510(k) Data Aggregation

    K Number
    K171404

    Validate with FDA (Live)

    Date Cleared
    2017-07-17

    (66 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BOSTON XO® (hexafocon A) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery). Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. Boston XO Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject: 1. cannot be adequately corrected with spectacle lenses 2. requires a rigid gas permeable contact lens surface to improve vision 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities Common causes of corneal distortion include but are not limited to corneal infections as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error. The lenses may be disinfected using a chemical disinfection (not heat) system only.

    BOSTON XO2® (hexafocon B) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery). BOSTON XO2 Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule. Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. Boston XO2 Scleral Contact Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject: 1. cannot be adequately corrected with spectacle lenses 2. requires a rigid gas permeable contact lens surface to improve vision 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO2 Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error. The lenses may be disinfected using a chemical disinfection (not heat) system only.

    Device Description

    The Boston XO and Boston XO2 Rigid Gas Permeable (RGP) Contact Lenses are fluoro silicone acrylate copolymer rigid gas permeable contact lenses available in a variety of tints and may contain an ultraviolet light absorber. The hemispherical shells are available in several lens designs including spherical, aspherical, toric, scleral and irregular cornea designs.

    AI/ML Overview

    The provided document is a 510(k) summary for the BOSTON XO®/XO2® Rigid Gas Permeable Contact Lenses. It outlines the indications for use and compares the device to predicate devices to demonstrate substantial equivalence, rather than detailing a study proving the device meets specific acceptance criteria with explicit performance metrics.

    Therefore, many of the requested details about acceptance criteria, study design, sample sizes, ground truth establishment, and expert involvement are not present in this type of regulatory document.

    However, based on the information provided, here's what can be extracted:

    Acceptance Criteria and Device Performance

    The document does not explicitly state numerical acceptance criteria for performance or a study proving those criteria are met. Instead, it relies on demonstrating substantial equivalence to already approved predicate devices. The "performance" in this context refers to the physical and chemical properties of the lens materials.

    The table below summarizes the physical comparison of the subject device (Boston XO & XO2 Scleral Lens) to predicate devices (BostonSight PD Prosthetic Device and Boston Scleral Lens). The acceptance criteria implicitly are that these properties are comparable to legally marketed predicate devices, and that the expanded indications for use for scleral lenses do not introduce new questions of safety or effectiveness.

    FeatureBoston XO (hexafocon A) - Reported PerformanceBoston XO2 (hexafocon B) - Reported PerformancePredicate Device (oprifocon A) - BostonSight PD Prosthetic DevicePredicate Device (oprifocon A) - Boston Scleral Lens
    Refractive Index1.4151.4241.4231.423
    Specific Gravity1.271.191.241.24
    Wetting Angle49°38°56°56°
    Hardness81788181
    Modulus1500 Mpa1160 Mpa1300 Mpa1300 Mpa
    Visible Light Transmittance> 92%> 95%> 95%> 95%
    Water Content<1%<1%<1%<1%
    Oxygen Permeability (Dk)1001418585

    Study Information (as inferable from the document):

    1. Sample sizes used for the test set and the data provenance:

      • Test set sample size: Not applicable/Not provided. The submission states, "No performance data is required" as "There are no changes to the Boston XO (hexafocon A) and Boston XO2 (hexafocon B) Rigid Gas Permeable contact lens material or lens designs as part of this application. These materials have been previously cleared for scleral lens designs. This application is for labeling revisions."
      • Data provenance: Not applicable, as detailed performance data from a new study is not provided. The comparison is based on existing, previously cleared material properties and indications of predicate devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. No new clinical study requiring ground truth establishment by experts is described. The comparison is against established properties and indications of previously approved devices.
    3. Adjudication method for the test set:

      • Not applicable. No new clinical study requiring adjudication is described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This document pertains to Rigid Gas Permeable Contact Lenses, not an AI-assisted diagnostic or interpretive device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This is a contact lens, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable. The "ground truth" here is the established safety and effectiveness of the predicate devices based on their approved indications and material properties.
    7. The sample size for the training set:

      • Not applicable. No new development of an algorithm or device requiring a training set is described.
    8. How the ground truth for the training set was established:

      • Not applicable. See point 7.

    In summary, this 510(k) submission primarily leverages the concept of substantial equivalence. The manufacturer argues that the Boston XO and XO2 Scleral Lenses have similar actions, technological characteristics, and present the same risks and benefits as the specified predicate devices (BostonSight PD Prosthetic Device and Boston Scleral Lens) when used for the expanded indications. Therefore, new performance data was not deemed necessary for this particular submission, which focuses on labeling revisions for existing materials.