(195 days)
The BEAR® (Bridge Enhanced ACL Repair) Implant is a bovine extracellular matrix collagen-based implant for treatment of anterior cruciate ligament (ACL) injuries. The BEAR® Implant is indicated for skeletally-mature patients at least 14 years of age with a complete rupture of the ACL, as confirmed by MRI. Patients must have an ACL stump attached to the tibia to construct the repair.
The BEAR® Implant (22 mm in diameter and 45mm in length) is cylindrical in shape and comprised of collagen and extracellular matrix derived from bovine connective tissue, which has been cleaned, disinfected and processed by a proprietary manufacturing method. The implant has been terminally sterilized by electron-beam irradiation and is intended to be used with up to 10 ml of autologous blood drawn during the surgical implantation procedure. The BEAR® Implant stabilizes the blood in the gap between the torn ligament ends. The BEAR® Implant is resorbed within 8 weeks and replaced with a fibrovascular repair tissue.
The provided text describes the clinical study (BEAR II) conducted to support the safety and effectiveness of the BEAR® (Bridge-Enhanced ACL Repair) Implant. It is important to note that this document is a regulatory submission for a medical device and not a study validating an AI-powered diagnostic device. Therefore, many of the requested criteria related to AI/ML device performance (e.g., ground truth establishment for training, expert consensus for test sets, MRMC studies, standalone algorithm performance) are not applicable to this submission.
However, I can extract and present the acceptance criteria and the study results relevant to the BEAR® Implant's clinical performance, as described in the provided text.
Acceptance Criteria and Reported Device Performance for BEAR® Implant
The acceptance criteria for the BEAR® Implant are primarily demonstrated through non-inferiority testing against a control (ACLR – ACL Reconstruction with autograft) for primary effectiveness endpoints and superiority/non-inferiority for secondary effectiveness endpoints. Safety parameters are also assessed for comparable rates between the BEAR group and the control group.
1. Table of Acceptance Criteria and Reported Device Performance
The core of the clinical acceptance is based on the primary effectiveness endpoints of the BEAR II study: IKDC Subjective Score and Instrumented AP Knee Laxity.
| Acceptance Criteria (Endpoint) | Null Hypothesis (for non-inferiority) | Reported Device Performance (Mean ± SD) | p-value | Meets Criteria? |
|---|---|---|---|---|
| Primary Effectiveness Endpoints (24 Months Post-Surgery) | ||||
| IKDC Patient Reported Score | True difference in means (BEAR - Control) ≤ -11.5 (clinically significant difference) | BEAR: 88.6 ± 13.4 | Control: 84.6 ± 13.3 | <0.001 |
| Mean Difference (95% CI) | 4.03 (-1.55, 9.61) | |||
| KT Instrumented AP Knee Laxity (mm) (Injured Knee - Non-Injured Knee) | True difference in means (BEAR - Control) ≥ 2.0 mm (clinically significant difference) | BEAR: 1.7 ± 3.2 | Control: 1.8 ± 2.8 | 0.001 |
| Mean Difference (95% CI) | -0.10 (-1.45, 1.25) | |||
| Safety Endpoints (Selected at 24 Months) | ||||
| Deep Joint Infection/Incision and Drainage of Deep Surgical Site Infection | Rates are equal between groups (null hypothesis in p-value calculation) | BEAR: 0% (0/65) | Control: 0% (0/35) | 1.000 |
| Evidence of Graft or Implant Rejection | Rates are equal between groups | BEAR: 0% (0/65) | Control: 0% (0/35) | 1.000 |
| Graft or Repair Failure | Rates are equal between groups | BEAR: 13.8% (9/65) | Control: 5.7% (2/35) | 0.320 |
| Additional Surgical Procedures Required on Study Knee | Rates are equal between groups | BEAR: 12.3% (8/65) | Control: 11.4% (4/35) | 1.000 |
Note: The table focuses on primary clinical endpoints. Other non-clinical (bench, animal) acceptance criteria are mentioned in the text for material characteristics, sterilization, and basic performance.
2. Sample Size and Data Provenance
- Test Set (Clinical Study - BEAR II):
- Sample Size:
- Modified Intent-to-Treat (mITT) population (primary analysis): 104 subjects (73 in BEAR group, 31 in Control group).
- As-Treated (AT) population (safety analysis): Same as mITT (65 in BEAR group, 35 in Control group after accounting for those who didn't undergo surgery and different N for some comparisons).
- Data Provenance: The BEAR II study was conducted at "one U.S. site" (United States). It was a prospective, randomized controlled trial.
- Sample Size:
3. Number of Experts for Ground Truth and Qualifications
- Not Applicable: This is a clinical trial for a medical implant, not an AI/ML diagnostic device requiring expert annotation for ground truth. Ground truth for clinical outcomes like IKDC score, laxity, safety events (e.g., re-tear, infection) is established directly through clinical assessments, patient-reported outcomes, and objective measurements in a controlled study environment.
- Clinicians Involved: Three surgeons performed the procedures in the BEAR II study. Their specific qualifications (e.g., years of experience) are not detailed in the provided text.
4. Adjudication Method for the Test Set
- Not Applicable in the traditional sense of AI/ML ground truth adjudication: Clinical trials follow detailed statistical analysis plans (SAPs) for handling data, missing values (e.g., multiple imputation used for primary effectiveness endpoints), and endpoint assessment. There isn't a "ground truth adjudication" process by multiple experts as would be seen for image annotations. Clinical events (e.g., graft failure) are defined by specific clinical criteria (e.g., pivot shift exam, Lachman exam, MRI findings, need for revision surgery).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No: An MRMC comparative effectiveness study is typically performed for imaging-based diagnostic devices to assess how AI assistance affects human reader performance. This study is a direct clinical comparison of two surgical interventions, not an imaging diagnostic device.
6. Standalone Performance
- Not Applicable: The BEAR® Implant is a physical medical device (implant) used in a surgical procedure, not a standalone algorithm. Its performance is intrinsically linked to its surgical implantation and the patient's biological response, rehabilitation, and long-term outcomes. The clinical study is the standalone performance evaluation in humans.
7. Type of Ground Truth Used
- Clinical Outcomes and Objective Measurements: The ground truth for the effectiveness and safety of the BEAR® Implant was established through:
- Patient-Reported Outcomes (PROs): IKDC Subjective Score, KOOS domains, ACL Return-to-Sport Index (RSI) score.
- Objective Clinical Measurements: KT-instrumented AP knee laxity, hamstring strength (hand-held dynamometer), hamstring to quadriceps ratio.
- Clinical Event Rates: Graft/repair failure (defined by objective examination findings, MRI, or revision surgery), infections, need for additional surgical procedures.
- Histopathology and Biomechanical Testing (Animal Studies): Used for design validation, resorption rates, and mechanical properties.
8. Sample Size for the Training Set
- Not Applicable: This device is not an AI/ML algorithm that requires a "training set" in the computational sense. The "training" for optimizing the device and procedure occurred during pre-clinical development (bench and animal testing) and an early feasibility clinical study (BEAR I), which precedes the pivotal BEAR II study.
- BEAR I (early feasibility): "x" patients, "x" of whom received the BEAR® Implant (specific numbers redacted as (b)(4) in the document).
9. How the Ground Truth for the Training Set Was Established
- Not Applicable (see point 8). For pre-clinical development:
- Bench Testing: Utilized standardized test methods with predefined acceptance criteria for material properties (e.g., collagen content, DNA content, density) and functional integrity (e.g., blood absorption, Keith needle test).
- Animal Studies (VIV-003, VIV-004): Ground truth was established through direct observation (gross necropsy), histological examination, and biomechanical testing of harvested tissues according to study protocols.