510(k) Data Aggregation

    K Number
    K233264

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-22

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anthogyr Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility.

    The cycle of sterilization, for Anthogyr INTEGRAL Guided Surgery Cassettes, uses pre-vacuum steam: 132 °C (270° F) during 4 minutes with 30 minutes drying time.

    Anthogyr INTEGRAL Guided Surgery Cassettes have been validated for a maximum load of with the associated instrument. The worst-case recommended load is: 886 g.

    The device dimension of Anthogyr INTEGRAL Guided Surgery Cassettes is 290x176x62 mm.

    The cassettes are not intended to be stacked during sterilization process.

    Device Description

    Anthoqyr cassettes are reusable rigid containers, comprising a case bottom (base), one or more removable inner tray base (tray) and a tray lid (lid). The trays are composed of brackets made of medical grade silicone, namely brackets used to maintain the Anthogyr dental instruments in place during the surgical or prosthetic procedure and during sterilization. The base and trays have markings and/or colors code to indicate either the surgical workflow, or the position of the instruments in the kit. The lid holds all the instruments securely in place during treatment. To facilitate the surgical procedure and the correct use and positioning of the instruments, the trays have instrument pictograms and/ or color-coded workflows printed on the surface. Surgical instruments of the Anthogyr Dental Implant System intended to be placed in the Anthogyr cassettes are used for bed preparation, placement, maintenance and explanation of the implants from Anthogyr Dental Implant System. These devices are all Class I exempt or already have class II pre-market notification clearance as described in 21 CFR 872.3980 (Endosseous dental implant accessories) and are not subject devices of this submission.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Anthogyr INTEGRAL Guided Surgery Cassettes. It does not describe a study involving an AI/software device that requires establishing ground truth by experts or evaluating human reader performance. Instead, it focuses on demonstrating substantial equivalence of a medical device (sterilization cassettes) to a predicate device through non-clinical performance testing.

    Therefore, many of the requested criteria, such as "Number of experts used to establish the ground truth," "Adjudication method," "MRMC comparative effectiveness study," and "standalone (i.e. algorithm only) performance" are not applicable to this document.

    However, I can extract the acceptance criteria and reported device performance for the non-clinical tests mentioned.

    Here's a breakdown of the requested information based on the provided text, indicating "Not Applicable" (N/A) for criteria that do not fit the nature of this submission:


    1. Table of Acceptance Criteria and Reported Device Performance

    Standard or Test MethodPurpose of the TestingAcceptance CriteriaReported Performance
    Custom• Test Soil: Blood Soil (BLSO) • Cleaning Method: Manual • Residuals Tested: Hemoglobin and ProteinVisual Inspection: No Visible Soil Hemoglobin Test: <2.2 µg/cm2 Protein Test: <6.4 µg/cm2Passed
    Custom• Test Soil: Blood Soil (BLSO) • Cleaning Method: Automated • Residuals Tested: Hemoglobin and ProteinVisual Inspection: No Visible Soil Hemoglobin Test: <2.2 µg/cm2 Protein Test: <6.4 µg/cm2Passed
    ISO 17665-1Sterilization validation, including sterilant penetration and drying timeAll Biological Indicators must be incubated for at least 7 days at 55-60°C. All positive controls for SAL testing must show characteristic growth of the indicator organism.Passed
    Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration StaffLife cycle (simulate usage) testingThe tested samples must withstand 100 cycles of use (cleaning, sterilization, and functional tests) without compromising their functionalitiesPassed
    ISO 10993-5 (Cytotoxicity)Cytotoxicity testingLess than 30% cell proliferation inhibitionPassed

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document states that "The worst-case cassettes were tested for performance." However, it does not specify an exact number of cassettes or samples used for each test.
    • Data Provenance: The tests are non-clinical (laboratory/bench-top), not derived from patient data. The manufacturer is Anthogyr, located in Sallanches, France, and the submission is by Straumann USA, LLC.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • N/A. Ground truth in this context refers to the defined acceptance criteria for engineering and materials performance, not expert consensus on medical images or diagnoses. The 'experts' would be the engineers and quality control personnel performing and evaluating the tests against established industry standards and internal protocols.

    4. Adjudication Method for the Test Set

    • N/A. Adjudication is typically for resolving discrepancies in human expert labeling or diagnoses. For non-clinical bench testing, the results are objectively measured against pre-defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • N/A. This is a medical device (sterilization cassette), not an AI/software device that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • N/A. This is a hardware medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • The ground truth for this device's performance is based on pre-defined acceptance criteria derived from:
      • Applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
      • International standards like ISO 17665-1 and ISO 10993-5.
      • Internal custom test protocols developed for cleaning and residual testing.
      • Functional requirements for the cassette's intended use.

    8. The Sample Size for the Training Set

    • N/A. This is not an AI/software device where a "training set" of data would be used to train a model. The "training set" concept is irrelevant here.

    9. How the Ground Truth for the Training Set Was Established

    • N/A. As explained above, there is no training set or AI model involved in the context of this device's submission.