510(k) Data Aggregation

    K Number
    K210264

    Validate with FDA (Live)

    Date Cleared
    2021-07-30

    (179 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC Catheter is indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. The maximum infusion flow rate is 5mL/second for power injection of contral venous pressure monitoring, it is recommended that a catheter lumen of 20gauge or larger be used.

    Device Description

    A family of peripherally inserted central catheters made from specially formulated and processed medical grade materials. Each PowerPICC® catheter has a kink resistant, reverse tapered design. Catheters are packaged in a tray with accessories for reliable long (greater than 30 days) or short (less than 30 days) term vascular access.

    AI/ML Overview

    The provided document is a 510(k) summary for the PowerPICC Catheter. It focuses on demonstrating substantial equivalence to a predicate device through various performance tests, rather than providing details of a study with acceptance criteria for an AI/ML powered device. Therefore, much of the requested information (sample sizes, ground truth establishment, MRMC studies, standalone performance, training set details) is not applicable or available in this document.

    Here's the relevant information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful completion of the listed tests, indicating that the device performed as expected under the specified conditions. The document states that the performance tests were conducted "in determining substantial equivalence" and that "the subject PowerPICC Family Catheters have been demonstrated to be substantially equivalent to the cited predicate and reference devices." This implies that the device met the acceptance criteria derived from these tests.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Biocompatibility TestingFree from biological hazard per ISO 10993-1.Confirmed free from biological hazard; health-based risk assessment performed.
    Leak TestCatheter assembly will not leak when distal end is occluded.Test confirmed non-leakage.
    Dimensional TestCompliance with dimensional specifications (OD/ID for single lumen, OD/lumen area for dual).Ensured compliance with dimensional specifications.
    Implantable LengthCompliance with dimensional specifications.Ensured compliance with dimensional specifications.
    Extension Leg LengthCompliance with dimensional specifications.Confirmed compliance with dimensional specifications.
    Catheter Collapse TestCatheter will not collapse under a vacuum during aspiration.Demonstrated non-collapse under vacuum.
    Burst TestCatheter burst pressure exceeds peak pressure at maximum flow conditions (distal end occluded).Confirmed burst pressure exceeds peak pressure.
    Hydraulic Catheter Burst TestCatheter burst pressure exceeds peak pressure at maximum flow conditions (distal end occluded).Confirmed burst pressure exceeds peak pressure.
    Power Injection ConditioningCatheter does not leak or burst during power injections at maximum indicated flow rate.Confirmed no leakage or burst.
    Gravity FlowAcceptable gravity flow performance for a full-length catheter.Demonstrated acceptable gravity flow.
    Luer to Extension Leg Tensile TestPeak tensile force exceeds minimum peak tensile force.Demonstrated peak tensile force exceeds minimum.
    Extension Leg to Trifurcation Tensile TestPeak tensile force exceeds minimum peak tensile force.Demonstrated peak tensile force exceeds minimum.
    Trifurcation to Shaft Tensile TestPeak tensile force exceeds minimum peak tensile force.Demonstrated peak tensile force exceeds minimum.
    Shaft Tensile TestEvaluate maximum catheter strain and modulus at break.Evaluated.
    RadiopacityDemonstrates catheter radio-detectability.Demonstrated radio-detectability.
    Tip TensilePeak tensile force exceeds minimum peak tensile force.Demonstrated peak tensile force exceeds minimum.
    Suture Wing Integrity TestMeasures maximum force suture wing can withstand prior to break.Measured maximum force.
    Priming VolumeMeasures volume required to prime a full-length catheter.Measured priming volume.
    OD SwellCatheter does not swell beyond twice the labeled OD during power injection.Confirmed no excessive swelling.
    Tip Stability TestCatheter tip remains in the same orientation during power injection.Confirmed tip stability.
    Guidewire Drag TestGuidewire can be removed without difficulty.Ensured guidewire removal without difficulty.
    Luer TestingLuer connectors meet requirements for Stress Cracking, Resistance to Separation (Axial Load, Unscrewing, Overriding).Confirmed compliance with luer connector requirements.
    ParticulateComplies with USP 788 for particulate matter.Demonstrated compliance with USP 788.

    2. Sample size used for the test set and the data provenance
    The document does not specify the sample sizes used for each test. It refers to "testing" in general. The data provenance is not mentioned, as these are in vitro and in vivo performance tests on the device itself, rather than studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. These are physical and biological performance tests, not clinical studies requiring expert ground truth for interpretation of outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as this is for physical and biological performance testing, not interpretation of data by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI/ML device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" for these tests are the established scientific and engineering principles, material properties, and regulatory standards (e.g., ISO 10993-1, ISO 10555-1, ASTM F640-12, ISO 10555-3, FDA Guidance on Premarket Notification for Intravascular Catheters, ISO 80369-7, USP 788). The device's performance is measured against these defined benchmarks.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/ML device.