT-NT nasal interface with the BT-V2S ventilator.

    The BT-NT is a nasal mask style patient circuit which attaches to the patients nose. It connects to the BT-V2S ventilator.

    The BT-NT is approximately 12cm wide and 2cm thick at its thickest point under the patient's nose. These dimensions do not include the 231 cm dual tubing that connects to the Ventilator. It is composed of various plastics.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device modification (BT-V2S Ventilator and accessories). It states that the device is substantially equivalent to a predicate device, meaning it does not claim to offer superior performance or new functionalities that require extensive studies to prove acceptance criteria. Instead, the focus is on demonstrating that the modification (a new nasal interface breathing circuit, BT-NT) does not introduce new safety or efficacy concerns and that the device remains compliant with existing standards.

    Therefore, the document does not contain the information requested in points 1-9 regarding the acceptance criteria, sample sizes, ground truth establishment, or expert involvement for a study proving device performance against specific acceptance criteria in the way one would expect for a novel AI/Diagnostic device.

    Here's a breakdown of what is available and why most of the requested information is absent:

    Explanation of Missing Information:

    This document describes a "Special 510(k): Device Modification" for a ventilator. The core claim is "substantial equivalence" to a predicate device (BT-V2). This means the manufacturer is asserting that the modified device is as safe and effective as the previously cleared device, not that it is a new device with novel performance claims requiring rigorous clinical trials or AI performance evaluations. The modification here is specifically a "new breathing circuit for the Breathe Technologies Ventilator (BT-V2S)."

    Therefore, the study performed is primarily bench testing and biocompatibility to prove that the modification does not negatively impact the existing performance and safety of the ventilator. This is fundamentally different from a study to establish novel performance metrics for AI or diagnostic accuracy where many of your questions would be highly relevant.

    Information from the Document (and limitations):

    1. Table of acceptance criteria and the reported device performance:

      • Acceptance Criteria (Implied): Compliance with
        • ISO 10993 (Biocompatibility)
        • FDA Draft Reviewer Guide for Ventilators (July 1995)
        • ASTM F1100 90 (1997), Standard Specification for Ventilators Intended for Use in Critical Care
        • IEC 60601 1 (1988), Amendment 1 (1991-11), Amendment 2 (1995): Medical electrical equipment - General Requirements for Safety
        • IEC 60601 2- 12 (2001-10): Medical electrical equipment - Particular requirements for the safety of lung ventilators - Critical care ventilators.
      • Reported Device Performance:
        • Biocompatibility data demonstrates that the device is in compliance with ISO 10993.
        • Bench testing, including a lung model, has demonstrated that the BT-V2S remains in compliance with the expectations of the medical community, the product labeling, and the listed Standards and Guidances.
        • The document states: "The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy."
    2. Sample size used for the test set and the data provenance: Not applicable. The "study" was bench testing (likely involving physical units of the ventilator and the new breathing circuit). There is no "test set" in the context of patient data, nor is there data provenance (e.g., country of origin, retrospective/prospective) for a bench test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in this context, refers to the compliance with established engineering and medical device standards, which are evaluated by accredited labs and engineers, not typically by clinical experts establishing diagnostic "ground truth."

    4. Adjudication method for the test set: Not applicable. There is no "adjudication" in the sense of resolving disagreements among experts on a test set. Compliance is determined by objective measurements against specified standard thresholds.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a ventilator, not an AI-powered diagnostic tool, and no MRMC study was performed or needed for this type of modification.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm-only device.

    7. The type of ground truth used: "Ground truth" was compliance with established engineering and medical device standards (ISO, ASTM, IEC) and the "expectations of the medical community" for ventilator performance, as verified through bench testing.

    8. The sample size for the training set: Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established: Not applicable, as there is no training set for this type of device.