    Intended Use

    The Pre-loaded Bio-Anchor is a bioabsorbable device with attached suture used to attach soft tissue to bone in arthroscopic or open procedures for the following indications: Shoulder (Bankart lesion repairs, SLAP lesion repairs, Acromio-clavicular separation repairs, Rotator cuff tear repairs, Capsular shift or capsulolabral reconstructions, Biceps tenodesis, Altoid repairs), Foot and Ankle (Hallux Valgus repairs, Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstructions, Midfoot reconstructions), Elbow, Wrist and Hand (Scapholunate ligament reconstructions, Ulnar or radial collateral ligament reconstructions, Tennis elbow repair, Biceps tendon reattachment), Knee (Extracapsular repairs and reattachments of: medial collateral ligament, lateral collateral ligament, posterior oblique ligament or joint capsule to tibia, joint capsule closure to anterior proximal tibia; Extracapsular reconstruction, Iliotibial band tenodesis; Patellar realignment and tendon repairs), Bladder Neck Suspension (Soft tissue fixation of the pubic bone for the purpose of bladder neck suspension for female urinary incontinence due to urethral hypermobility).

    Device Description

    The Pre-loaded Bio-Anchor is an injection molded Poly (L-lactic) acid suture anchoring device with an attached nonabsorbable braided polyester surure USP size #2. The device is circular in shape with three cylindrical ribs perpendicular to the long axis. The suture is passed through the eyelet of the anchor and attached to a wire by shrink tubing.

    AI/ML Overview

    The provided text is a 510(k) Summary of Safety and Effectiveness for a medical device (Pre-loaded Bio-Anchor Absorbable Suture Anchor). This type of document is a premarket notification to the FDA, demonstrating that a new device is "substantially equivalent" to a legally marketed predicate device.

    It is crucial to understand that a 510(k) submission primarily focuses on demonstrating substantial equivalence, not necessarily on presenting original clinical studies with acceptance criteria, sample sizes, and ground truth establishment in the way one might find for a novel, high-risk device requiring PMA (Premarket Approval).

    Therefore, the information you've requested about specific acceptance criteria, detailed study designs, and performance metrics (like effect size for MRMC studies or standalone algorithm performance) is not typically found in a 510(k) summary like this one.

    Instead, 510(k) submissions typically reference:

    • Performance data from the predicate device: The new device is compared to a previously cleared device.
    • Bench testing: To demonstrate physical and mechanical properties.
    • Biocompatibility testing: To ensure safety when in contact with the body.
    • Sterilization validation: To confirm the device can be consistently sterilized.

    Based on the provided text, here's what can be extracted and what cannot:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or detailed reported device performance in the way a clinical study report would. It states: "Testing has been done to prove safety and effectiveness of the devices." The primary "performance" discussed is the demonstration of substantial equivalence to predicate devices in design, function, and intended use.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the summary. Since the submission focuses on substantial equivalence and refers to "testing," it's likely referring to bench-top mechanical testing, biocompatibility testing, and sterilization validation, not large-scale clinical trials that would have clearly defined test sets in terms of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. The assessment for a suture anchor is generally based on mechanical properties and biological interactions, not interpretation of images or patient data that would require "ground truth" established by human experts in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. As explained above, the "test set" and ground truth establishment do not align with methods requiring adjudication by experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable to this type of device. This device is a physical implant (suture anchor), not an AI diagnostic tool or software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical medical implant, not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied in AI/diagnostic studies is not directly applicable here. For a mechanical device like a suture anchor, "ground truth" would be related to material properties meeting specifications, biomechanical performance (e.g., pull-out strength), and biocompatibility, which are verified through engineering and biological testing, not clinical "outcomes data" in the same sense as a drug or complex diagnostic. The basis for safety and effectiveness is largely demonstrated through comparison to predicates and testing of physical properties.

    8. The sample size for the training set

    This information is not applicable/not provided. Training sets are relevant for AI algorithms, which this device is not.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. Training sets and ground truth establishment methods are primarily for AI/machine learning models.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964805

    Validate with FDA (Live)

    Device Name
    BIO-ANCHOR
    Manufacturer
    Date Cleared
    1997-06-09

    (192 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bio-Anchor is a bioabsorbable device used for soft tissue to bone fixation in orthopedic applications except for anterior or posterior cruciate ligament repair or reconstruction.

    Shoulder:

    1. Bankart lesion repairs
    2. SLAP lesion repairs
    3. Acromio-clavicular separation repairs
    4. Rotator cuff tear repairs
    5. Capsular shift or capsulolabral reconstructions
    6. Biceps tenodesis
    7. Deltoid repairs

    Foot and Ankle:

    1. Hallux Valgus repairs
    2. Medial or lateral instability repairs/reconstructions
    3. Achilles tendon repairs/reconstructions
    4. Midfoot reconstructions

    Elbow, Wrist and Hand:

    1. Scapholunate ligament reconstructions
    2. Ulnar or radial collateral ligament reconstructions
    3. Tennis elbow repair
    4. Biceps tendon reattachment

    Knee:

    1. Extra-capsular repairs and reattachments of:
      a. medial collateral ligament
      b. lateral collateral ligament
      c. posterior oblique ligament or joint capsule to tibia
      d. joint capsule closure to anterior proximal tibia
    2. Extracapsular reconstruction, iliotibial band tenodesis
    3. Patellar realignment and tendon repairs
    Device Description

    The Bio-Anchor is a PLLA suture anchoring device. The device is cylindrical in shape with three circular ribs perpendicular to the long axis. The first and second ribs are the same size, with the proximal, third rib wider to allow for a centrally located eyelet running perpendicular to the long axis of the device. Freely passing suture is then threaded onto the Bio-Anchor by the surgeon. The lack of permanent attachment of the suture to the Bio-Anchor allows the use of sliding knots. Once the Bio-Anchor is implanted, the ends of the suture are passed through the soft tissue and tied for reattachment. Synthetic non-absorbable, polyester USP #0, #1, or #2 suture is recommended for use with this device.

    The material used for this device is:

    Anchor - Poly (L-lactic) acid (PLLA)

    AI/ML Overview

    The provided text is a 510(k) Summary of Safety and Effectiveness for the Linvatec Bio-Anchor™ Absorbable Suture Anchor. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a performance study with quantitative results.

    Therefore, many of the requested categories regarding acceptance criteria, study design, sample sizes, ground truth establishment, and MRMC studies cannot be answered directly from the provided document. The document describes the device, its intended use, and its similarities and dissimilarities to predicate devices. It refers to a "chart of similarities and dissimilarities" to demonstrate substantial equivalence, which is a qualitative comparison rather than a formal performance study with acceptance criteria.

    Below is an attempt to address your request based only on the information provided in the input, with many fields explicitly marked as "Not provided in the document."

    Acceptance Criteria and Device Performance

    Since this is a 510(k) submission primarily focused on substantial equivalence, specific numeric acceptance criteria for performance are not explicitly stated in the provided text. The "performance" demonstrated is typically compliance with mechanical and material properties similar to the predicate device. The table below represents a qualitative comparison used for substantial equivalence rather than quantitative acceptance criteria.

    Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (as described in the document)
    Intended Use Equivalence: Device's purpose should be similar to predicate.New Product (Bio-Anchor): Soft tissue to bone fixation in orthopedic applications except for anterior or posterior cruciate ligament repair or reconstruction.Predicate (Preloaded Soft Tissue Anchor): Soft tissue to bone fixation.Predicate (Bio-Anchor Absorbable Suture Anchor - K955486): Soft tissue to bone fixation.
    Material Equivalence: Device material should be similar to or already established predicate.New Product (Bio-Anchor): Poly (L-lactic) acid.Predicate (Preloaded Soft Tissue Anchor): Titanium Alloy.Predicate (Bio-Anchor Absorbable Suture Anchor - K955486): Poly (L-lactic) acid.Note: The new Bio-Anchor's material is identical to the original Linvatec Bio-Anchor (K955486).
    Sterilization Method Equivalence: Sterilization method should be standard and similar to predicate or well-established.New Product (Bio-Anchor): ETO Sterilization.Predicate (Preloaded Soft Tissue Anchor): Gamma Sterilization.Predicate (Bio-Anchor Absorbable Suture Anchor - K955486): ETO Sterilization.
    Size Equivalence (where applicable): Device sizes should be comparable or within a reasonable range.New Product (Bio-Anchor): 3.5mm diameter, Suture: USP Sizes #0 to #2.Predicate (Preloaded Soft Tissue Anchor): 2.5mm - 5.2mm diameter, Suture: USP sizes #0 to #2.Predicate (Bio-Anchor Absorbable Suture Anchor - K955486): 3.5mm diameter, Suture: USP Sizes #0 to #2.
    Mechanism of Action/Design Equivalence: Device design and how it achieves its function should be similar.New Product (Bio-Anchor): Cylindrical shape with three circular ribs, central eyelet for freely passing suture. Lack of permanent suture attachment allows sliding knots.Predicate (Preloaded Soft Tissue Anchor): Not explicitly detailed, but implied to achieve soft tissue to bone fixation. The Preloaded predicate would imply a pre-attached suture, a dissimilarity noted. The Bio-Anchor predicate (K955486) would be very similar in design as it shares the same name and material.

    Study Details

    1. Sample size used for the test set and the data provenance: Not provided in the document. The document describes a comparison to predicate devices for substantial equivalence, not a performance study with a defined test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/Not provided. Ground truth establishment for a clinical test set is not described as part of this 510(k) submission.
    3. Adjudication method for the test set: Not applicable/Not provided.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is for a medical device (suture anchor), not an AI-assisted diagnostic or therapeutic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithm.
    6. The type of ground truth used: Not applicable/Not provided. The basis for approval is substantial equivalence to a legally marketed predicate device, primarily based on design, materials, and intended use, rather than a clinical performance study with "ground truth" as typically understood in diagnostic device evaluation.
    7. The sample size for the training set: Not applicable/Not provided. This is a medical device, not an AI model requiring a training set.
    8. How the ground truth for the training set was established: Not applicable/Not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K963369

    Validate with FDA (Live)

    Device Name
    BIO-ANCHOR
    Manufacturer
    Date Cleared
    1997-05-22

    (268 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bladder Neck Suspension
    The Linvatec Bio-Anchor is indicated for bladder neck suspension for female urinary incontinence due to urethral hypermobility.

    Device Description

    The Bio-Anchor is a PLLA suture anchoring device. The device is cylindrical in shape with three circular ribs perpendicular to the long axis. The first and second ribs are the same size, with the proximal, third rib wider to allow for a centrally located eyelet running perpendicular to the long axis of the device. Freely passing suture is then threaded onto the Bio-Anchor by the surgeon. The lack of permanent attachment of the suture to the Bio-Anchor allows the use of sliding knots. Once the Bio-Anchor is implanted, the ends of the suture are passed through the soft tissue and tied for reattachment. Synthetic non-absorbable, polyester USP #0, #1, or #2 suture is recommended for use with this device.
    The material used for this device is:
    Anchor - Poly (L-lactic) acid (PLLA)
    The Bio-Anchor Absorbable Suture Anchor is available in sizes ranging from 2.5mm to 5.2mm diameter.

    AI/ML Overview

    The provided text is a 510(k) Summary of Safety and Effectiveness for a medical device (Bio-Anchor Absorbable Suture Anchor) submitted to the FDA in 1997. This type of document is a premarket notification for demonstrating substantial equivalence to a legally marketed predicate device, not a study designed to prove the device meets specific acceptance criteria based on performance data.

    Therefore, the document does not contain the acceptance criteria, reported device performance, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC study results, standalone algorithm performance, or details on how ground truth was established for training and test sets as typically found in a clinical study report.

    Here's why this information is missing and an explanation based on the document type:

    This 510(k) submission focuses on demonstrating substantial equivalence to existing, legally marketed devices. It does this by:

    • Describing the device: Its material (PLLA), shape, and mechanism of action.
    • Stating its intended use: Bladder neck suspension for female urinary incontinence due to urethral hypermobility.
    • Identifying predicate devices: Linvatec Revo/Mini Revo, Mitek GII Anchor and Absorbable Anchor, Zimmer Statak Device, Zimmer Resorbable Soft Tissue Attachment Device, and Vesica Medical, Inc. Bone Anchor System.
    • Comparing the new device to predicates: A table highlights similarities and dissimilarities in materials, intended use, sterilization, and sizes. The key similarity emphasized is the material (PLLA) with the Zimmer Resorbable device and the intended use with several predicates.

    The FDA's review and approval (as indicated by the letter in Section 4 and 5) signify that they found the device substantially equivalent to predicates and thus determined it to be safe and effective for its stated intended use, based on the information provided and comparisons to established devices. It does not imply that a new clinical study with specific acceptance criteria was performed or reported in this document.

    If such a study were conducted, its results would typically be included as part of the submission, detailing methods, endpoints, statistical analyses, and performance metrics. However, for a 510(k) requiring substantial equivalence, extensive comparative performance data against specific acceptance criteria is often not required if equivalence can be demonstrated through other means (e.g., material, design, and intended use similarity to predicate devices).

    Ask a Question

    Ask a specific question about this device

    K Number
    K955486

    Validate with FDA (Live)

    Device Name
    BIO-ANCHOR
    Manufacturer
    Date Cleared
    1996-02-29

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1