Found 2 results

510(k) Data Aggregation

    K Number
    K200697

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-23

    (281 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Adult Patients
    AtriCure's Cryo2 cryo2 cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
    The Cryo2 cryolCE cryo-ablation probes are also intended for use to temporarily block pain by ablating peripheral nerves.
    For Adolescent Patients
    The Cryo2 cryolCE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.
    1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.
    For Adult Patients
    AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use performed by freezing target tissues, creating an inflammatory response (cryonecrosis) for blocking pain by temporarily ablating peripheral nerves.
    For Adolescent Patients
    The cryolCE cryoSPHERE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization1 in adolescent patients of at least 12 years of age.
    1 Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology.

    Device Description

    The cryosurgical handpiece, or cryo-ablation probe, utilized in the AtriCure Cryosurgical System is a hand held, single use, cryosurgical instrument intended for the cryosurgical treatment of cardiac surgery; it is also intended for use in blocking pain by temporarily ablatinq peripheral nerves. The cryosurgical handpiece uilizes a high-pressure cryogen (nitrous oxide, N20) to freeze target tissues, creating an inflammatory response, and ultimately, cryonerosis. The cryosurgical handpiece provides below -40°C, the temperature at which intracellular ice formation occurs which is considered lethal to cells. When high pressure nitous oxide is supplied to the cryprobe via the AtriCure Cryo Module, rapid cooling is achieved via the Joule-Thompson effect. The end effector, or cryolip, of the probe is malleable to allow access to varying anatomy and anatomical sites.
    The cryo-ablation probes are comprised of a cryotip end effector, shaft, handle, thermocouple, inlet tube, and exhaust tube. The cryotip consists of an aluminum boiler and three infices distributed throughout the cryotip internally to provide uniform cooling. The 4-mm diameter, cryotip is malleable throughout its 10-cm length. A supplied forming tool can be cryotip into the desired form. The cryolip is attached to an insulated rigid shaft that allows the surgeon to adjust the exposed cryolip up to 10 cm in therapeutic length. A thermocouple is affixed to the proximal external surface of the temperatures on the console. The handle is attached to the shaft. Inlet and exhaust tubes and thermocouple wire pass through the handle and connect to the AtriCure Cryo Module (ACM).
    The cryoICE PROBE is a sterile, single-use cryosurgical instrument.
    The form tool facilitates bending of the malleable tip.
    This PROBE was designed for treatment of cardieving controlled temperatures down to -50° to -70°C; it can also be used to temporarily block pain by ablating peripheral nerves in adolescent patients of at least 12 years of aqe.
    The cryolCE cryoSPHERE cryoablation (CRYOS/CRYOS-L) probe is a single use cryablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.
    The PROBE is a single-use device offered in two configurations: standard length probe shaft (CRYOS), and extended length probe shaft (CRYOS-L). The probe shaft is malleable and supports forming by the supplied TOOL. The PROBE features a spherical 8mm cryoablation tip.

    AI/ML Overview

    This document is a 510(k) summary for AtriCure CryoICE cryo-ablation probes (Cryo2, CryoS, CryoS-L). It focuses on the expanded indication for use in adolescent patients aged 12 and above for temporary pain unique to its use in intercostal nerves under direct visualization.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it states that clinical evidence demonstrated the use of the cryoablation probes in adolescent patients (≥ 12 years) for intercostal nerve ablation was "as safe and effective" as its use in adults. The evaluation was based on the following endpoints, implying these were the performance areas assessed:

    Acceptance Criteria (Implied)Reported Device Performance
    Safety (regarding intraoperative or post-operative complications)Demonstrated to be as safe as adult use.
    Effectiveness (regarding intraoperative or post-operative complications)Demonstrated to be as effective as adult use.
    Incidence of adverse eventsComparable to adult use.
    Incidence of cryoablation associated numbnessComparable to adult use.
    Incidence of neuropathic painComparable to adult use.
    Incidence of chest wall numbnessComparable to adult use.
    Incidence of post-operative pneumothoraxComparable to adult use.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states that AtriCure partnered with the UCSF Stanford Pediatric Device Consortium (PDC) to evaluate clinical evidence, which comprised clinical publications. The exact sample size from these publications is not specified in this summary. The data provenance (country of origin, retrospective/prospective nature) is also not explicitly stated in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the 510(k) summary. The evaluation relies on published clinical evidence rather than a dedicated expert panel for ground truth establishment within this specific submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the 510(k) summary. Given that the evidence is based on clinical publications, any adjudication would have been part of the original studies, and details are not reiterated here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as this device is a physical medical instrument (cryoablation probe) and not an AI-assisted diagnostic or interpretative tool. Therefore, the concept of human readers improving with AI assistance is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable as the device is a physical cryoablation probe and not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the comparison of safety and effectiveness was based on outcomes data from the clinical publications. Specifically, it assessed "intraoperative or post-operative complications," "adverse events observed," and specific post-ablation conditions like "cryoablation associated numbness," "neuropathic pain," "chest wall numbness," and "post-operative pneumothorax." The comparison was against the known safety and effectiveness in adult patients.

    8. The sample size for the training set

    This document describes a premarket notification (510(k)) for a medical device and relies on existing clinical evidence. There is no "training set" in the context of machine learning for this type of device submission. The clinical evidence used for assessment is analogous to a validation set.

    9. How the ground truth for the training set was established

    As there is no "training set" in the context of machine learning for this device, this question is not applicable. The clinical evidence used for evaluation was derived from existing clinical publications, where the "ground truth" (e.g., presence of complications, pain levels) would have been established within those individual studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180138

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-02-15

    (29 days)

    Product Code
    Regulation Number
    882.4250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AtriCure's cryolCE cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.

    Device Description

    The cryosurgical handpiece, or cryo-ablation probe, utilized in the AtriCure Cryosurgical System is a hand held, single use, cryosurgical instrument intended for the cryosurgical treatment of cardiac arrhythmias during cardiac surgery; it is also intended for use in blocking pain by temporarily ablating peripheral nerves. The cryosurgical handpiece utilizes a high-pressure cryogen (nitrous oxide, N2O) to freeze target tissues, creating an inflammatory response, and ultimately, cryonecrosis. The cryosurgical handpiece provides probe temperatures below -40°C, the temperature at which intracellular ice formation occurs which is considered lethal to cells. When high pressure nitrous oxide is supplied to the cryoprobe via the AtriCure Cryo Module, rapid cooling is achieved via the Joule-Thompson effect. The end effector, or cryotip, of the probe is malleable to allow access to varying and anatomical sites. The cryo-ablation probes are comprised of a cryotip end effector, shaft, handle, thermocouple, inlet tube, and exhaust tube. The cryotip consists of an aluminum boiler and three internal inlet orifices distributed throughout the cryotip internally to provide uniform cooling. The 4-mm diameter, cryotip is malleable throughout its 10-cm length. A supplied forming tool can be used to bend the desired form. The cryotip is attached to an insulated rigid shaft that allows the surgeon to adjust the length of the exposed cryotip up to 10 cm in therapeutic length. A thermocouple is affixed to the proximal external surface of the shaft to display real time temperatures on the console. The handle is attached to the shaft. Inlet and exhaust tubes and thermocouple wire pass through the handle and connect to the AtriCure Cryo Module (ACM).

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the AtriCure cryoICE cryo-ablation probe (CRYO2). It is a submission to the FDA for a modified version of an already cleared device. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving the performance of a new device against clinical acceptance criteria in the way AI/ML devices often require.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not applicable or available in this document for the following reasons:

    • This is a 510(k) for a modified device, not a novel AI/ML device. The focus is on demonstrating that the modified device is as safe and effective as a legally marketed predicate device, not on proving new clinical performance metrics against defined acceptance criteria.
    • The device is a cryo-ablation probe, not AI/ML software. The performance data described is non-clinical bench testing (mechanical reliability, transit, shelf-life, cryogen performance/thermal insulation), which is standard for hardware modifications to medical devices. It does not involve algorithms, human readers, or clinical ground truth in the context of diagnostic or prognostic AI.
    • "Acceptance criteria" here refer to regulatory requirements for demonstrating substantial equivalence, not performance metrics. The conclusion states that the modified device is "equivalent to the previously cleared AtriCure cryoICE cryoablation probe (CRYO2) as there is no change to intended use, operating principals, or function of the device." This "equivalency" is the primary acceptance criterion for a 510(k) of this nature.

    Based on the provided document, I cannot generate the table or specific details requested because the information simply isn't present, as it pertains to a different type of device and regulatory submission.