    K Number
    K211261

    Validate with FDA (Live)

    Date Cleared
    2021-05-28

    (32 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Axis Charcot Fixation System in diameters of 4.5 to 8.5mm is indicated for reconstruction procedures, non-unions and fusions of bones in the foot and ankle including the metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus; specific examples include: medial and lateral column fusion resulting from neuropathic osteoarthropathy (Charcot).

    Device Description

    The Axis Charcot Fixation System consists of 5.5, 6.5 and 7.5mm cannulated, titanium alloy fixation beams and accessories used for midfoot reconstruction. The modified device adds additional sizes, 4.5 and 8.5mm cannulated, titanium alloy fixation beams, previously cleared under Extremity Medical's 4.5 to 8.5 Screw System (K171018). The additional sizes offer the surgeon options for placement based on patient anatomy.

    AI/ML Overview

    The provided text is a 510(k) summary for the Axis Charcot Fixation System, a medical device. It does not include information about AI/ML device performance, acceptance criteria, or studies involving human readers, as it describes a bone fixation system.

    Therefore, most of the requested information cannot be extracted from this document as it pertains to AI/ML device evaluation.

    However, based on the provided text, I can infer the following:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner typical for AI/ML performance. Instead, it refers to "engineering analyses evaluating mechanical strength" and "pullout strength". The performance determined was "substantially equivalent to the predicate devices for the intended use."

    Acceptance Criteria (Inferred from testing type)Reported Device Performance
    Mechanical strength of the smallest beam diameter (4.5mm) is substantially equivalent to predicate.Demonstrated substantial equivalence to predicate devices.
    Pullout strength of the shortest, smallest beam diameter is substantially equivalent to predicate.Demonstrated substantial equivalence to predicate devices.
    The largest beam diameter (8.5mm) is no worst case than the predicates.No additional safety and effectiveness concerns presented.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided. The testing described is mechanical, not involving patient data or a "test set" in the context of an AI/ML device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The "ground truth" for this device would be established through engineering standards and physical measurements, not expert consensus on medical images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Mechanical testing does not involve adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was performed or is relevant to this device. This device is a physical bone fixation system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This device is a physical bone fixation system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is based on engineering standards and physical measurements of mechanical properties (strength, pullout force).

    8. The sample size for the training set

    This information is not applicable and not provided. There is no AI/ML model being "trained" for this device.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. There is no AI/ML model being "trained" for this device.

    In summary, the provided document describes a 510(k) submission for a physical medical device (bone fixation system) and does not contain information related to AI/Machine Learning device performance or ground truth establishment in the context of diagnostic or interventional AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171018

    Validate with FDA (Live)

    Date Cleared
    2017-07-13

    (99 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Axis Charcot Fixation System:
    The Axis Charcot Fixation System in diameters of 5.5, 6.5 and 7.5mm is indicated for reconstruction procedures, non-unions and fusions of bones in the foot and ankle including the metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus; specific examples include: medial and lateral column fusion resulting from neuropathic osteoarthopathy (Charcot).
    4.5 to 8.5 Screw System:
    The 4.5 to 8.5mm Screw System is intended for fixation arthrodesis of the metatarsal-cuneiform, navicular-cuneiform, metatarsal-cuboid, talonavicular, and calcaneocuboid joints.

    Device Description

    Axis Charcot Fixation System
    The Axis Charcot Fixation System consists of 5.5, 6.5 and 7.5mm cannulated, titanium alloy fixation beams and accessories used for midfoot reconstruction.
    4.5 to 8.5 Screw System
    The 4.5 to 8.5 diameter screws consists of cannulated, titanium alloy fixation screws for use in bone reconstruction, osteotomy, arthrodesis and fracture repair and fixation in the foot.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    This document is a 510(k) premarket notification for the "Axis Charcot Fixation System, 4.5 to 8.5mm Screw System." It's important to note that 510(k) submissions typically demonstrate substantial equivalence to a predicate device, rather than de novo approval requiring extensive clinical efficacy studies. Therefore, the "acceptance criteria" here are framed around demonstrating comparable performance to existing, legally marketed devices.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance (vs. Predicate)
    Material EquivalenceMaterial of manufactureTi-6Al-4V (equivalent to predicate devices)
    Screw Size OfferingDiameters (5.5, 6.5, 7.5mm for Axis Charcot; 4.5 to 8.5mm for Screw System)Equivalent to predicate devices (WMT Salvation Beams and Bolts, Smith & Nephew 6.5 and 8.0mm Cannulated Screws; Extremity Medical Screw and Washer System, Paragon28 Monster Screw System)
    Mechanical PerformancePull-out strengthBench testing performed and compared to predicate device
    Static bendingBench testing performed and compared to predicate device
    Dynamic bendingBench testing performed and compared to predicate device
    Design CharacteristicsOptional accessory washer/nutDesign differences in the clip of the washer/nut do not introduce new issues of safety or effectiveness compared to predicate washers/nuts.
    Intended Use EquivalenceIndications for Use"substantially equivalent" to predicate devices based on indications for use.
    Principle of Operation(Implicitly assessed)"substantially equivalent" to predicate devices based on principles of operation.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "No clinical testing was performed." Therefore, there is no "test set" in the sense of patient data.

    • Sample Size for Mechanical Testing: Not explicitly stated in the summary, but typical for bench testing, multiple samples of each device configuration would be tested.
    • Data Provenance: Not applicable as no clinical data was collected. The bench testing would have been conducted by Extremity Medical, LLC or a contracted lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical testing was performed, there was no ground truth to establish from patient data using expert review. The "ground truth" for the mechanical testing would be the raw data collected from the testing equipment, and the comparison point would be the performance of the predicate device (likely from published data or internal testing).

    4. Adjudication method for the test set

    Not applicable. No expert adjudication of clinical data was performed.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone fixation system (screws), not an AI-powered diagnostic tool. MRMC studies are used for evaluating diagnostic performance in image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a medical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical testing:

    • Ground Truth: The physical properties and performance measurements (e.g., pull-out force, bending stiffness, fatigue life) derived directly from the bench tests.
    • Comparison Basis: The performance characteristics of the identified predicate devices, likely obtained through literature, publicly available predicate device information, or internal testing for comparison purposes.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device. The design and manufacturing processes are informed by engineering principles, material science, and the performance characteristics of existing devices, rather than a "training set" in the context of data-driven models.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.