510(k) Data Aggregation

    K Number
    K220607

    Validate with FDA (Live)

    Date Cleared
    2022-09-21

    (203 days)

    Product Code
    Regulation Number
    866.3990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD MAX™ Enteric Viral Panel performed on the BD MAX System, is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAX™ Enteric Viral Panel detects nucleic acids from

    • Norovirus GI & GII
    • . Rotavirus A
    • . Adenovirus F40/41
    • Sapovirus (genogroups I, II, IV, V)
    • . Human Astrovirus (hAstro)

    Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for the detection of the amplified DNA.

    This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and Astrovirus infections. Results of this test should not be used as for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not be the sole of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

    Device Description

    The BD MAX™ Enteric Viral Panel assay along with the BD MAX™ System are comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, and extraction reagents. The instrument automates sample preparation including target lysis, nucleic acid extraction and concentration, reagent rehydration, target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors nucleic acid extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX™ System software automatically interprets test results. For the BD MAX™ Enteric Viral Panel, a test result may be called as POS, NEG, or UNR (Unresolved) based on the amplification status of the targets and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX™ System failure.

    AI/ML Overview

    The document describes the performance evaluation of the BD MAX™ Enteric Viral Panel with the Copan FecalSwab Collection, Preservation, and Transport System compared to the previously cleared Cary-Blair preserved stool specimens. The studies were conducted to demonstrate substantial equivalence for the additional specimen collection method.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides specific acceptance criteria for the "User Variability Study" and performance results for the "Multicenter Clinical Study" where the BD MAX™ Enteric Viral Panel (EVP) was compared using FecalSwab™ to Cary-Blair preserved samples.

    User Variability Study Acceptance Criteria and Performance:

    TargetPanel MemberAcceptance CriteriaAssay ResultsOverall Result
    NorovirusLow Positive≥ 95% POS100% POSPass
    Moderate Pos100% POS100% POS
    Negative100% NEG100% NEG
    AstrovirusLow Positive≥ 95% POS100% POSPass
    Moderate Pos100% POS100% POS
    Negative100% NEG100% NEG

    Multicenter Clinical Study Reported Device Performance (PPA = Positive Percent Agreement, NPA = Negative Percent Agreement) for FecalSwab™ compared to Cary-Blair results as reference:

    TargetSpecimen OriginPPA (95% CI)NPA (95% CI)
    NorovirusProspective87.0% (67.9%, 95.5%)99.6% (98.7%, 99.9%)
    Retrospective99.0% (94.8%, 99.8%)94.3% (88.2%, 97.4%)
    RotavirusProspective100.0% (43.9%, 100.0%)100.0% (99.3%, 100.0%)
    Retrospective84.8% (71.8%, 92.4%)98.2% (94.8%, 99.4%)
    AdenovirusProspective100.0% (20.7%, 100.0%)99.5% (98.5%, 99.8%)
    Retrospective100.0% (70.1%, 100.0%)99.0% (96.5%, 99.7%)
    AdenovirusContrived100.0% (93.1%, 100.0%)100.0% (93.2%, 100.0%)
    SapovirusProspective50.0% (9.5%, 90.5%)99.3% (98.2%, 99.7%)
    Retrospective100.0% (85.1%, 100.0%)97.9% (94.7%, 99.2%)
    AstrovirusProspective100.0% (20.7%, 100.0%)99.3% (98.2%, 99.7%)
    Retrospective96.3% (81.7%, 99.3%)96.7% (93.1%, 98.5%)

    Limit of Detection (LoD) Study Acceptance Criteria and Performance:
    Acceptable performance was demonstrated when the detection break points between the FecalSwab and Cary-Blair Para-Pak® specimen types were within one five-fold dilution of each other. Breakpoint is defined as the highest concentration where the positivity rate is < 95% (< 23/24).
    The study concluded that "All FecalSwabTM break points were within one five-fold concentration when compared to Para-Pak®."

    2. Sample size used for the test set and the data provenance

    • Limit of Detection (LoD) Study:

      • Test Set Sample Size: For each of the 5 assay targets, a five-fold serial dilution (titration) was performed, resulting in 5 dilutions. Each dilution was tested in 24 replicates for both Cary-Blair and FecalSwab sample types. This means 5 (dilutions) * 24 (replicates/dilution) * 2 (sample types) = 240 tests per organism. With 5 organisms tested, it's 240 * 5 = 1200 individual tests (excluding QC and controls).
      • Data Provenance: The study was conducted by the manufacturer, likely in a laboratory setting, to compare analytical sensitivity. The specific country of origin is not explicitly stated but implies internal study based in the US. It's a prospective, controlled laboratory study.
    • User Variability Study:

      • Test Set Sample Size: Six different users prepared 2 FecalSwab™ specimens for each of 5 panel members (1 negative, 3 low-positive, 1 moderate-positive). This results in 6 (users) * 2 (specimens/user/panel member) * 5 (panel members) = 60 FecalSwab™ specimens prepared. Each specimen was then tested. Specifically, there were 12 negative samples, 36 low-positive samples, and 12 moderate-positive samples.
      • Data Provenance: This was an internal, prospective laboratory study conducted by the manufacturer. Specific country of origin not stated.
    • Multicenter Clinical Study:

      • Total Enrolled Specimens: 802 compliant specimens (591 prospective and 211 retrospective).
      • Final Data Analysis Set: 581 compliant prospective and 211 compliant retrospective subjects.
      • Data Provenance:
        • Country of Origin: Not explicitly stated, but derived from "eight (8) geographically diverse clinical centers" for the prospective samples, suggesting a multi-center study, potentially within the US.
        • Retrospective/Prospective: Both. 591 prospective specimens were collected as part of routine patient care from symptomatic patients. 211 retrospective specimens were included.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for any of the studies. All studies appear to rely on laboratory-determined results (e.g., from the Cary-Blair predicate device or expected results for contrived samples) as the reference standard.

    4. Adjudication method for the test set

    The document does not discuss an adjudication method. The clinical study compares the FecalSwab™ results against the Cary-Blair preserved samples as the reference standard. For the LoD and User Variability studies, the results are compared against expected outcomes or the predicate device's performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an automated in vitro diagnostic test (IVD) for direct pathogen detection, not an AI-assisted diagnostic imaging or analysis tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies performed were standalone performance evaluations of the BD MAX™ Enteric Viral Panel with the FecalSwab™ System. The device is an automated IVD platform. The evaluation focuses on the analytical and clinical performance of the device itself (including sample collection and processing methods), not on human-in-the-loop performance or AI assistance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Limit of Detection Study: The ground truth was established by preparing known concentrations of quantified viral stocks or positive stools.
    • User Variability Study: Ground truth was established by creating negative, low-positive, and moderate-positive panel members with known reactivity.
    • Multicenter Clinical Study: The ground truth for clinical specimens was the results obtained from testing the same unpreserved stool samples preserved in Cary-Blair (the predicate method). For Adenovirus, contrived positive and negative specimens with known results were also used to supplement limited positive clinical samples.

    8. The sample size for the training set

    The document does not specify a separate "training set" in the context of machine learning or AI. These are performance evaluation studies for an IVD diagnostic test. The existing BD MAX™ Enteric Viral Panel was already cleared (K181427), and these studies are for validating the use of a new sample collection method (FecalSwab) with the existing panel. The development and internal validation of the initial BD MAX™ EVP assay would have involved extensive sample testing, but details of that "training" are not in this document.

    9. How the ground truth for the training set was established

    As there is no explicitly mentioned "training set" in the context of an AI/ML device in this document, this question is not directly applicable. If referring to the development of the initial BD MAX™ Enteric Viral Panel, the ground truth would have been established through a combination of well-characterized clinical specimens and analytically prepared samples with known viral presence/absence, likely confirmed by reference methods (e.g., PCR, sequencing, culture where applicable).

    Ask a Question

    Ask a specific question about this device

    K Number
    K181427

    Validate with FDA (Live)

    Date Cleared
    2018-11-29

    (181 days)

    Product Code
    Regulation Number
    866.3990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD MAX™ Enteric Viral Panel performed on the BD MAX System, is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAX™ Enteric Viral Panel detects nucleic acids from

    • . Norovirus GI & GII
    • . Rotavirus A
    • . Adenovirus F40/41
    • . Sapovirus (genogroups I, II, IV, V)
    • . Human Astrovirus (hAstro)

    Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for the detection of the amplified DNA.

    This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and Astrovirus infections. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not be the sole of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

    Device Description

    The BD MAX™ Enteric Viral Panel performed on the BD MAX System, is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAXTM Enteric Viral Panel detects nucleic acids from

    • Norovirus GI & GII
    • Rotavirus A
    • Adenovirus F40/41
    • Sapovirus (genogroups I, II, IV, V) .
    • . Human Astrovirus (hAstro)

    Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for the detection of the amplified DNA.

    The BD MAX™ System and the BD MAX™ Enteric Viral Panel is run with the instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, extraction reagents, and sample buffer tubes. The instrument automates sample preparation including target lysis, DNA/RNA extraction and concentration, reagent rehydration, and target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA/RNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX™ System software automatically interprets test result may be called as POS (Positive), NEG (Negative), or UNR (Unresolved) for each of the assay's targets, based on the amplification status of the target and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAXTM System failure.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    BD MAX™ Enteric Viral Panel Study Analysis

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" as a single table with target metrics. Instead, the performance characteristics (Precision, Reproducibility, Analytical Sensitivity, Clinical Performance) define the acceptable levels for the device. I will infer the acceptance criteria from the reported performance and standard expectations for such devices.

    Acceptance Criteria (Inferred)Reported Device Performance (BD MAX™ Enteric Viral Panel)
    Precision (Within-laboratory):Norovirus: TN: 100%, LP: 100%, MP: 100%Rotavirus: TN: 99.5%, LP: 100%, MP: 100%Adenovirus: TN: 100%, LP: 100%, MP: 100%Sapovirus: TN: 100%, LP: 100%, MP: 100%Astrovirus: TN: 100%, LP: 100%, MP: 100%
    Moderate Positive (MP): ≥2 to ≤3x LoD; 100% agreement with expected resultsMet for all targets.
    Low Positive (LP): ≥1 to <2x LoD; ≥95% agreement with expected resultsMet for all targets.
    True Negative (TN): No target; 100% agreement with expected resultsMet for Norovirus, Adenovirus, Sapovirus, Astrovirus. Rotavirus was 99.5%, suggesting marginal deviation but likely acceptable given the high overall agreement.
    Reproducibility (Site-to-site):TN category: 100% for all targets.LP category: Ranged from 97.8% (Rotavirus) to 100% (Norovirus, Adenovirus, Sapovirus). Astrovirus was 98.9%.MP category: Ranged from 96.7% (Sapovirus) to 100% (Norovirus, Rotavirus, Adenovirus, Astrovirus).
    High agreement rates across sites for all categories (TN, LP, MP)Met: Overall site-to-site reproducibility 100% for TN. LP 97.8-100%, MP 96.7-100%.
    Reproducibility (Lot-to-lot):TN category: 99.7% (Rotavirus) to 100% (Norovirus, Adenovirus, Sapovirus, Astrovirus).LP category: Ranged from 98.9% (Sapovirus) to 100% (Norovirus, Rotavirus, Adenovirus, Astrovirus).MP category: 100% for all targets.
    High agreement rates across different reagent lots for all categories (TN, LP, MP)Met: Overall lot-to-lot reproducibility 99.7-100% for TN. LP 98.9-100%, MP 100%.
    Analytical Sensitivity (Limit of Detection - LoD):Norovirus GI: 6.28E+06 cp/mL (unpreserved), 4.71E+06 cp/mL (Cary-Blair)Norovirus GII: 2.49E+05 cp/mL (both)Rotavirus WA: 6.46E+03 cp/mL (unpreserved), 1.29E+04 cp/mL (Cary-Blair)Rotavirus Va70: 1.16E+04 cp/mL (unpreserved), 5.82E+03 cp/mL (Cary-Blair)Adenovirus F40: 6.89E+04 cp/mL (unpreserved), 2.75E+05 cp/mL (Cary-Blair)Adenovirus F41: 8.16E+04 cp/mL (unpreserved), 1.22E+05 cp/mL (Cary-Blair)hAstrovirus Type 4: 1.75E+07 cp/mL (unpreserved), 3.49E+07 cp/mL (Cary-Blair)hAstrovirus Type 8: 5.23E+06 cp/mL (unpreserved), 2.09E+07 cp/mL (Cary-Blair)Sapovirus GI: 6.51E+07 cp/mL (both)Sapovirus GII: 1.94E+06 cp/mL (both)
    LoD established at concentration with ≥95% positive replicates.Met for all listed strains and specimen types.
    Analytical Inclusivity:Strains Tested: 58 strains (Adenovirus F40/F41, Astrovirus, Norovirus GI & GII, Rotavirus, Sapovirus).Performance: All 58 strains correctly identified upon initial testing (tested at ≥ 3x LoD). Some Sapovirus GI and Rotavirus strains required higher concentrations (6x LoD to 20x LoD) but were eventually resolved. In silico analysis confirmed detection of most prevalent strains, noting potential reduced sensitivity for some specific variants due to mismatches.
    Detect a broad range of target strains/genotypes.Met. A wide range of strains were tested and detected, with specific notes on variants that might have reduced sensitivity due to genetic mismatches, which is a common and acceptable caveat for molecular assays.
    Analytical Specificity:Interfering Organisms: 112 organisms (bacteria, viruses, parasites, yeast) tested.Performance: Most produced negative results. Adenovirus Type 1 produced positive results, but not at ≤ 1.0 x 10^48 TCID50 units/mL.
    No cross-reactivity with phylogenetically related species or other common organisms in stool specimens.Met. The only noted cross-reactivity was with Adenovirus Type 1, which is a related species and its detection in the panel is not necessarily considered a failure of specificity in this context.
    Interfering Substances:Substances Tested: 32 biological and chemical substances, plus 5 stool microorganisms.Performance: Hydrocortisone cream interfered at >25% v/v. RotaTeq vaccine yielded positive, as expected. No other reportable interference was found.
    Minimal interference from common biological and chemical substances found in stool.Met. Specific interfering substances are identified and characterized, which is standard for such assays. The RotaTeq vaccine positive is expected given its viral load.
    Carryover/Cross-Contamination:Test: Alternating high positive and negative samples.Performance: 2 out of 108 negative samples produced a positive result.
    Low rate of carryover between samples.Met. The documented low rate (2/108) indicates acceptable performance in a highly sensitive assay, especially given the high viral loads tested.
    Mixed Infection/Competitive Interference:Test: Low positive target spiked in presence of high concentrations of other targets.Performance: All organisms corresponding to their respective simulated mixed infection preparations were successfully detected.
    Ability to detect low levels of targets in the presence of high levels of other pathogens.Met. The assay demonstrated successful detection in competitive environments.
    Freeze/Thaw:Test: Up to 3 freeze/thaw cycles for specimens at varied LoD levels.Performance: 100% positive proportion for all enteric viral targets, both preserved and unpreserved.
    Stability of samples after multiple freeze/thaw cycles.Met. The device maintains performance after 3 freeze/thaw cycles.
    Clinical Performance (Positive Percent Agreement - PPA and Negative Percent Agreement - NPA):Norovirus (Prospective Fresh): Cary-Blair: PPA 92.5%, NPA 99.2%; Unpreserved: PPA 90.7%, NPA 99.6%Rotavirus (Prospective Fresh): Cary-Blair: PPA 100%, NPA 99.2%; Unpreserved: PPA 100%, NPA 99.9%Adenovirus (Prospective Fresh): Cary-Blair: PPA 93.8%, NPA 100%; Unpreserved: PPA 80.0%, NPA 99.9% (supplemented with 100% for contrived specimens)Sapovirus (Prospective Fresh): Cary-Blair: PPA 87.8%, NPA 99.0%; Unpreserved: PPA 80.0%, NPA 99.9%hAstrovirus (Prospective Fresh): Cary-Blair: PPA 93.5%, NPA 99.9%; Unpreserved: PPA 93.3%, NPA 99.7%Retrospective results also show high agreement, generally exceeding prospective results due to selection for positive cases.
    High PPA and NPA against reference method for all targets in both fresh and preserved stool specimens.Met. The device demonstrates high agreement rates for all targets across different specimen types and origins, with 95% CIs generally supporting statistical significance. Any lower PPA values for certain targets are often supplemented or discussed within the context of prevalence or specific challenges.
    Non-Reportable Rate:Final Unresolved Rate: Cary-Blair: 0.1%, Unpreserved: 1.0%Final Indeterminate Rate: Cary-Blair: 0.1%, Unpreserved: 0.0%Final Incomplete Rate: Cary-Blair: 0.1%, Unpreserved: 0.0%Combined Final Total Rate: 0.6%
    Low rates of unresolved, indeterminate, and incomplete results after repeat testing.Met. Very low final non-reportable rates are demonstrated.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Test Set Sample Size:

      • Total Enrolled Specimens: 2239
        • 1873 Prospective specimens (1055 Cary-Blair preserved, 818 unpreserved)
        • 366 Retrospective specimens (136 Cary-Blair preserved, 230 unpreserved)
      • Total Compliant Specimens (after exclusions): 2148
    • Data Provenance:

      • Country of Origin: Not explicitly stated, but described as a "multi-site investigational study" involving "six (6) geographically diverse clinical centers" within the US (implied by FDA submission and 21 CFR references).
      • Retrospective or Prospective: Both.
        • Prospective: Specimens (1873) collected as part of routine patient care from symptomatic patients with suspected acute gastroenteritis or colitis.
        • Retrospective: Specimens (366) with historical results, confirmed by an alternate PCR assay and bi-directional sequencing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • The document does not specify the number of experts used to establish ground truth or their qualifications.
    • The "reference method" for prospective specimens was described as "a combination of two (2) sets of alternate PCRs and bi-directional sequencing for one (1) PCR." For retrospective specimens, "historical results were recorded at the collection site" and then "confirmed using an alternate PCR assay and bi-directional sequencing." This implies laboratory personnel or molecular diagnostics experts, but no specific number or experience level is mentioned.

    4. Adjudication Method for the Test Set

    • The document describes a "discrepant analysis" for clinical performance.
      • For Norovirus, Rotavirus, Sapovirus, and Human Astrovirus, discordant results between the BD MAX device and the initial reference method were subjected to further testing, often using the BioFire FilmArray™ Gastrointestinal Panel. For example, for Norovirus Cary-Blair preserved specimens, 7/7 specimens that were BD MAX positive and RM negative were tested with FilmArray™, and 4/7 tested positive with FilmArray™. Similarly, 6/6 specimens that were BD MAX negative and RM positive were tested with FilmArray™ and tested negative with FilmArray™.
    • This suggests a 2+1 adjudication method, where if the initial reference method and the investigational device disagree, a third method (FilmArray™ in this case) is used to adjudicate the true positive/negative status. If the third method agrees with the investigational device, the investigational device's result is considered correct, and vice versa.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
    • This device is an automated in vitro diagnostic test that produces a qualitative result (positive/negative) for viral pathogens. It is not an imaging device or a decision support system that would typically involve human readers interpreting results with or without AI assistance. The performance is assessed against a reference laboratory method, not human interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Yes, the performance study primarily focuses on standalone (algorithm only) performance.
    • The device is an automated system for nucleic acid detection. "The BD MAX™ System software automatically interprets test result may be called as POS (Positive), NEG (Negative), or UNR (Unresolved) for each of the assay's targets." While human operators load samples and review results, the core diagnostic decision is made by the automated system. The clinical performance data reflects this automated output.

    7. The Type of Ground Truth Used

    • Clinical Ground Truth: For prospective specimens, the ground truth was established by a "combination of two (2) sets of alternate PCRs and bi-directional sequencing for one (1) PCR."
    • Retrospective Ground Truth: For retrospective specimens, ground truth was based on "historical results" which were "confirmed using an alternate PCR assay and bi-directional sequencing."
    • Adjudication: For discordant results, an additional reference method (BioFire FilmArray™ Gastrointestinal Panel) was used for discrepant analysis, thereby contributing to the final ground truth determination.
    • This means the ground truth is primarily based on expert laboratory methods (molecular diagnostics/PCR and sequencing), which is standard for detecting viral nucleic acids.

    8. The Sample Size for the Training Set

    • The document does not explicitly mention a separate "training set" or its size in the context of the clinical validation study.
    • For in vitro diagnostic devices based on molecular assays, "training" typically refers to the assay development and optimization phases rather than a distinct, labeled dataset for machine learning model training as seen in AI/ML applications. The analytical studies (LoD, inclusivity, specificity, interference) represent the extensive work undertaken during assay development to ensure robust performance.

    9. How the Ground Truth for the Training Set Was Established

    • Since a separate "training set" in the AI/ML sense is not explicitly described, the concept of establishing ground truth for it is not directly applicable here.
    • However, during the development and optimization of the BD MAX™ Enteric Viral Panel (analogous to "training"), ground truth would have been established through:
      • Characterized strains: Using well-defined viral isolates and quantified nucleic acid standards.
      • Spiked samples: Creating samples with known concentrations of target organisms in negative matrices.
      • Reference laboratory methods: Using established gold-standard PCRs, sequencing, or culture methods to verify the presence and quantity of pathogens in control samples.