Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233893

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-01-25

    (48 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AlphaVent Knotless SP PEEK Anchor is intended to be used for soft-tissue to bone fixation in the shoulder, foot and ankle, knee, hand and wrist, elbow, and hip in skeletally mature pediatric and adult patients. It is indicated for use in the following procedures:

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair

    Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Quadriceps Tendon Repair, Secondary or Adjunct Fixation for ACL/PCL Reconstruction or Repair, Meniscal Root Repair, MPFL Repair/Reconstruction

    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Reconstruction, Lateral Epicondylitis Repair

    Hip: Capsular Repair, Acetabular Labral Repair, Gluteus Medius Repair, Proximal Hamstring Repair

    Device Description

    The AlphaVent Knotless SP PEEK Anchors (herein referred to as the subject device(s)) are bone anchors with a screw-in mechanism. The subject device system is comprised of a poly-ether-etherketone (PEEK) eyelet and anchor body, pre-assembled onto a disposable stainless-steel inserter, which enables insertion of the anchor into bone via self-punching or after creation of a pilot hole. The devices are single use, provided sterile, and are packaged in sterile barrier systems (SBS). The AlphaVent Knotless SP PEEK Anchors are sterilized via Ethylene Oxide to a Sterility Assurance Level of 106. The devices are intended to be used in a professional healthcare facility.

    AI/ML Overview

    This document is a 510(k) summary for the AlphaVent Knotless SP PEEK Anchor, a medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with acceptance criteria and device performance in the way typically seen for a diagnostic or AI-driven device.

    Therefore, many of the requested categories for describing acceptance criteria and a study proving device performance are not applicable or cannot be extracted from this type of regulatory submission. This document highlights benchtop testing for product safety and equivalence, not clinical performance against ground truth.

    Here's an attempt to answer the questions based only on the provided text, indicating when information is not present:


    Acceptance Criteria and Device Performance Study (Based on 510(k) Summary)

    This 510(k) summary focuses on demonstrating substantial equivalence of the AlphaVent Knotless SP PEEK Anchor to a legally marketed predicate device (Arthrex SwiveLock Anchor, K203495) and a reference device (Stryker AlphaVent PEEK Suture Anchor, K231093). The "study" described is a series of non-clinical benchtop tests, a biological risk assessment, and bacterial endotoxin testing, designed to show the new device performs at least as safely and effectively as the predicate.

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device PerformanceComments
    Ultimate Tensile Strength (UTS)Equivalent or higher UTS as compared to predicate devicesDemonstrated equivalent or higher UTS as compared to the predicate devicesNon-clinical benchtop testing. No specific numerical values provided for criteria or performance; reported comparatively.
    Insertion TestingSuccessful insertionSuccessful insertionNon-clinical benchtop testing. No specific metrics provided.
    Biological Risk AssessmentMet requirements of ISO10993-1Met requirements of ISO10993-1
    Bacterial Endotoxin Testing (LAL test)< 20 EU/deviceMet the limit acceptance criterion of < 20 EU/deviceBased on FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile''.

    2. Sample sizes used for the test set and the data provenance

    • Sample size for test set: Not specified in the provided text for any of the benchtop tests.
    • Data provenance: The tests are non-clinical benchtop tests, implying laboratory-controlled conditions. There is no mention of country of origin of data or whether it's retrospective or prospective, as these terms typically apply to patient-level clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. For this device (a mechanical anchor used for soft-tissue to bone fixation), "ground truth" is established through engineering and biological performance standards, not through expert human interpretation of data/images. The tests evaluate physical properties and biocompatibility.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical data (e.g., image interpretation). This submission details benchtop engineering and biological tests, not clinical performance requiring human expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. An MRMC study is relevant for diagnostic devices, especially those incorporating AI. This document concerns a physical surgical anchor. There is no AI component mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a component of a surgical procedure, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the non-clinical tests, the "ground truth" is defined by established engineering standards, material specifications, and biological safety guidelines (e.g., ISO10993-1, USP <85>). The performance is measured against these objective criteria rather than subjective human interpretation or clinical outcomes data presented in this document.

    8. The sample size for the training set

    • Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Not applicable. (See point 8).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1