Search Results
Found 8 results
510(k) Data Aggregation
(197 days)
The system is intended to be used by a trained/qualified physician in a hospital or clinical setting for ultrasound evaluation of fetal, abdominal, pediatric, small organ (thyroid, testes, breast), neonatal cephalic, musculoskeletal (conventional and superficial), cardiac (adult and pediatric), peripheral vascular, trans-rectal, obstetrics/ gynecology and urology applications in B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, DPA, PWD, CWD, Combined (B, M, CFM, CPA, DPA, PWD, XBeam, Panoscope), and others (3D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, THI, MFI, ECG, VS Flow, Color M, DICOM).
The SIUI Apogee 1000 series (Apogee 1000/ Apogee 1000Neo/ Apogee 1000Lite/ Apogee 1000Exp/ Apogee 1000B/W/ Apogee 1 series (Apogee 1T/ Apogee 1T/ Apogee 1G) Digital Color Ultrasound Imaging System is capable of the following operating modes: B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, DPA, PWD, CWD, Combined(B, M, CFM, CPA, DPA, PWD, CWD, XBeam, Panoscope), and others (3D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, THI, TDI, DPA, MFI, ECG, VS Flow, Color M, DICOM). The system is designed for use in linear, convex, phased array scanning modes and supports linear, convex, phased array and endocavity (trans-vaginal and trans-rectal) transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities. This system is a Track 3 device and the software used in Apogee 1000 series / Apogee 1 series Digital Color Ultrasound Imaging System is Ultrasound Software by SIUI and is based on the predicate device. The Level of Concern for the Ultrasound Software is Moderate.
The provided document is a 510(k) summary for the SIUI Apogee 1000 series/Apogee 1 series Digital Color Doppler Ultrasound Imaging System. It describes the device, its intended use, and its comparison to a predicate device (SIUI Apogee 2300, K173000) to demonstrate substantial equivalence.
Based on the content, here's an analysis of the acceptance criteria and study information:
Description of Acceptance Criteria and Device Performance
The document does not explicitly state specific, quantifiable acceptance criteria (e.g., minimum sensitivity, specificity, or image quality metrics) for the device's performance. Instead, it relies on demonstrating substantial equivalence to a legally marketed predicate device (Apogee 2300, K173000) through compliance with recognized standards and a comparison of technological characteristics.
The "device performance" reported is the demonstration that the subject device (Apogee 1000 series/Apogee 1 series) is comparable to the predicate device in terms of:
- Intended Use: Both devices have the same intended use.
- Technological Characteristics: They share the same PC module, digital signal processing module, and ultrasound front-end module. The overall user interface is substantially unchanged.
- Functions: While some models of the subject device may lack the 4D imaging mode present in the predicate, they are similar in "other ultrasonic diagnostic functions, imaging modes, file storage management, compatible peripherals, measurement and calculation functions."
- Safety and Effectiveness: Non-clinical tests demonstrate that the subject device performs as well as, and is substantially equivalent with respect to safety and effectiveness of the predicate device.
Table of Acceptance Criteria and Reported Device Performance
Given the nature of the 510(k) submission for a diagnostic ultrasound system based on substantial equivalence, specific numerical acceptance criteria (like those for AI/CADe devices) are not typically presented in this format. The "acceptance criteria" here are implicitly meeting the regulatory requirements for substantial equivalence by demonstrating comparable technical characteristics and complying with relevant standards.
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Safety: Device meets recognized safety standards. | Complies with IEC 60601-1 (Basic safety and essential performance), IEC 60601-1-2 (EMC), IEC 60601-2-37 (Ultrasonic medical diagnostic), ISO 14971 (Risk management), ISO 10993 (Biological evaluation), IEC 62304 (Software life cycle), IEC 60601-1-6 (Usability), IEC 62366-1 (Usability engineering), ISO 15223-1 (Symbols), ISO 13485 (Quality management systems). |
| Effectiveness: Device performs as intended. | The subject device performs "as well as" and is "substantially equivalent" to the predicate device (Apogee 2300, K173000) in terms of technical features, general functions, applications, and indications for use. Similarities include: shared PC, digital signal processing, and ultrasound front-end modules; substantially unchanged user interface; comparable ultrasonic diagnostic functions, imaging modes (with the exception of 4D imaging on some subject device models), file storage, peripherals, and measurement/calculation functions. The non-clinical tests confirmed compliance with relevant standards. These tests demonstrate the device's ability to function effectively for its intended diagnostic applications. |
| Substantial Equivalence: Device is equivalent to predicate. | Non-clinical tests demonstrate that the subject device is "as safe, as effective, and performs as well as the predicate." Differences (e.g., cosmetic design, no 4D mode on some models, new probe models) "have no impact on safety or efficacy" and "do not raise any new potential or increased safety risks." |
Study Details
-
Sample size used for the test set and the data provenance:
- The document describes non-clinical testing to demonstrate compliance with various international standards (e.g., electrical safety, EMC, software lifecycle, biological evaluation).
- It does not mention a "test set" in the context of clinical image data or patient data. The evaluation is primarily based on engineering and performance standards applied to the device itself, rather than testing its diagnostic output against a specific dataset.
- Data provenance is not applicable as this is not a study involving clinical data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. There is no mention of an external "test set" with ground truth established by experts. The "ground truth" implicitly comes from adherence to established engineering and medical device standards.
-
Adjudication method for the test set:
- Not applicable. No "test set" or adjudication process described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done.
- This submission is for a general-purpose ultrasound imaging system, not an AI/CADe (Computer-Aided Detection/Diagnosis) device designed to assist human readers or perform automated analysis. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not relevant to this submission.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No, a standalone algorithm performance study was not done.
- This device is an imaging system, not an algorithm, and is intended for use by a "trained/qualified physician" (human-in-the-loop).
-
The type of ground truth used:
- For the non-clinical testing, the "ground truth" is implied by the technical specifications outlined in the international standards (e.g., electrical parameters, mechanical safety, software functionality, biocompatibility). It's not a clinical ground truth like pathology or patient outcomes.
-
The sample size for the training set:
- Not applicable. This is not an AI/ML device that requires a "training set."
-
How the ground truth for the training set was established:
- Not applicable. No training set is mentioned.
Ask a specific question about this device
(196 days)
The system is intended to be used by a trained qualified physician in a hospital or clinical setting for ultrasound evaluation of fetal, abdominal, pediatric, small organ (thyroid, testes, breast), neonatal cephalic, musculoskeletal (conventional and superficial), cardiac (adult and pediatric), peripheral vascular, trans-rectal, obstetrics/ gynecology and urology applications in B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, CWD, CWD, Combined (B, M, CFM, CPA, DPA, PWD, CWD, XBeam, Panoscope), and others (3D/4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, THI, MFI, ECG, VS Flow, Color M, DICOM).
The SIUI Apogee 6500/ Apogee 6300/ Apogee 6200 Digital Color Ultrasound Imaging System capable of the following operating modes: B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, DPA, PWD, CWD, Combined (B, M, CFM, CPA, DPA, PWD, CWD, XBeam, Panoscope), and others (3D/4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, THI, MFI, ECG, VS Flow, Color M, DICOM).
The system is designed for use in linear, convex, phased array scanning modes and supports linear, convex, phased array and endocavity (trans-vaginal and trans-rectal) transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities.
This system is a Track 3 device and the software used in Apogee 6500 / Apogee 6300/Apogee 6200 Digital Color Ultrasound Imaging System is Ultrasound Software by SIUI and is based on the predicate device. The Level of Concern for the Ultrasound Software is Moderate.
The provided text is a 510(k) summary for the Apogee 6500, Apogee 6300, and Apogee 6200 Digital Color Doppler Ultrasound Imaging Systems. This document aims to demonstrate substantial equivalence to a legally marketed predicate device (Apogee 5500).
Based on the provided information, the device is an ultrasound imaging system, not an AI/ML-driven device, and therefore the majority of the requested information (acceptance criteria, specific study details for AI/ML performance, ground truth, expert adjudication, MRMC studies) is not applicable or not provided in this type of regulatory document.
Here's an breakdown of the available information in response to your request, and why some parts cannot be answered from this document:
1. A table of acceptance criteria and the reported device performance:
This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting specific performance metrics against pre-defined acceptance criteria for a novel AI/ML algorithm. The performance is assessed by comparison to the predicate device and compliance with relevant international standards.
- Acceptance Criteria (Implied for Substantial Equivalence): The device is substantially equivalent if it has the same intended use, similar technological characteristics, and does not raise new questions of safety or effectiveness compared to the predicate device.
- Reported Device Performance:
- Intended Use: Same as the predicate device (Apogee 5500, K160853).
- Technical Characteristics: Basic and key technical features (Operation Principle, Operation Controls, System Setup, Operation Modes, Measurement, Calculation and Report item Cine and File Management) are the same as the predicate.
- Acoustic Power Levels: Meet FDA limits, same as the predicate device.
- Materials of Probes: Same as the predicate device.
- Probes: Similar probes as the predicate, with two new transducers (P5FC and V6LF) whose applications fall within the predicate's indications for use.
- Safety and Efficacy: Differences between the new device and the predicate have no impact on safety or efficacy and do not raise new potential or increased safety risks. The new device is equivalent in performance to existing legally marketed devices.
2. Sample size used for the test set and the data provenance:
- Not applicable / Not provided. As this is an ultrasound imaging system (not an AI/ML diagnostic algorithm), the performance is assessed through engineering tests, compliance with safety standards, and comparison of technical specifications to the predicate device. There is no "test set" in the context of image classification or detection algorithms with associated ground truth labels to be tested.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. Ground truth establishment using expert review is relevant for AI/ML performance studies, which are not described in this 510(k) summary for an ultrasound imaging system.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. Adjudication is relevant for resolving discrepancies in expert labeling of imaging data for AI/ML algorithm development/testing, which is not the subject of this document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. The document explicitly states: "Clinical testing is not necessary for the Apogee 6500/ Apogee 6300/ Apogee 6200 Digital Color Doppler Ultrasound Imaging System in order to demonstrate substantial equivalence to the predicate device." MRMC studies evaluating human reader performance with and without AI are not mentioned or required for this type of device clearance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable / Not provided in the context of diagnostic performance. This device is an imaging system, not a diagnostic algorithm. Its "performance" is assessed by its imaging capabilities, safety, and compliance with standards, not a standalone diagnostic accuracy.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable / Not provided. The concept of "ground truth" as used in AI development is not relevant to the substantial equivalence demonstration for this general-purpose ultrasound system. Its performance evaluation relies on engineering verification, compliance with safety standards (e.g., IEC 60601 series), and direct comparison of specifications and intended use to the predicate device.
8. The sample size for the training set:
- Not applicable / Not provided. This is not an AI/ML algorithm requiring a training set. The "software" mentioned (Ultrasound Software by SIUI) is based on the predicate device's software and its "Level of Concern" is Moderate, implying it was developed and validated through standard software engineering practices for medical devices (e.g., IEC 62304), not through machine learning training on a dataset.
9. How the ground truth for the training set was established:
- Not applicable / Not provided. As there is no AI/ML training set, the establishment of ground truth for such a set is not relevant.
In summary, this 510(k) pertains to a new generation of an existing ultrasound imaging system, demonstrating its substantial equivalence to a previously cleared model. The evaluation focuses on hardware, software (system-level, not AI/ML), safety standards, and technical specifications, rather than clinical performance metrics typically associated with AI/ML-driven diagnostic aids.
Ask a specific question about this device
(270 days)
The Apogee is a gas conserver intended as a delivery device for medical grade oxygen from a high-pressure oxygen cylinders. Apogee is an ambulatory device which allows patients to ambulate longer than they would with continuous flow regulator on the same cylinder. The Apogee is intended to be used in the hospital, healthcare facilities, or home care environments.
The proposed device, Apogee, senses each nostril independently through a bifurcated cannula and responds to the dynamic changes in nasal air-flow making a breath-to-breath determination of where to deliver the next bolus of oxygen: to the left, the right, or both nostrils. The Apogee Device is composed of the main components as detailed below: Control Unit, Control Board (Regulated Voltage Supply), Embedded Processor, User supplied Demand Cannula, The Apogee Bypass Cannula and Restrictor and Cylinder Tubing.
This document is an FDA 510(k) summary for the Apogee oxygen conserver. It describes the device, its intended use, and compares it to a predicate device (Inspired Technologies Model 350G). The information provided focuses on the device's design, technology, and non-clinical performance testing rather than clinical study results involving human patients or complex AI algorithms with specific performance metrics like sensitivity, specificity, or AUC.
Therefore, many of the requested elements for acceptance criteria and study details (like sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone AI performance, and training set details) are not applicable or not present in this type of regulatory document concerning a mechanical-electronic medical device like an oxygen conserver. The "performance data" section primarily refers to engineering and safety bench testing, not clinical performance studies with diagnostic accuracy targets for AI.
Here's a breakdown based on the provided document:
Acceptance Criteria and Device Performance for Apogee Oxygen Conserver
This 510(k) summary focuses on demonstrating substantial equivalence to a predicate device through engineering and safety testing. It does not contain specific "acceptance criteria" in the context of diagnostic performance (e.g., sensitivity, specificity, AUC) for an AI algorithm. Instead, the "acceptance criteria" are implied by compliance with recognized standards and comparable performance to the predicate device in bench tests.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a non-AI, mechanical-electronic device, the acceptance criteria are not typically expressed as diagnostic accuracy metrics. Instead, they relate to engineering specifications and performance in bench testing to ensure safety and effectiveness comparable to the predicate. The document highlights the following:
| Acceptance Criteria (Implied by Standards/Predicate Comparison) | Reported Device Performance | Comments |
|---|---|---|
| Electrical Safety & EMC Compliance(e.g., AAMI ANSI ES 60601-1, IEC 60601-1-11, IEC 60601-1-2) | Complies with AAMI ANSI ES 60601-1: 2005 + A1: 2012, IEC 60601-1-11, and IEC 60601-1-2: 2014. | Demonstrated compliance with relevant electrical safety and electromagnetic compatibility standards. |
| Software Verification & Validation(for "moderate" level of concern software) | Verification and validation testing conducted. | Software testing completed for safety and functionality. |
| Mechanical Integrity & Durability(e.g., storage, aging, drop, durability) | Testing performed related to storage, effects of aging, drop, and durability. | Device withstands expected physical stresses. |
| Biocompatibility(for patient-contacting materials - permanent duration) | Testing included: - Gas emission VOC- Particulate Matter (PM2.5)- Inorganic gases CO, CO2 and Ozone- Toxicological Risk Assessment | Materials are safe for patient contact. |
| Oxygen Bolus Delivery Equivalence(compared to predicate, volume and time) | Bench testing demonstrated that both devices (Apogee and predicate) delivered an equivalent amount of oxygen bolus to the patient in the same time. | Key functional performance matched the predicate. |
| Trigger Sensitivity(compared to predicate) | Bench testing performed, details not quantified but implied comparable to predicate (-0.3 cmH2O for Apogee vs. +0.03 to -0.2 cmH2O for predicate). | Demonstrated a similar ability to sense patient inhalation. |
| Volume Flow Performance | Bench testing performed. | Functional flow characteristics validated. |
| Performance Pre- and Post-Cleaning | Bench testing performed. | Device maintains performance after cleaning. |
| Breath Range Capability(e.g., up to 35 BPM for predicate) | Apogee: Up to 40 BPM | Apogee performs across a relevant physiological breath rate range, slightly better than predicate. |
| Patient Oxygen Outlet Pressure(e.g., 19-25 psig for predicate) | Apogee: 19-30 psig | Comparable and within safe operating range. |
| Water Ingress Protection(e.g., IPX1 for predicate) | Apogee: IP22 | Improved liquid ingress capability (IP22 for Apogee vs. IPX1 for predicate). No differences in safety or effectiveness found due to this improvement. |
Study Proving Device Meets Acceptance Criteria:
The study described is a series of non-clinical bench tests and compliance assessments against recognized industry standards, rather than a diagnostic accuracy or comparative effectiveness study involving human subjects or AI-assisted diagnostic tasks.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not specified in terms of patient data. The "test set" refers to physical units of the device tested in a lab setting. The number of units tested is not provided, but it would typically be a small number of engineering samples.
- Data Provenance: Not applicable in the context of patient data. This involves laboratory bench testing of physical device units, not patient data sets. The testing was presumably conducted by the manufacturer or contracted labs.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: Not applicable. Ground truth for an oxygen conserver's performance in bench testing is based on engineering specifications, calibrated measurement equipment, and industry standards, not expert interpretation of medical images or clinical data.
- Qualifications of Experts: Not applicable. The "ground truth" for this device's performance is objective measurement against engineering specifications.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not applicable. This is not a study where human interpretations are being adjudicated.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- MRMC Study: No. This type of study is typically performed for diagnostic devices (e.g., imaging AI) where human readers' performance is evaluated with and without an AI assist. This document concerns a therapeutic mechanical device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Performance: The "software" in this device controls the mechanical function (sensing breath and delivering oxygen bolus). Its "standalone performance" is implicitly assessed through the bench testing described (e.g., bolus delivery accuracy, trigger sensitivity). It's not a diagnostic algorithm with standalone performance metrics like sensitivity/specificity.
7. The Type of Ground Truth Used:
- Type of Ground Truth: The ground truth for this device's performance is based on engineering specifications, physical measurements from calibrated instruments, and compliance with recognized national and international standards for medical device safety and performance (e.g., AAMI ANSI ES 60601-1, IEC 60601-1-11, ISO 80601-2-67).
8. The Sample Size for the Training Set:
- Sample Size for Training Set: Not applicable. This is a non-AI mechanical device. The "software" mentioned is embedded control software, not a machine learning model that requires a training set in the conventional sense of AI development.
9. How the Ground Truth for the Training Set Was Established:
- Ground Truth for Training Set: Not applicable. As there's no machine learning training set, no such ground truth establishment occurred. The control logic for the device's functions (e.g., breath sensing, oxygen delivery) is derived from engineering design principles, physiological understanding, and validated algorithms for precise control, not from a data-driven training process.
Ask a specific question about this device
Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculoskeletal, cardiac, peripheral vascular, transvaginal and trans-rectal applications in B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, PWD, CWD, Combined (B, M, CFM, CPA, PWD, CWD, XBeam, Panoscope, and others (3D, 4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, ECG , VS Flow, Color M and DICOM).
The SIUI Apogee 2100/ Apogee 2300 is a Digital Ultrasound Imaging System capable of the following operating modes: B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, PWD, CWD, Combined (B, M, CFM, CPA, PWD, CWD, XBeam, Panoscope), others (3D, 4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, ECG, VS Flow, Color M, DICOM). The system is designed for use in linear, convex, phased array and 3D scanning modes and supports linear, convex, phased array, 3D and endocavity (trans-vaginal and trans-rectal) transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities.
Here's a breakdown of the acceptance criteria and study information, based on the provided text, with a crucial caveat: The document describes a 510(k) submission for an ultrasound imaging system, which primarily focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific clinical performance acceptance criteria through the types of studies you've listed (e.g., MRMC, standalone AI performance).
Therefore, many of your requested points about clinical studies, ground truth establishment, sample sizes for training/test sets, expert qualifications, and adjudication methods are not present in this type of regulatory document because they are not typically required for a 510(k) clearance of an ultrasound imaging system. The information provided heavily emphasizes hardware and software feature equivalence and safety standards.
Acceptance Criteria and Device Performance
The document doesn't explicitly state "acceptance criteria" in terms of clinical performance metrics like sensitivity, specificity, accuracy, or reader agreement. Instead, the acceptance criteria for this 510(k) submission are implied through demonstrating substantial equivalence to predicate devices in terms of:
- Indications for Use: The new device's intended clinical applications must be substantially equivalent to or within the scope of the predicate devices.
- Technological Characteristics: The new device's core technology, operating modes, and transducers must be substantially equivalent.
- Safety and Effectiveness: Compliance with recognized medical device safety standards and acoustic output limits substituted for direct clinical evidence of safety and effectiveness for a new device.
Table of Acceptance Criteria and Reported Device Performance (as implied by the 510(k) process):
| Acceptance Criterion (Implied for 510(k) Clearance) | Reported Device Performance (Summary from Document) |
|---|---|
| 1. Indications for Use Equivalence | - Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculoskeletal, cardiac, peripheral vascular, transvaginal and trans-rectal applications. - Operating Modes: B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, PWD, CWD, Combined (B, M, CFM, CPA, PWD, CWD, XBeam, Panoscope), others (3D, 4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, ECG, VS Flow, Color M, DICOM). - Comparison to Predicate: "All the Indications for Use of the Apogee 2100/ Apogee 2300 are included in the indications for use of predicate SIUI Apogee 1200." and "The Apogee 2100/ Apogee 2300 have three more imaging functions than the predicate SIUI Apogee 5500, namely VS Flow, Color M and DICOM." Additional new configurable transducers (L10LN, C6LN, U5LN) and 5 additional exams (Neonatal Cephalic, Trans-vaginal and Trans-rectal, Gynecology and Urology) are identical to Apogee 5500. |
| 2. Technological Characteristics Equivalence | - Apogee 2100/2300 vs. Apogee 1200: "mainly have cosmetic changes," "same in their working principle and internal modules," "use the same PC module, digital signal processing module, and ultrasound front-end module." Control panel and display adjusted for ergonomics. - Apogee 2100/2300 vs. Apogee 5500: "similar imaging modes," "similar diagnostic ultrasound applications," "similar technical parameters and hardware implementation principle" for transducers. |
| 3. Safety (Acoustic Output & General Safety Standards Compliance) | - Tested per FDA Guidance (Sept 2008) for Diagnostic Ultrasound Systems and Transducers. - Acoustic output measured and calculated per NEMA UD 2:2004. - Conforms to IEC 60601-1, ISO10993-5, ISO 10993-10, and IEC 60601-1-2. |
Study Information (Based on the 510(k) document):
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable/not provided. This document describes a 510(k) submission based on technical and performance equivalence and safety standards, not a clinical study with a specific test set of patient data with ground truth.
- Data Provenance: Not applicable/not provided for a clinical test set. The submission relies on engineering tests, compliance with standards, and comparison to existing cleared devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. No ground truth establishment by clinical experts is described for a test set of patient data.
-
Adjudication method for the test set:
- Not applicable. No clinical test set requiring adjudication is described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study was not done, as this is an ultrasound imaging system regulatory submission, not an AI diagnostic algorithm for which such studies are commonly required. The device itself is the imaging system, not an AI feature providing interpretive assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is not an AI algorithm. Its performance is inherent to its imaging capabilities, which are assessed against technical specifications and predicate device equivalence.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable for clinical ground truth. The "ground truth" in this context is adherence to engineering specifications, safety standards, and functional equivalence to predicate devices, verified through technical testing.
-
The sample size for the training set:
- Not applicable. There is no mention of an algorithm being trained on a dataset.
-
How the ground truth for the training set was established:
- Not applicable. No training set is mentioned.
In summary: The provided document is a 510(k) premarket notification for an ultrasound imaging system. The "study" referenced is largely a technical and engineering comparison study to predicate devices and a compliance assessment against recognized safety and performance standards. It does not involve a clinical study with patients, expert readers, or AI performance metrics as would be seen for a diagnostic AI device submission.
Ask a specific question about this device
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal, Abdominal, Pediatric, Small Organ (Thyroid, Testes, Breast), Neonatal Cephalic, Musculoskeletal (Conventional and Superficial), Cardiac (Adult and Pediatric), Peripheral Vascular, Trans-rectal, Obstetrics/ Gynecology and Urology applications.
The SIUI Apogee 5500/ Apogee 5300 /Apogee 5800 is a Digital Ultrasound Imaging System capable of the following operating modes: B-Mode(B, 2B, 4B), M-Mode, CFM, CPA, PWD, CWD, Combined(B, M, CFM, CPA, PWD, CWD, XBeam, Panoscope), and other(3D, 4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, ECG).The system is designed for use in linear, convex, phased array and 3D scanning modes and supports linear, convex, phased array, 3D and endocavity (trans-vaginal and trans-rectal) transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities.
This document describes the premarket notification for the SIUI Apogee 5500, Apogee 5300, and Apogee 5800 Digital Color Doppler Ultrasound Imaging Systems. It focuses on demonstrating substantial equivalence to predicate devices rather than providing specific acceptance criteria and performance data for a novel AI device or software. Therefore, much of the requested information regarding acceptance criteria, sample sizes for test and training sets, expert qualifications, and adjudication methods is not directly available in the provided text.
Here is an analysis based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly those required to demonstrate substantial equivalence to previously cleared predicate devices, as per FDA's 510(k) pathway. The reported performance is primarily functional similarity and adherence to safety standards, rather than specific quantitative metrics.
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Safe and Effective for Intended Use | The device demonstrates substantial equivalence in intended use, technological characteristics, imaging modes, measurement functions, and safety standards to predicate devices. Testing was performed per FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" (September 2008). The device conforms to applicable medical device safety standards, such as IEC 60601-1, ISO10993-5, and ISO 10993-10. Acoustic output is measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004. |
| Similar Intended Use | Subject Device IFU: Fetal, Abdominal, Pediatric, Small Organ (Thyroid, Testes, Breast), Neonatal Cephalic, Musculoskeletal (Conventional and Superficial), Cardiac (Adult and Pediatric), Peripheral Vascular, Trans-vaginal, Trans-rectal, Obstetrics/Gynecology and Urology applications. |
| Predicate Apogee 3800 IFU (K110841): Abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular applications in B, M, PWD, Color Doppler and 3D imaging modes. | |
| Predicate Acuson X700 IFU (K141846): Cardiac (Adult, Pediatric), Trans-esophageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The subject device's applications are covered by the Acuson X700. | |
| Similar Technological Characteristics (Imaging Modes & Functions) | Subject Device Modes: B-Mode (B, 2B, 4B), M-Mode, CFM, CPA, PWD, CWD, Combined (B, M, CFM, CPA, PWD, CWD, XBeam, Panoscope), and other (3D, 4D, Trapezoidal /Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, ECG). |
| Predicate Apogee 3800 (K110841): Same basic operating modes (B, M, PWD, Color Doppler, 3D imaging modes) as the subject device, with the subject device adding new modes like XBeam, Panoscope, 4D, Trapezoidal / Extended Sector Imaging, Elastography, Anatomical M-mode, TDI, MFI, and ECG. These new modes were tested and proved no impact on device safety and effectiveness. | |
| Predicate Acuson X700 (K141846): The new imaging modes and applications in the subject device are similar to those in the Acuson X700. | |
| Similar Transducers and Applications Connected | Subject Device Transducers: Linear, convex, phased array, 3D and endocavity (trans-vaginal and trans-rectal) transducers. Specifically mentioned additional transducers: V6LC, ECBP, L10LC, C6LC, U5LC. |
| Predicate Apogee 3800 (K110841): The subject device adds V6LC, ECBP, L10LC, C6LC, U5LC transducers, which are identical to those available on the Acuson X700 (K141846). | |
| Predicate Acuson X700 (K141846): The transducers of the subject device and Acuson X700 have similar technical parameters and hardware implementation principle. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly mention a "test set" in the context of an AI/algorithm performance study. The evaluation appears to be based on engineering validation and verification testing of the ultrasound system's performance characteristics (e.g., acoustic output, image quality checks) and comparison to predicate devices, rather than a clinical study with a defined patient test set for an algorithm.
- Sample Size: Not specified for any performance testing beyond general system validation.
- Data Provenance: Not applicable from the provided text, as it doesn't describe a study with patient data for algorithm evaluation. The company (SIUI) is based in China.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided because the submission is for an ultrasound imaging system, not an AI algorithm that would typically require expert-established ground truth on a test set of medical images. The "ground truth" for an ultrasound system typically relates to its physical performance characteristics, image quality, and diagnostic accuracy as determined by established medical principles and comparison to predicate devices.
4. Adjudication Method
Not applicable, as there is no mention of a study involving human reader interpretation or ground truth establishment that would require an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This type of study would be relevant for evaluating the impact of AI assistance on human reader performance, which is not the subject of this 510(k) submission.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No. This submission is for a full ultrasound imaging system, not a standalone AI algorithm. While the system incorporates "new imaging modes" such as XBeam, Panoscope, 3D, 4D, Elastography, etc., these are features of the imaging system itself, not separate AI algorithms. The document states these new modes "were tested and proved no impact on device safety and effectiveness," which implies technical validation rather than a standalone clinical performance study typically associated with AI.
7. Type of Ground Truth Used
The ground truth for the device's performance is implicitly based on:
- Engineering specifications and measurements: Adherence to established standards for acoustic output (NEMA UD 2), electrical safety (IEC 60601-1), and biocompatibility (ISO 10993-5, ISO 10993-10).
- Comparison to predicate devices: The "ground truth" for substantial equivalence is the validated performance and safety profile of the legally marketed predicate devices (SIUI Apogee 3800 and SIEMENS Acuson X700).
- Qualitative assessment: Clinical utility and image quality are likely assessed by medical professionals during internal validation, but no specific study details are provided.
8. Sample Size for the Training Set
Not applicable. This is not a submission for an AI algorithm that involves a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable for the same reason as point 8.
Ask a specific question about this device
(107 days)
AMS Pelvic Floor Repair System is intended for use where the connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological and gastroenterological anatomy. This includes but is not limited to the following procedures: pubourethral support, including urethral slings for the treatment of incontinence, vaginal wall prolapse repairs including anterior and posterior wall repairs, vaginal suspension, reconstruction of the pelvic floor and tissue repair.
The Apogee and Perigee Systems with IntePro Lite and InteXen LP are sterile, single use procedure kits that consist of stainless steel, curved needle passers and an implantable mesh assembly.
The provided text is a 510(k) summary statement and related FDA correspondence for a surgical mesh device. The document states that the device components were "tested for biocompatibility and performance requirements and found to be substantially equivalent to the predicate devices." However, it does not provide specific quantitative acceptance criteria or a detailed study description that proves the device meets those criteria.
Therefore, the requested information, which typically involves detailed performance metrics, sample sizes, expert involvement, and ground truth establishment, is not present in the provided text.
Here's a breakdown of why each point cannot be answered:
- A table of acceptance criteria and the reported device performance: The document states "tested for biocompatibility and performance requirements and found to be substantially equivalent to the predicate devices." It does not provide a table with specific criteria (e.g., tensile strength, pore size, elasticity) or the numerical results from these tests.
- Sample sized used for the test set and the data provenance: Not mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned. The testing described is for biocompatibility and performance, not involving human interpretation of medical data where experts would establish ground truth.
- Adjudication method for the test set: Not applicable for the type of testing described (biocompatibility and performance).
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is about a surgical mesh, not an AI-powered diagnostic or assistive tool.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is for a physical medical device.
- The type of ground truth used: Not applicable in the context of the described physical device testing. The "ground truth" for material properties would be the established standards or values of the predicate device.
- The sample size for the training set: Not applicable. This is not a machine learning device.
- How the ground truth for the training set was established: Not applicable. This is not a machine learning device.
Ask a specific question about this device
(120 days)
The AMS Pelvic Floor Repair System is intended for use where the connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological, and gastroenterological anatomy. This includes but is not limited to the following procedures: pubourethral support, including urethral slings, vaginal wall prolapse repairs including anterior and posterior wall repairs, vaginal suspension, reconstruction of the pelvic floor and tissue repair.
The Apogee and Perigee prolapse repair systems are part of the AMS Pelvic Floor Repair System family of devices. They consist of a permanently-implanted mesh assembly and non-implanted surgical instruments (needle passers) that are used as aids to place the mesh assembly in the pelvic floor.
This K081710 510(k) submission describes a modification to an existing surgical mesh device. The core of the submission revolves around demonstrating that the modified device (with a PC coating) is "substantially equivalent" to its predicate device (without the coating). This type of submission rarely includes detailed studies with acceptance criteria in the way a new device would. Instead, the "acceptance criteria" are implicitly met if the new device performs similarly enough to the predicate device.
Here's the breakdown of the information requested, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) for a modified device, the "acceptance criteria" are not explicitly stated for clinical performance as they might be for a novel device. Instead, the study aimed to show "substantial equivalence" to a predicate device.
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance |
|---|---|
| Biocompatibility: Demonstrated no adverse biological reactions due to the new PC coating. | Components of Apogee and Perigee systems were tested for biocompatibility and found to be substantially equivalent to the predicate device. |
| Mechanical Performance (Tensile Strength): Maintained similar tensile strength in both weft and warp directions compared to the predicate. | Tensile testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Cycle): Maintained similar cyclic loading performance in both weft and warp directions compared to the predicate. | Cycle testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Elongation): Maintained similar elongation characteristics in both weft and warp directions compared to the predicate. | Elongation testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Stiffness): Maintained similar stiffness characteristics in both weft and warp directions compared to the predicate. | Stiffness testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Dimensional Characteristics (Fiber Diameter): Maintained similar fiber diameter compared to the predicate. | A comparison of the dimensional characteristics was completed and included fiber diameter. Results indicated substantial equivalence. |
| Dimensional Characteristics (Pore Size): Maintained similar pore size compared to the predicate. | A comparison of the dimensional characteristics was completed and included pore size. Results indicated substantial equivalence. |
| Dimensional Characteristics (Mesh Density): Maintained similar mesh density compared to the predicate. | A comparison of the dimensional characteristics was completed and included mesh density. Results indicated substantial equivalence. |
| Dimensional Characteristics (Mesh Thickness): Maintained similar mesh thickness compared to the predicate. | A comparison of the dimensional characteristics was completed and included mesh thickness. Results indicated substantial equivalence. |
| Indications for Use: The modified device has the same intended use as the predicate device. | The Indications for Use for the Apogee® and Perigee® Systems with PC Coated IntePro® Lite™ are identical to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not specify sample sizes for mechanical testing or biocompatibility studies. The studies were likely bench-top (mechanical) and potentially in-vitro/in-vivo (biocompatibility) tests.
- Sample Size for Test Set: Not specified.
- Data Provenance: Not specified, but generally, such tests are conducted in a laboratory setting, typically in the country of the manufacturer or a contracted testing facility. It's retrospective in the sense that the data is generated for the submission, not from a clinical trial.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of 510(k) submission, focused on material and mechanical equivalence for a modified surgical mesh, does not involve expert-established ground truth in the way a diagnostic AI device would. The "ground truth" for the mechanical tests would be the objective measurements themselves against established engineering standards or direct comparison to the predicate device's measured properties. For biocompatibility, it refers to standardized testing results.
4. Adjudication Method for the Test Set
Not applicable. This is not a study involving human interpretation or clinical outcomes requiring adjudication. The tests described are objective, physical, and biological assessments.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. This submission is for a surgical mesh, not a diagnostic device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done
Not applicable. This submission is for a surgical mesh, which is a physical implant, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the tests mentioned (tensile, cycle, elongation, stiffness, dimensional characteristics, and biocompatibility) would be based on:
- Objective physical measurements: Using standardized testing methodologies and equipment to quantify mechanical properties and dimensions.
- Biocompatibility standards and assays: Results from validated in-vitro and potentially in-vivo tests comparing the new material's biological response to known standards or the predicate material.
8. The Sample Size for the Training Set
Not applicable. This submission is for a physical medical device (surgical mesh), not an AI algorithm, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. No training set was used.
Ask a specific question about this device
(42 days)
The Cynosure Apogee Elite laser is indicated for permanent hair reduction and the treatment of vascular lesions, pigmented lesions, and wrinkles.
755 nm:
The Cynosure Apogee Elite Dermatological Laser is indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin.
It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.
1064 nm:
The Cynosure Apogee Elite Dermatological laser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café' au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques.
The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
Cynosure Apogee Elite laser (755nm), having an Alexandrite crystal rod as the lasing medium, with a wavelength of 755 nm. The 1064 nm wavelength is generated from the Nd:YAG laser head. Laser activation is by either finger or foot switch. Overall weight of the laser is 45 Kg, and the size is 113x53x84 cm (HxWxD). Electrical requirement is 230 VAC, 50A, 50-60 Hz, single phase.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Cynosure Apogee Elite Laser:
This 510(k) summary does not include any acceptance criteria or a study proving the device meets said criteria. Instead, it relies on substantial equivalence to predicate devices.
Let's break down the reasons and why this document doesn't fit the request:
-
A table of acceptance criteria and the reported device performance:
- Not found. This document explicitly states "Nonclinical Performance Data: none" and "Clinical Performance Data: none."
-
Sample size used for the test set and the data provenance:
- Not found. Since no studies were conducted or provided, there is no test set or data provenance to report.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not found. No test set means no ground truth establishment by experts.
-
Adjudication method for the test set:
- Not found. No test set means no adjudication method.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a laser system, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or is relevant here.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a hardware medical laser system, not a software algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc.):
- Not found. No ground truth was established as no performance data was provided for the device.
-
The sample size for the training set:
- Not found. No training set was used as no new performance data was generated for this 510(k). The device is cleared based on equivalence.
-
How the ground truth for the training set was established:
- Not found. No training set means no establishment of ground truth for it.
Summary of the document's approach:
The Cynosure Apogee Elite Laser's 510(k) submission relies on demonstrating substantial equivalence to already legally marketed devices (Cynosure Apogee TKS, Cynosure Acclaim). This means the manufacturer argues that their new device is as safe and effective as existing predicate devices because it shares the "same principle of operation, the same wavelength and essentially the same power range and the same indications for uses."
Therefore, this document does not contain the information requested about specific acceptance criteria or performance studies demonstrating the device meets those criteria. It is a clearance based on comparison rather than new performance data.
Ask a specific question about this device
Page 1 of 1