Search Results
Found 3 results
510(k) Data Aggregation
The BoneSource® Hydroxyapatite Cement (HAC) Expanded Kit is indicated for use in the repair The Donesourgical burr holes, contiguous craniotomy cuts and other cranial defects with a surface or nourcement but notes) e defect. BoneSource is also indicated for augmentation or arestoration of bony contour in the craniofacial skeleton including the fronto-orbital, malar and mental areas.
The BoneSource® Hydroxyapatite Cement (HAC) Expanded Kit contains the previouslycleared BoneSource hydroxyapatite cement powder along with a new prefilled syringe containing sodium phosphate solution and a mixing spatula. Kits are being provided in sizes ranging from 5 g to 50 g.
The provided text is related to the 510(k) submission for BoneSource® Hydroxyapatite Cement (HAC) Expanded Kit. However, it does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria.
The document states:
- Proprietary Name: BoneSource® Hydroxyapatite Cement (HAC) Expanded Kit
- Common Name: Hydroxyapatite Cement (HAC)
- Intended Use: Repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects, and augmentation/restoration of bony contour in the craniofacial skeleton (fronto-orbital, malar, mental areas).
- Predicate Device: BoneSource Hydroxyapatite Cement (K953339, K864537 and K970400) and Norian® CRS™ Craniofacial Repair System (K973789).
- Device Description: It's an expanded kit containing previously cleared BoneSource hydroxyapatite cement powder, a new prefilled syringe with sodium phosphate solution, and a mixing spatula.
The 510(k) summary only describes the device and its intended uses, and asserts substantial equivalence to predicate devices. It does not include details on performance testing, acceptance criteria, or specific studies that demonstrate the device meets any performance criteria. Therefore, I cannot generate the requested table or answer the specific questions about studies, sample sizes, ground truth, or expert involvement.
Ask a specific question about this device
(73 days)
The Osteogenics BoneSource® Hydroxyapatite Cement (HAC) is a self-setting, calcium phosphate cement intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects with a surface area no larger than 25cm per defect.
The Osteogenics BoneSource® Hydroxyapatite Cement (HAC) is a self-setting, calcium phosphate cement that hardens in an aqueous environment at body temperature and consists solely of calcium phosphate compounds. The cement is mixed freshly at the time of implantation and can be applied directly into the defect to fill the void. BoneSource® remains malleable for approximately 20 minutes during which time it can be contoured to custom-fit the defect. The setting reaction is complete after four hours. BoneSource® is fabricated by combining two phosphates of calcium (i.e., dicalcium phosphate and tetracalcium phosphate). At the time of use, BoneSource® is combined with sterile water which results in an isothermic setting reaction yielding pure HAC after four hours. Under in-vitro conditions (37°C), the product hardens in approximately 20 minutes. After four hours, the reaction is complete, yielding pure hydroxyapatite with no significant dimensional changes and no by-products. The compressive strength of BoneSource® is ≥ 50 MPa. The BoneSource® paste during the setting reaction has been determined to be in the range of 6.5 to 8.5. As a consequence of its apatitic nature, the product is highly compatible with both soft and hard tissue. Approximately 45% of the implant volume consists of micropores (the remainder is solid) with an average pore size of <1 micron in diameter. This small pore diameter serves to inhibit the passage of infection-causing microorganisms. BoneSource® is supplied sterile and is intended for single use only.
Here's an analysis of the provided text, focusing on acceptance criteria and supporting studies for the BoneSource® Hydroxyapatite Cement:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a bulleted or numbered list with corresponding "reported performance." Instead, it describes various tests and their outcomes, implying that successful outcomes from these tests constituted the basis for acceptance. I have inferred the acceptance criteria based on the described testing and conclusions.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| In-Vitro Characterization | |
| Complete setting reaction to pure hydroxyapatite within 4 hours (X-ray Diffraction) | X-ray diffraction patterns taken before hardening and after hardening demonstrate that the setting reaction is complete after four hours. The device becomes totally apatitic with no by-products present. |
| pH of paste during setting range 6.5 to 8.5 | The pH of the BoneSource® paste during setting has been determined to be in the range of 6.5 to 8.5, verified by laboratory testing and consistent with published literature. |
| No significant dimensional changes upon hardening | Under in-vitro conditions (37°C), BoneSource® hardens in approximately 20 minutes with no significant dimensional changes. Measurements of molds vs. specimens, and 50g samples in acrylic cylinders, showed negligible change. |
| Compressive strength ≥ 50 MPa | The compressive strength of BoneSource® is ≥ 50 MPa. |
| Biocompatibility (Inferred: Must be non-toxic, non-irritant, non-mutagenic, non-pyrogenic, and well-tolerated at implant site) | |
| In Vitro Cytotoxicity (Agarose Overlay) - Non-toxic to L-929 cells | Non-toxic |
| Guinea Pig Delayed Contact Sensitization Test - Non-irritant | Non-irritant: none of the animals in the study showed abnormal clinical signs. |
| Salmonella Mammalian Mutagenicity Assessment/Ames Assay - Non-mutagenic | Non-mutagenic |
| Limulus Amebocyte Lysate (LAL) Endotoxin Test - Non-pyrogenic | Non-pyrogenic: no detectable level of endotoxin. |
| USP Intramuscular Implant Test - Non-toxic effect at implant site | Non-toxic effect: difference between average encapsulation score for BoneSource® and negative control implants was zero for each animal and overall. |
| 30-Day HAC Cranial Implant Test (dogs) - Good tissue integration, no infection/toxicity, osteogenesis | Defects filled with fibrin, fibrous tissue, and new bone. New bone extended from edges. Osteogenesis evident at all sites. No evidence of infection or toxicity of surrounding bone. |
| Animal Studies (Inferred: Demonstrate safety, effectiveness, bone integration, and resorption/replacement) | |
| Hydroxyapatite Cement: Basic Chemistry and Histologic Properties (cats) - No toxicity, extrusion, migration, infection, fibrous encapsulation; bone growth into and replacement of implant | No toxic reactions, extruded implants, disk migrations, wound infections, nor fibrous encapsulation found. Bone growth occurred into implants; disks resorbed and replaced by bone. |
| Obliteration and Reconstruction of the Cat Frontal Sinus - Progressive replacement by bone, no complications, good integration, no volume loss/contour change, infection resistance | Progressive replacement with bone over time (CT/radiographs). No postoperative mortalities, complications, wound infections, or extrusions. Excellent integration, no loss of volume or significant change in implant contour. Ability to resist infection demonstrated. |
| Experimental Hydroxyapatite Cement Cranioplasty (cats) - No wound infections, structural failures; well-tolerated; progressive replacement by new bone/soft tissue without shape/volume change | No wound infections or structural failures observed; implants well tolerated histologically. Progressive replacement of cement by new bone and soft tissue without change in shape or volume. New bone comprised 77.3% (HAC) and 64.7% (HAC/autogenous bone) of replacement tissue. |
| Clinical Effectiveness (Inferred: Acceptable implant survival rate, minimal volume loss, device-related explants) | |
| Implant survival rate (based on volume loss > 10% or explantation as failure) | Cumulative survival rate of 81.32% (with 11 implants not yet evaluated at 24 months and 31 not yet due for 24-month follow-up). This reflects incidences of volume loss in excess of 10% for four implants and all explants as failures (though none were considered device-related by investigators). |
| Clinical Safety (Inferred: Laboratory values within normal/clinically non-significant ranges, acceptable complication/adverse event rates) | |
| Laboratory Values (Ca, Cl, Na, K, Bicarbonate, Phosphate) - Maintain within clinically normal ranges or demonstrate non-significant changes | Calcium: statistically significant changes at 12 and 24 months, but mean within-patient percent change averaged only 3.21% and 4.21% respectively, not clinically significant. Min/max levels well within handbook range. Sodium: statistically significant change at 24 months, but mean within-patient change averaged 1.01%, not clinically significant. Min/max levels well within handbook range. Potassium: statistically significant changes at 12 and 24 months, but none clinically significant. Chloride, bicarbonate, phosphate: no statistical significance. Six patients had slightly high chloride early post-op, but dropped to non-significant levels. |
| Complications (defect-specific & patient-specific) - Favorable rates, non-device related | Majority of implants presented no defect-specific complications. Majority of patients presented with no patient-related complications. |
| Adverse Events/Death - Not device-related | Five adverse events reported: two deaths from cancer (not device-related); one stroke after tumor removal with pneumonia (not device-related); one hematoma (not specified if device-related, but overall conclusion states adverse events not device-related when administered as intended); one CSF leak with BoneSource® injected through nostril (protocol violation), resulting in explantation. Overall conclusion: None device-related when administered as intended. |
| Substantial Equivalence | Demonstrated substantial equivalence to CranioPlastic™ (PMMA) in design and function. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Clinical Test Set:
- Sample Size: 103 patients presenting 175 cranial defects.
- Data Provenance: Prospective, non-randomized clinical investigational study conducted under an approved Investigational Device Exemption. Country of origin is not explicitly stated but implied to be the US, given the FDA 510(k) submission.
-
Animal Studies Test Set:
- Hydroxyapatite Cement: Basic Chemistry and Histologic Properties: 9 cats.
- Hydroxyapatite Cement: Obliteration and Reconstruction of the Cat Frontal Sinus: 9 cats.
- Experimental Hydroxyapatite Cement Cranioplasty: 6 cats for primary study, plus 3 additional cats for controls.
- 30-Day HAC Cranial Implant Test (Biocompatibility): 2 dogs (10 defects total, 8 with BoneSource®, 2 controls).
- Data Provenance: Prospective animal studies. No country of origin specified.
-
In-Vitro Testing: Sample sizes for specific in-vitro tests (e.g., number of specimens for dimensional verification, compressive strength) are not explicitly detailed, but implied to be sufficient for scientific assessment.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
-
Clinical Study:
- Effectiveness (Radiographic Evaluation): The assessment was done by "radiographic evaluation (x-ray or CAT Scan)." The number and qualifications of the readers or experts who performed these evaluations are not specified within this summary.
- Safety (Complications/Adverse Events): "Considered by the Investigator and Medical Monitor." This implies at least two experts (Investigator, Medical Monitor). Their specific qualifications (e.g., number of years experience) are not provided, but they would be medical professionals overseeing the clinical trial.
- Histopathology (30-Day HAC Crania Implant Test in dogs): A "blinded histopathological evaluation" was performed. The number and qualifications of the histopathologist(s) are not specified.
-
Animal Studies: The histological and radiographic evaluations in the cat studies were conducted, but the specific number and qualifications of the experts (e.g., veterinary pathologists, radiologists) who established the ground truth are not mentioned.
4. Adjudication Method for the Test Set
- Clinical Study: Not explicitly stated. For "Effectiveness" evaluation (volume loss), the method for combining or adjudicating multiple radiographic interpretations (if more than one reader was used) is not described. For "Safety" related to device-relatedness of outcomes, it states "considered by the Investigator and Medical Monitor," implying a consensus or adjudicated decision between at least these two individuals.
- 30-Day HAC Cranial Implant Test (dogs): A "blinded histopathological evaluation" was performed, but the adjudication method (e.g., if multiple pathologists disagreed) is not detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?
- No, an MRMC comparative effectiveness study involving AI assistance for human readers was not done. This document describes the safety and effectiveness of a medical device (a bone cement), not an AI diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not applicable here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not Applicable. This document describes a physical medical device (bone cement), not a diagnostic algorithm.
7. The Type of Ground Truth Used
- Clinical Study:
- Effectiveness: Based on radiographic evaluation (x-ray or CAT Scan) for implant stability (volume loss) and evaluation of explants (physical assessment).
- Safety: Based on laboratory data (blood chemistry) and direct clinical observations and records of postoperative complications and adverse events, interpreted by medical professionals.
- Animal Studies:
- Histology: Direct examination of tissue samples for bone growth, resorption, tissue reactions, etc. (e.g., "Examination of decalcified and undecalcified sections," "blinded histopathological evaluation").
- Imaging: Computed tomography and radiographs.
- Gross observation: Direct visual assessment of implant site, wound, etc.
- In-Vitro Tests: Based on laboratory measurements and analytical techniques (e.g., X-ray Diffraction, pH meter, micrometers for dimensional changes, mechanical testing for compressive strength).
8. The Sample Size for the Training Set
- Not Applicable. This document describes the testing of a physical medical device, not a machine learning model. There is no concept of a "training set" in this context. The animal and in-vitro studies could be analogized to preclinical foundational research, but not a dataset for training an algorithm.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no training set for an algorithm, this question is not relevant to the provided text.
Ask a specific question about this device
(346 days)
The Osteogenics BoneSource™ Hydroxyapatite Cement (HAC) is a self-setting, calcium phosphate cement intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects with a surface area no larger than 25cm2 per defect.
The Osteogenics BoneSource™ Hydroxyapatite Cement (HAC) is a self-setting, calcium phosphate cement that hardens in an aqueous environment at body temperature and consists solely of calcium phosphate compounds. The cement is mixed freshly at the time of implantation and can be applied directly into the defect to fill the void. BoneSource™ remains malleable for approximately 20 minutes during which time it can be contoured to custom-fit the defect. The setting reaction is complete after four hours. BoneSource™ is fabricated by combining two phosphates of calcium (i.e., dicalcium phosphate and tetracalcium phosphate). At the time of use, BoneSource™ is combined with sterile water which results in an isothermic setting reaction yielding pure HAC after four hours. Under in-vitro conditions (37℃), the product hardens in approximately 20 minutes. After four hours, the reaction is complete, yielding pure hydroxyapatite with no significant dimensional changes and no by-products. The compressive strength of BoneSource™ is ≥ 50 MPa. The pH of the BoneSource™ paste during the setting reaction has been determined to be in the range of 6.5 to 8.5. As a consequence of its apatitic nature, the product is highly compatible with both soft and hard tissue. Approximately 45% of the implant volume consists of micropores (the remainder is solid) with an average pore size of < 6 microns in diameter. This small pore diameter serves to inhibit the passage of infection-causing microorganisms. BoneSource™ is supplied sterile and is intended for single use only.
Here's a breakdown of the acceptance criteria and the studies performed for the BoneSource™ Hydroxyapatite Cement (HAC), based on the provided document:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance | Study Proving Performance |
|---|---|---|
| Mechanical Properties | ||
| Compressive Strength | ≥ 50 MPa | In-vitro Testing |
| Material Properties | ||
| Setting Time (at 37°C) | ~20 minutes to harden; complete reaction in 4 hours | In-vitro Testing (Dimensional Verification, X-ray Diffraction) |
| pH during setting | 6.5 to 8.5 | pH Value Determination (In-vitro) |
| Dimensional Changes | No significant dimensional changes | Dimensional Verification (In-vitro) |
| By-products after setting | No by-products; becomes totally apatitic | X-ray Diffraction (In-vitro) |
| Micropore Diameter | < 6 microns (inhibits infection-causing microorganisms) | Device Description (stated property) |
| Biocompatibility | ||
| In Vitro Cytotoxicity | Non-toxic to L-929 cells | In Vitro Cytotoxicity Using Agarose Overlay |
| Guinea Pig Delayed Contact Sensitization | Non-irritant | Guinea Pig Delayed Contact Sensitization Test |
| Salmonella Mammalian Mutagenicity | Non-mutagenic | Salmonella Mammalian Mutagenicity Assessment/Ames Assay |
| Endotoxin Level | Non-pyrogenic (no detectable level of endotoxin) | Limulus Amebocyte Lysate (LAL) Endotoxin Test |
| USP Intramuscular Implant Test | Non-toxic effect (zero difference in encapsulation score) | USP Intramuscular Implant Test |
| Host tissue acceptance | Well accepted by host tissue with no evidence of local or systemic adverse effects | Animal Testing (various studies), Biocompatibility Testing, Clinical Testing |
| Infection Resistance | Able to resist infection (due to pore size and successful use in non-sterile setting) | Animal Testing (Cat Frontal Sinus study), Device Description, Clinical Testing (Surgical Site Infection rate favorable) |
| Clinical Performance (Safety & Effectiveness) | ||
| Cumulative Survival Rate | 97% | Clinical Testing (Effectiveness - Life-Table Method) |
| Explantation due to volume loss > 10% (failures) | 4 out of 175 implants (2.2%) considered failures, attributed to inadequate drainage | Clinical Testing (Effectiveness) |
| Explantation > 25% volume loss | 20 implants, but none device-related (attributed to patient conditions like tumor recurrence) | Clinical Testing (Effectiveness) |
| Adverse effect on laboratory values (Calcium, Sodium, Potassium, Chloride, Bicarbonate, Phosphate) | No clinically significant adverse effects; values within normal ranges after initial postoperative period | Clinical Testing (Safety - Laboratory Values) |
| Defect-specific complications (e.g., edema, tenderness, seroma, hematoma) | Majority (77.59%) had no complications; most complications not device-related and none serious/life-threatening | Clinical Testing (Safety - Complications) |
| Patient-specific complications (e.g., headache, sinusitis, infection) | Majority (66.02%) had no complications; most complications not device-related or anticipated for cranial surgery; surgical site infection rate favorable (2.9% for explantation vs 5% for acrylics) | Clinical Testing (Safety - Complications) |
| Device-related adverse events | None when administered as intended (deaths from cancer, stroke, hematoma, CSF leak/extrusion not considered device-related or due to protocol violation) | Clinical Testing (Safety - Adverse Events/Death) |
Detailed Study Information:
2. Sample Size for Test Set and Data Provenance:
- In-vitro Testing: Not specified, but laboratory-based testing.
- Animal Testing:
- "Basic Chemistry and Histologic Properties": 9 cats.
- "Obliteration and Reconstruction of the Cat Frontal Sinus": 9 cats.
- "Experimental Hydroxyapatite Cement Cranioplasty": 6 cats for main study, 3 additional cats for controls.
- Biocompatibility Testing:
- Guinea Pig Delayed Contact Sensitization Test: Not specified.
- USP Intramuscular Implant Test: Not specified.
- 30-Day HAC Cranial Implant Test: 2 dogs (10 defects).
- Clinical Testing:
- Patients: 103 patients.
- Cranial Defects: 175 cranial defects.
- Data Provenance: Prospective, multi-site clinical investigation conducted under an approved Investigational Device Exemption (IDE). Countries of origin are not explicitly stated, but it's a US FDA submission, implying US-based studies.
3. Number of Experts and Qualifications for Ground Truth (Test Set):
- Clinical Testing (Effectiveness - Radiographic Evaluation): The effectiveness was determined by "volume loss at each postoperative interval." While radiographic evaluation (x-ray or CAT Scan) was used, the number and qualifications of experts performing these evaluations are not specified.
- Clinical Testing (Safety - Explantation Review): "Medical Monitor" and "Investigator" reviewed explantations and their relation to the device. Specific numbers and qualifications are not provided.
- 30-Day HAC Cranial Implant Test (Animal Study): "A blinded histopathological evaluation of the defects was performed." The number and qualifications of the histopathologists are not specified.
4. Adjudication Method (Test Set):
- The document does not explicitly describe an adjudication method for reconciling differing opinions (e.g., 2+1, 3+1). For clinical evaluations and histopathology, it primarily mentions evaluations being performed or reviewed by individuals/roles (Medical Monitor, Investigator, histopathologist).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was explicitly performed comparing human readers with and without AI assistance, as this device does not involve AI. The comparison is between BoneSource™ (HAC) and a predicate device (CranioPlastic™ / PMMA).
6. Standalone Performance Study (Algorithm only without human-in-the-loop):
- This question is not applicable as the device is a medical implant (cement), not an AI algorithm.
7. Type of Ground Truth Used (Test Set):
- In-vitro Testing: Scientific measurements (e.g., X-ray diffraction patterns, pH measurements, dimensional measurements, compressive strength tests).
- Animal Testing:
- Healing/Integration: Histological examination, Computed Tomography, Radiographs, gross observation at sacrifice.
- Complications: Gross observation at sacrifice, histological examination.
- Biocompatibility Testing: Standardized assays (e.g., cytotoxicity, sensitization, mutagenicity, endotoxin tests), histological evaluation.
- Clinical Testing:
- Effectiveness: Radiographic assessment of volume loss (x-ray/CAT scan), evaluation of explanted devices.
- Safety: Laboratory blood chemistry values, observation and reporting of postoperative complications and adverse events by clinical staff and investigators.
8. Sample Size for Training Set:
- This question is not applicable as the device is a physical implant, not an AI algorithm that requires a training set. The term "training set" is not relevant in this context.
9. How Ground Truth for Training Set was Established:
- This question is not applicable for the same reason as point 8.
Ask a specific question about this device
Page 1 of 1