Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172717

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-25

    (259 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Automatic Tissue Processing Unit is used in medical procedures involving the harvesting and transferring of autologous adipose tissue. The Automatic Tissue Processing Unit is used for concentrating adipose tissue harvested with a legally marketed lipoplasty system. The Automatic Tissue Processing Unit is intended for use in the following surgical specialties when the concentration of harvested adipose tissue is desired.

    • Neurosurgery Gastrointestinal Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery Arthoscopic Surgery
    Device Description

    The fat harvested by liposuction, called as lipoaspirates, contains fat tissues, bloody impurities and tumescent solution. The harvested lipoaspirates are not appropriate for autologous fat transfer without removal of the bloody impurities and tumescent solution containing anesthetic agent, anti-coagulant agent, antibiotics etc. In addition, the harvested fat tissues are not homogenous and contain uneven fibrotic tissues. So, for the purpose of in autologous fat transfer surgical technique. ACPU kit is designed and developed for harvest of the fine homologous adipose tissue in a sterile and closed way. ACPU needs to be centrifuged by using exclusive centrifuge (ACS Combo-A-Centrifuge/Product code: JQC) and the kits are single-use only.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) summary regarding the Automatic Tissue Processing Unit, structuring the information according to your requested points:

    Device: Automatic Tissue Processing Unit (ACPU-100/ACPU-200)

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the successful results of the specified tests, indicating that the device meets predetermined standards for each parameter. The "Requirements" column serves as the acceptance criteria.

    Test ItemAcceptance Criteria (Requirements)Reported Device Performance (Results)
    AppearanceNo defects, flaws, breakage, or contamination of foreign objects.Pass
    MeasurementWithin ±5% of stated dimensions.Pass
    LeakageNo leakage.Pass
    Residuals of cleaning liquidLess than 20 ml.Pass
    PackagingAdhesive strength not less than 3N (ASTM F-88).Pass
    Nucleated Cell CountMore than 3.86x10^5 per ml of adipose tissue."3.86 over" - Implies greater than or equal to, thus meeting the criteria.
    Nucleated Cell ViabilityMore than 71.4%."71.4% over" - Implies greater than or equal to, thus meeting the criteria.
    Extraction test:
    Appearance (sample prep extract)No foreign material.Pass
    pH (Difference)≤ 1.5Pass
    KMnO4 Reducing agents (Difference in titres)≤ 2.0 mlPass
    Evaporating residue (Difference in extractables)≤ 1.0 mgPass
    Heavy metal (as Pb)Not darker than standard solution.Pass
    UV-vis Spectrum (Difference in absorbance 250nm-350nm)≤ 0.1Pass
    Biocompatibility Tests:
    Cytotoxicity (ISO 10993-5)Pass (implicitly defined by standard)Pass
    Skin Sensitization (ISO 10993-10)Pass (implicitly defined by standard)Pass
    Intracutaneous Reactivity Test (ISO 10993-10)Pass (implicitly defined by standard)Pass
    Acute Systemic Toxicity Test (ISO 10993-11)Pass (implicitly defined by standard)Pass
    Pyrogen Test (ISO 10993-11, Annex F)Pass (implicitly defined by standard)Pass
    Hemolysis Test (ISO 10993-4)Pass (implicitly defined by standard)Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for the "bench tests." The tests listed are primarily physical, chemical, and biological characterization tests performed on the device itself or its materials, rather than a clinical human subject test set. The data provenance is not specified, but these are laboratory tests rather than patient data.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Since the tests performed are bench tests (biocompatibility, mechanical, sterility, extraction, cell counts/viability), the concept of "ground truth" established by human experts in a diagnostic or clinical context is not applicable. The ground truth for these tests is based on objective laboratory measurements and established scientific standards (e.g., ISO guidelines, ASTM standards, chemical analysis techniques).

    4. Adjudication Method for the Test Set

    Adjudication methods like "2+1" or "3+1" are relevant for clinical studies where human interpretation of medical images or outcomes needs consensus. For the bench tests described, adjudication is not applicable. The "results" are objective measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The document describes bench testing to demonstrate functionality and substantial equivalence to a predicate device, not a human reader study.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is an Automatic Tissue Processing Unit, a mechanical/biological processing system, not an AI algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance data in this submission consists of:

    • Objective physical measurements: Dimensions, leakage, adhesive strength.
    • Chemical analysis results: Residuals of cleaning liquid, pH, KMnO4 reducing agents, evaporating residue, heavy metals, UV-vis spectrum.
    • Biological assay results: Nucleated cell count and viability.
    • Standardized biological test outcomes: Biocompatibility tests (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, hemolysis) performed according to ISO standards, where "Pass" is the objective outcome based on the standard's criteria.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The data presented is for validation and verification of the physical device's performance.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1