510(k) Data Aggregation

    K Number
    K221036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-01

    (85 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide Warming is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide Warming is non-sterile, clear and water based personal lubricant. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms only. It is not compatible with polyurethane condoms. The device is composed of water, glycerin, propylene glycol, polyquaternium-7 and citric acid.

    AI/ML Overview

    The provided document describes the K221036 submission for Astroglide Warming, a personal lubricant. It does not describe a study involving an AI/ML device and human readers. Therefore, I cannot fulfill all aspects of your request as it pertains to AI/ML device performance, ground truth, expert opinions, or MRMC studies.

    However, I can extract the acceptance criteria and the summary of non-clinical performance testing for the Astroglide Warming device itself.

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Test Method)Reported Device Performance
    Absence of particulate matterNo particlesMet specification (implied by "met the specifications at all time points")
    ColorColorlessMet specification (implied by "met the specifications at all time points")
    ClarityClearMet specification (implied by "met the specifications at all time points")
    OdorOdorlessMet specification (implied by "met the specifications at all time points")
    Viscosity175 - 330 cPMet specification (implied by "met the specifications at all time points")
    Osmolality2600 – 2800 mOsm/kg, dilution factor of 5Met specification (implied by "met the specifications at all time points")
    pH3.5-5.5Met specification (implied by "met the specifications at all time points")
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Met specification (implied by "met the specifications at all time points")
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Met specification (implied by "met the specifications at all time points")
    Presence of pathogenic organisms (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans)Absent (USP<62>)Met specification (implied by "met the specifications at all time points")
    Antimicrobial effectivenessMeets USP<51> acceptance criteria for Category 2 products.Met specification (implied by "met the specifications at all time points")
    Biocompatibility: CytotoxicityNon-cytotoxic (per ISO 10993-5:2009/(R)2014)Non-cytotoxic
    Biocompatibility: Sensitization and IrritationNon-sensitizing, non-irritating (per ASTM D6355)Non-sensitizing, non-irritating
    Biocompatibility: Systemic ToxicityNot systemically toxic (per ISO 10993-11:2017)Not systemically toxic
    Shelf Life (8.5 months)Met all specifications in Table 1Met the specifications at all time points during accelerated aging study.
    Condom Compatibility: Natural Rubber Latex CondomsCompatible (per ASTM D7661-18)Compatible with natural rubber latex condoms.
    Condom Compatibility: Polyisoprene CondomsCompatible (per ASTM D7661-18)Compatible with polyisoprene condoms.
    Condom Compatibility: Polyurethane CondomsNot compatible (per ASTM D7661-18)Not compatible with polyurethane condoms.

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical performance testing for a personal lubricant, not an AI/ML device. Therefore, the concept of "test set" in the context of AI/ML is not applicable here. The testing involved samples of the product itself.

    • Biocompatibility testing: Not specified number of samples, but refers to ISO and ASTM standards.
    • Shelf Life Accelerated Aging Study: Not specified number of samples. The study duration was 3 months at 40°C.
    • Condom Compatibility: Not specified number of samples, but refers to ASTM D7661-18.
    • Data provenance: Not explicitly stated, but assumed to be from internal lab testing conducted by or for BioFilm, Inc. It is retrospective in the sense that the tests were completed before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a physical product, not an AI/ML device requiring expert ground truth for image or data interpretation.

    4. Adjudication method for the test set:

    Not applicable. This is a physical product, not an AI/ML device requiring adjudication for ambiguous cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a physical product, not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical product, not an AI/ML algorithm. The "standalone" performance here refers to the device's inherent physical and chemical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this product is based on established industry standards and validated test methods as defined by organizations like ISO, ASTM, and USP. For example:

    • Physical and Chemical Properties: The specifications (e.g., viscosity, pH, osmolality, color, odor, clarity, absence of particulate matter) are based on direct measurements against defined ranges.
    • Microbial Limits: Determined by specific USP <61> and <62> methods.
    • Antimicrobial Effectiveness: Determined by USP <51> methods.
    • Biocompatibility: Determined by ISO 10993-5, ISO 10993-11, and ASTM D6355 standards.
    • Condom Compatibility: Determined by ASTM D7661-18.

    8. The sample size for the training set:

    Not applicable. This is a physical product, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.