Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K090539

    Validate with FDA (Live)

    Date Cleared
    2009-10-30

    (245 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiPAP AutoSV Advanced is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.

    Device Description

    The Respironics BiPAP AutoSV Advanced is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

    The BiPAP AutoSV Advanced pressure control contains various which are used to configure positive pressure therapies. With these controls, the device delivers minimum pressure support determined by the PSmin control. The device may automatically provide additional pressure support with inspiratory pressures between PSmin and PSmax to normalize patient ventilation during sleep disordered breathing events. Note: When PSmin = PSmax, this is equivalent to traditional bi-level therapy.

    The BiPAP AutoSV Advanced is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

    The BiPAP AutoSV Advanced Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing and a method of venting exhaled gases.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study proving the device meets them in the format requested. The document is a 510(k) summary for the BiPAP AutoSV Advanced, focusing on its substantial equivalence to predicate devices and general performance modifications.

    However, based on the text, I can extract the following relevant information:

    Key Takeaways from the Text:

    • Clinical Validation: The manufacturer performed "clinical validation to ensure the efficacy of the algorithm was not affected" due to modifications to the AutoSV algorithm.
    • Bench Data: Other device modifications were validated using "bench data" including waveform performance, triggering data, and overall event detection and control data.
    • Equivalence Claim: The testing "confirmed that the BiPAP AutoSV Advanced performs equivalently to the device predicate BiPAP AutoSV (K063540). All tests were verified to meet the required acceptance criteria."

    Given the limited information, I cannot accurately fill out many of the requested fields. Here's a structured response based on what is available and what cannot be determined:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "All tests were verified to meet the required acceptance criteria" and that the device "performs equivalently to the device predicate BiPAP AutoSV (K063540)." However, it does not provide a specific table of these criteria or quantitative reported performance values for each criterion.

    Acceptance Criteria (Example - Not explicitly stated in document)Reported Device Performance (Example - Not explicitly stated in document)
    Event detection accuracy (e.g., % sensitivity/specificity for apneas)"Improved event detection" compared to predicate. Equivalent to predicate.
    Pressure control stability/accuracy"Use of EPAP Control for pressure control when central events are detected." Equivalent to predicate.
    Response time of backup rate"Improving the response time of the backup rate." Equivalent to predicate.
    Waveform performanceMet required acceptance criteria; equivalent to predicate.
    Triggering performanceMet required acceptance criteria; equivalent to predicate.
    Overall event detection and controlMet required acceptance criteria; equivalent to predicate.

    2. Sample Size and Data Provenance:

    • Sample Size for Test Set: Not specified. The document mentions "clinical validation" but does not give the number of patients or cases.
    • Data Provenance: Not specified. It's unclear if the validation data was prospective or retrospective, or its country of origin.

    3. Number of Experts and Qualifications for Ground Truth:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method:

    • Adjudication Method: Not specified.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • MRMC Study Done?: No information provided. The document focuses on performance equivalence to a predicate device, not on human reader improvement with AI assistance.

    6. Standalone Performance Study:

    • Standalone Study Done?: Likely yes, as implied by "bench data" and "clinical validation to ensure the efficacy of the algorithm was not affected." However, specific metrics for standalone performance (e.g., sensitivity, specificity for event detection) are not provided. The comparison is made against a predicate device.

    7. Type of Ground Truth Used:

    • Type of Ground Truth: Not specified explicitly. For clinical validation of a ventilator algorithm, it would typically involve physiological measurements and potentially expert interpretation of physiological signals (e.g., polysomnography data for sleep disordered breathing events). For bench testing, it would be against known physical models or simulated conditions.

    8. Sample Size for Training Set:

    • Sample Size: Not specified. The document mentions modifications to an existing algorithm ("AutoSV algorithm") but does not detail a training set for a de novo algorithm development.

    9. How Ground Truth for Training Set was Established:

    • How Ground Truth Established: Not specified. As it's an update to an existing algorithm, it's possible that historical data and expert-defined criteria from the predicate device's development were used, but this is conjecture.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090248

    Validate with FDA (Live)

    Device Name
    BIPAP AUTOSV
    Date Cleared
    2009-03-04

    (30 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiPAP AutoSV is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.

    Device Description

    The Respironics BIPAP AutoSV is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

    The BiPAP AutoSV pressure control that contains various which are used to configure positive pressure therapies. With these controls, the device delivers minimum pressure support determined by the EPAP and IPAP Min controls. The device may automatically provide additional pressure support with inspiratory pressures between IPAP Min and IPAP Max to normalize patient ventilation during sleep disordered breathing events. Note: When EPAP < IPAP Min = IPAP Max, this is equivalent to traditional bi-level therapy.

    The BIPAP AutoSV is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

    The BiPAP AutoSV Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modified device, the Respironics BiPAP AutoSV. The submission claims substantial equivalence to previously cleared predicate devices. The "study" mentioned refers to design verification tests performed on the device to ensure it meets required acceptance criteria. However, the document does not contain detailed information about a clinical study with a detailed methodology, acceptance criteria, and specific performance results in the format you requested for "proving" the device meets acceptance criteria.

    Based on the provided text, here's what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Design Verification Tests for risk analysis and product requirementsAll tests were verified to meet the required acceptance criteria.
    Compliance with applicable standards referenced in "Guidance for the Content of Pre-market Submissions for Software in Medical Devices," May 2006.The modified device complies with these standards.
    No impact on the safety and effectiveness of the device compared to the predicate.Respironics has determined that the modifications have no impact on the safety and effectiveness of the device.

    Missing Information:

    • Specific quantitative acceptance criteria (e.g., "pressure accuracy must be within +/- X cmH2O").
    • Specific quantitative performance results from the design verification tests (e.g., "pressure accuracy was Y cmH2O").

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. The document mentions "design verification tests" but does not detail the number of devices or scenarios tested.
    • Data Provenance: Not specified. It's internal testing conducted by Respironics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for design verification tests is typically established through engineering specifications and regulatory standards, not expert consensus on patient data.

    4. Adjudication method for the test set:

    • Not applicable. This typically refers to clinical study data adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This type of study is not mentioned in the document. The device is a ventilator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The document implies that the device (which contains a microprocessor controller) underwent standalone testing as part of the design verification process. The "algorithm" here would be the control logic of the ventilator. However, specific details about protocol for standalone algorithm performance are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For design verification tests, the "ground truth" would be the engineering specifications and regulatory standards that the device is required to meet.

    8. The sample size for the training set:

    • Not applicable in the context of this document. This refers to AI/ML model training, which is not the focus of this submission. The device is microprocessor-controlled but not described as using machine learning that would require a "training set" in the common sense of the term for AI products.

    9. How the ground truth for the training set was established:

    • Not applicable (see point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K063540

    Validate with FDA (Live)

    Device Name
    BIPAP AUTOSV
    Manufacturer
    Date Cleared
    2007-02-22

    (90 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiPAP AutoSV is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.

    Device Description

    The Respironics BiPAP AutoSV is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

    The BiPAP AutoSV pressure control that contains various which are used to configure positive pressure therapies. With these controls, the device delivers minimum pressure support determined by the EPAP and IPAP Min controls. The device may automatically provide additional pressure support with inspiratory pressures between IPAP Min and IPAP Max to normalize patient ventilation during sleep disordered breathing events. Note: When EPAP < IPAP Min = IPAP Max, this is equivalent to traditional bi-level therapy.

    The BiPAP AutoSV is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

    The BiPAP AutoSV Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases..

    AI/ML Overview

    The provided 510(k) summary for the BiPAP AutoSV device does not contain the detailed information necessary to fully address all aspects of your request regarding acceptance criteria and the study that proves the device meets them. This document is a summary and primarily focuses on demonstrating substantial equivalence to predicate devices through
    bench testing and general comparisons, rather than a detailed report of clinical study outcomes or specific acceptance criteria for performance metrics.

    However, based on the provided text, here's what can be extracted and what information is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary states: "Bench testing has confirmed that the BiPAP AutoSV performs equivalently to the device predicate VPAP Adapt (K051364). All tests were verified to meet the required acceptance criteria."

    This statement confirms that acceptance criteria exist and were met, but the document does not specify what those criteria are, nor does it provide a table of performance metrics. It only broadly mentions "waveform performance data, triggering data, alarms data, and overall event diction and control data."

    Acceptance CriteriaReported Device Performance
    Not specified in the provided document.Equivalently performed to predicate VPAP Adapt (K051364) in waveform performance, triggering, alarms, and overall event detection and control data.
    Specific thresholds for performance related to waveform, triggering, alarms, and event detection.Met all required acceptance criteria for these tests.

    2. Sample Size Used for the Test Set and Data Provenance

    The document refers to "comprehensive bench testing."

    • Sample Size: The sample size for the test set is not specified. This typically refers to the number of devices or test conditions used in bench testing.
    • Data Provenance: The data provenance is bench testing, meaning it's laboratory-generated data from Device A versus Device B comparison, rather than human subject data. The "country of origin" for the data is implicitly where the testing was conducted, likely the manufacturer's facilities (Respironics, Inc. 1001 Murry Ridge Lane Murrysville, PA 15668, USA). This testing is retrospective in the sense that it compares a new device against an already cleared predicate device's performance characteristics.

    3. Number of Experts and Their Qualifications for Ground Truth

    • This device's submission relies on bench testing for proving equivalence, not clinical studies involving human patients where expert-established ground truth would be relevant for diagnostic or therapeutic accuracy. Therefore, information about the number or qualifications of experts used to establish ground truth is not applicable in this context as described in the summary.

    4. Adjudication Method for the Test Set

    • Since the testing described is primarily bench testing comparing device performance parameters, an "adjudication method" in the sense of resolving discrepancies between expert interpretations (e.g., 2+1, 3+1 for clinical evaluations) is not applicable. The "adjudication" would be based on objective measurements against pre-defined engineering or performance specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done according to this 510(k) summary. The document explicitly states the equivalence was demonstrated through "comprehensive bench testing." MRMC studies are typically used for diagnostic devices to assess the impact of AI on human reader performance, which doesn't fit the context of this ventilatory support system's submission as described.

    6. Standalone (Algorithm Only) Performance

    • The summary describes the device as a "microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels." The "AutoSV algorithm" is mentioned as a component. While the algorithm is a key part of the device's functionality, the performance testing described is for the integrated device ("BiPAP AutoSV performs equivalently"). Therefore, a standalone algorithm-only performance study is not explicitly mentioned or implied. The performance is assessed at the device level.

    7. Type of Ground Truth Used

    • For the bench testing, the "ground truth" would be the established performance specifications and data from the predicate device (Resmed VPAP Adapt, K051364). The BiPAP AutoSV's performance data (waveform, triggering etc.) was compared against these established benchmarks to demonstrate "equivalence." It is not pathology, outcomes data, or expert consensus in a clinical diagnostic sense.

    8. Sample Size for the Training Set

    • This 510(k) submission is for a medical device (ventilator) that incorporates an algorithm. It is not an AI/ML diagnostic or prognostic software submission where a "training set" of patient cases for algorithm development would typically be disclosed as a distinct component. The document does not provide information on a "training set" for the AutoSV algorithm. Algorithm development often involves internal testing and refinement, but details about specific training data are not part of this 510(k) summary.

    9. How Ground Truth for the Training Set Was Established

    • As a "training set" for an explicit AI/ML algorithm is not described in this 510(k) summary, the method for establishing its ground truth is not applicable/not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050759

    Validate with FDA (Live)

    Device Name
    BIPAP AUTO
    Manufacturer
    Date Cleared
    2005-06-15

    (83 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiPAP Auto delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea. The device is for use in the home or hospital/institutional environment on adult patients.

    Device Description

    The Respironics BiPAP Auto is a microprocessor controlled blower based bi-level positive pressure system that delivers two different positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. Two modes are being introduced, which include the Auto Bi-Level mode and Split Night Auto Bi-Level mode. Bi-Level therapy with automatically adjusting IPAP and EPAP levels is delivered in the Auto Bi-Level mode. The Split Night Auto Bi-Level mode limits the pressure during the first portion of the night to allow diagnostic recording; thereafter the pressure responds to the patient's needs to provide an assessment of therapeutic requirements. Pressure relief upon exhalation (Bi-Flex feature) may be enabled during Auto Bi-Level therapy.

    The BiPAP Auto is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, an exhalation device, and a patient interface device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BiPAP Auto device, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided document primarily focuses on establishing "substantial equivalence" to predicate devices rather than setting explicit quantifiable acceptance criteria against specific performance metrics for the new Auto Bi-Level and Split Night Auto Bi-Level modes. However, the document does outline the clinical efficacy criteria for the Auto Bi-Level mode.

    Acceptance Criteria (for Clinical Efficacy)Reported Device Performance
    Primary: As effective at determining optimal pressures as manually titrated, bi-level positive airway pressure.Demonstrated (Clinical Efficacy Study)
    Secondary: Maintain lower average treatment pressures (average IPAP and average EPAP) as compared to manually titrated fixed pressure bi-level therapy.Demonstrated (Clinical Efficacy Study)
    Secondary: AHI (Apnea-Hypopnea Index) will remain below 10 during polysomnography.Demonstrated (Clinical Efficacy Study)

    Note: The document states that the "Clinical Efficacy Study; demonstrated that the Auto Bi-Level Mode (auto-adjusting bi-level positive airway pressure) of the BiPAP Auto device is as effective at determining optimal pressures as manually titrated, bi-level positive airway pressure," and that "The following performance indicators additionally demonstrate the clinical efficacy of the Auto Bi-Level therapy: Auto Bi-Level therapy maintains lower average treatment pressures (as defined by average IPAP and average EPAP as compared to manually titrated fixed pressure bi-level therapy), and the AHI will remain below 10 during polysomnography." This indicates that all these criteria were met.

    Study Information

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided text.
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective). It simply refers to a "Clinical Efficacy Study."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The clinical efficacy study compares the device's auto-adjusted pressures to "manually titrated, bi-level positive airway pressure." The manual titration itself implies expert involvement, but the number and qualifications of those experts are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study as typically understood for human readers interacting with AI was not described. The study was a "Clinical Efficacy Study" comparing the device's auto-adjusting mode to manually titrated therapy. This is a comparison between an automated therapy mode and a traditional manual therapy, not a study of human readers assisted by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the "Bench Testing BiPAP Auto vs. REMstar Auto with C-Flex" assessed the event detection and control capability of the BiPAP Auto. This would be a standalone evaluation of the algorithm's performance in a controlled environment.
    • The "Clinical Efficacy Study" also assessed the Auto Bi-Level Mode which is an algorithm-driven therapy, comparing its effectiveness against manual titration. While patients are involved, the core evaluation is of the algorithm's ability to determine optimal pressures and maintain AHI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the "Clinical Efficacy Study," the "ground truth" or comparator for optimal pressure determination was "manually titrated, bi-level positive airway pressure." This implies expert-driven titration. The AHI remaining below 10 during polysomnography suggests polysomnography results were used as an objective measure.

    8. The sample size for the training set:

    • This information is not provided in the document. The device's "Auto therapy algorithm" was added, and it's stated that the "Split Night mode (split night study feature) is based on the Split Night Auto CPAP mode (split night study feature) currently cleared by the FDA in K041010 (REMstar Auto with C-Flex CPAP System)." This suggests prior development and possibly training from the predicate REMstar Auto.

    9. How the ground truth for the training set was established:

    • This information is not provided in the document. Given that the algorithm and features are adapted from the predicate device (REMstar Auto with C-Flex), the training and ground truth establishment would likely have occurred during the development of that earlier device. The submission focuses on the modifications and equivalence.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1