Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K240215

    Validate with FDA (Live)

    Date Cleared
    2024-06-28

    (154 days)

    Product Code
    Regulation Number
    882.4950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BLUE 400 and BLUE 400 S are accessories to the surgical microscope and allow the fluorescence observation of fluorophores with an excitation peak between 400 nm and the fluorescence emission observation comprising the spectrum in a spectral band of 620 - 710 nm.

    The ZEISS BLUE 400 and BLUE 400 S are surgical microscope accessories used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery.

    Device Description

    The BLUE 400 and BLUE 400 S are fluorescence accessories to qualified surgical microscopes, intended to allow intraoperative viewing of malignant glioma tissue grade III and IV under fluorescence. The overall system is comprised of excitation (illumination) and emission (observation) filters to detect fluorescence and are optimized in conjunction with the drug to pass light between 620 – 710 nanometers. The BLUE 400 S filters allow the surgical microscope to produce excitation light in a wavelength range covering at least 400 - 410 nanometers that excites an approved optical imaging agent and enables the surgeon to observe the emitted fluorescent signal in the oculars or on a display. Fluorescence of marked brain tissue helps visualization of tissue associated with Grade III & IV glioma during neurosurgeries.

    Compared to the blue visualization of the surrounding non-fluorescent tissue in the BLUE 400 image, BLUE 400 S is designed to visualize the surrounding nonfluorescent tissue more similar to white light impression, while tumor visualization of grade III and IV glioma remains consistent. With the visualization of non-fluorescent anatomy in an almost white light impression, BLUE 400 S is expected to allow PplX visualization with less frequent switching between fluorescence and white light imaging modes.

    BLUE 400 and BLUE 400 S can be installed only into qualified ZEISS surgical microscopes. For these accessories to be used with a qualified ZEISS surgical microscope, the critical components of the surgical microscope need to fulfill the clinically relevant parameters for the Indications for Use of BLUE 400 and BLUE 400 S.

    AI/ML Overview

    The provided FDA 510(k) summary (K240215) describes the Carl Zeiss Meditec AG BLUE 400 and BLUE 400 S accessories to surgical microscopes for fluorescent visualization of grade III and IV gliomas.

    Here's an analysis of the acceptance criteria and study information:


    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on demonstrating substantial equivalence to a predicate device (BLUE 400, K211346) through technical and performance testing, rather than defining explicit clinical acceptance criteria in terms of sensitivity, specificity, or other diagnostic measures for identifying gliomas. The acceptance criteria used are in the form of "Passed" results for various technical and functional tests.

    Test DescriptionAcceptance Criteria (Implied by "Passed" result)Reported Device Performance
    Brightness of the fluorescence ocular imageImage brightness of ZEISS fluorescence target at ocular plane at 250 mm working distance meets specified target. (Specific value not provided here, but "Passed")Passed
    IrradianceExcitation light density in the object plane meets specified target. (Specific value not provided here, but "Passed")Passed
    Spatial resolution of the ocular imageSpatial resolution measured with test target in white light mode at min/max magnification and 200mm working distance meets specified target. (Specific value not provided here, but "Passed")Passed
    Excitation wavelength (of the microscope)Excitation wavelength range of PpIX (400 nm to 410 nm) is covered by both BLUE 400 and BLUE 400 S options. (Specific quantitative range achieved for subject device: BLUE 400: 400-430nm; BLUE 400 S: 398-457nm for 50% edges)Passed
    Excitation filterOptical filter specification of excitation filter meets requirements.Passed
    Emission wavelength (of the microscope - ocular image)Design review/measurement of spectrum at ocular plane meets requirements. (Specific quantitative range achieved for subject device: BLUE 400: >450nm; BLUE 400 S: 540-728nm for 50% edges)Passed
    Emission wavelength (of the microscope - video image)Design review/measurement of spectrum at ocular plane meets requirements.Passed
    Emission filterOptical filter specification of emission filter meets requirements.Passed
    Non-mirrored video imageVisual inspection with test target in white light mode confirms non-mirrored image.Passed
    Non-rotated video imageVisual inspection with test target in white light mode confirms non-rotated image.Passed
    Non-deformed video imageVisual inspection of geometric distortions of a test target with a circle in white light mode shows no significant deformation.Passed
    Centered video imageVisual inspection and measurement with a test target in white light mode confirms centered image.Passed
    Photometric resolution of video imageGrey value resolution test with photometric resolution test target in white light mode meets requirements.Passed
    Signal-to-noise ratio of the video image (sensitivity)Signal-to-noise ratio of video image of a fluorescent target at a given signal value meets requirements.Passed
    Latency of the video imageVideo latency in white light mode meets requirements.Passed
    Spatial resolution of the video imageSpatial resolution measured with test target in white light mode meets requirements.Passed
    Spectrum of the Illumination Source (TS1)Irradiance spectrum (250 nm - 1020 nm, mW/cm²) of illumination source measured and verified with spectrometer prior to excitation filter module application.Passed
    Maximum Power and Irradiance of the Illumination Source (TS2)Maximum output power and irradiance of illumination sources measured and verified with power meter at end of microscope light guide prior to excitation filter module application.Passed
    Irradiance Spectrum of the Excitation Light and Spectral Response of the Excitation Filter (TS3)Irradiance spectrum (250 nm - 1020 nm) of illumination light after excitation filter module measured; 50% decrease edges of blue excitation peak calculated and found acceptable.Passed
    Maximum Excitation Power and Power Density (TS4)Maximum power (mW) and power density (mW/cm²) of excitation light measured at multiple working distances/zoom settings. Subject device measurements comparable to predicate device.Passed
    Optical Path Loss (TS5)Detectable light output and total losses in relation to device working distance and zoom setting, calculated by dividing output signal at eyepiece by illumination signal at focal plane for the same zoom setting, found acceptable.Passed
    Spectrum of the Emission Filter (TS6)Spectrum (350 nm – 1050 nm) of emission filter integrated in surgical microscope measured; 50% edge of spectrum calculated and found acceptable.Passed
    Homogeneity of the Excitation Light at the Focal Point (TS7)Reflected signal from white sheet of paper at 30 cm working distance imaged, and intensity profile calculated to demonstrate homogeneity of excitation light, found acceptable.Passed
    System Sensitivity (TS8)BLUE 400: Fluorescence signal in eyepiece for ZEISS BLUE 400 fluorescent target at 22.5 cm working distance comparable to predicate device. BLUE 400 S: Ratio of reconstructed fluorescence signal to reconstructed remission spectrum (corresponding to fluorescence to emission light ratio) found acceptable. (Specific details of "acceptable" criteria are not provided)Passed
    Pre-Operative Phantom Test (TS9)ZEISS BLUE 400 test phantom (with one fluorescent area) suitable for pre-operative checks of a surgical microscope; imaged by camera and observed through eyepiece.Passed
    Spectrum of the Camera Filter (TS10)Spectrum at camera interface measured to demonstrate camera filter blocks near infrared and infrared leakage of excitation light to the camera.Passed
    Special Controls TestingPerformance with and without cover glass met defined specifications.Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical performance testing (bench testing) using "ZEISS fluorescence target," "spatial test target," "photometric resolution test target," "test target with a circle," "white sheet of paper," and "ZEISS BLUE 400 test phantom."

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient for bench validation of optical and system parameters. These are physical components being tested, not patient samples.
    • Data Provenance: The tests are described as bench/non-clinical system testing. This indicates the data was generated in a lab setting by the manufacturer (Carl Zeiss Meditec AG, Germany, based on manufacturer details). It is not derived from patient data.
    • Retrospective/Prospective: Not applicable as it's bench testing, not clinical studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable. This submission relies on technical bench testing against predefined engineering specifications and comparison to a legally marketed predicate device. The "ground truth" for these tests is the physical measurement of optical properties and system functions, validated against engineering requirements, not clinical expert consensus on patient data.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies for establishing ground truth (e.g., determining disease presence in an image) based on multiple readers. This submission describes bench testing where the outcome is a "Passed" result based on meeting physical specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The submission focuses on non-clinical technical equivalence and performance of the device's optical and system functions. It does not evaluate human reader performance with or without AI assistance. The device itself (BLUE 400/BLUE 400 S) is a filter accessory, not an AI-powered diagnostic tool aiming to improve reader interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No, a standalone algorithm performance study was not done. This device is an accessory to a surgical microscope. It enhances visualization for a human surgeon; it is not a standalone algorithm that provides diagnoses or interpretations. The software verification testing mentioned is for the device's control software, not a diagnostic algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for the performance testing is engineering specifications and measurements of optical, electrical, and mechanical properties. For example, excitation wavelength range is validated against the known excitation peak of PpIX, and image properties (resolution, brightness, lack of deformation) are validated against defined standards for surgical microscopes. The comparison to the predicate device also serves as a benchmark for equivalence.

    8. The Sample Size for the Training Set

    • Not applicable. The BLUE 400 and BLUE 400 S are physical filter accessories for a surgical microscope, not an AI/machine learning device that requires a training set. The software mentioned is for the device's operation and control, not for image analysis or diagnostic inference that would necessitate a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As the device is not an AI/ML system requiring a training set, there is no ground truth established for such a set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211346

    Validate with FDA (Live)

    Device Name
    BLUE 400
    Date Cleared
    2022-07-22

    (445 days)

    Product Code
    Regulation Number
    882.4950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BLUE 400 is an accessory of the surgical microscope and allows the fluorescence observation of fluorophores with an excitation peak between 400 mm and the fluorescence emission observation comprising the spectrum in a spectral band of 620 - 710 nm.

    The ZEISS BLUE 400 is a surgical microscope accessory used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery.

    Device Description

    The BLUE 400 is an accessory to the Zeiss surgical microscopes (OPMI PENTERO 800, OPMI PENTERO 900, and KINEVO 900), intended to allow intraoperative viewing of malignant glioma tissue under fluorescence. The BLUE 400 accessory is entirely composed of optical filters: the "Excitation" filter and the "Emission" filters. The Excitation filter is designed to filter all light wavelengths except 400 - 470 nanometers and is optimized to pass light between 400 - 410 nanometers. The Emission filters are designed to filter all light wavelengths except 430 - 800 nanometers and is optimized to pass light between 620 - 710 nanometers.

    When installed in the surgical microscopes (class I), the BLUE 400 introduces optical filters to the illumination and viewing optical paths. The BLUE 400 includes installation of a software license that facilitates use of the accessory. After the SW license is installed, the user has the option to switch from the normal white light mode of the surgical microscope to the BLUE 400 mode.

    The BLUE 400 accessory, when installed in the surgical microscopes, is intended to be used in conjunction with an approved optical imaging agent that is excited mainly in the wavelength range of 400 – 410 nanometers and fluoresces in the wavelength range of 620 - 710 nanometers.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BLUE 400 device, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in a quantitative manner (e.g., specific thresholds for irradiance or power). Instead, they are implied by the "Passed" result for each test, indicating that the device met the defined specifications for each performance parameter. The study is a bench performance test comparing the subject device (BLUE 400) to a predicate device (Leica FL400).

    TestAcceptance Criteria (Implied)Reported Device Performance
    Spectrum of the Illumination SourceIrradiance spectrum (250 nm - 1020 nm, mW/cm²) verified and assessed prior to excitation filter application.Passed
    Maximum Power and Irradiance of the Illumination SourceMaximum output power and irradiance measured and verified prior to excitation filter application.Passed
    Irradiance Spectrum of the Excitation Light and Spectral Response of the Excitation FilterIrradiance spectrum (250 nm - 1020 nm) of illumination light after excitation filter passage measured; 50% decrease edges of blue excitation peak calculated.Passed
    Maximum Excitation Power and Power DensityMaximum power (mW) and power density (mW/cm²) of excitation light measured at various working distances and zoom settings and compared to the predicate device.Passed
    Optical Path LossOptical path loss calculated by dividing output signal (eyepiece without emission filter) by illumination signal (focal plane with spectrometer).Passed
    Spectrum of the Emission FilterSpectrum (350 nm - 1050 nm) of the emission filter (integrated into microscope) measured; 50% edge of the spectrum calculated.Passed
    Homogeneity of the Excitation Light at the Focal PointIntensity profile of reflected signal from white paper imaged by camera demonstrated homogeneity.Passed
    System SensitivityFluorescence signal in the eyepiece of the subject device compared to the predicate device using a fluorescent target.Passed
    Pre-Operative Phantom TestSuitability of the ZEISS BLUE 400 test phantom for pre-operative checks of KINEVO 900 and OPMI PENTERO 900 demonstrated via camera imaging and eyepiece observation.Passed
    Spectrum of Camera FilterSpectrum at camera interface measured to demonstrate camera filter blocks near infrared and infrared leakage of excitation light.Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench performance testing. Therefore, the "test set" in the traditional sense of patient data is not applicable. The testing involves:

    • Test Sets: The devices themselves (subject device BLUE 400 and predicate device Leica FL400) and various optical components (filters, light sources, specialized targets/phantoms).
    • Data Provenance: The tests were conducted internally by Carl Zeiss Meditec AG, as part of their 510(k) submission. This is a prospective bench study. No external data sources or patient data are mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This was a bench performance study assessing physical and optical properties, not a clinical study requiring expert interpretation of patient data to establish ground truth. The "ground truth" was established by the physical measurements and calculations against defined specifications.

    4. Adjudication Method for the Test Set

    Not applicable. This was a direct measurement and comparison bench study, not a clinical study requiring adjudication of expert readings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an accessory (an optical filter) to a surgical microscope, not an AI or imaging diagnostic tool that would typically involve a multi-reader multi-case study for diagnostic accuracy or human performance improvement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself is an optical filter system for a surgical microscope. It does not have an "algorithm" in the sense of an AI or software that performs standalone interpretation. Its function is to modify light for improved visualization by a human surgeon. Therefore, standalone algorithm performance is not applicable. The document does mention Software verification testing was performed in accordance with FDA Guidance to demonstrate the software (for the license installation and mode switching) is performing as intended. This is analogous to a standalone performance check for the software component, but not for an interpretative algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the bench performance tests, the "ground truth" was established by:

    • Physical measurements: Using calibrated instruments like spectrometers, power meters, and thermopiles to measure optical properties (irradiance spectrum, power, density, emission spectrum).
    • Calculations: Such as calculating the 50% decrease edges of spectral peaks and optical path loss.
    • Comparison to predefined specifications: The results were evaluated against specific technical requirements and specifications for each test.
    • Comparison to a predicate device: For certain tests like maximum excitation power density and system sensitivity, the performance of the subject device was compared directly to the predicate device (Leica FL400).

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1