Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K181768

    Validate with FDA (Live)

    Date Cleared
    2018-08-20

    (48 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamine, Phencyclidine. Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine500 ng/mL
    Oxazepam300 ng/mL
    Cocaine150 ng/mL
    Marijuana50 ng/mL
    Methamphetamine500 ng/mL
    Morphine300 ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    EDDP300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL

    Configuration of the AssureTech Panel Dip Tests and the AssureTech Quick Cup Tests can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    For in vitro diagnostic use only.

    Device Description

    The AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Panel Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch

    AI/ML Overview

    The document describes the performance characteristics and studies for the AssureTech Panel Dip Tests and AssureTech Quick Cup Tests, which are qualitative drug detection devices for human urine.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative drug tests are typically based on the correct classification of samples, especially around the cutoff concentration. The reported device performance is presented as the percentage of correct results at various concentrations relative to the cutoff. For the purpose of this table, "Acceptance Criteria" will be interpreted as the expectation of high accuracy, particularly at concentrations significantly above or below the cutoff, and reasonable performance around the cutoff. The data provided primarily focuses on the "lay-user study" for all drugs listed.

    Drug (Cut-off level)Test TypeSample ConcentrationAcceptance Criteria (Implied)Reported Performance (Percentage of correct results)
    Amphetamine (AMP - 500 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (132 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (262 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (381 ng/mL)100% Negative100% Negative (20/20)
    +25% Cutoff (637 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (765 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (884 ng/mL)100% Positive100% Positive (20/20)
    Secobarbital (BAR - 300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (80 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (160 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (229 ng/mL)100% Negative90% Negative (18/20)
    +25% Cutoff (368 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (447 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (528 ng/mL)100% Positive100% Positive (20/20)
    Buprenorphine (BUP - 10 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (2.4 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (4.6 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (7.7 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (13.2 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (15.2 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (16.9 ng/mL)100% Positive100% Positive (20/20)
    Oxazepam (BZO - 300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (78 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (157 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (220 ng/mL)100% Negative100% Negative (20/20)
    +25% Cutoff (382 ng/mL)100% Positive100% Positive (20/20)
    +50% Cutoff (461 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (531 ng/mL)100% Positive100% Positive (20/20)
    Cocaine (COC - 150 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (39 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (80 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (115 ng/mL)100% Negative100% Negative (20/20)
    +25% Cutoff (192 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (221 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (268 ng/mL)100% Positive100% Positive (20/20)
    EDDP (300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (73 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (154 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (231 ng/mL)100% Negative90% Negative (18/20)
    +25% Cutoff (376 ng/mL)100% Positive100% Positive (20/20)
    +50% Cutoff (464 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (533 ng/mL)100% Positive100% Positive (20/20)
    Methylenedioxy-methamphetamine (MDMA - 500 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (122 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (265 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (386 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (633 ng/mL)100% Positive100% Positive (20/20)
    +50% Cutoff (768 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (890 ng/mL)100% Positive100% Positive (20/20)
    Methamphetamine (MET - 500 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (134 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (259 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (383 ng/mL)100% Negative90% Negative (18/20)
    +25% Cutoff (637 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (771 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (889 ng/mL)100% Positive100% Positive (20/20)
    Morphine (MOR - 300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (78 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (154 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (229 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (369 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (457 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (530 ng/mL)100% Positive100% Positive (20/20)
    Methadone (MTD - 300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (73 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (146 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (232 ng/mL)100% Negative90% Negative (18/20)
    +25% Cutoff (379 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (446 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (521 ng/mL)100% Positive100% Positive (20/20)
    Oxycodone (OXY - 100 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (28 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (48 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (79 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (129 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (147 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (179 ng/mL)100% Positive100% Positive (20/20)
    Phencyclidine (PCP - 25 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (8 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (14 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (21 ng/mL)100% Negative90% Negative (18/20)
    +25% Cutoff (30 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (39 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (48 ng/mL)100% Positive100% Positive (20/20)
    Propoxyphene (PPX - 300 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (71 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (152 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (228 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (382 ng/mL)100% Positive100% Positive (20/20)
    +50% Cutoff (455 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (522 ng/mL)100% Positive100% Positive (20/20)
    Nortriptyline (TCA - 1000 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (265 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (514 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (763 ng/mL)100% Negative100% Negative (20/20)
    +25% Cutoff (1279 ng/mL)100% Positive100% Positive (20/20)
    +50% Cutoff (1531 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (1782 ng/mL)100% Positive100% Positive (20/20)
    Marijuana (THC - 50 ng/mL)Lay User-100% Cutoff (0 ng/mL)100% Negative100% Negative (20/20)
    -75% Cutoff (14 ng/mL)100% Negative100% Negative (20/20)
    -50% Cutoff (27 ng/mL)100% Negative100% Negative (170/170)
    -25% Cutoff (39 ng/mL)100% Negative95% Negative (19/20)
    +25% Cutoff (60 ng/mL)High Positive95% Positive (19/20)
    +50% Cutoff (78 ng/mL)100% Positive100% Positive (40/40)
    +75% Cutoff (84 ng/mL)100% Positive100% Positive (20/20)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Analytical Performance (Precision, Specificity, Interference, Effect of Urine Specific Gravity and pH):

      • Sample Size: For precision studies, samples were prepared at various concentrations (-100% to +100% cutoff). Each concentration was tested with 3 lots of devices, 2 runs per day for 25 days, resulting in 50 individual tests per concentration per lot (2 runs/day * 25 days = 50). This totals approximately 400 tests per drug per lot per device type (8 concentrations * 50 tests).
      • For Interference, Specificity, and Effect of Urine Specific Gravity/pH studies, urine samples were tested using three lots of each device at 25% below and 25% above Cut-Off levels (and drug-free urine for interference). Specific sample numbers are not explicitly stated for these sub-studies but imply multiple tests per lot.
      • Data Provenance: The analytical studies were performed in-house, meaning within the manufacturer's or an affiliated laboratory. The data is thus prospective for the purpose of these device validations. The "country of origin" is not explicitly stated for the source of base urine samples, but the manufacturer is based in Hangzhou, China. The drugs were "spiked" into negative samples.
    • Comparison Studies (with GC/MS):

      • Sample Size: 80 unaltered clinical samples (40 negative and 40 positive) for each drug (AMP500, COC150, MET500) were used for each device (Panel Dip and Quick Cup).
      • Data Provenance: The samples were "unaltered clinical samples," implying these were real-world samples. The study was performed in-house. The country of origin of these clinical samples is not specified, but given the manufacturer's location, they are likely from China or a region where the manufacturer has access to clinical samples. This is a retrospective analysis against a gold standard.
    • Lay-user study:

      • Sample Size: 310 lay persons for each device format (Panel Dip and Quick Cup). Urine samples were prepared at various concentrations (-100%, -75%, -50%, -25%, +25%, +50%, +75% of the cutoff). The exact breakdown of how these 310 lay users were distributed across the different concentrations is shown in the tables (e.g., for AMP, 20 samples at -100% Cutoff, 170 at -50% Cutoff, etc., which sums to 310).
      • Data Provenance: The urine samples were "spiked drugs into drug free-pooled urine specimens." This indicates the samples were laboratory-prepared, not unaltered clinical samples. The study was conducted at "three intended user sites," but their location (country) is not specified. This is a prospective study using simulated clinical samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Analytical Performance (Precision, Specificity, Interference): Ground truth established by LC/MS (Liquid Chromatography/Mass Spectrometry), which is an analytical chemical method, not by human experts. The drug concentrations in spiked samples were confirmed by LC/MS.
    • Comparison Studies (with GC/MS): Ground truth established by GC/MS (Gas Chromatography/Mass Spectrometry), which is the gold standard confirmatory method for drug testing. This is a laboratory-based analytical method, not human expert interpretation.
    • Lay-user study: Ground truth established by LC/MS to confirm the drug concentrations in the spiked urine samples.

    It's important to note that for these in vitro diagnostic devices, the "experts" in establishing ground truth are typically the highly precise analytical instruments (GC/MS, LC/MS) not human readers as would be the case for imaging studies.

    4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

    • Analytical Performance & Lay-user study: No adjudication method explicitly described as the ground truth was an objective measurement by LC/MS. The results are compared directly to the known spiked concentrations confirmed by LC/MS.
    • Comparison Studies (with GC/MS): No adjudication method explicitly described. The device results were compared directly against the GC/MS results. The "Viewers" (human operators) for the dip cards/quick cups for AMP, COC, and MET showed some discordant results with GC/MS, but there is no mention of a process to adjudicate these discrepancies among the viewers or with the GC/MS result.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. The device is a rapid test (lateral flow immunoassay) which relies on visual interpretation of lines, not complex image analysis or AI.
    • The "Comparison Studies" involved three "laboratory assistants" (referred to as Viewer A, B, C) manually interpreting the device results against GC/MS. This is a multi-reader study, but it's not a comparative effectiveness study of human readers with vs. without AI. It assesses the consistency of human interpretation of the device results.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. The devices are qualitative lateral flow immunoassays designed for human visual interpretation. They are not AI algorithms; their performance is intrinsically tied to human reading. The "Performance Characteristics" section covers the analytical performance of the device components themselves, independent of human interpretation to a logical degree (e.g., cross-reactivity, stability).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The ground truth used for all performance studies (Analytical Performance, Comparison Studies, and Lay-user study) was analytical confirmation by either LC/MS or GC/MS. These are highly sensitive and specific laboratory methods considered the gold standard for drug detection and quantification in urine.

    8. The sample size for the training set

    • The document describes the performance of a rapid diagnostic test device (immunoassay), not a machine learning or AI algorithm that typically requires a distinct training set. Therefore, there is no specific "training set" as understood in the context of AI/ML development mentioned for the device itself.
    • The device's internal parameters (e.g., antibody binding characteristics) would be developed and optimized by the manufacturer during product development, but this process doesn't constitute a "training set" in the AI sense.

    9. How the ground truth for the training set was established

    • As there is no "training set" for an AI/ML algorithm mentioned, this question is not applicable. The device's performance is based on its inherent chemical and immunological properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180349

    Validate with FDA (Live)

    Date Cleared
    2018-04-05

    (56 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamine, Phencyclidine. Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine1000 ng/mL
    Oxazepam300 ng/mL
    Cocaine300 ng/mL
    Marijuana50 ng/mL
    Methamphetamine1000 ng/mL
    Morphine300 ng/mL or 2000 ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    EDDP300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL

    Configuration of the AssureTech Panel Dip Tests and the AssureTech Quick Cup Tests can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    For in vitro diagnostic use only.

    Device Description

    The AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Panel Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch

    AI/ML Overview

    The provided document describes the validation of the AssureTech Panel Dip Tests and AssureTech Quick Cup Tests for qualitative and simultaneous detection of various drugs in human urine. The study presented focuses on the analytical performance, comparison with a predicate device, and a lay-user study.

    Here's an analysis of the acceptance criteria and study proving the device meets them:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in a dedicated table, but are implied by the results of the precision studies and lay-user studies demonstrating high percentages of correct results around the cutoff concentrations. For the lay-user study, the implicitly accepted performance is 90-100% correct results across various drug concentrations, particularly at and near the cutoff levels.

    Here's a table summarizing the reported device performance, based on the lay-user study, for each tested drug, focusing on the "The percentage of correct results (%)" column. Note that the precision studies showed similar performance for different lots and concentration ranges.

    Table of Acceptance Criteria (Implied) and Reported Device Performance (Lay-User Study)

    Drug (Identifier)Cut-off level (ng/mL)Sample Conc. (% of Cutoff)Implied Acceptance Criteria (Min. % Correct)Reported Device Performance (Min. % Correct across all concentrations tested)
    Amphetamine (AMP)1000-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Secobarbital (BAR)300-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Cocaine (COC)300-100%, -75%, -50%100%100%
    -25%90%90%
    +25%90%95%
    +50%, +75%100%100%
    Buprenorphine (BUP)10-100%, -75%, -50%100%100%
    -25%90%(Missing data)
    +25%90%(Missing data)
    +50%, +75%100%100%
    Methamphetamine (MET)1000-100%, -75%, -50%100%100%
    -25%90%90%
    +25%90%95%
    +50%, +75%100%100%
    Methadone (MTD)300-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Morphine (MOR)300/2000-100%, -75%, -50%100%100%
    -25%90%90%
    +25%90%90%
    +50%, +75%100%100%
    Oxycodone (OXY)100-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Phencyclidine (PCP)25-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Marijuana (THC)50-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Oxazepam (BZO)300-100%, -75%, -50%100%100%
    -25%90%90%
    +25%90%100%
    +50%, +75%100%100%
    Methylenedioxy-methamphetamine (MDMA)500-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    Nortriptyline (TCA)1000-100%, -75%, -50%100%100%
    -25%90%95%
    +25%90%90%
    +50%, +75%100%100%
    EDDP300-100%, -75%, -50%100%100%
    -25%, +25%90%95%
    +50%, +75%100%100%
    d-Propoxyphene (PPX)300-100%, -75%, -50%100%100%
    -25%90%100%
    +25%90%95%
    +50%, +75%100%100%

    Note on Buprenorphine (BUP) results: The table for BUP in the document has missing values ("С Г لك ه" and blank for "No. of Negative"). This prevents reporting the exact percentages for those concentrations. However, the existing data for BUP still shows high performance where reported.

    2. Sample Size Used for the Test Set and Data Provenance

    • Lay-user Study (Primary Test Set):

      • Sample Size: 310 lay persons for each device format (Panel Dip Tests and Quick Cup Tests). For each drug, there were varying numbers of samples at different concentrations: 20 samples at -100%, -75%, -25%, +25%, +75% cutoff, and 170 samples at -50% cutoff, and 40 samples at +50% cutoff.
      • Data Provenance: The origin of the urine samples (e.g., country of origin) is not specified. The study was conducted "in-house" (meaning by the manufacturer or their designated testing facility) and appears to be prospective in nature, as samples were prepared by spiking drugs into negative urine specimens, aliquoted, and then distributed blindly to users.
    • Comparison Studies (Clinical Samples):

      • Sample Size: For each drug, 80 unaltered clinical samples (40 negative and 40 positive). These were tested by three laboratory assistants for each device.
      • Data Provenance: The origin (e.g., country) of these clinical samples is not specified. They are described as "clinical samples," implying they were collected from real patients, making this a retrospective evaluation of existing samples.
    • Precision Studies:

      • Sample Size: For each drug concentration (-100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off, +100% cut off), tests were performed in two runs per day for 25 days, for a total of 50 tests per concentration per lot. With 3 lots tested, this amounts to 150 tests per concentration.
      • Data Provenance: Samples were "prepared by spiking drug in negative samples," indicating these were lab-prepared samples, making it a prospective controlled study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Lay-user Study Ground Truth: The ground truth for the lay-user study samples was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is a gold standard analytical method for drug concentration determination. The document states: "The concentrations of the samples were confirmed by LC/MS." No human experts are explicitly mentioned as establishing the ground truth for this set beyond the technical personnel operating the LC/MS.

    • Comparison Studies Ground Truth: The ground truth for the clinical samples in the comparison studies was also established by LC/MS. The document states: "The samples were blind labeled and compared to LC/MS results." Similarly, no human expert consensus or interpretation was required for establishing the ground truth for these samples, as LC/MS provides a quantitative, objective measure.

    • Precision Studies Ground Truth: The ground truth for these spiked samples was also established by LC/MS to confirm the prepared drug concentrations.

    4. Adjudication Method for the Test Set

    • Lay-user Study: No adjudication method is described. Each lay participant evaluated their assigned blind-labeled sample independently. The results were then compared to the LC/MS ground truth.

    • Comparison Studies: No adjudication method is described. Three laboratory assistants independently ran the samples, and their results were compared directly to the LC/MS results. Discordant results are individually listed but no process for reconciling them or for a consensus reading is mentioned using these three viewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.

    • No MRMC study involving human readers improving with AI vs. without AI assistance was done. This device is a rapid, lateral flow immunochromatographic assay, not an AI-assisted diagnostic tool requiring human interpretation with or without AI. The "readers" in the comparison study are laboratory assistants interpreting the dip/cup test results, not radiologists or similar specialists interpreting complex images, and no AI component is involved for assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.

    • Not applicable. The device itself is the "algorithm" in a sense, as it performs the chemical reaction to yield a positive or negative result. The studies assess the performance of this device. There is no separate, purely computational algorithm. The device's performance is inherently "standalone" in how it chemically processes the sample and displays a result, which is then visually interpreted by a human user (whether a lab assistant or a layperson).

    7. The Type of Ground Truth Used

    • The primary type of ground truth used for both the precision studies, comparison studies, and the lay-user study was analytical confirmation by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate and objective method for determining the concentration of drugs in a sample, which then defines whether a sample is truly positive or negative relative to a specific cutoff.

    8. The Sample Size for the Training Set

    • Not applicable / Not explicitly stated. This is a chemical assay, not a machine learning model, so there isn't a "training set" in the conventional AI/ML sense. The device's components and parameters are developed through R&D, not through training on a dataset. The results presented are for validation and performance evaluation.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no "training set" for an AI/ML model for this type of device, the concept of establishing ground truth for a training set doesn't apply. The device's underlying chemistry and immunoassay principles are based on established scientific knowledge.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1