510(k) Data Aggregation

    K Number
    K243349

    Validate with FDA (Live)

    Date Cleared
    2026-01-23

    (452 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    0.079 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuardian™ Remote Monitoring System detects and monitors cardiac arrhythmias in ambulatory patients, when prescribed by a physician or other qualified healthcare professional.

    The BodyGuardian Remote Monitoring System is intended for use with adult and pediatric patients who are at least 29 days old in clinical and non-clinical settings to collect and transmit electrocardiogram (ECG) and other health parameters to healthcare professionals for monitoring and evaluation. Health parameters, such as heart rate and ECG data, are collected from external devices such as ECG sensors.

    The BodyGuardian Remote Monitoring System does not provide any diagnosis.

    Device Description

    The BodyGuardian Remote Monitoring System (BGRMS) is a system for recording and analyzing ECG data for cardiac arrhythmias to assist healthcare professionals, including ECG technicians at 24/7 attended analysis centers in evaluating a patient's cardiac health. Reports are generated for clinician review, that provide analysis and summary of the ECG data collected during a patient's monitoring study. Both the predicate and proposed devices, feature a modular design inclusive of outpatient cardiac telemetry (commonly called mobile cardiac telemetry (MCT)), cardiac event monitor and connected/non-connected Holter modalities. Components in the system external to the software include ECG monitors, electrodes, mobile phones and apps.

    The BGRMS System includes the following main components:

    • ECG monitor – a patient worn device for ECG waveform data collection and transmission, utilized with compatible electrodes
    • Mobile App – applications that execute on an off-the-shelf (OTS) smartphone to communicate with the ECG monitor and the PatientCare Server for collection and transmission of data
    • PatientCare – server software responsible for receiving, storing, analyzing, and displaying and reporting data gathered from the ECG monitors; includes the ECG analysis algorithm BeatLogic™
    • AI-Based Device Software Functionality (AI-DSF) – Automated classification of continuous
      ECG based on the proprietary BeatLogic™ AI algorithm. BeatLogic consists of an ensemble of deep neural networks (DNNs), trained on real-world patient data and post-processing logic that combines the DNN output to produce individual beat, rhythm, and waveform classifications. This output is intended to be reviewed and confirmed by healthcare professionals to assist in diagnosis.
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the BodyGuardian Remote Monitoring System (BGRMS v3.0) contains information on the device's acceptance criteria and study to prove it.

    Acceptance Criteria and Reported Device Performance

    The clinical validation results met all predefined acceptance criteria, though the specific criteria are not explicitly detailed in the provided document beyond "substantially equivalent performance for BeatLogic." The performance was assessed by evaluating the Sensitivity and Positive Predictive Value (PPV) for key rhythms. While specific numerical values for the acceptance criteria are not given, the reported device performance is stated as meeting these unspecified criteria.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantially equivalent performance for BeatLogic algorithmMet predefined acceptance criteria
    Acceptable Sensitivity for key rhythmsAchieved (specific values not provided in document)
    Acceptable Positive Predictive Value (PPV) for key rhythmsAchieved (specific values not provided in document)
    Consistent arrhythmia detection performance across subgroupsDemonstrated consistent performance across compatible ECG device configurations and accessory types, gender, age, US geographic region, and indication for monitoring.

    Details of the Study Proving Device Meets Acceptance Criteria

    1. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated, but described as "real-world, randomly selected ECG records" with a demographic breakdown of 48.6% Female, 39.2% Male, 12.2% unknown gender, 50.1% < 65 years of age, 49.8% ≥ 65 years of age, and 0.1% unknown age.
    • Data Provenance: "Real-world patient data" with representation across "US geographic region," indicating data from the United States. The data is retrospective as it was used to train and validate the algorithm, selected to reflect various algorithm outputs, compatible ECG device configurations, accessory types, and demographic factors.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The document states that the BeatLogic™ AI algorithm's "output is intended to be reviewed and confirmed by healthcare professionals to assist in diagnosis," but it does not specify how the ground truth for the test set was established or the number/qualifications of experts involved in this process.

    3. Adjudication Method for the Test Set

    This information is not provided in the document.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • Was it done?: No, an MRMC comparative effectiveness study is not explicitly mentioned. The study focuses on the standalone performance of the AI algorithm (BeatLogic™).
    • Effect size of human readers improvement: Not applicable, as an MRMC study was not described.

    5. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Was it done?: Yes. The performance measurement of the BeatLogic™ algorithm involved evaluating Sensitivity and PPV, which are metrics typically used for standalone algorithm performance against a ground truth. The document explicitly states, "Performance of the algorithm was assessed by evaluating the Sensitivity and Positive Predictive value (PPV) for key rhythms across different patient subgroups." Furthermore, it mentions that the algorithm's "output is intended to be reviewed and confirmed by healthcare professionals to assist in diagnosis," implying that the performance reported is that of the algorithm prior to human review.

    6. The Type of Ground Truth Used

    The ground truth annotations were established based on "ground truth annotations on real-world ECG data." The method of establishing these annotations (e.g., expert consensus, pathology, outcomes data) is not explicitly stated. However, the context of cardiac arrhythmia detection strongly suggests ground truth would be established by qualified cardiologists or electrophysiologists.

    7. The Sample Size for the Training Set

    • Sample Size: Not explicitly stated, but described as "real-world, randomly selected ECG records" that ensured "representation across algorithm outputs, compatible ECG device configurations and accessory types and demographic factors encompassing patient age, gender, geographic location, and indication for monitoring."

    8. How the Ground Truth for the Training Set was Established

    The document states that the BeatLogic™ AI algorithm consists of "deep neural networks (DNNs), trained on real-world patient data." However, the specific method for establishing the ground truth for this training data is not explicitly provided. It is implied that this involved annotations on "real-world patient data," but the process for generating these annotations (e.g., expert review, automated processes) is not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192732

    Validate with FDA (Live)

    Date Cleared
    2020-03-26

    (181 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuardian System detects and monitors cardiac arrhythmias in ambulatory patients, when prescribed by a physician or other qualified healthcare professional. Not for use with patients requiring attended, in-hospital monitoring for life threatening arrhythmias.

    The Preventice BodyGuardian Remote Monitoring System is intended for use with adult and pediatric patients in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug in devices such as ECG sensors, weight scales, blood pressure meters and pulse oximeters, and parameters such as ECG, heart rate, body weight, temperature, respiration rate, blood pressure, and SpO2.

    The Preventice BodyGuardian Remote Monitoring System does not provide any diagnosis.

    Device Description

    Arrhythmia Detector and Alarm; Patient Physiological Monitor (with arrhythmia detection)

    The predicate device's ECG Unit is a small, ambulatory cardiac monitor that records and transmits ECG data. The subject device's ECG Unit is also a small, ambulatory cardiac monitor that records and transmits ECG data.

    Both products, the predicate and subject devices, provide a data hub function that connects to commercially available devices such as weight scales, blood pressure monitors, pulse oximeters, ECG Units and other plug-in devices.

    For both devices, the predicate and subject devices, data is transmitted to an external device or server, depending model. If not a server, this device sends the data to a remote computer server that allows healthcare professionals to access and review the data.

    AI/ML Overview

    This document is a 510(k) summary for the BodyGuardian Remote Monitoring System (K192732). It focuses on demonstrating substantial equivalence to a predicate device (BodyGuardian Remote Monitoring System, K151188) rather than providing detailed acceptance criteria and performance data from a new clinical study.

    Here's an analysis based on the provided text, addressing your points where possible, and noting where information is explicitly stated as not present (e.g., clinical testing was not performed):

    Key Takeaway: The submission relies on non-clinical testing and comparison to a predicate device, asserting that the technology is not new and therefore a clinical study was not necessary. This means many of your requested details about acceptance criteria directly tied to a clinical performance study are not available in this document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of quantitative acceptance criteria for clinical performance (e.g., sensitivity, specificity for arrhythmia detection) because no clinical study was conducted for this submission. The "acceptance criteria" discussed are implicitly related to demonstrating substantial equivalence through non-clinical means.

    The non-clinical "performance testing" mentioned includes:

    • EMC and electrical safety testing
    • Electrical and mechanical safety testing
    • System safety testing
    • Software verification and validation
    • Performance testing (general statement, no specific metrics provided)
    • Predicate device comparison tests
    • Usability Testing
    • Biocompatibility for patient contact materials

    The document states, "We conclude that the results of testing show the Preventice BG RMS to be substantially equivalent to the predicate device." This implies the performance observed during these non-clinical tests met the unstated internal acceptance criteria for substantial equivalence.


    2. Sample size used for the test set and data provenance:

    • Test Set Sample Size: Not applicable for clinical performance, as "a clinical study was not considered necessary."
    • Data Provenance: Not applicable for clinical performance. The non-clinical tests would have been performed in a lab setting by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • Not applicable as no clinical study with a test set requiring expert ground truth establishment for arrhythmia detection was performed.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable for the same reason as above.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC or comparative effectiveness study involving human readers and AI assistance was conducted or described. The device's primary function is described as detecting and monitoring, rather than assisting human readers in interpreting images or patterns.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The document implies the device has an algorithmic component for "detects and monitors cardiac arrhythmias." However, no specific standalone performance metrics (e.g., sensitivity, specificity, accuracy) for this algorithm are provided, nor is a dedicated "standalone study" described in the context of clinical performance. The focus is on the device as a whole system.

    7. The type of ground truth used:

    • For the non-clinical "performance testing," the ground truth would typically be established based on engineering specifications, known signal inputs, and reference standards for electrical, mechanical, and software verification. For biocompatibility, it would be based on ISO standards. For usability, it would involve direct observation and user feedback against defined usability goals. No clinical ground truth (e.g., expert consensus, pathology, outcome data) for arrhythmia diagnosis is mentioned.

    8. The sample size for the training set:

    • Not applicable, as no mention of a training set for an AI/ML algorithm (in the context of clinical performance) is made. The device is referred to using existing technology ("This technology is not new").

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.

    Summary of what the document does state regarding validation:

    The validation for this 510(k) submission primarily relies on:

    • Bench Testing: EMC, electrical/mechanical safety, system safety, software V&V, general performance testing, and predicate device comparison tests.
    • Biocompatibility: Testing to ensure patient contact materials are safe.
    • Quality Assurance Measures: FMEA, Design FMEAs, Performance Requirements Testing (including Final System Verification and Validation Testing), ISO 60601 Testing.
    • Usability Testing: Performed, including a pediatric usability study to show "no new hazards or risk associated with pediatric use."
    • Comparison to Predicate Device (BodyGuardian Remote Monitoring System, K151188): The core argument for substantial equivalence is that the new device has "the same technological characteristics" and "the same intended uses" as the predicate device. "There are no fundamental differences between their technological characteristics."
    • Waiver of Clinical Testing: Explicitly states, "This technology is not new, therefore a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing."
    Ask a Question

    Ask a specific question about this device

    K Number
    K151188

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-09-17

    (136 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuardian System detects and monitors cardiac arrhythmias in ambulatory patients, when prescribed by a physician or other qualified healthcare professional. Not for use with patients requiring attended, In-hospital monitoring for life threatening arrhythmias.

    The Preventice BodyGuardian Remote Monitoring System is intended for use with adult patients in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters. The Preventice BodyGuardian Remote Monitoring System does not provide any diagnosis.

    Device Description

    The BodyGuardian System detects and monitors cardiac arrhythmias in ambulatory patients, when prescribed by a physician or other qualified healthcare professional. Not for use with patients requiring attended, In-hospital monitoring for life threatening arrhythmias.

    The Preventice BodyGuardian Remote Monitoring System is intended for use with adult patients in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters. The Preventice BodyGuardian Remote Monitoring System does not provide any diagnosis.

    AI/ML Overview

    Here’s an analysis of the provided information regarding the BodyGuardian Remote Monitoring System, focusing on acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The provided document does not explicitly state quantitative acceptance criteria or device performance in the typical format of sensitivity, specificity, accuracy, etc., for arrhythmia detection. Instead, it relies on a comparison to predicate devices and results of non-clinical bench testing.

    Therefore, the "acceptance criteria" can be inferred as demonstration of substantial equivalence to predicate devices through various non-clinical tests, ensuring safety and effectiveness.

    Acceptance Criteria (Inferred)Reported Device Performance
    Functional Equivalence to Predicate Devices: The device should perform an equivalent function as the predicate devices in detecting and monitoring cardiac arrhythmias and transmitting health parameters.The BodyGuardian System is stated to have the "same technological characteristics as the predicate devices in that all devices are intended to detect, transmit, and report digital data using similar computerized technology." It also has the "same intended uses as the predicate devices in that all devices are intended for ambulatory monitoring."
    Electrical and Mechanical Safety Testing: The device must meet established safety standards.EMC and electrical safety testing, electrical and mechanical safety testing, and system safety testing were conducted.
    Software Verification and Validation: The software components must be verified and validated.Software verification and validation were performed.
    Performance Testing: The device's overall performance must be acceptable.Performance testing was conducted.
    Predicate Device Comparison Tests: Direct comparison testing against predicate devices.Predicate device comparison tests were conducted.
    Usability Testing: The device must be user-friendly and safe for its intended users.Usability Testing was conducted.
    Wireless Transmission Data Quality and Coexistence: The wireless communication must be reliable and not interfere with other devices.Wireless transmission data quality and wireless coexistence testing were conducted.
    Biocompatibility for Patient Contact Materials: Materials in contact with the patient must be biocompatible.Biocompatibility testing for patient contact materials (Makrolon Versaflex) was performed.
    Quality Assurance Measures: Development process must follow quality standards (e.g., FMEA).Failure Mode Effects Analysis/Hazard Analysis (FMEA) and Design FMEAs for mechanical and RF designs were applied. Performance Requirements Testing including Final System Verification and Validation Testing, ISO 60601 Testing, 3rd Edition were completed.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "Clinical Testing: This technology is not new, therefore a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing."

    Therefore, there was no clinical test set (and thus no sample size, data provenance) used to demonstrate the device's performance for arrhythmia detection in this submission. The evaluation was based solely on non-clinical (bench) testing and comparison to predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set was used to establish ground truth for this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No clinical studies, including MRMC studies, were performed or required for this submission. The device is a "Remote Monitoring System" that collects and transmits data; it "does not provide any diagnosis." There is no mention of AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe a standalone performance study for an algorithm related to arrhythmia detection. The device is described as a system that "detects and monitors cardiac arrhythmias" and collects data from "ECG sensors," but the validation focuses on the system's overall function and equivalence to predicates, not the independent performance of an arrhythmia detection algorithm against ground truth. Since it states the device "does not provide any diagnosis," it suggests the detection is a flag for healthcare professionals to review, rather than a diagnostic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as no clinical data or ground truth was established for performance evaluation in this submission. The ground truth for the non-clinical testing would have been defined by the test protocols (e.g., simulated signals for electrical safety, known material properties for biocompatibility).

    8. The sample size for the training set

    Not applicable, as no clinical data or machine learning models requiring a training set are described in this submission.

    9. How the ground truth for the training set was established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121197

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-10

    (113 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuardian System detects and monitors non-lethal cardiac arrhythmias in ambulatory patients, when prescribed by a physician or other qualified healthcare professional. The BodyGuardian System continuously records, stores and periodically transmits the following physiological data to a remote computer server for up to 30 days at a time:
    • ECG
    • Heart rate (including HR variability and HR reliability)
    • Respiration rate
    • Activity

    Device Description

    The BodyGuardian System is an ambulatory cardiac monitoring system prescribed by healthcare providers. It monitors and records a patient's electrocardiographic (ECG) data, heart rate, respiration rate and activity level. The complete system consists of components that collect data, send the data to a remote Preventice computer server, store the data in secure databases, and present the data for review by healthcare professionals.

    AI/ML Overview

    The provided documentation presents a 510(k) summary for the BodyGuardian System. While it lists non-clinical testing performed, it explicitly states "Clinical Testing: Not applicable." Therefore, the document does not contain information about acceptance criteria or a study proving the device meets those criteria based on clinical performance.

    However, based on the provided text, here's what can be extracted regarding the device and its testing:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria or detailed reported device performance in terms of clinical accuracy (e.g., sensitivity, specificity for arrhythmia detection) as no clinical testing was deemed applicable.

    However, the "Non-Clinical Testing" section implies performance was evaluated against internal benchmarks or standards during the following:

    Acceptance Criteria (Implied)Reported Device Performance (Implied from non-clinical testing)
    Compliance with EMC and electrical safety standardsPassed EMC and electrical safety testing
    Accurate ECG performanceECG performance testing conducted
    Accurate activity level measurementActivity level measurement validated
    Accurate respiration rate measurementRespiration rate measurement validated
    Software functions correctly and reliablySoftware verification and validation completed
    Biocompatibility of materialsBiocompatibility testing conducted

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as no clinical test set was used for performance evaluation according to the document. The non-clinical testing would have used internal test samples, but details on their size or provenance are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical test set with expert-established ground truth was used for performance evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document states "Clinical Testing: Not applicable," which means no MRMC study, human-in-the-loop performance, or AI assistance was evaluated in a clinical context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document implies standalone testing of components like "ECG performance," "Activity level measurement," and "Respiration rate measurement" during non-clinical testing. However, this is not a comprehensive "standalone" evaluation of the arrhythmia detection algorithm in a clinical context, as clinical testing was not applicable. Details of these individual test methodologies and results are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the ground truth would have been established through controlled electrical signals, mechanical simulations, reference devices, or chemical analysis, depending on the specific test (e.g., calibrated ECG simulators for ECG performance, known activity levels for activity validation). However, the document does not specify these details.

    8. The sample size for the training set

    Not applicable. The document does not describe the development or training of an AI/ML algorithm that would require a training set in the context of clinical performance. The "software verification and validation" would have involved internal testing against coding standards and functional requirements, but not a "training set" in the machine learning sense.

    9. How the ground truth for the training set was established

    Not applicable, as no training set (for AI/ML algorithm) is mentioned or implied.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070235

    Validate with FDA (Live)

    Date Cleared
    2007-05-18

    (113 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuard infusion Pump system is designed for infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes, of administration, including intravenous, subcutaneous, percutaneous, intraarterial, epidural, enteral, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound site). The system is intended for patients who require maintenance medications, analgesics, PCA therapy, parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids therapy in hospital and home care environments.

    Device Description

    The BodyGuard Infusion system includes: Infusion Pump Battery Charger Infusion Sets (Microset/ BodySet) Bolus cable (optional) Drop sensor (optional)

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter for the BodyGuard Infusion Pump System and primarily focuses on the regulatory approval and indications for use. It does not provide details about acceptance criteria, study methodologies, sample sizes, expert qualifications, or ground truth establishment relevant to the performance of the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042696

    Validate with FDA (Live)

    Date Cleared
    2005-03-11

    (162 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuard infusion Pump system is designed for infusion of continuous or intermittent delivery at medically-controlled infusion rates through clinically acceptable routes of administration (e.g. intravenous, subcutaneous, percutaneous, intraarterial, epidural, enteral, in close proximity to nerves, and into an arterial, regional, or surgical wound site). The system is intended for patients who require maintenance medications, analgesics, parenteral and enteral nutrition fluids, chemotherapeutic agents or PCA therapy; parenteral fluids therapy in hospital and home care environments.

    Device Description

    The BodyGuard Infusion system includes: Infusion Pump Battery Charger Infusion Sets (Microset/ BodySet) Bolus cable (optional) Drop sensor (optional)

    AI/ML Overview

    I regret to inform you that the documents you provided do not contain the specific information needed to fulfill your request.

    The provided text is an FDA 510(k) clearance letter for the "BodyGuard Infusion Pump System." This letter confirms that the device is substantially equivalent to a predicate device and can be marketed. However, it does not include details about:

    • Acceptance criteria for device performance.
    • A study describing performance results against those criteria.
    • Sample sizes, data provenance, ground truth establishment, expert qualifications, or adjudication methods.
    • Information regarding MRMC studies, standalone performance, or training set details.

    The letter focuses on the regulatory clearance process and the device's indications for use, not the detailed performance studies or the methodology behind them.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or answer the other specific questions based on the input text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031749

    Validate with FDA (Live)

    Date Cleared
    2003-08-21

    (77 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BodyGuard Infusion system is designed to transfer medication and fluids intravenously. The system is intended for patients who require maintenance medications, PCA therapy, Parenteral nutritional fluids and general I.V. fluids therapy in hospital and home care environments.

    Device Description

    The BodyGuard Infusion system includes: Infusion Pump Battery Charger Infusion Sets (Microset/ BodySet female luer) PCA cable (optional) Charging cable (optional)

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for the "BodyGuard Infusion System". It declares the device substantially equivalent to a predicate device. However, this document does not contain any information regarding specific acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications in the way that would be necessary to fill out the requested table and study details.

    The letter explicitly states: "FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies." It is primarily focused on regulatory clearance based on substantial equivalence, not on a detailed clinical performance study report.

    Therefore, I cannot extract the requested information from the provided text. The document is administrative in nature, confirming regulatory compliance rather than detailing clinical validation or performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K000441

    Validate with FDA (Live)

    Device Name
    BODYGUARD
    Manufacturer
    Date Cleared
    2000-04-11

    (61 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device