510(k) Data Aggregation

    K Number
    K242108

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-16

    (29 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ambu® aScope™ 5 Uretero is a sterile, single-use, flexible, digital video ureteroscope intended to be used for endoscopic access and visual quidance in the upper urinary tract. Ambu® aScope™ 5 Uretero is intended to be used with the compatible Ambu displaying unit and can be used in conjunction with endoscopic instruments via its working channel. Ambu® aScope™ 5 Uretero is intended for patients requiring retrograde (transurethral) and/or antegrade (percutaneous) ureteroscopy procedures for visualization and examination with a flexible ureteroscope, and for removal of renal and ureter calculi.

    Ambu® aView™ 2 Advance is intended to display live imaging data from compatible Ambu visualization devices.

    Device Description

    The Ambu® aScope™ 5 Ureteroscopy System is a combination of an endoscope, Ambu® aScope™ 5 Uretero and a displaying unit, the Ambu® aView™ 2 Advance.

    The Ambu® aScope™ 5 Uretero is a sterile, digital video ureteroscope. The Ambu® aScope™ 5 Uretero can be used in retrograde (transurethral) and/or antegrade (percutaneous) ureteroscopy procedures for providing endoscopic access and visual quidance to and in the upper urinary tract.

    The Ambu® aScope™ 5 Uretero is available in one size and can be operated by either left or right hand. The Ambu® aScope™ 5 Uretero can be used with endoscopic accessories. The working channel system allows the passage of endoscopic instillation/ suction of fluids. The Ambu° aScope™ 5 Uretero is intended to be used with a compatible Ambu displaying unit, Ambu° aView™ 2 Advance.

    The Ambu® aView™ 2 Advance, also referred to as displaying unit, is a non-sterile, reusable digital monitor intended to display live imaging data from Ambu visualization devices. The product consists of a 12.8″ LCD screen. The displaying unit is powered by a rechargeable lithium-ion battery and includes a power supply with region-specific power cable.

    The Ambu® aView™ 2 Advance has the following physical and performance characteristics:

    • · Can process and display live imaging data from Ambu® aScope™ 5 Uretero to a monitor
    • · Can record, store and transport image data from Ambu® aScope™ 5 Uretero
    • Is a portable device with an integrated monitor, and the possibility to connect to an external monitor
    AI/ML Overview

    This FDA 510(k) summary (K242108) is a clearance for an updated version of the Ambu® aScope™ 5 Uretero and Ambu® aView™ 2 Advance, primarily focusing on additional compatibility between these two devices. It does not provide a study to prove the device meets acceptance criteria in the traditional sense of a clinical or comparative effectiveness study against a human baseline. Instead, it describes non-clinical performance testing to demonstrate that the updated device combination remains safe and effective and substantially equivalent to its predicate devices.

    Therefore, many of the requested categories (like MRMC study, expert ground truth, sample size for test sets/training sets in a clinical context) are not applicable or cannot be extracted from this specific document, as the summary focuses on technical performance and equivalence,
    not a study of diagnostic accuracy or human performance improvement.

    Here's a breakdown of the information that can be extracted or inferred:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document lists various performance characteristics that were tested. The overall reported performance is that the device "performed as expected and met the test specifications set." Specific numerical acceptance criteria or performance metrics are not
    provided in this summary.

    Acceptance Criteria CategoryReported Device Performance (as stated in document)
    Optical performancePerformed as expected and met test specifications
    - Field of view and Direction of viewPerformed as expected and met test specifications
    - Sharpness and Depth of fieldPerformed as expected and met test specifications
    - Image intensity uniformityPerformed as expected and met test specifications
    - Geometric distortionPerformed as expected and met test specifications
    - ResolutionPerformed as expected and met test specifications
    - Color performancePerformed as expected and met test specifications
    - Noise performancePerformed as expected and met test specifications
    - Dynamic rangePerformed as expected and met test specifications
    Software verification testingPerformed as expected and met test specifications
    Electrical Safety (IEC 60601-1:2005+A1:2012+A2:2020 Ed. 3.2 and IEC 60601-2-18:2009 Ed. 3.0)Performed as expected and met test specifications
    Electromagnetic Compatibility (IEC 60601-1-2:2014+A1:2020 Ed. 4.1)Performed as expected and met test specifications

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This document describes non-clinical performance testing of the device itself (hardware and software), not a study with patient data. Therefore, the concepts of "sample size" for a test set (in a clinical context) or "data provenance" (country/retrospective/prospective) are not applicable. The testing was likely conducted in a controlled lab environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    N/A. As this involved technical performance testing of the device, not a diagnostic or clinical evaluation, the concept of "experts" establishing ground truth in a clinical sense is not relevant here. Engineering and testing personnel would have verified performance against technical specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    N/A. Adjudication methods like 2+1 or 3+1 are used for clinical studies with expert reviewers. This was technical performance testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This document does not mention or describe an MRMC comparative effectiveness study. The device is a ureteroscope system, not an AI-assisted diagnostic tool for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    N/A. This is a medical device for direct visualization, not an algorithm being evaluated for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the non-clinical tests would be the established engineering and optical specifications/standards (e.g., specific measurable values for resolution, field of view, electrical safety limits set by IEC standards, etc.) that the device was designed to meet.

    8. The sample size for the training set:

    N/A. This document pertains to the clearance of a physical medical device and its compatibility, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    N/A. Not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233630

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-24

    (224 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aScope 5 Uretero is a sterile, single-use, flexible, digital video ureteroscope intended to be used for endoscopic access and visual quidance in the upper urinary tract.

    aScope 5 Uretero is intended to be used with the compatible Ambu displaying unit and can be used in conjunction with endoscopic instruments via its working channel.

    aScope 5 Uretero is intended for patients requiring retrograde (transurethral) and/or antegrade (percutaneous) ureteroscopy procedures for visualization and examination with a flexible ureteroscope, and for removal of renal and ureter calculi.

    The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.

    Device Description

    The Ambu® aScope™ 5 Ureteroscopy System is a combination of an endoscope, Ambu® aScope™ 5 Uretero and a displaying unit, the Ambu® aBox™ 2.

    The Ambu® aScope™ 5 Uretero is a sterile, digital video ureteroscope. The Ambu® aScope™ 5 Uretero can be used in retrograde (transurethral) and/or antegrade (percutaneous) ureteroscopy procedures for providing endoscopic access and visual guidance to and in the upper urinary tract.

    The Ambu® aScope™ 5 Uretero is available in one size and can be operated by either left or right hand. The Ambu® aScope™ 5 Uretero can be used with endoscopic accessories. The working channel system allows the passage instruments, instillation/ suction of fluids. The aScope 5 Uretero is intended to be used with a compatible Ambu displaying unit - Ambu® aBox™ 2.

    The Ambu® aBox™ 2, also referred to as displaying unit, is a non-sterile, reusable digital monitor intended to display live imaging data from Ambu visualization devices. The product consists of a base unit with a 12.8″ LCD screen mounted on the top. The device is powered by an integrated power supply and comes with country specific power cables.

    The Ambu® aBox™ 2 has the following physical and performance characteristics:

    • · Can process and display live imaging data from Ambu® aScope™ 5 Uretero to a monitor
    • · Can record, store and transport image data from Ambu® aScope™ 5 Uretero
    • Is a portable device with an integrated monitor, and the possibility to connect to an external monitor
    AI/ML Overview

    The provided text pertains to a 510(k) premarket notification for the Ambu® aScope™ 5 Uretero and Ambu® aBox™ 2. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing a de novo clinical trial to prove a new device meets specific performance criteria.

    Therefore, the document describes verification and validation testing against predefined specifications and established standards, not a comparative clinical study with acceptance criteria in the manner requested for an AI/device performance study.

    Based on the provided text, the device's performance was evaluated through a series of non-clinical (bench) tests to demonstrate that it performs as expected and meets its predefined specifications. There is no mention of an AI component requiring a study to prove improved human reader performance with AI assistance, nor is there information about ground truth established by experts, sample sizes for training or test sets in the context of an AI model.

    The document explicitly states: "All included bench tests, which have been designed to evaluate substantial equivalence as well as the product's conformity to established quality and performance measures, have been successfully conducted, and have passed the predefined acceptance criteria."

    Here's an attempt to answer your questions based only on the provided text, highlighting what is and is not present:


    Acceptance Criteria and Study for Ambu® aScope™ 5 Uretero and Ambu® aBox™ 2

    The acceptance criteria and "study" described in this 510(k) are primarily focused on non-clinical verification and validation testing to demonstrate substantial equivalence to predicate devices, rather than a clinical trial for an AI-powered device. The device in question is a flexible digital video ureteroscope and its display unit. There is no indication of an AI component in this submission.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists categories of tests performed and states that the device "met the test specifications set" and "passed the predefined acceptance criteria." However, it does not provide a specific table with quantitative acceptance criteria values or detailed reported performance values for each test. Instead, it offers a high-level summary of the types of tests conducted.

    Category of TestAcceptance Criteria (General)Reported Device Performance (General)
    Verification Tests"Test specifications set""Performed as expected and met the test specifications set."
    Insertion portion dimensions and performance"Test specifications set""Performed as expected and met the test specifications set."
    Working channel performance"Test specifications set""Performed as expected and met the test specifications set."
    Bending performance"Test specifications set""Performed as expected and met the test specifications set."
    Irrigation performance"Test specifications set""Performed as expected and met the test specifications set."
    Suction performance"Test specifications set""Performed as expected and met the test specifications set."
    Functionality"Test specifications set""Performed as expected and met the test specifications set."
    Optical Performance Tests"Test specifications set" (e.g., ISO 8600-3)"Performed as expected and met the test specifications set."
    Field of view (ISO 8600-3)"Test specifications set""Performed as expected and met the test specifications set."
    Direction of view (ISO 8600-3)"Test specifications set""Performed as expected and met the test specifications set."
    Sharpness and Depth of field"Test specifications set""Performed as expected and met the test specifications set."
    Image intensity uniformity"Test specifications set""Performed as expected and met the test specifications set."
    Geometric distortion"Test specifications set""Performed as expected and met the test specifications set."
    Resolution"Test specifications set""Performed as expected and met the test specifications set."
    Color performance"Test specifications set""Performed as expected and met the test specifications set."
    Noise performance"Test specifications set""Performed as expected and met the test specifications set."
    Dynamic range"Test specifications set""Performed as expected and met the test specifications set."
    Safety & Other TestsRelevant ISO/IEC/ASTM standards"Performed as expected and met the test specifications set."
    Photobiological safety (IEC 62471)"IEC 62471""Performed as expected and met the test specifications set."
    Transportation study (ASTM D4169)"ASTM D4169""Performed as expected and met the test specifications set."
    Sterilization validation (ISO 11135)"ISO 11135""Performed as expected and met the test specifications set."
    Stability study (ASTM F1980)"ASTM F1980""Performed as expected and met the test specifications set."
    Sterile packaging integrityN/A (implied by sterile packaging)"Performed as expected and met the test specifications set."
    Biocompatibility (ISO 10993)"ISO 10993-1, 5, 23, 10, 11""Assessed according to relevant standards."
    Electrical Safety (IEC 60601-1, 60601-2-18)"IEC 60601-1 and IEC 60601-2-18""Performed as expected and met the test specifications set."
    EMC (IEC 60601-1-2)"IEC 60601-1-2""Performed as expected and met the test specifications set."

    2. Sample Size for Test Set and Data Provenance:

    The document describes bench testing, not clinical studies with patient data. Therefore, the concept of "sample size for the test set" in the context of patient data is not applicable here. The "test set" would refer to the number of devices or components tested during verification and validation. The document does not specify the number of units tested for each category. Data provenance (country of origin, retrospective/prospective) is not applicable to these bench tests.

    3. Number of Experts and Qualifications for Ground Truth:

    Not applicable. The ground truth for this device's performance is based on engineering specifications and recognized industry standards (e.g., ISO, IEC, ASTM), not expert consensus from medical image review.

    4. Adjudication Method for the Test Set:

    Not applicable. This was not a study involving human interpretation of medical images or data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a MRMC comparative effectiveness study was not performed. The device is a ureteroscope and a display unit, not an AI diagnostic tool designed to assist human readers.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. This device does not have an AI algorithm with standalone performance.

    7. Type of Ground Truth Used:

    The "ground truth" for this submission is based on engineering specifications, design requirements, and adherence to international performance and safety standards (e.g., ISO 8600-3 for optical performance, IEC 60601 for electrical safety, ISO 11135 for sterilization, ISO 10993 for biocompatibility). There is no "ground truth" derived from expert consensus, pathology, or outcomes data in the context of diagnostic accuracy.

    8. Sample Size for the Training Set:

    Not applicable. This device does not involve training an AI model, so there is no training set.

    9. How Ground Truth for the Training Set Was Established:

    Not applicable. As above, there is no AI training set.