    K Number
    K180834

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-10

    (133 days)

    Product Code
    Regulation Number
    870.5900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician with prescription privileges determines that temperature therapy is necessary or desirable.

    Indications for use for the Altrix system include:

    a. Maintain pre-set body temperature as determined by the physician

    b. Maintain normal body temperature during surgical procedures

    c. For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn units, and medical/surgical units

    d. Adult and pediatric patients

    e. Monitoring and controlling patient temperature

    f. Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients

    Device Description

    Altrix components include the controller, reusable hose set(s), thermal transfer devices, patient temperature probes and reusable adaptor cable(s). The controller regulates water temperatures between 4.0 - 40.0° C and circulates the heated or cooled water via hose set(s) through the thermal transfer device(s). A graphical display provides the user an interface for selecting desired water or patient temperature settings, operating modes, help menus and other key parameters. Visual indicators are displayed to inform the user of system status or when the user must confirm a setting selection. The system's water temperature and flow outputs are monitored to ensure optimal system operation.

    AI/ML Overview

    Note: The provided text is a 510(k) summary for a medical device (Altrix Precision Temperature Management System). It describes the device, its intended use, and the basis for its substantial equivalence to a predicate device. This document focuses on regulatory approval and does not contain detailed information about a clinical study or performance evaluation with specific acceptance criteria and outcome metrics that would typically be found in a study report.

    Therefore, for aspects like acceptance criteria expressed numerically, sample sizes for test sets, expert involvement in ground truth establishment, MRMC studies, and detailed standalone performance, the information is not present in the provided text. The document primarily discusses bench testing for software modifications and mentions "acceptable results" without quantifying them.


    Here's an attempt to structure the answer based on the provided text, highlighting where information is absent and where inferences might be drawn from the context of regulatory submission:

    The provided document describes the 510(k) submission for the Altrix Precision Temperature Management System, focusing on modifications to the device's software, minor components, and labeling. The core of the submission is to demonstrate substantial equivalence to a previously cleared predicate device (Altrix Precision Temperature Management System - K152266).

    Given the nature of a 510(k) summary for modifications, the "study" proving the device meets acceptance criteria is primarily an engineering verification and validation process, not a clinical trial with human subjects testing the primary efficacy of the device.


    Acceptance Criteria and Device Performance (Inferred from Regulatory Context)

    Since this 510(k) is for modifications, the acceptance criteria would revolve around demonstrating that the modifications do not adversely affect the device's safety and effectiveness and that the modified device continues to meet the performance specifications established for the original predicate device. The "reported device performance" is described qualitatively as "successfully completed" and "acceptable results."

    Acceptance Criteria (Inferred from 510(k) Submission for Modifications)Reported Device Performance (From Text)
    Software Modifications:
    Compliance with ANSI AAMI IEC 62304:2006 (Medical device software - Software life cycle processes) and related guidances."Software modifications were completed using a software development life cycle process in accordance with FDA recognized standard 13-32, ANSI AAMI IEC, 62304:2006, Medical device software - Software life cycle processes, and related guidances."
    Meet specified requirements through non-clinical bench testing."Non-clinical bench testing was successfully completed to verify the software modifications meet specified requirements."
    Validate user needs with acceptable results."Software modifications were also validated to user needs with acceptable results."
    Other Modifications (minor components, labeling):
    Meet specified requirements through verification."Other modifications were also successfully verified to meet specified requirements."
    Overall Performance:
    Maintain substantial equivalence to the predicate device."The results of the non-clinical bench testing concluded that the subject device is substantially equivalent to the predicate device."
    No adverse impact on safety or effectiveness due to modifications.Implied by the conclusion of substantial equivalence.

    Study Details:

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not specified quantitatively. The testing involved "non-clinical bench testing." This typically implies testing on devices themselves rather than a dataset of patient images or readings.
      • Data Provenance: Not applicable in the traditional sense of patient data. The "testing" refers to the verification and validation of software and hardware changes on the device itself.
      • Retrospective or Prospective: Not applicable as it's bench testing, not a clinical data study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. As this was bench testing for software and minor component modifications, the "ground truth" would be established by engineering specifications, validated test protocols, and potentially expert review of the software design and functionality, rather than human experts labeling medical data.
    3. Adjudication method for the test set:

      • Not Applicable. There is no mention of human adjudication for the test results described, as the testing focuses on internal device performance against specifications.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study was not conducted. This device is a thermal regulating system, not an imaging AI diagnostic aid, and the submission is for device modifications, not a de novo AI product. The text does not mention any AI components assisting human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Partially Applicable / Assumed. The "Software modifications were also validated to user needs with acceptable results" and "Non-clinical bench testing was successfully completed to verify the software modifications meet specified requirements" implies testing the algorithm's performance in isolation to ensure it functions as intended based on its design specifications. However, specific standalone performance metrics (e.g., accuracy, precision for a measurement) are not provided in this summary.
    6. The type of ground truth used:

      • Engineering Specifications / Functional Requirements. For software and hardware modifications, the "ground truth" is defined by the device's design specifications, functional requirements, and the expected output under various conditions, as opposed to clinical outcomes or expert consensus on patient data.
    7. The sample size for the training set:

      • Not Applicable. This is a 510(k) for modifications to a traditional medical device, not a machine learning or AI model development that requires a "training set" of data.
    8. How the ground truth for the training set was established:

      • Not Applicable. (See point 7).
    Ask a Question

    Ask a specific question about this device

    K Number
    K152266

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-02

    (479 days)

    Product Code
    Regulation Number
    870.5900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician with prescription privileges determines that temperature therapy is necessary or desirable.

    Indications for use for the Altrix system include:
    a. Maintain pre-set body temperature as determined by the physician
    b. Maintain normal body temperature during surgical procedures
    c. For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn units, and medical/surgical units
    d. Adult and pediatric patients
    e. Monitoring and controlling patient temperature
    f. Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients

    Device Description

    The Altrix Precision Temperature Management System components include the controller, reusable hose set(s), thermal transfer devices, patient temperature probes and reusable adaptor cable(s). The controller regulates water temperatures between 4.0° C (39.2° F) and 40.0° C (104.0° F) and circulates the heated or cooled water via hose set(s) through the thermal transfer device(s). A graphical display provides the user an interface for selecting desired water or patient temperature settings, operating modes, help menus and other key parameters. Visual indicators are displayed to inform the user of system status or when the user must confirm a setting selection. The system's water temperature and flow outputs are monitored to ensure optimal system operation.

    The controller can supply water to an individual or multiple thermal transfer devices simultaneously with each of these circuits monitored separately. Three operating modes are available to facilitate patient care: Automatic, Manual and Monitor. The patient temperature probe is used with the controller to provide closed loop feedback for automatic patient temperature management and monitoring. The controller alarms activate visual and audible indications for when safety parameters are exceeded or it detects system function or performance irregularities. The Altrix system also includes the ability to provide a patient temperature output reference signal to be connected to a non-specific third party device/system.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Altrix Precision Temperature Management System. It primarily focuses on demonstrating substantial equivalence to a predicate device (Medi-Therm Hyper/Hypothermia System) through a comparison of technological characteristics and non-clinical bench testing.

    Therefore, the acceptance criteria and study information requested typically associated with clinical performance studies (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for training) are not fully available or directly applicable in the same way as they would be for a device that relies on clinical performance data for its clearance.

    Here's a breakdown of the information that is available, and what is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the sense of predefined thresholds for clinical performance metrics (like sensitivity, specificity, accuracy) using a clinical dataset. Instead, the "acceptance criteria" for this premarket notification are demonstrated through:

    • Substantial Equivalence: The primary acceptance criterion is that the Altrix device is substantially equivalent to the predicate device (Medi-Therm Hyper/Hypothermia System) in terms of indications for use, technological characteristics, safety, and performance.
    • Compliance with Standards: The device must comply with relevant recognized safety and performance standards.
    • Successful Bench Testing: Non-clinical bench testing must successfully verify the device's function and performance to specified engineering requirements.

    Table of Performance (derived from "Technological Characteristics and Substantial Equivalence Summary" and "Bench and Standards Testing"):

    Category / CriterionAcceptance Criteria (Predicate or Standard)Reported Device Performance (Altrix)
    Indications for UseSame as Medi-ThermSame as Medi-Therm
    Temperature Control Range - Water4 - 42° C (Medi-Therm)4.0 - 40.0° C
    Controller Accuracy - Water Temperature± 0.8° C (Medi-Therm)± 0.3° C
    Display Accuracy - Water Temperature± 0.3° C (Medi-Therm)± 0.2° C
    Settings Range - Patient Temperature30 - 41° C (Medi-Therm)32.0 - 38.0° C
    Controller Accuracy - Patient Temperature± 0.5° C (Medi-Therm)± 0.3°C (25.0° - 45.0° C)
    Patient Temperature Display Accuracy± 0.3° C (Medi-Therm)± 0.3° ( 25.0° - 45° C)
    High Water Temperature Limit (Software)43° C (Medi-Therm)>42.5° C
    Water Temperature Deviation AlarmNo (Medi-Therm)Yes; ±0.8° C
    General Safety Standard ComplianceUL 416 (Medi-Therm)ANSI/AAMI ES60601-1:2005(R)2012 and A1 2012, C1: 2009/(R)2012 and A2:2010/(R)2012 (and other specific IEC/ISO standards listed)
    EMC/EMI Standard ComplianceEN60601-1-2 (Medi-Therm)IEC60601-1-2
    Bench Testing (General)Successful verification of functions and performance to specified requirementsSuccessfully verified warming/cooling performance, mechanical/thermal overload, controller abuse, environmental, fluid intrusion, packaging. Disinfection validated.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Not Applicable in a clinical sense. The device relies on non-clinical bench testing, not a clinical "test set" of patient data. The provenance of the bench testing data would be from Stryker Medical's internal labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not Applicable. No clinical test set with human-established ground truth was used for this 510(k) clearance, as clinical testing was deemed not required.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted diagnostic device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a temperature management system, not an algorithm, and its performance is evaluated through engineering and functional testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the non-clinical bench testing, the "ground truth" refers to engineering specifications, physical measurements (e.g., temperature, flow rates), and compliance with recognized industry standards. These are objective, measurable criteria.

    8. The sample size for the training set

    • Not Applicable. There is no mention of a training set as this is not a machine learning/AI device.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set was used.