Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143571

    Validate with FDA (Live)

    Date Cleared
    2015-01-23

    (37 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Linearity LQ Cystatin-C is an assayed quality control material intended to simulate human patient samples for use in determining linearity, calibration, and the verification of reportable range for the Cystatin-C analyte.

    The Linearity LQ Cystatin-C is for In Vitro Diagnostic use only.

    Device Description

    The Audit® MicroControls™ Linearity LQ Cystatin-C product is an in-vitro diagnostic device consisting of one set of five levels of liquid, linearity/QC material. For the set there are five levels labeled A, B, C, D and E. The set contains 1ml for each level. The set contains the Cystatin-C analyte and additives in human and bovine based serum. Materials of human origin used in the manufacture of this linearity set have been tested using FDA approved methods and are found to be non-reactive for HbsAg and antibodies to HCV and HIV-1/2.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    This document is a 510(k) Premarket Notification for a Linearity LQ Cystatin-C quality control material (device). The goal of the submission is to demonstrate substantial equivalence to a predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/StudyAcceptance CriteriaReported Device Performance
    Shelf Life (Unopened)Product claims: 3 years, when stored unopened at 2-8°C.Accelerated stability studies were conducted to establish the shelf life stability claims. Acceptance criteria were met to support the product claims of 3 years, when stored unopened at 2-8°C. (Real-time studies are ongoing).
    Open Vial StabilityProduct claims: Once a vial has been opened, the product will be stable for 30 days when stored tightly capped at 2-8°C.Real-time stability studies conducted at the end of accelerated stability studies. Acceptance criteria were met to support the product claims of 30 days when stored tightly capped at 2-8°C.
    Target Ranges for Cystatin-CTarget ranges for each of the 5 levels were calculated as ± 10% of the target value. (Implied: Measured values should fall within these ranges).The mean value of the Cystatin-C analyte was used to establish target concentration value at each level. The target ranges for Cystatin-C (in mg/L) for each level are:- Level A: 0.584-0.714 (Target 0.649)- Level B: 2.09-2.56 (Target 2.33)- Level C: 3.69-4.51 (Target 4.10)- Level D: 5.08-6.21 (Target 5.65)- Level E: 6.44-7.87 (Target 7.15)
    Non-reactivity for Infectious Disease MarkersHuman-origin materials used must be tested using FDA approved methods and found to be non-reactive for HbsAg and antibodies to HCV and HIV-1/2.Materials of human origin used in the manufacture of this linearity set have been tested using FDA approved methods and are found to be non-reactive for HbsAg and antibodies to HCV and HIV-1/2.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a quality control material used for linearity, calibration verification, and reportable range for a specific analyte (Cystatin-C). It's not a diagnostic device with a "test set" in the traditional sense of patient data.

    • Test Set Description: The "test set" here refers to the actual device (Linearity LQ Cystatin-C) which comes in five levels (A, B, C, D, E) of a liquid material. Each level contains 1ml.
    • Sample Size:
      • For stability studies: Vials from two lots of finished product were used for real-time stability (stored at 2-8℃ and -80℃). Measurements were taken at two different time points. The analyte values from these vials were tested in duplicate.
      • For value assignment: The Cystatin-C analyte was measured multiple times.
    • Data Provenance: The document does not specify the country of origin of the data. The studies were conducted by Aalto Scientific, Ltd., located in Carlsbad, CA. These are prospective studies performed to characterize the device's stability and assigned values.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (quality control material) does not typically involve human "experts" establishing a ground truth for a test set in the way a diagnostic algorithm for medical images would. Instead, the ground truth (target concentration values) is established through technical laboratory measurements.

    • Number of Experts: Not applicable.
    • Qualifications: Not applicable. The "ground truth" (target values) is established through analytical testing methods on a specific instrument.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, this is not a diagnostic device where patient cases are reviewed and adjudicated. The "ground truth" (target analyte concentrations) for the quality control material is inherent to its formulation and confirmed through laboratory measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    Not applicable. This is a quality control material, not an AI-powered diagnostic device, and therefore no MRMC study was conducted or relevant.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a laboratory quality control material, not an algorithm or AI.


    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth used for this device, the Audit® MicroControls™ Linearity LQ Cystatin-C, is analytical reference measurement. Specifically:

    • Analyte Value Assignment: "The mean value of the Cystatin-C analyte was used to establish target concentration value at each level." This refers to the measured concentration on a reference instrument (Beckman Immage) using a specific reagent.
    • Stability: Measured values at different time points and storage conditions are compared against initial or expected values established through analytical testing.

    8. The Sample Size for the Training Set

    Not applicable. This is a quality control material, not a machine learning algorithm, and therefore does not have a "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this type of device, the concept of establishing ground truth for it does not apply.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1