Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K140591

    Validate with FDA (Live)

    Date Cleared
    2014-05-14

    (68 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PhaSeal system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal system also prevents microbial ingress.

    Device Description

    The PhaSeal® System is a sterile single-use closed system drug transfer device. The closed transfer of liquid takes place through a double membrane utilizing self-sealing elastomeric membranes, tightly fitted together through a bayonet fitting on all PhaSeal components. A single lumen cannula perforates the double membranes for the transfer of liguid. When the cannula is retracted the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols and spills and also minimizing the risk of microbial contamination.

    The PhaSeal Injector is one component of the PhaSeal system. It is a luer device adaptor that may be fitted to a syringe or IV tubing. It is used to gain dry and leak-proof access to a drug container or administration device which has been sealed with a PhaSeal Protector or Connector. Liquid transfer takes place through tightly fitting elastomeric double membranes to minimize exposure to potentially hazardous drug aerosols and spills occurring during reconstitution, administration and disposal processes.

    AI/ML Overview

    This document describes the performance and acceptance criteria for the BD PhaSeal Closed System Transfer Device - Injector, as summarized from the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines design verification tests to demonstrate substantial equivalence to the predicate device. The acceptance criteria are implicit in the "Equivalent to Predicate" or "Per ISO 10993" results for each characteristic.

    CharacteristicTest PerformedAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityPer ISO 10993Compliance with ISO 10993Per ISO 10993
    Membrane LeakageDye Leak TestLeakage performance equivalent to predicateEquivalent to Predicate
    Membrane FragmentationFragmentation TestFragmentation performance equivalent to predicateEquivalent to Predicate

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the "Dye Leak Test" or "Fragmentation Test." It mentions "Design Verification tests were performed based on the risk analysis performed," suggesting a controlled testing environment, but no specific sample numbers are given. The data provenance is internal to BD Medical - Medical Surgical Systems, likely conducted in a laboratory setting for device validation. It is a prospective study in the sense that the tests were performed specifically to verify the modified device's performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the given text. The tests appear to be objective performance tests (dye leak, fragmentation) rather than expert-dependent assessments requiring ground truth establishment by human experts. Biocompatibility (ISO 10993) is likely assessed by trained laboratory personnel in accordance with the standard, not necessarily clinical "experts" in the sense of radiologists or similar.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests performed (Biocompatibility, Membrane Leakage, Membrane Fragmentation) are objective measurements or adherence to standards, not subjective assessments requiring adjudication by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This information is not applicable. The device is a Closed System Transfer Device (CSTD), a physical medical device, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study with human readers and AI assistance is irrelevant to its purpose and performance evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This information is not applicable. The device is a physical product and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used:

    For the performance tests:

    • Biocompatibility: The "ground truth" would be adherence to the established scientific and regulatory standards outlined in ISO 10993 for biological evaluation of medical devices.
    • Membrane Leakage & Fragmentation: The "ground truth" or reference for performance is the predicate device (BD PhaSeal® Connector, Injector, Protector - K123213). The goal was to prove "equivalence," meaning the modified device should perform as well as the predicate device in these specific functional tests.

    8. The Sample Size for the Training Set:

    This information is not applicable. The described study is a design verification for a physical medical device, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as above (no training set).

    Ask a Question

    Ask a specific question about this device

    K Number
    K130197

    Validate with FDA (Live)

    Date Cleared
    2013-02-27

    (30 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PhaSeal system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor contractions outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal system also prevents microbial ingress.

    Device Description

    The PhaSeal® System is a sterile single-used closed system drug transfer device. The closed transfer of liguid takes place through a double membrane utilizing self-sealing elastomeric membranes, tightly fitted together through a bayonet fitting on all PhaSeal components. A single lumen cannula perforates the double membranes for the transfer of liquid. When the cannula is retracted the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols and spills and also minimizing the risk of microbial contamination.

    The PhaSeal Protector is one component of the PhaSeal system. It is a drug vial adapter that is fitted to the drug vial and is used as a docking station between the drug vial and the BD PhaSeal Injector. In addition the Protector equilibrates the pressure difference which occurs when fluid or air is added or removed from the drug vial. Liquid transfer takes place through tightly fitting elastomeric double membranes to minimize exposure to potentially hazardous drug aerosols and spills occurring during reconstitution, administration and disposal processes.

    AI/ML Overview

    The provided 510(k) summary (K130197) does not contain information related to acceptance criteria, device performance, or any studies involving human readers or AI algorithms.

    This document describes a medical device called the "BD PhaSeal Closed System Transfer Device - P55," which is a sterile, single-use closed system drug transfer device. The 510(k) summary focuses on the device's description, intended use, and substantial equivalence to a predicate device (BD PhaSeal® Connector, Injector, Protector - K123213).

    Therefore, I cannot provide the requested information, including:

    1. A table of acceptance criteria and the reported device performance: This information is not present in the provided text. The document describes the function of the device (airtight, leakproof, prevents transfer of contaminants, etc.) but does not quantify performance against specific criteria.
    2. Sample size used for the test set and the data provenance: Not applicable as no such study is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no such study is described.
    4. Adjudication method for the test set: Not applicable as no such study is described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not applicable as this is a physical medical device, not an AI or imaging diagnostic tool.
    6. If a standalone performance study was done: While there would undoubtedly be internal testing by the manufacturer for a device of this type (e.g., leak tests, microbial ingress tests), the provided 510(k) summary does not detail these studies' acceptance criteria, methodologies, or results. The summary focuses on the regulatory submission process.
    7. The type of ground truth used: Not applicable as no ground truth for diagnostic or interpretative accuracy is relevant or discussed for this device.
    8. The sample size for the training set: Not applicable as this is not a machine learning device.
    9. How the ground truth for the training set was established: Not applicable as this is not a machine learning device.

    In summary, the provided text details a regulatory submission for a physical medical device and does not include the type of performance study data or acceptance criteria typically associated with AI/ML devices or diagnostic tools.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123213

    Validate with FDA (Live)

    Date Cleared
    2013-01-09

    (86 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PhaSeal system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal system also prevents microbial ingress.

    Device Description

    The PhaSeal® System is a sterile single-used closed system drug transfer device. The closed transfer of liquid takes place through a double membrane utilizing self-sealing elastomeric membranes, tightly fitted together through a bayonet fitting on all PhaSeal components. A single lumen cannula perforates the double membranes fro the transfer of liquid. When the cannula is retracted the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols and spills and also minimizing the risk of microbial contamination.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    Item#Performance Specification: Acceptance CriteriaStatus of BD PhaSeal® System (Reported Performance)
    1Leakproof ConnectionsNo Leaks (Fluorescein Test)
    2Airtight ConnectionsNo Visible Smoke (TiCl4 Test)
    3Microbial IngressNo Ingress at the Protector or Connector

    Description of the Study Proving Acceptance Criteria

    The study described is a series of non-clinical tests designed to demonstrate that the BD PhaSeal® Closed System Drug Transfer Device meets the specified performance criteria for leakproofness, airtightness, and microbial ingress. The summary states, "As there is no change to the subject device in comparison to the predicate devices, the performance data provided represent the performance of both the predicate and subject device of this 510(k)." The tests were conducted to substantiate its use as a Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System (ONB product code).

    Details on the Tests:

    • Leakproof Connections: Tested using a Fluorescein Test.
    • Airtight Connections: Tested using a TiCl4 Test (which detects vapor/smoke).
    • Microbial Ingress: Tested to ensure no ingress at the Protector or Connector.

    Additional Information Not Present in the Document:

    The provided 510(k) summary focuses on demonstrating substantial equivalence through technological characteristics and a summary of performance tests. It does not contain the following information typically found in detailed study reports:

    1. Sample size used for the test set and the data provenance: Not specified. The document only mentions "additional tests referenced in the table" without detailing sample sizes or the origin (country, retrospective/prospective) of the data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for these types of physical performance tests (leak, airtight, microbial ingress). Ground truth is established by the test method itself, not expert consensus.
    3. Adjudication method for the test set: Not applicable for these performance tests. The outcome is typically a pass/fail based on direct observation (e.g., presence/absence of leaks, smoke, or microbial growth).
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a medical device for drug transfer, not an imaging or diagnostic AI product, so MRMC studies are not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI/algorithm-driven device.
    6. The type of ground truth used: The ground truth for these tests is defined by the objective results of the specified testing methodologies (Fluorescein Test, TiCl4 Test, and microbial ingress testing). For example, "no leaks" is the ground truth from the fluorescein test.
    7. The sample size for the training set: Not applicable. This device does not use machine learning or require a training set.
    8. How the ground truth for the training set was established: Not applicable. This device does not use machine learning or require a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1