Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K181565

    Validate with FDA (Live)

    Date Cleared
    2018-07-13

    (29 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AQUARIUS 8600 is intended for use by a qualified doctor or technologist on both adult and paediatric patients for taking diagnostic radiographic exposures of all body parts of the AQUARIUS 8600 provides digital image capture and is intended to replace radiographic film/screen. The x-ray tube and associated equipment are not provided with the proposed sensor. Prescription use only.

    The AQUARIUS 8600 is not intended for mammography.

    Device Description

    The Aquarius 8600 is a digital radiography sensor which automatically collects x-ray images from an x-ray source. The Aquarius 8600 sensor (flat panel type) collects x-rays and digitizes the images for their transfer and display to a computer. The sensor does not have an x-ray source, which is provided by independent manufacturers. The sensor includes with a flat panel for x-ray acquisition and digitization and a computer (including proprietary processing software) for processing, annotating and storing x-ray images.

    AI/ML Overview

    This is a 510(k) premarket notification for the AQUARIUS 8600 Digital Radiography Sensor. The core of this submission is to demonstrate the substantial equivalence of the modified device to a previously cleared predicate device.

    Based on the provided text, the acceptance criteria and the study proving the device meets those criteria are established through a non-clinical performance evaluation and image comparison rather than a human-in-the-loop clinical trial or a detailed algorithm performance study with predefined metrics like sensitivity/specificity. This is typical for certain types of device modifications or new devices where the primary claim is equivalence in image quality or basic functionality to an existing device.

    Here's the breakdown of the information requested, based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of numerical performance thresholds (e.g., specific sensitivity, specificity, or image quality scores). Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to predicate devices in terms of:

    Acceptance Criteria CategoryReported Device Performance and Justification
    Image Quality and Characteristics"The Aquarius 8600 produces images of similar quality and characteristics that are equivalent to those of the both the Aquarius 8600 and BIOK4600 predicate devices."
    Hardware Performance"Because the sensor hardware is the same as the Aquarius 8600 predicate device hardware, the sensor has the same performance, biocompatibility, effectiveness, thermal, electrical and mechanical safety and is substantially equivalent to the predicate device."
    Software Functionality"The Aquarius 8600 software functionality is equivalent to the original BIOK4600 predicate device, except for the Sensor Driver interface application which reads the images from the Aquarius 8600 flat panel instead of the BIOK 4600 sensor."
    Conformity to Standards"The design, development and production of the sensor conforms to 892.1680 and ISO 13485 quality systems."
    Technical Specifications (Physical/Operational)Spatial resolution: 3.9 lp/mmOptical resolution: 3.9 lp/mmAcquisition to display time: < 2 secNetwork interface: 1000 mbps LANImage and file formats: DICOM compliantPower: 24Vdc, 1.9A max (18W typ)These specs implicitly met by equivalence to predicate.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document mentions "Test images have been submitted along with the equivalent images from the Aquarius 8600 predicate device." However, it does not specify the number of images or cases in this test set.
    • Data Provenance: The provenance (country of origin, retrospective/prospective) of these "test images" is not specified in the provided text. It's likely that these were images generated internally or from previous testing since the hardware is effectively the same as a predicate.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts in a formal ground truth establishment process for a test set. The assessment appears to be based on a direct comparison of physical image attributes and technical specifications against the predicate devices, rather than a clinical interpretation by experts.

    4. Adjudication method for the test set

    Given that formal expert review to establish ground truth is not mentioned, there is no stated adjudication method. The comparison seems to be based on the manufacturer's internal assessment of image similarity and technical equivalence.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or reported. This device is a digital radiography sensor (hardware and basic image acquisition/processing software), not an AI-powered diagnostic aid for interpretation. Therefore, the concept of human readers improving with AI assistance is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A "standalone" algorithm performance study in the sense of sensitivity, specificity, and other performance metrics calculated by an algorithm on a dataset was not performed or reported. The evaluation focused on the technical equivalence of the image acquisition system to previously cleared devices. The "algorithm" here is primarily within the proprietary processing software for image acquisition and digitization, not a diagnostic AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically used for AI diagnostics (e.g., for disease detection) is not applicable here. The "ground truth" in this context is implicitly the established performance and image characteristics of the predicate devices. The new device is considered "equivalent" if its images and functionality align with what the predicate devices produce.

    8. The sample size for the training set

    This device is not an AI algorithm that undergoes "training" on a dataset in the typical machine learning sense. It's a digital radiography sensor with associated image processing software. Therefore, there is no concept of a training set in the provided document.

    9. How the ground truth for the training set was established

    As there is no training set for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173273

    Validate with FDA (Live)

    Date Cleared
    2017-11-09

    (28 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts on both adult and pediatic patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography

    Device Description

    This device is a medical x-ray image acquisition device. X-rays generated by X-ray generator/ tube that penetrate patient's body are converted to a digital file by the detector. After that the detector sends this digital file to a pc where the companion software has been installed. This software was cleared in our previous submission, K170202. A monitor displays this image. Images can then be transferred via the DICOM protocol. This device is nearly identical to the predicate but now the digital panels have Wi-Fi in addition to Ethernet interfaces. The Aquarius 8600 will be marketed in two possible configurations: Aquarius 8600 1717WCl (tethered or wireless), 17 x 17 inch flat panel as a retrofit package with Magellan software). Aquarius 8600 1417WCl (tethered or wireless), 14 x 17 inch flat panel as a retrofit package with Magellan software).

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a stationary x-ray system (Aquarius 8600 1417WCI; Aquarius 8600 1717WCI). It describes the device, its similarity to a predicate device, and the non-clinical tests conducted to establish substantial equivalence.

    Based on the provided information, no specific acceptance criteria or a dedicated study proving the device meets those criteria are explicitly detailed in a format that would allow for a complete answer to all parts of your request.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Aquarius 8600 1417TC/1717TC, K170202) and references a cleared flat panel detector (K160810). The testing described is primarily non-clinical.

    Here's an analysis of what information can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance

    No explicit table of acceptance criteria is provided. The document states "Performance characteristics" are similar to the predicate and provides some technical specifications (MTF, DQE, Panel Resolution, Pixel Size) for both the predicate and the new device. It notes that the new panels have "slightly better MTF" and "slightly lower DQE," implying these performance characteristics were measured against the predicate's performance rather than predefined acceptance criteria.

    CharacteristicPredicate Device Performance (Aquarius 8600 1417TC/1717TC K170202)New Device Performance (Aquarius 8600 1417WCI; Aquarius 8600 1717WCI)Acceptance Criteria
    Panel CommunicationTethered Gigabit EthernetTethered Gigabit Ethernet or Wireless via previously cleared panel (K160810)Not explicitly stated, implied to be equivalent or enhanced wireless capability is acceptable.
    Sensor TypeAmorphous Silicon, TFTAmorphous Silicon, TFTSame as predicate, implying acceptance.
    ScintillatorCsI: TICsI: TISame as predicate, implying acceptance.
    Panel Resolution2816 x 3328 or 3328 x 33282304 x 2800 or 3072 x 3072Not explicitly stated, "NOT A MEANINGFUL DIFFERENCE" is asserted.
    Panel Size14x17 inches or 17x17 inches14x17 inches or 17x17 inchesSame as predicate, implying acceptance.
    Pixel Size127 μm150 μm (14x17) or 139 μm (17x17)Not explicitly stated, "NOT A MEANINGFUL DIFFERENCE" is asserted.
    Resolution (lp/mm)3.9 lp/mm3.59 lp/mm or 3.33 lp/mmNot explicitly stated, "NOT A MEANINGFUL DIFFERENCE" is asserted.
    Image Depth14 bits14 bitsSame as predicate, implying acceptance.
    Preview Image2 secondsLess than 3 seconds"EQUIVALENT" asserted, implying acceptance for slight increase.
    MTF @1 lp/mm14 x 17: 0.60817 x 17: 0.51714 x 17: 0.66717 x 17: 0.706"The new panels have slightly better MTF," implying this improved performance is acceptable.
    DQE (0)14 x 17: 0.7417 x 17: 0.6814 x 17: 0.4017 x 17: 0.53"The new panels have slightly lower DQE," implying this reduced performance is deemed acceptable for substantial equivalence.
    Safety/EMCEN/IEC 60601-1, SafetyEN/IEC 60601-1-2 EMCSAME (Tested to IEC 60601-1:2012, IEC 60601-1-2:2007)Compliance with standards, implying acceptance.

    2. Sample size used for the test set and the data provenance

    The document mentions "Clinical images were provided," but explicitly states "these images were not necessary to establish substantial equivalence based on the modifications to the device... but they provide further evidence... that the complete system works as intended."

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as the clinical images were not presented as part of a formal study with ground truth established by experts for performance evaluation against a specific clinical claim. They were "further evidence" of system functionality.

    4. Adjudication method for the test set

    Not applicable, as no formal clinical study with ground truth and expert adjudication is described for the purpose of establishing device performance against acceptance criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an x-ray imagine acquisition device, not an AI-powered diagnostic tool. No AI component is mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an x-ray imagine acquisition device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as no formal clinical study with ground truth for performance evaluation is described. For the non-clinical bench testing, the "ground truth" would be the physical properties and measurements of the device components themselves, following established engineering and physics principles.

    8. The sample size for the training set

    Not applicable. This is a hardware device (x-ray panel and system), not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a hardware device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171169

    Validate with FDA (Live)

    Date Cleared
    2017-10-12

    (174 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts on both adult and pediatic patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.

    Device Description

    The Aquarius 8600 1717TG and Aquarius 8600 1417TG are digital flat panels, specifically termed solid state digital X-Ray detector. This technology couples a scintillator with an a-Si TFT sensor, and through integration with a radiographic imaging system, x-ray images can be captured and digitalized. The resulting RAW files are DICOM 3.0 compatible allowing image files to be processed by IDC Magellan software.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Performance Metric)Reported Device Performance (Aquarius 8600 1417TG & 1717TG)Predicate Device Performance (Aquarius 8600 1717TC)Notes
    Detective Quantum Efficiency (DQE) (0)0.4460.684The proposed devices show slightly reduced DQE performance at all spatial frequencies compared to the predicate, with the difference increasing with spatial frequency. This is attributed to the change in scintillator.
    Modulation Transfer Function (MTF) at 1 lp/mm0.596 (1417TG), 0.585 (1717TG)0.502The proposed devices show similar but slightly improved MTF response compared to the predicate device.
    Modulation Transfer Function (MTF) at 2 lp/mm0.290 (1417TG), 0.283 (1717TG)0.230The proposed devices show similar but slightly improved MTF response compared to the predicate device.
    Modulation Transfer Function (MTF) at 3 lp/mm0.142 (1417TG), 0.144 (1717TG)0.104The proposed devices show similar but slightly improved MTF response compared to the predicate device.
    Modulation Transfer Function (MTF) at 3.5 lp/mm0.095 (1417TG), 0.103 (1717TG)0.083The proposed devices show similar but slightly improved MTF response compared to the predicate device.
    Noise Power Spectrum (NPS) at 0 lp/mm8.568.10The proposed devices have a similar noise performance profile at spatial frequencies compared to the predicate device.
    Noise Power Spectrum (NPS) at 1 lp/mm4.213.20The proposed devices have a similar noise performance profile at spatial frequencies compared to the predicate device.
    Noise Power Spectrum (NPS) at 2 lp/mm1.431.20The proposed devices have a similar noise performance profile at spatial frequencies compared to the predicate device.
    Noise Power Spectrum (NPS) at 3 lp/mm0.640.60The proposed devices have a similar noise performance profile at spatial frequencies compared to the predicate device.
    Noise Power Spectrum (NPS) at 3.5 lp/mm0.520.55The proposed devices have a similar noise performance profile at spatial frequencies compared to the predicate device.
    Environmental, electrical, mechanical safetyAll testing passed (based on IEC 60601-1, IEC 60601-1-2)Not explicitly compared here, but predicate likely met similar standardsCompliance with international standards for medical electrical equipment and electromagnetic compatibility.
    Software lifecycle and validationDocumented lifecycle based on IEC 62304, full verification, validation, and regression testing performedNot explicitly compared here, but predicate likely met similar standardsAdherence to FDA guidance for software in medical devices, indicating robust software development practices.
    Diagnostic image quality (visual assessment)Images diagnostically similar, and slightly superior, to the predicate device(Reference: predicate images)Laboratory images using phantoms were reviewed by a certified Radiological Technologist.

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a number of images or cases in the provided text for the DQE, MTF, NPS, or visual assessment tests. For the visual assessment, it mentions "laboratory images using phantoms were acquired." These are likely a set of standardized phantom images.
      • Data Provenance: The data for DQE, MTF, and NPS are from "measured" curves, implying direct testing of the device hardware. The visual assessment used "laboratory images using phantoms." This suggests the data is prospectively generated from controlled laboratory settings, not from patient data. The country of origin for the testing is not specified, but the applicant company is located in Canada.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: One (a single Radiological Technologist).
      • Qualifications: "A Radiological Technologist certified in the United States of America and Canada." Specific experience level (e.g., years) is not provided.
    3. Adjudication method for the test set:

      • No adjudication method is described for the visual assessment. A single expert made the determination without mention of a consensus or tie-breaking process. For the DQE, MTF, and NPS, these are quantitative measurements that do not require expert adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. The submission states: "No clinical testing was performed for this special 510(k) submission." The visual assessment mentioned was a comparison of phantom images by a single technologist, not a clinical study involving multiple readers assessing patient cases. Also, this device is an X-ray detector, not an AI-assisted diagnostic tool, so improvement with AI assistance is not applicable in this context.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is not applicable as the device is a digital X-ray detector, not a standalone algorithm. Its performance is measured directly through physical parameters (DQE, MTF, NPS) and its ability to produce diagnostic images.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the DQE, MTF, and NPS measurements, the "ground truth" is generally considered to be the intrinsic physical properties of the detector, measured against established standards and methodologies.
      • For the visual assessment, the "ground truth" was the subjective opinion of a single certified Radiological Technologist comparing images from the proposed device to the predicate device using phantoms.
    7. The sample size for the training set:

      • Not applicable/Not provided. This device is a hardware component (digital flat panel detector) and associated software for image acquisition and processing. It does not employ machine learning or AI models that require specific "training sets." The IDC Magellan software underwent "full system level verification, validation and regression testing" as part of its development, but this is software testing, not ML model training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no specific "training set" in the context of machine learning model development for this device. The software validation relies on established software engineering principles and testing against specifications inherent to its function.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171175

    Validate with FDA (Live)

    Date Cleared
    2017-10-05

    (167 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.

    Device Description

    The Aquarius 8600 1417WC is a digital flat panel (specifically termed solid state digital X-Ray detector), with the option for users to use the detector in either a tethered or wireless mode. This technology couples a scintillator with an a-Si TFT sensor, and through integration with a radiographic imaging system, x-ray images can be captured and digitalized. The resulting RAW files are DICOM 3.0 compatible allowing image files to be processed by IDC Magellan software.

    AI/ML Overview

    This 510(k) submission describes the Aquarius 8600 1417WC, a digital flat panel X-ray detector. It is a modification of a previously cleared device (Aquarius 8600 1717TC) and integrates a 510(k)-cleared tethered/wireless flat panel detector with existing software and workstation components. The submission focuses on demonstrating substantial equivalence to the predicate device.

    Here's an analysis of the provided text in the context of acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly define "acceptance criteria" in a quantitative manner for specific diagnostic tasks. Instead, it compares the technological characteristics and performance metrics of the proposed device (Aquarius 8600 1417WC) with those of the predicate device (Aquarius 8600 1717TC), aiming to demonstrate "similar" or "better" performance, thereby supporting substantial equivalence.

    CharacteristicAcceptance Criterion (Implicitly "Similar to or Better Than Predicate")Reported Device Performance (Aquarius 8600 1417WC)Predicate Performance (Aquarius 8600 1717TC)
    DQE (0)≥ Predicate DQE (0)0.8170.684
    MTF (1 lp/mm)≥ Predicate MTF (1 lp/mm)0.5570.517
    MTF (2 lp/mm)≥ Predicate MTF (2 lp/mm)0.2800.230
    MTF (3 lp/mm)≥ Predicate MTF (3 lp/mm)0.1570.123
    MTF (3.5 lp/mm)≥ Predicate MTF (3.5 lp/mm)0.1180.088
    NPS Profile"Similar noise performance profile"Reported NPS curves (e.g., 19.15 at 0 lp/mm)Reported NPS curves (e.g., 8.01 at 0 lp/mm)
    Resolution"Same or better resolution performance" (based on MTF comparison)"Same or better resolution performance"Referenced by MTF values
    Image Quality (Diagnostic Similarity)Produces images diagnostically similar to the predicate device"produces images that are diagnostically similar"Implicitly the images from the predicate device
    Environmental, Electrical, Mechanical SafetyAll testing passed based on IEC 60601-1 and IFC 60601-1-2All testing passedNot explicitly stated, but implied as predicate is already cleared
    Software Lifecycle/ValidationDocumented lifecycle, design, requirements, verification, validation, regression testing per FDA/IEC guidanceDocumentation and testing performedNot explicitly stated, but implied as predicate is already cleared

    2. Sample size used for the test set and the data provenance:

    • Sample Size for DQE, MTF, NPS: Not explicitly stated as a numerical count of images or measurements. These are physical characteristics of the detector itself, typically measured under controlled laboratory conditions, not on a "test set" of patient images in the traditional sense.
    • Sample Size for Diagnostic Similarity: "Laboratory images using phantoms were acquired with the proposed Aquarius 8600 1417WC detector and compared to images acquired with the Aquarius 1717TC predicate device." The number of phantoms or images is not specified.
    • Data Provenance: The DQE, MTF, and NPS measurements are laboratory-derived data. The "laboratory images using phantoms" are also laboratory-derived. There is no indication of country of origin for the data; it would likely be from the manufacturer's testing facilities. The tests are non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: "A Radiological Technologist certified in the United States of America and Canada has reviewed the images (included in this submission)..." This indicates one radiological technologist was involved.
    • Qualifications of Experts: "Radiological Technologist certified in the United States of America and Canada." (No mention of years of experience or specialization beyond general radiology).

    4. Adjudication method for the test set:

    • No formal adjudication method (e.g., 2+1, 3+1) is described for the image review to establish diagnostic similarity. It appears to be a single reviewer's assessment.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. The device is a digital X-ray detector, not an AI-assisted diagnostic tool. The comparison is between the performance of the proposed detector and a predicate detector, not between human readers with and without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, in spirit, for the detector's physical performance. The DQE, MTF, and NPS measurements represent the standalone performance of the detector hardware. While a human is involved in analyzing the output (e.g., in phantom image review), the core metrics (DQE, MTF, NPS) characterize the intrinsic performance of the detector itself, independent of a diagnostic interpretation by a human.

    7. The type of ground truth used:

    • For DQE, MTF, NPS: These are objective physical performance metrics measured under controlled laboratory conditions using established methodologies. The "ground truth" is the accurately measured physical response of the detector.
    • For Diagnostic Similarity: The "ground truth" is essentially the predicate device's image quality, which is already considered diagnostically acceptable. The expert's role was to confirm that the proposed device's images (of phantoms) were "diagnostically similar" to those of the predicate device. This is a form of expert consensus/comparison against an established benchmark rather than an independent "ground truth" for disease detection.

    8. The sample size for the training set:

    • Not applicable. This submission is for a digital X-ray detector, which captures images. It does not describe an AI/machine learning model that would require a "training set" of images to learn from. The software (Magellan 3) processes and displays images but is not an AI algorithm in the context of typical training sets.

    9. How the ground truth for the training set was established:

    • Not applicable. As no training set for an AI/ML model is described, there's no mention of how ground truth for such a set would be established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170202

    Validate with FDA (Live)

    Device Name
    Aquarius 8600
    Date Cleared
    2017-02-22

    (30 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.

    Device Description

    The Aquarius 8600 1717TC are digital flat panels, specifically termed solid state digital X-Ray detector. This technology couples a scintillator with an a-Si TFT sensor, and through integration with a radiographic imaging system, x-ray images can be captured and digitalized. The resulting RAW files are DICOM 3.0 compatible allowing image files to be processed by IDC Magellan software.

    The Aquarius 8600 1717TC and 1417TC represents a modification of our own predicate device cleared under K070079, X3C Digital Radiographic Detector.

    The Aquarius 8600 1717TC and 1417TC integrates the 510(k) cleared flat panel detectors (K122173 and K122919) with IDC Magellan software and workstation. There were no changes made to the cleared panels, workstation or software. All components were integrated and tested to make the Aquarius 8600 1717TC and 1417TC medical devices.

    The Aquarius 8600 will be marketed in two possible configurations:
    Aquarius 8600 1717TC (tethered, 17 x 17 inch flat panel as a retrofit package with Magellan software) referred to as "1717TC" throughout the rest of this document
    Aquarius 8600 1417TC (tethered. 14 x 17 inch flat panel as a retrofit package with Magellan software) referred to as "1417TC" throughout the rest of this document.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Aquarius 8600, a digital flat panel X-ray detector. However, it does not include information about a study that proves a device meets acceptance criteria related to AI/Machine Learning device performance, as there is no AI component mentioned in this submission. The submission is for a radiographic imaging system and focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data (DQE, MTF, NPS), safety testing, and software lifecycle documentation.

    Therefore, many of the requested items (e.g., sample size for test set, data provenance, ground truth establishment, expert adjudication, MRMC study, standalone performance against ground truth, training set details) are not applicable to this specific submission, as they relate to the validation of an AI/ML algorithm rather than a traditional medical imaging hardware component.

    I will fill in the relevant information that is present in the document and explicitly state where information is not applicable or not provided.


    Acceptance Criteria and Device Performance (Aquarius 8600)

    The acceptance criteria here are implicitly defined by demonstrating "substantial equivalence" to a predicate device (IDC X3C Digital Radiographic Detector) through comparable performance metrics and safety standards. The study primarily relies on non-clinical (bench) testing.

    1. Table of Acceptance Criteria (based on predicate equivalence) and Reported Device Performance

    Characteristic / Acceptance Criteria (vs. Predicate)Predicate Value (X3C)Aquarius 8600 (1717TC & 1417TC) Performance
    Detective Quantum Efficiency (DQE)
    DQE (0)0.6410.684
    Acceptance Comment:Similar performance at all spatial frequencies, with proposed device having slightly better DQE(0).
    Modulation Transfer Function (MTF)
    MTF at 1 lp/mm0.3850.517
    MTF at 2 lp/mm0.1340.230
    MTF at 3 lp/mm0.0750.123
    MTF at 3.5 lp/mm0.0630.063
    Acceptance Comment:Same or better resolution performance at all spatial frequencies.
    Noise Power Spectrum (NPS)
    NPS at 0 lp/mm2.218.01
    NPS at 1 lp/mm0.983.30
    NPS at 2 lp/mm0.821.27
    NPS at 3 lp/mm0.820.64
    NPS at 3.5 lp/mm0.810.58
    Acceptance Comment:Similar noise performance profile at spatial frequencies, but specific values differ.
    Environmental, Electrical, Mechanical SafetyIEC 60601-1 ed. 3.0Passed IEC 60601-1 ed. 3.0
    EMC TestingIEC 60601-1-2:2007Passed IEC 60601-1-2:2007
    Software LifecycleIEC 62304 ed. 1.1:2015Documented as per IEC 62304 ed. 1.1:2015
    Image Quality (Subjective Review)Baseline with X3CDiagnostically similar, slightly superior, using phantoms

    Note: The increase in NPS values for the Aquarius 8600 at lower spatial frequencies compared to the X3C detector is noted in the table but the document states they have "a similar noise performance profile". This suggests that while raw values may differ, the overall expected noise characteristics for diagnostic purposes were deemed acceptable for substantial equivalence.

    2. Sample size used for the test set and the data provenance:

    • Test Set: Non-clinical (phantom) images were used. The document does not specify an exact number of phantom images or specific test scenarios beyond mentioning "Laboratory images using phantoms were acquired."
    • Data Provenance: The testing was conducted by Imaging Dynamics Company, Ltd. (IDC) (Canada). The nature of the non-clinical testing (phantoms) means there's no patient data provenance to specify. The study is a pre-market submission for substantial equivalence, relying on bench tests, rather than a clinical trial with retrospective/prospective patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: One expert.
    • Qualifications: A "Radiological Technologist certified in the United States of America and Canada." (Note: This is a technologist, not a radiologist, and their role was to review images for diagnostic similarity, not to establish a clinical "ground truth" from patient data.)

    4. Adjudication method for the test set:

    • Adjudication Method: None explicitly mentioned or applicable beyond the single radiologic technologist's review of phantom images. This was not a multi-reader clinical study requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. The submission explicitly states: "No clinical testing was performed for this special 510(k) submission." The evaluation was primarily based on non-clinical performance parameters (DQE, MTF, NPS) and a subjective review of phantom images by one technologist.
    • Effect Size of Human Readers Improvement: Not applicable, as no MRMC study with human-in-the-loop AI assistance was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable. This device is a digital X-ray detector, which is hardware, not an AI algorithm. Its "performance" is measured by physics-based parameters (DQE, MTF, NPS) derived from phantom images, not by an algorithm's diagnostic output against a clinical ground truth.

    7. The type of ground truth used:

    • Type of Ground Truth: For the quantitative performance metrics (DQE, MTF, NPS), the "ground truth" is established by the standardized measurement methodologies described in relevant international standards and best practices for X-ray detector characterization. For the qualitative image review, the "ground truth" was a subjective assessment by a certified technologist comparing images from the new device to the predicate using phantoms. No clinical "ground truth" (e.g., pathology, clinical outcomes) from patient data was used.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This is a hardware device; there is no AI/ML algorithm that requires a "training set" in the context of this 510(k) submission.

    9. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1