510(k) Data Aggregation

    K Number
    K110879

    Validate with FDA (Live)

    Date Cleared
    2011-08-19

    (142 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The JuggerKnot™ Mini Soft Anchors are intended to be used for soft tissue to bone fixation with indications for use in:

    Shoulder Bankart repair

    Foot and Ankle Midfoot Reconstruction, Hallux valgus reconstruction

    Hand and Wrist

    Ulnar or lateral collateral ligament reconstruction, Repair/reconstruction of collateral ligaments, flexor and extensor tendon at the PIP (proximal interphalangeal, DIP (distal interphalangeal), and MCP (metacarpal interphalangeal) joints for all digits, Scapholunate ligament reconstruction.

    Device Description

    The JuggerKnot™ Mini Soft Anchors consist of a coreless sleeve structure and suture. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the JuggerKnot™ Mini Soft Anchors, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    Mechanical FixationEquivalent fixation strength to predicate devices"The test results indicate that the Biomet Sports Medicine JuggerKnot™ Mini Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use."

    Note: The document explicitly states "None provided as a basis for substantial equivalence" under "Clinical Testing." This indicates that biological safety and efficacy were demonstrated through non-clinical testing and comparison to predicate devices, rather than human trials.

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated. The summary mentions "mechanical pullout testing" where the JuggerKnot™ Mini Soft Anchors were compared to two predicate devices. This implies a sample size of anchors for each device type subjected to the pullout test, but the exact number isn't quantified.
    • Data Provenance: The testing was "Non-clinical laboratory testing," indicating it was conducted in a controlled lab environment. The country of origin is not specified, but the applicant "Biomet Sports Medicine" is located in Warsaw, Indiana, USA, suggesting the testing likely occurred in the US or under US guidance. The testing is retrospective in the sense that it's performed on manufactured devices, not on living subjects over time.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable in the context of this device and study. The "ground truth" for non-clinical mechanical testing is typically established by engineering standards and validated testing protocols, not by expert consensus in the way clinical diagnostic image interpretation would be. The outcome (fixation strength) is objectively measured by testing equipment.

    4. Adjudication Method for the Test Set

    This question is not applicable. As the testing is mechanical and quantitative (fixation strength), there is no "adjudication" of expert opinions. The measurements from the mechanical tests are the primary data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical cases. This submission focuses on the mechanical performance of a soft tissue fixation device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is not an AI/algorithm-based diagnostic device. The "device" itself is a physical implant. The "standalone" performance here refers to the mechanical performance of the anchor itself when tested in a controlled laboratory setting. The non-clinical testing described is essentially demonstrating the "standalone" mechanical performance of the anchors.

    7. The Type of Ground Truth Used

    The ground truth used for the non-clinical testing was objective mechanical measurements of fixation strength, as determined by standardized laboratory pullout tests.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" in the context of this 510(k) submission for a mechanical device. The device's design is based on engineering principles and similarity to predicate devices, not on machine learning from a dataset.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable for the reasons stated in point 8.