Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K243343

    Validate with FDA (Live)

    Device Name
    BD CTGCTV2
    Date Cleared
    2025-04-22

    (179 days)

    Product Code
    Regulation Number
    866.3393
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD CTGCTV2 assay incorporates automated DNA extraction and real-time polymerase chain reaction (PCR) for the direct, qualitative detection of DNA from:

    • Chlamydia trachomatis (CT)
    • Neisseria gonorrhoeae (GC)
    • Trichomonas vaginalis (TV)

    The assay may be used for detection of CT, GC and/or TV DNA in patient- or clinician-collected vaginal swab specimens (in a clinical setting) and male and female urine specimens. The assay may also be used for the detection of CT and GC DNA in endocervical swab and Liquid-Based Cytology (LBC) specimens in ThinPrep® PreservCyt® Solution using an aliquot that is removed prior to processing for the ThinPrep® Pap test. The assay may also be used for the detection of CT and GC DNA in clinician-collected rectal and oropharyngeal swab specimens.

    The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial, gonococcal, and/or trichomoniasis urogenital disease and chlamydial and gonococcal extragenital infection.

    The BD CTGCTV2 assay is available for use with the BD MAX™ System (urogenital specimens) or the BD COR™ System (urogenital and extragenital specimens), as described above.

    Device Description

    The BD CTGCTV2 assay, performed on the BD COR™ System (hereafter referred to as BD CTGCTV2), is designed for use with the applicable BD Molecular specimen collection and transport devices for male and female urine, rectal swabs, oropharyngeal swabs, vaginal swabs, endocervical swabs, and LBC specimens (PreservCyt®). Specimens are collected and transported to the testing laboratory using their respective transport devices under conditions of time and temperature that have been determined to maintain the integrity of the target nucleic acids.

    The BD COR™ MX Instrument, when combined with the BD COR™ PX Instrument, is to be used for automated sample preparation, extraction, and purification of nucleic acids from multiple specimen types, as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based real-time PCR for simultaneous and differential detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.

    The BD CTGCTV2 assay extraction reagents are dried in 96-well microtiter plates that contain binding magnetic affinity beads and Sample Processing Control (SPC). Each tube is capable of binding and eluting sample nucleic acids. The SPC monitors the integrity of the reagents and the process steps involved in DNA extraction, amplification and detection, as well as for the presence of potential assay inhibitors.

    The BD CTGCTV2 assay liquid reagent plate includes Wash, Elution and Neutralization buffers. The beads (described above), together with the bound nucleic acids, are washed and the nucleic acids are eluted by a combination of heat and pH. When performed on BD COR™ System, there is an additional buffer to rehydrate the dried extraction mix. Eluted DNA is neutralized and transferred to the Amplification reagent (described below) to rehydrate the PCR reagents. After reconstitution, the BD COR™ System dispenses a fixed volume of PCR-ready solution containing extracted nucleic acids into the BD PCR Cartridge. Microvalves in the BD PCR Cartridge are sealed by the system prior to initiating PCR in order to contain the amplification mixture and thus prevent evaporation and contamination.

    The BD CTGCTV2 assay is comprised of two targets for Chlamydia trachomatis (detected on the same optical channel), two targets for Neisseria gonorrhoeae (detected on two different optical channels) and one target for Trichomonas vaginalis (detected on one optical channel). Only one Chlamydia trachomatis target is required to be positive in order to report a positive result. Both Neisseria gonorrhoeae targets are required to be positive in order to report a positive result.

    The amplified DNA targets are detected using hydrolysis (TaqMan®) probes, labeled at one end with a fluorescent reporter dye (fluorophore), and at the other end, with a quencher moiety. Probes labeled with different fluorophores are used to detect the target analytes in different optical channels of the BD COR™ System. When the probes are in their native state, the fluorescence of the fluorophore is quenched due to its proximity to the quencher. However, in the presence of target DNA, the probes hybridize to their complementary sequences and are hydrolyzed by the 5'-3' exonuclease activity of the DNA polymerase as it synthesizes the nascent strand along the DNA template. As a result, the fluorophores are separated from the quencher molecules and fluorescence is emitted. The BD COR™ System monitors these signals at each cycle of the PCR and interprets the data at the end of the reaction to provide qualitative test results for each analyte (i.e., positive or negative).

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the BD CTGCTV2 assay detail its performance in detecting Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in extragenital specimens (rectal and oropharyngeal swabs).

    Here's an analysis based on your request:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the BD CTGCTV2 assay are implicitly demonstrated through its clinical performance results, where the assay's sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA) for extragenital specimens are compared against a Composite Comparator Algorithm (CCA). The FDA's clearance indicates that these performance metrics met the necessary standards for substantial equivalence.

    Table of Acceptance Criteria and Reported Device Performance:

    While explicit numerical acceptance criteria (e.g., "PPA must be >= X%") are not directly stated in the provided text, the reported performance measures are the ones that met the FDA's requirements for clearance.

    MetricTarget/Condition (Implicit Acceptance Criteria)Reported Device Performance (BD CTGCTV2)
    Chlamydia trachomatis (CT) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use100% (86.2–100% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Neisseria gonorrhoeae (GC) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use92.8% (85.8–96.5% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.5% (99.1–99.7% CI)
    Chlamydia trachomatis (CT) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use97.7% (93.5–99.2% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.4% (99.0–99.7% CI)
    Neisseria gonorrhoeae (GC) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use95.8% (89.7–98.4% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Non-Reportable Rate (Total CT and GC)Reasonably low for clinical utility (e.g., <5%)0.6% (0.4–0.9% CI) initial test

    Study that Proves the Device Meets the Acceptance Criteria:

    The key study proving the device meets the acceptance criteria is the Clinical Performance Study for Extragenital Specimens.

    1. Sample Size Used for the Test Set and Data Provenance:

    • Total Subjects Consented: 2,439
    • Eligible Subjects: 2,375
    • Total Eligible Specimens (test set): 4,652 (oropharyngeal and rectal swabs, considered for randomization)
    • Eligible Specimens for Testing: 4,579 (2,303 oropharyngeal swabs and 2,276 rectal swabs)
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the mention of "eight geographically diverse sites" suggests a multi-site clinical study, likely within the US, given FDA clearance.
      • Retrospective or Prospective: Prospective, as specimens were collected from subjects enrolled in the study. The study involved consenting subjects and collecting specimens specifically for this evaluation.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document.
    • The ground truth was established by a Composite Comparator Algorithm (CCA), not direct human expert interpretation of images or other subjective data.

    3. Adjudication Method for the Test Set:

    • The adjudication method used for the ground truth (CCA) was based on the concurrence of at least 2 out of 3 commercially available NAAT assays. This means it was a "2 out of 3" agreement model for establishing the positive or negative status of a specimen.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a fully automated in vitro diagnostic (IVD) assay that detects nucleic acids, not an AI-assisted imaging device requiring human interpretation. Therefore, the concept of "human readers improving with AI assistance" does not apply.

    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study. The BD CTGCTV2 assay is an automated system (BD COR™ System) that performs sample preparation, extraction, amplification, and detection. Its performance "without human-in-the-loop" is precisely what the clinical performance study measures, specifically in comparison to a CCA, which also represents an objective, algorithm-based comparator.

    6. The Type of Ground Truth Used:

    • The ground truth used was a Composite Comparator Algorithm (CCA), consisting of results from three commercially available Nucleic Acid Amplification Tests (NAATs).

    7. The Sample Size for the Training Set:

    • The document does not specify the sample size for the training set. The provided clinical performance data (4,579 specimens) pertains to the test set used for validation. Given that this is an IVD assay, the "training" analogous to machine learning would relate to the development and optimization of the assay's reagents, protocols, and analytical parameters, rather than a distinct "training set" of patient data in the typical sense of AI/ML model development.

    8. How the Ground Truth for the Training Set Was Established:

    • As the document does not explicitly discuss a "training set" in the context of machine learning, it also does not describe how a ground truth for such a set was established. For an IVD assay, the "ground truth" during development (analogous to training) would typically involve well-characterized reference strains, spiked samples with known concentrations, and analytical validation studies to optimize assay parameters, which are alluded to in the "Analytical Performance Evaluation" section (e.g., LoD, Inclusivity, Cross-Reactivity, Microbial Interference).
    Ask a Question

    Ask a specific question about this device

    K Number
    K210585

    Validate with FDA (Live)

    Device Name
    BD CTGCTV2
    Date Cleared
    2022-05-10

    (438 days)

    Product Code
    Regulation Number
    866.3393
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD CTGCTV2 assay incorporates automated DNA extraction and real-time polymerase chain reaction (PCR) for the direct, qualitative detection of DNA from:

    • Chlamydia trachomatis (CT)
    • . Neisseria gonorrhoeae (GC)
    • . Trichomonas vaginalis (TV)

    The assay may be used for detection of CT, GC and/or TV DNA in patient- or clinician-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens. The assay may also be used for the detection of CT and GC DNA in endocervical swab and Liquid-Based Cytology (LBC) specimens in ThinPrep PreservCyt Solution using an aliquot that is removed prior to processing for the ThinPrep Pap test.

    The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease, gonococcal urogenital disease and/or trichomoniasis.

    The BD CTGCTV2 assay is available for use on the BD MAX System or the BD COR System.

    Device Description

    As with the existing BD CTGCTV2 for BD MAX System, K182692, the BD COR PX/MX (BD COR) high throughput system conducts sample extraction steps to isolate and concentrate DNA which is then amplified to detect specific sequences for diagnostic purposes.

    The BD COR System is designed to allow the user to place clinical specimens directly into designated transport racks to be loaded into the System. Once the specimens are loaded, the System will perform the necessary pre-analytical steps such as vortexing, aliquoting into a molecular tube with the correct diluent, sorting/grouping of the secondary samples for testing by assay, pre-warming and cooling of the sample (where required), and transport of the sample into a molecular analyzer, where extraction, amplification and detection will take place.

    Additionally, the steps of ordering tests on the instrument for specific samples will be managed directly by the user interaction with the Laboratory Information System (LIS), which communicates with the instrument.

    Once the clinical specimens are received in the laboratory and loaded into the transport racks, the user will not be required to directly handle the specimen again prior to result reporting and removal from the system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the BD CTGCTV2 assay, based on the provided document:

    Acceptance Criteria and Device Performance

    The core of this submission focuses on demonstrating the substantial equivalence of the BD CTGCTV2 assay when run on the BD COR System to its previously cleared performance on the BD MAX System. Therefore, the acceptance criteria are implicitly tied to demonstrating comparable analytical and clinical performance between the two platforms.

    Key Performance Metrics (Implicit Acceptance Criteria) and Reported Device Performance:

    Performance MetricAcceptance Criteria (Implicit: Equivalence to BD MAX)Reported Device Performance on BD COR System (Test Set)
    Within-Laboratory PrecisionHigh percentage agreement with expected results across different target concentrations (Moderate Positive, Low Positive, High Negative, True Negative) and low CV for Ct scores.PreservCyt Samples:
    - CT (MP): 100% Correct. LP: 98.6% Correct. HN: 38.9% Positive. TN: 100% Negative.
    - GC (MP): 95.8% Correct. LP: 93.1% Correct. HN: 43.1% Positive. TN: 100% Negative.
    Urine Samples:
    - CT (MP): 100% Correct. LP: 100% Correct. HN: 54.2% Positive. TN: 100% Negative.
    - GC (MP): 98.6% Correct. LP: 100% Correct. HN: 44.4% Positive. TN: 100% Negative.
    - TV (MP): 100% Correct. LP: 100% Correct. HN: 37.5% Positive. TN: 100% Negative.
    Variance Component Analysis (PreservCyt Ct.Scores): Total CV ranged from 1.23% (GC2 MP) to 4.50% (GC1 LP).
    Variance Component Analysis (Urine Ct.Scores): Total CV ranged from 1.69% (CT MP) to 3.56% (TV LP).
    Multi-Site ReproducibilityHigh percentage agreement with expected results across different sites and low overall CV for Ct scores.PreservCyt Samples (Overall across 3 sites):
    - TN: 100%. HN: 38.9% to 48.1% positive. LP: 91.7% to 98.1% correct. MP: 98.1% to 100% correct. Overall CV (%) for Ct.Score results ranged from 1.75% to 4.15%.
    Urine Samples (Overall across 3 sites):
    - TN, LP, MP: 100%. HN: 37.0% to 58.3% positive. Overall CV (%) for Ct.Score results ranged from 1.74% to 4.18%.
    Analytical Sensitivity (LoD) EquivalenceDifference in Mean Ct.Score between BD COR and BD MAX should ideally be close to zero, with 95% CI covering zero, indicating equivalent analytical sensitivity.Vaginal Swabs: Difference in Mean Ct.Score (BD COR - BD MAX) for various targets and concentrations generally close to zero, with 95% CIs mostly crossing zero, indicating equivalence. Largest difference: -0.67 for GC2, 95% CI (-0.857, -0.494).
    Urine: Differences mostly close to zero. Largest difference: -1.10 for GC2, 95% CI (-1.375, -0.842).
    Manually Converted LBC: Differences mostly close to zero. Largest difference: 0.64 for GC1, 95% CI (0.348, 0.927).
    COR PX Converted LBC: Differences mostly close to zero. Largest difference: 0.71 for GC1, 95% CI (0.316, 1.100).
    Cross-Contamination RateA very low cross-contamination rate, typically aiming for near 0% of negative samples yielding false positives when run near high positive samples.Two false positive results (0.37%, 95% CI: 0.10-1.34%) out of 540 negative samples tested when interspersed with high positive Chlamydia trachomatis samples.
    Clinical Agreement (PPA & NPA)High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) between the BD COR System and the reference BD MAX System (ideally >95% with tight CIs).CT: Average PPA: 100%, NPA: 100%.
    GC: Average PPA: 97.6% (95% CI: 95.6%, 99.1%), NPA: 100%. (Individual sites ranged from 95.5% to 99.1% PPA).
    TV: Average PPA: 99.7% (95% CI: 99%, 100%), NPA: 98.5% (95% CI: 96.3%, 100%). (Individual sites ranged from 99.0% to 100% PPA and 97.3% to 99.1% NPA).
    Deming Regression for Ct.ScoreSlope close to 1 and intercept close to 0 between BD COR and BD MAX Ct.Scores, indicating a linear and equivalent relationship. Bias estimates also close to zero.CT: Slope 1.06 (0.99, 1.13), Intercept -1.85 (-3.88, 0.18). Bias estimates mostly low, ranging from -0.22 to 0.93.
    GC1: Slope 1.02 (0.99, 1.06), Intercept -0.06 (-0.99, 0.87). Bias estimates mostly positive, ranging from 0.47 to 0.92.
    GC2: Slope 1.03 (0.99, 1.07), Intercept -0.72 (-1.76, 0.33). Bias estimates mostly positive, ranging from 0.01 to 0.70.
    TV: Slope 1.09 (0.98, 1.19), Intercept -2.25 (-5.30, 0.80). Bias estimates mostly positive, ranging from 0.03 to 1.68.
    Non-Reportable RateLow non-reportable rate (including Unresolved, Indeterminate, Incomplete), indicating reliable assay operation.Combined Target (Total Initial Rate): 2.4% (31/1298) with 95% CI (1.7%, 3.4%). After retesting (Final Rate), this dropped to 0.0% (0/1297) with 95% CI (0.0%, 0.3%). This includes one non-reportable due to a non-readable label and 26 indeterminate results due to a consumable positioning issue (which were retested successfully).

    Study Information: BD CTGCTV2 Assay on BD COR System

    This submission pertains to the BD CTGCTV2 assay being used on the BD COR System. The key study is a clinical agreement study and analytical performance studies designed to demonstrate that the performance of the assay on the BD COR System is equivalent to its already cleared performance on the BD MAX System (K182692).

    1. A table of acceptance criteria and the reported device performance:
    * See table above.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    * Analytical Performance (Precision & Reproducibility Test Set):
    * Precision (within-laboratory): 72 replicates for each target (CT, GC, TV) at each of 4 concentration levels (MP, LP, HN, TN) for PreservCyt samples and 4 concentration levels for Urine samples. Total N is not explicitly stated as a single number but is derived. For example, for CT in PreservCyt, it's 72 * 4 = 288 data points.
    * Reproducibility (multi-site): For each target and concentration level, 36 replicates per site across 3 sites (36 * 3 = 108 total replicates per target/level).
    * Analytical Sensitivity Confirmation: 4 panel members (A, B, C, D) created with 1.5x LoD and 3x LoD for various target organisms and strains in pooled female urine, pooled vaginal swab, and pooled PreservCyt LBC matrix. The study compared performance between BD COR and BD MAX. The exact number of replicates for each panel member for this confirmation study is not explicitly stated as a total N, but implied to be sufficient for statistical comparison (mean Ct.Scores and 95% CIs are reported).
    * Cross-Contamination: 540 positive samples and 540 negative samples for a total of 1080 samples.
    * Clinical Agreement Study (Test Set):
    * Sample Size: 433 independent panel members.
    * Provenance: "Remnant urine specimens from the previous clinical trial for BD CTGCTV2 on BD MAX as well as urine specimens obtained from both internal and external collections were used for the comparison study." This suggests a retrospective collection of remnant urine specimens combined with potentially prospective collections from internal and external sources to create the clinical panels. The country of origin is not explicitly stated, but given the FDA submission, it can be inferred to be primarily US-based or at least compliant with US regulations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    * For the Precision, Reproducibility, Analytical Sensitivity, and Cross-Contamination studies: The ground truth was established by the known concentrations or presence/absence of organisms in contrived samples or spiked matrices. These are analytical studies, not clinical studies requiring expert interpretation.
    * For the Clinical Agreement Study: The BD MAX System results served as the reference/ground truth. The positive or negative status of a panel member was defined by "≥2 out of 3 evaluable results obtained on the BD MAX." This is a comparator method, not direct expert consensus on patient samples. Therefore, no human experts were used to establish the ground truth for the clinical agreement study beyond the definition of the BD MAX as the reference standard.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    * For the Clinical Agreement Study: The "ground truth" (reference comparator result from BD MAX) was established by "≥2 out of 3 evaluable results obtained on the BD MAX". This functions as a form of "consensus" or adjudication among multiple runs (3 aliquots) on the reference system.
    * For the analytical studies (precision, reproducibility), ground truth was based on known concentrations, so no adjudication by a panel of experts was necessary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) assay that detects nucleic acids. It's an automated molecular system, not an AI-assisted diagnostic tool that human readers would interpret. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Yes, this is a standalone performance study in the context of an IVD assay. The BD CTGCTV2 assay on the BD COR System is an automated system for qualitative detection of DNA. The studies described (analytical and clinical agreement) evaluate the performance of this automated system directly, without human interpretation of its diagnostic output. The output itself (positive/negative) is the final result.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    * Analytical Studies (Precision, Reproducibility, Cross-Contamination, Analytical Sensitivity): The ground truth was based on known concentrations of purified organisms or specific strains spiked into negative matrices. This is an analytical ground truth.
    * Clinical Agreement Study: The ground truth was the result from the legally marketed predicate device, the BD MAX System, determined by a "consensus" of ≥2 out of 3 evaluable results from the BD MAX. This is a (predicate) comparator ground truth.

    8. The sample size for the training set:
    * The document describes studies for validation and equivalency demonstration of the BD CTGCTV2 assay on the BD COR System compared to the BD MAX System. It does not mention "training sets" in the context of machine learning or AI models.
    * For IVD assays, "training" typically refers to the initial development and optimization of the assay performed by the manufacturer. The data used for this developmental phase is not typically detailed in 510(k) summaries, which focus on formal validation studies.
    * The assay itself incorporates "automated DNA extraction and real-time PCR," which are well-established molecular biology techniques, not typically "trained" in the AI sense.

    9. How the ground truth for the training set was established:
    * As noted above, the document does not describe a "training set" in the context of an AI/machine learning model. Therefore, this question is not applicable. The development of the PCR assay and its parameters would have involved extensive laboratory work by the manufacturer, but the "ground truth" for those developmental phases would be based on well-characterized materials and samples, similar to the analytical studies described for validation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1