Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181221

    Validate with FDA (Live)

    Date Cleared
    2018-11-30

    (207 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.

    Device Description

    The BD PhaSeal™ Optima Closed System Transfer Device (CSTD) is a sterile single-use closed system drug transfer device intended for the reconstitution and transfer of antineoplastic or other hazardous drugs in the healthcare setting. The system is comprised of four devices-Protector, Injector, Connector, and Infusion Adapter. The closed transfer of liquid drugs takes place through a double membrane utilizing self-sealing elastomeric membranes that are tightly fitted together through the collet-style fitting on each of the BD PhaSeal™ Optima system devices. During use, the single lumen cannula of the Injector perforates the double membranes for the transfer of liquids. When the cannula is retracted the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, leaks and spills. The BD PhaSeal™ Optima system prevents microbial ingress for up to 168 hours. Performance of the self-sealing membrane has been substantiated up to 10 penetrations.

    AI/ML Overview

    The BD PhaSeal™ Optima Closed System Transfer Device is an airtight and leakproof closed system drug transfer device (CSTD) designed to prevent environmental contaminants from entering the system and to prevent drug vapor concentrations from escaping, thus minimizing exposure to drug vapor, aerosols, and spills. It also prevents microbial ingress for up to 168 hours.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance TestAcceptance Criteria / StandardReported Device Performance
    Microbial ingress testingPer FDA guidance document, "Intravascular Administration Sets Premarket Notification Submissions [510(k)]" (presumably demonstrating no microbial ingress for up to 168 hours).PASS
    Airtight connectionsUtilizing TiCl4 vapor test (specific acceptance criteria for vapor leakage not detailed, but likely related to no detectable vapor escape).PASS
    Leakproof connectionsUtilizing Fluorescein test (specific acceptance criteria for liquid leakage not detailed, but likely related to no detectable liquid escape).PASS
    Protector assembly force and removal force testingNot explicitly stated, but implies forces within acceptable ranges for proper device function and user interaction.PASS
    Protector expansion chamber burst testingNot explicitly stated, but implies the chamber can withstand internal pressure without bursting.PASS
    Packaging integrity and shelf life testingPer ASTM D4169-16 (Performance Testing of Shipping Containers), ASTM F88/F88M-15 (Seal Strength), ASTM F2096-11 (Gross Leak Detection). Acceptance criteria would involve maintaining package integrity and sterility over the shelf life.PASS
    Hazardous vapor containmentUsing alcohol residuals level criterion (specific residual level not detailed, but implies levels below a hazardous threshold).PASS
    Ethylene Oxide (EO) residuals testingPer ISO 10993-7 (Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals). Acceptance criteria would be below specified limits.PASS
    EO sterilization validationPer ISO 11135 (Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process). Acceptance criterion: Sterility Assurance Level (SAL) of 10-6.PASS
    Pyrogenicity - bacterial endotoxin test (LAL)Acceptance criteria not explicitly stated, but implies endotoxin levels below a specified threshold.PASS
    Biocompatibility: CytotoxicityPer ISO 10993-5. Acceptance criteria: non-cytotoxic.All materials are biocompatible.
    Biocompatibility: SensitizationPer ASTM F2148 and ISO 10993-10. Acceptance criteria: non-sensitizing.All materials are biocompatible.
    Biocompatibility: Intracutaneous ReactivityPer ISO 10993-10 and USP 39-NF 34 <88>. Acceptance criteria: no significant irritation or reactivity.All materials are biocompatible.
    Biocompatibility: Acute Systemic ToxicityPer ISO 10993-11. Acceptance criteria: no systemic toxicity.All materials are biocompatible.
    Biocompatibility: Pyrogenicity (material-mediated rabbit pyrogen)Per USP <151> and ISO 10993-11. Acceptance criteria: no pyrogenic response.All materials are biocompatible.
    Biocompatibility: HemolysisPer ISO 10993-4. Acceptance criteria: non-hemolytic.All materials are biocompatible.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document does not explicitly state the specific sample sizes used for each performance test. It only indicates that "Performance testing was conducted" and that the "Results of these tests demonstrate that the subject device is substantially equivalent to the predicate device." The data provenance is not specified beyond being generated by the device manufacturer, Becton, Dickinson and Company. The studies appear to be retrospective in nature, performed by the manufacturer to demonstrate product performance for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Not applicable. This device is a medical device (Closed System Transfer Device), not an AI-powered diagnostic or assistive tool that would require expert-established ground truth for its performance assessment. The tests conducted are laboratory-based and measure physical, chemical, and biological properties according to established standards.

    4. Adjudication Method for the Test Set:

    Not applicable. See point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    Not applicable. This is not an AI-powered device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This is not an AI-powered device. The "standalone" performance here refers to the device's inherent mechanical and material performance.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is established by the pre-defined acceptance criteria set by recognized international (ISO, ASTM) and national (FDA, USP) standards and guidance documents. For example, for microbial ingress, the ground truth is the absence of microbial contamination as defined by the FDA guidance. For airtight connections, the "ground truth" is the non-detection of vapor using the TiCl4 test. For biocompatibility, the ground truth is that the materials do not elicit adverse biological responses as defined by ISO 10993 standards.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1