Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K212491

    Validate with FDA (Live)

    Date Cleared
    2021-09-02

    (24 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Air Relax Pro Model AR-4.0 is intended for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Pro Model AR-4.0 simulates kneading and stroking of tissues by using an inflatable garment.

    Device Description

    This submission is for the Diode Art Engineering Air Relax Pro Model AR-4.0. The Air Relax Pro Model AR-4.0 is a powered inflatable tube massager. It is intended to temporarily relieve minor muscle aches and pains, and to temporarily increase circulation to the treated areas. It simulates manual kneading and stroking of tissues by use of an inflatable pressure cuff. The device is to be used by people who are in good health.

    The device is a Class II, type BF applied part that receives power a separately approved external IEC 60601-1 compliant power supply or optional battery pack.

    The Air Relax Pro Model AR-4.0 consists of an air compressor unit with a control system, an inflatable "garment" (arms, legs and hips), plastic air tubing with a proprietary connector for connecting the device to the garment. A description of each of these components is provided below. The hip garment is also referred to as "shorts".

    The user interface is a front panel display and buttons.

    The Air Relax Pro Model AR-4.0 contains an air compressor with a system control that that allows the user to adjust the amount of air coming from the air compressor and going to the individual segments of the inflatable garment.

    There is no electrical contact with the user and the device does not transfer or detect energy to or from the user.

    The user interface of the Air Relax Pro Model AR-4.0 provides for starting and stopping the massage treatment, allows for adjusting time and intensity (pressure) of the treatment. The device also provide a proprietary keyed connector to the tubing which connects to the garment. The tubing connector at the garment is also proprietary. The proprietary connectors ensure that neither the device, tubing nor garment can be misconnected to any other device or garment.

    Pressure selection is performed by pressing pressure button multiple times. Pressure level is selectable between 40 and 170 mmHg

    There are four modes (Progressive, Sequential, Drain and Overlay) that determine the inflation sequence of the cells within the garments.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Air Relax Pro Model AR-4.0, a powered inflatable tube massager. It focuses on demonstrating substantial equivalence to a predicate device (Air Relax Model AR-3.0) rather than providing detailed acceptance criteria and a study proving those criteria for a novel device.

    Therefore, much of the requested information regarding detailed acceptance criteria for specific performance metrics and a dedicated study to prove these against novel criteria is not present in the provided document. The document primarily highlights the similarities to a predicate device and relies on compliance with general medical device standards.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria for the Air Relax Pro Model AR-4.0's performance in terms of its intended therapeutic effect (e.g., specific thresholds for "temporary relief of minor muscle aches" or "temporary increase in circulation").

    Instead, the document establishes "acceptance" by demonstrating substantial equivalence to a predicate device (Air Relax Model AR-3.0) through a comparative table (page 5). The criteria for acceptance are implicitly that the new device is as safe and effective as the predicate, meaning any differences do not raise new safety or efficacy concerns.

    Here's a summary of the comparative performance and specifications, which serve as the implicit "acceptance criteria" through equivalence:

    Feature/CriteriaAir Relax Pro Model AR-4.0 (Subject Device)Air Relax Plus Model AR-3.0 (Predicate Device)Comment (Acceptance Status)
    Indications for UseIdenticalIdenticalIdentical (Accepted)
    OTC or PrescriptionOTCOTCIdentical (Accepted)
    Environment of UseClinics, hospital, athlete training, homeClinics, hospital, athlete training, homeIdentical (Accepted)
    Compliance with StandardsES 60601-1, IEC 60601-1-2, IEC 60601-1-11ES 60601-1, IEC 60601-1-2, IEC 60601-1-11Identical (Accepted)
    Mode of OperationSequential/PeristalticSequential/PeristalticIdentical (Accepted)
    Power100~240V, 50/60Hz100~240V, 50/60HzIdentical (Accepted)
    Device Pressure Range40 - 170 mmHg60 - 170 mmHgSimilar (Predicate has a higher minimum pressure). Comment: "Similar to predicate except low pressure, it does not affect safety and effectiveness." (Considered Acceptable)
    Garments materialNylon with a Polyurethane LaminateNylon with a Polyurethane LaminateIdentical (Accepted)
    Leg Attachment SizesSize "2": 34 X 15.35, "3": 38.2 X 15.35, "4": 42 X 15.35Size "2": 34 X 15.35, "3": 38.2 X 15.35, "4": 42 X 15.35Identical in size and materials. "Number of Inflatable garment segments differs." (Considered Acceptable due to no impact on safety and effectiveness)
    Arm Attachment SizesSize "1": 34 X 15.35, "2": 39.3 X 15.35Size "1": 34 X 15.35, "2": 39.3 X 15.35Identical in size and materials. "Number of Inflatable garment segments differs." (Considered Acceptable due to no impact on safety and effectiveness)
    Shorts Attachment SizesSize "1": 20.8 X 32.5, "2": 24.6 X 32.5Size "1": 20.8 X 32.5, "2": 24.6 X 32.5Identical in size and materials. "Number of Inflatable garment segments differs." (Considered Acceptable due to no impact on safety and effectiveness)
    Number of Inflatable Garment Segments64Different. Comment: "No impact on safety and effectiveness" (Considered Acceptable)
    Weight2.2 kg (4.85 pounds)1.67kg (3.7 pounds)Different. Comment: "Weight is different but it does not affect safety and effectiveness." (Considered Acceptable)
    Dimensions (W x H x D)7.7" x 7.1" x 10.4"9.5" x 4.7" x 7.5"Different. Comment: "Size is different but it does not affect safety and effectiveness." (Considered Acceptable)
    Housing Materials and ConstructionsMolded ABS enclosureMolded ABS enclosureIdentical (Accepted)
    Patient contactNon-conductive garmentsNon-conductive garmentsIdentical (Accepted)
    Safety FeaturesButton at control unit allows stop/pauseButton at control unit allows stop/pauseIdentical (Accepted)
    Modes"PROG", "SEQT", "OVLAY", "DRAIN""P", "S", "F", "Target"Similar. Comment: "Similar to predicate. Differences do not affect safety and effectiveness." (Considered Acceptable) Note: "OVLAY" vs "F" seem functionally similar, "DRAIN" vs "Target" seem different but deemed not to affect safety/effectiveness.
    Treatment Duration5-95 minutes15 or 30 minutesDifferent. Comment: "Extended duration for subject device but predicate time can be re-enabled any number of times so no significant difference." (Considered Acceptable)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states:

    • "No animal testing was performed" (page 8)
    • "No clinical testing was performed" (page 9)

    This indicates that there was no specific test set of patient data used to prove the device's performance against clinical acceptance criteria. The evaluation was primarily based on:

    • Engineering bench testing (controls, indicators, performance)
    • Compliance with recognized electrical and medical device safety standards (AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11).
    • Demonstration of substantial equivalence to a legally marketed predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Since no clinical test set was used, there was no need for experts to establish ground truth from patient data. The "ground truth" for substantial equivalence was established by the FDA's review process comparing the device to the predicate based on technical specifications and declared compliance with standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical test set or human interpretation was involved that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical therapy massager, not an AI-assisted diagnostic or therapeutic tool that involves human readers or interpretation of medical images. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical therapy massager, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the purpose of this 510(k), the "ground truth" for demonstrating safety and effectiveness relies on:

    • Established performance of the predicate device: The predicate (Air Relax Model AR-3.0) is already legally marketed and presumed safe and effective for its indications.
    • Compliance with recognized industry standards: The device's adherence to standards like IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-1-11 (home healthcare environment) provides "ground truth" for its electrical and mechanical safety and basic performance characteristics.
    • Bench testing: Verifying the function of controls, indicators, and overall performance through engineering tests.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211460

    Validate with FDA (Live)

    Date Cleared
    2021-05-25

    (14 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment.

    Device Description

    This submission is for the Diode Art Engineering Air Relax Plus Model AR-3.0. The Air Relax Plus Model AR-3.0 is a powered inflatable tube massager. It is intended to temporarily relieve minor muscle aches and pains, and to temporarily increase circulation to the treated areas. It simulates manual kneading and stroking of tissues by use of an inflatable pressure cuff. The device is to be used by people who are in good health.

    The device is a Class II, type BF applied part that receives power from a separately approved external IEC 60601-1 compliant power supply or optional battery pack.

    The Air Relax Plus Model AR-3.0 consists of an air compressor unit with a control system, an inflatable "garment" (arms, legs and hips), plastic air tubing with a proprietary connector for connecting the device to the garment A description of each of these components is provided below. The hip garment is also referred to as "shorts".

    The user interface is a front panel display and buttons.

    The Air Relax Plus Model AR-3.0 contains an air compressor with a system control that allows the user to adjust the amount of air coming from the air compressorand going to the individual segments of the inflatable garment.

    There is no electrical contact with the user and the device does not transfer or detect energy to or from the user.

    The user interface of the Air Relax Plus Model AR-3.0 provides for starting and stopping the massage treatment, allows for adjusting time and intensity (pressure) of the treatment. The device also provides a proprietary keyed connector to the tubing which connects to the garment. The tubing connector to the garment is also proprietary. The proprietary connectors ensure that neither the device, tubing nor garment can be misconnected to any other device or garment.

    Pressure selection is performed by pressing pressure button multiple times. There are four pressure levels with approximate pressure levels as below:

    Level 1: 60 mmHg Level 2: 100 mmHg Level 3: 140 mmHg Level 4: 170 mmHg

    There are three modes (Progressive, Sequential, Full massage P, S and F) that determine the inflation sequence.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Air Relax Plus Model AR-3.0, comparing it to a predicate device (Air Relax Model AR-1.0/AR-2.0, K191441) and another reference device (Rapid Reboot Compression Therapy System, K182668, inferred). The document focuses on demonstrating substantial equivalence for regulatory clearance rather than providing a detailed study proving the device meets specific acceptance criteria through clinical or performance metrics against pre-defined thresholds.

    Therefore, many of the requested details about acceptance criteria, specific study design, sample sizes, expert ground truth, adjudication methods, multi-reader multi-case studies, and standalone algorithm performance are not applicable or not explicitly provided in this type of regulatory submission. The goal of a 510(k) is to show the new device is as safe and effective as a legally marketed predicate, not necessarily to prove absolute effectiveness against novel criteria with dedicated clinical trials.

    The document primarily relies on bench testing and comparison to the predicate device's established safety and performance profile.

    Here's a breakdown of the information that can be extracted, and what is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The concept of "acceptance criteria" in the context of a 510(k) for a device like this is typically demonstrated by meeting recognized standards and by showing similar performance characteristics to the predicate device. Explicit quantitative acceptance criteria with numerical targets for clinical performance are not presented. Instead, the "acceptance criterion" is implicitly demonstrating that the new device is substantially equivalent to the predicate and complies with relevant safety standards.

    Criterion TypeAcceptance Criterion (Implicit)Reported Device Performance (Comparative)
    Indications for UseIdentical to predicate device for temporary relief of minor muscle aches and pains and temporary increase in circulation.Identical: "The Air Relax Plus Model AR-3.0 is intended for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax Plus Model AR-3.0 simulates kneading and stroking of tissues by using an inflatable garment."
    ClassificationClass II Device, IRP (21 CFR 890.5650)Identical: Class II Device, IRP (21 CFR 890.5650)
    OTC or PrescriptionOver-The-Counter UseIdentical: OTC
    Environment of UseClinics, hospital, athlete training, and home environments.Identical: Clinics, hospital, athlete training, and home environments.
    Compliance with StandardsES 60601-1, IEC 60601-1-2, IEC 60601-1-11Identical: ES 60601-1, IEC 60601-1-2, IEC 60601-1-11. Bench testing confirmed compliance.
    Mode of OperationSequential/Peristaltic, with options for various inflation sequences.Identical principle: Sequential/Peristaltic. Subject device has modes "P", "S", "F", and "Target" (user selects specific chamber). Predicate has "A", "B", "C", and "Manual" (user selects specific chamber). These are described as "Identical to predicate except mode names differ."
    PowerSafe and effective operation under specified power.Different but acceptable: Subject device: 100~240V, 50/60Hz. Predicate: 120V, 60Hz. "Although the power rating of the subject device is different, both devices comply with ANSI/AAMI ES60601-1, so the difference does not affect safety and effectiveness."
    Device Pressure RangeOperational pressure range for therapeutic effect.Lower range than predicate: Subject device: 60 - 170 mmHg. Predicate: 0 - 230 mmHg. The document does not state this lower range affects safety or effectiveness given the intended use.
    Garment MaterialSafe and durable materials for patient contact.Identical: Nylon with a Polyurethane Laminate.
    Garment SizesAppropriate sizes for various body parts.Leg Garment: Identical to predicate in size, construction, and materials. Arm Garment: Yes (Long: 13.8" x 33.5") vs. Predicate "No" / Reference "Yes" (Regular: 18" x 38", Long: 18" x 44"). Described as "Similar in size and construction. Materials are identical to K191441." Hip Garment: Yes (Large: 24.5" x 31.5") vs. Predicate "No" / Reference "Yes" (Regular: 26" x 32", Large: 26" x 35"). Described as "Similar in size and construction. Materials are identical to K191441."
    Safety FeaturesUser control to stop/pause therapy.Identical: "Button at control unit allows user to stop or pause therapy session at any time."
    Weight & DimensionsComparable to predicate for similar use.Identical: Weight 1.67 kg (3.7 pounds), Dimensions 9.5" x 4.7" x 7.5".
    Housing MaterialsDurable and safe enclosure.Identical: Molded ABS enclosure.
    Patient ContactNon-conductive materials.Identical: Non-conductive garments.
    Treatment TimeUser-selectable therapy duration.Similar but limited: Subject device: 15, 30 minutes. Reference device: 10, 20, 30 minutes with option to add 10 mins. "Similar to K182668 except time is limited to 30 minutes in subject device."

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • No clinical or animal testing was performed to generate a "test set" in the traditional sense for evaluating diagnostic or predictive performance.
      • Bench testing was done to verify controls, indicators, and performance, but the sample size (e.g., number of units tested, number of cycles) is not provided.
    • Data Provenance: Not applicable, as no clinical/animal data was generated. Bench testing would be internal lab data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. There was no clinical study requiring expert ground truth for a test set.

    4. Adjudication method for the test set

    • Not applicable. There was no clinical study requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. The device is a massager, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable, as this is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. No ground truth in the clinical sense was required for this type of device and regulatory submission. The "ground truth" for proving substantial equivalence relies on documentation of design, materials, and bench test performance against established safety standards, and direct comparison to the predicate device.

    8. The sample size for the training set

    • Not applicable. No training set was used as this device does not involve machine learning or AI.

    9. How the ground truth for the training set was established

    • Not applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191441

    Validate with FDA (Live)

    Date Cleared
    2021-03-05

    (645 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Air Relax/ Compressible Limb Sleeve System (Model: AR-1.0) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax/ Compressible Limb Sleeve System (Model: AR-1.0, AR-2.0) simulates kneading and stroking of tissues by using an inflatable garment.

    Device Description

    AR-1.0, AR-2.0is combined with leg, foot, garments. Leg garment has four air chambers operated by STEP3 controller and foot has one air chamber operated by STEP1. STEP3 controls a leg garment compressing sequentially in order of foot, calf and thigh and then injecting fresh air inside a leg garment through holes of garment to prevent from generating sweat by long wear of a garment.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "Air Relax, Compressible Limb Sleeve System (Model: AR-1.0, AR-2.0)". It is a submission to the FDA for market clearance, arguing for substantial equivalence to predicate devices.

    Based on the provided text, the device did not undergo a clinical study or performance study to prove it meets specific acceptance criteria based on human or algorithm performance metrics. The submission argues for substantial equivalence based on non-clinical testing and comparison to legally marketed predicate devices.

    Therefore, many of the requested items related to "acceptance criteria" and "study that proves the device meets the acceptance criteria" in terms of performance metrics (like sensitivity, specificity, or human improvement with AI assistance) cannot be answered directly from this document because such a study was explicitly not performed.

    Here's a breakdown of the information that can be extracted and an explanation of why other requested information is not present:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: For this 510(k) submission, the "acceptance criteria" are related to demonstrating substantial equivalence through adherence to recognized standards for safety and electrical performance, and through a comparative analysis against predicate devices. The implicit acceptance criterion is that the device is "as safe and effective" as the legally marketed predicates.
    • Reported Device Performance: The "performance" reported is primarily in terms of compliance with non-clinical standards and the comparison of technical specifications to predicate devices. No specific diagnostic accuracy or AI performance metrics are reported.
    Acceptance Criteria (Implicit for SE)Reported Device Performance (Compliance/Comparison)
    Electrical Safety (IEC 60601-1:2014)Device met IEC 60601-1:2014 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
    Electromagnetic Compatibility (EMC – IEC 60601-1-2:2016)Device met IEC 60601-1-2:2016 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
    Home Healthcare Environment Requirements (IEC 60601-1-11:2015)Device met IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
    Biocompatibility (EN ISO 10993-5:2009 for Cytotoxicity)Device met EN ISO 10993-5:2009 Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
    Biocompatibility (EN ISO 10993-10:2010 for Irritation and Skin Sensitization)Device met EN ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization.
    Substantial Equivalence in Intended UseThe device's intended use ("temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Relax/ Compressible Limb Sleeve System (Model: AR-1.0, AR-2.0) simulates kneading and stroking of tissues by using an inflatable garment") is stated to be "Same as K182668 (Primary predicate) and K193354 (Secondary predicate)".
    Substantial Equivalence in Principle of Operation, Technological Characteristics, and Safety/EffectivenessThe document concludes that the "subject device has same intended use and principle of operation, similar technological characteristics as that of predicate devices." It also states, "the differences between these parameters would not adversely impact the safety and effectiveness" and that the device "is at least as safe and effective as the all predicate devices and technologically comparable to the reference device and doesn't raise any new safety and/or effectiveness concerns."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. The document explicitly states: "No clinical study is included in this submission." Therefore, there is no "test set" in the sense of patient data for evaluating diagnostic or intervention performance. The "tests" mentioned are non-clinical (electrical safety, EMC, biocompatibility).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. As no clinical study was performed, there was no test set requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical study or test set for performance evaluation was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a "Powered Inflatable Tube Massager," not an AI-powered diagnostic or AI-assisted interpretation device. The submission explicitly states no clinical study was done, and therefore no MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical device (massager), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. No clinical study or performance evaluation requiring ground truth data was performed. The "ground truth" for this submission are the established safety and performance standards (IEC, ISO) and the characteristics of the predicate devices.

    8. The sample size for the training set

    • Not applicable. This is a physical device, not a machine learning model. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    • Not applicable. As above, no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173158

    Validate with FDA (Live)

    Device Name
    Air Relax
    Date Cleared
    2019-05-30

    (608 days)

    Product Code
    Regulation Number
    890.5650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UAM-8100, UAM-8100N is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as : Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.

    Device Description

    UAM-8100, UAM-8100N is a system to improve the blood circulation of patients and User can use it personally under their doctor's indication and guidance. UAM-8100, UAM-8100N is combined with leg and foot garments. The leg garment has four air chambers operated by a controller. The leg garment is compressed sequentially in order of foot, calf, and thigh, and then fresh air is injected inside the leg garment through holes

    AI/ML Overview

    The provided document is a 510(k) premarket notification decision letter from the FDA for the "Air Relax/Compressible Limb Sleeve System Model No. UAM-8100, UAM-8100N." This type of document is a regulatory approval, not a clinical study report.

    Therefore, the information required to answer your questions about acceptance criteria and a study proving device performance (tables, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set details) is not present in this document.

    This document primarily focuses on demonstrating substantial equivalence to a predicate device (K130385) based on non-clinical tests (biocompatibility, electrical safety, electromagnetic compatibility, and software verification/validation) and comparing design specifications. It does not detail clinical effectiveness studies with specific performance metrics or acceptance criteria for those metrics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1