510(k) Data Aggregation

    K Number
    K043495

    Validate with FDA (Live)

    Device Name
    BABYPAC
    Date Cleared
    2005-04-22

    (126 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The babyPAC 100 is a portable gas powered transport ventilator, suitable for use in an MRI environment up to 3 Testa, which features a battery powered integrated electronic pressure alarm unit. It is designed for use by qualified medical caregivers, paramedics and other trained personnel, for ambulance, hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency. It can be used on neonates and infants up to a bodyweight of 44lb (20 kg).

    Device Description

    The babyPAC ventilator consists of a control module with a disposable conventional Y patient circuit. The babyPAC ventilator is a pressure generator which depends solely on the pressure of the supply gas for its operation. The babyPAC 100 additionally incorporates an integrated electronic pressure alarm unit which becomes operational during Controlled Mandatory Ventilation. This unit provides a warning of significant changes which may occur in the patient's ventilation. Loss of battery power for the alarm is signaled to the user but will have no effect on the ventilation performance of the babyPAC ventilator, nor affect the mechanically operated alarms and performance on the babyPAC ventilator (K970158) already on the market.

    The babyPAC 100 portable ventilator is designed for use as an emergency and transport ventilator in ambulances and hospitals, and can be used in an MRI environment up to 3 Teslas. It is intended for transportation and for resuscitation of infants (above approx 11lb [5 kg]) and neonates up to 44 lb (20 kg).

    The control module of the babyPAC ventilator is rugged by virtue of its thick section structural foam plastic case and the controls for the gauge and internal pneumatics are recessed to minimise the possibilities of damage and inadvertent operation.

    The babyPAC 100 portable ventilator consists of a control module and the following items: 22mm disposable Polyethylene respirator hose/ 22mm OD female connector and Paediatric 'Y' piece connector

    The module weighs 8.25 lb (3.75 kilograms)

    The module control panel has the following features:

    • Adjustable Relief Pressure Control, range 12 to 80 cm H2O.
    • Inspiratory Time Control, range 0.25 to 2.0 seconds.
    • Expiratory Time Control, range 0.25 to 4.0 seconds
    • Inspiratory Pressure Control, range 12 to 70 cm H2O, with click action warning at and above 40 cm H2O.
    • PEEP/ CPAP Control, range 0 to 20 cm H2O with click action warning at and above 10 cm H2O.
    • IMV Control, range Expiratory time 2.5 to 40 seconds with CPAP Pressure level and oxygen concentration as selected.
    • Patient Inflation Pressure Manometer, range -10 to +100 cm H2O.
    • Variable Oxygen Concentration, range 45 (approx) to 100% with oxygen only as supply, 21 to 70 % with oxygen and air supplies
    • Supply Gas Failure Alarm indicators- One for air, the other for oxygen, these mechanically operated visual alarms give a warning that the supply gas has dropped to a pressure at which the ventilator will no longer be operating to specification (< 35 psi). With low pressure they show red, with adequate pressure they show white. Any visible red indicates that the supply should be changed. As the pressure falls the display will begin to oscillate from white to partial red as the supply pressure falls to the lower threshold level.

    The visual indication will be accompanied by an electronically generated medium priority (to EN 475: 1995) audible warning.

    • Electronic alarm bezel indicating
      • High Pressure/ Constant Positive Inflation Pressure Indicator Flashes Red LED with audible alarm at set relief pressure and with continuous positive pressure
      • Inflation Indicator Flashes Green LED every time inflation pressure rises through 10 cm H2O.
      • Low Pressure/ Disconnect Indicator Flashes Yellow LED with audible alarm if pressure does not rise through 10 cm H2O within eight seconds
      • Silence button silences audible alarm for 60 seconds. Flashes Orange LED to indicate to the operator that the audible alarm is silenced.
      • Single Gas Use Indicator Flashes a Green LED as a burst of 3 flashes every 30 seconds, whenever the ventilator is being used with only a single gas input (ie oxygen or air only). In the event that the single gas fails (ie. where the audible gas failure alarm has activated) the Low Pressure/ Disconnect Indicator will activate together with the audible alarm.
      • Low battery indicator Flashes Yellow LED with audible alarm.
    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about acceptance criteria and the study proving device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Safety TestingCompliance with EN794-3 (Lung Ventilators - Part 3 Particular requirements for emergency and transport ventilators) 1999"The ventilator passes all of these tests and met all requirements of the standards."
    Compliance with EN60601-1 (Medical Electrical Equipment - Part 1 General requirements for safety) 1990"The ventilator passes all of these tests and met all requirements of the standards."
    Environmental TestingCompliance with EN 60601-1-2: 1993"The results demonstrated that the babyPAC 100 complied with the guidelines and standards and that they performed within their specifications and functional requirements."
    Compliance with EN794-3: 1999"The results demonstrated that the babyPAC 100 complied with the guidelines and standards and that they performed within their specifications and functional requirements."
    Performance (Functional Equivalence to Predicate K970158)Delivered Tidal Volume, Frequency, Inspiration times, and Expiration time parameters are the same as the predicate (babyPAC K970158)."All measurements were within the specified tolerances of the ventilators," indicating "no functional difference."
    Performance (Functional Equivalence to Predicate K020899)Safe for use in an MRI environment up to 3 Teslas, and the performance of the electronic integrated alarms system is the same as the predicate (paraPAC 'medic' P200 K020899)."No functional difference... for safe use in an MRI environment up to 3 Teslas and operation of the integrated electronic alarms."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a 'sample size' in terms of patient data or a specific number of devices tested for each parameter. The testing described is typically design verification and validation testing rather than a clinical trial with a patient test set.

    • Test Set Sample Size: Not explicitly stated in terms of patient numbers or specific device units. The wording suggests testing of the ventilator's entire range and across its parameters.
    • Data Provenance: The document does not specify the country of origin of the data. Given the "Submitter" and "Address" information (Smiths Medical International Ltd., UK), it is highly probable the testing was conducted in the UK or a region complying with the cited European standards (EN). The data is part of a submission for regulatory clearance (510(k)), implying it is prospective with respect to the regulatory approval process, but the detailed testing itself would have been retrospective if conducted on existing manufactured units as part of the product development lifecycle.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (experts for ground truth determination) is not applicable or provided in this document. The "ground truth" for the device's performance is established by direct measurement against engineering specifications and regulatory standards, not by expert consensus on clinical data. The tests refer to objective measurements (e.g., Tidal Volume, Frequency, alarm activation) against defined tolerances and compliance with standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human interpretation where adjudication would be necessary. Performance is measured against objective technical standards and predicate device performance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is a ventilator, and the testing focuses on its engineering performance and safety compliance, not on human interpretation of medical images or data where MRMC studies are typically employed.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, a standalone performance study was done. The entire "Performance Data" section describes the device's performance in isolation (e.g., meeting EN standards, comparison of physical output parameters like tidal volume). The device itself is not an "algorithm" in the modern AI sense, but its functional performance was tested standalone without human intervention affecting the measurement of its core parameters. The electronic alarm unit's operation was also tested standalone in its comparison to the predicate.

    7. Type of Ground Truth Used

    The ground truth used is primarily based on:

    • Engineering Specifications and Tolerances: For parameters like Tidal Volume, Frequency, Inspiration/Expiration times, relief pressure, PEEP/CPAP, etc.
    • Regulatory and International Standards: E.g., EN794-3, EN60601-1, EN60601-1-2. Compliance with these standards forms the ground truth for safety and environmental performance.
    • Predicate Device Performance: For demonstrating substantial equivalence, the performance of the babyPAC (K970158) and paraPAC 'medic' P200 (K020899) served as the established ground truth for comparison.

    8. Sample Size for the Training Set

    Not applicable. This device is a medical ventilator, not an AI/ML algorithm that requires a "training set" in the computational sense. The design and validation are based on engineering principles and regulatory testing, not statistical learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970158

    Validate with FDA (Live)

    Device Name
    BABYPAC
    Manufacturer
    Date Cleared
    1998-04-28

    (467 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The babyPAC pneumatically powered ventilator is specifically designed for use by qualified medical caregivers, paramedics and other trained personnel for patient ventilation during respiratory distress or insufficiency in the rescue, pre-hospital and alternative site arenas.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for a device called "babyPAC." It confirms that the device is substantially equivalent to legally marketed predicate devices and can be marketed.

    This document does not contain the information necessary to describe:

    • Acceptance criteria and reported device performance.
    • Details about a study proving the device meets acceptance criteria (sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods).
    • Information on MRMC comparative effectiveness studies or standalone algorithm performance.
    • Training set details.

    Therefore, I cannot fulfill your request based on the given input.