Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K171451

    Validate with FDA (Live)

    Date Cleared
    2017-08-08

    (83 days)

    Product Code
    Regulation Number
    886.3920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ahmed® Glaucoma Valve Model FP8 is indicated for the management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medication, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.

    Device Description

    The Ahmed® Glaucoma Valve Model FP8 (AGV-FP8) is a valved aqueous drainage implant designed to regulate intraocular pressure in eyes suffering from intractable glaucoma. The Ahmed® device is comprised of a silicone drainage tube that is connected to a valve mechanism. This valve mechanism is the same in the predicate AGV-FP7. The valve mechanism consists of a silicone sheet folded and pressed between two complimentary polypropylene plates. The valve mechanism is securely positioned in a pocket inside of a silicone endplate that serves to distribute the aqueous humor from the anterior chamber of the eye over the surface of the endplate. The valve in the AGV-FP8 and the predicate AGV-FP7 behaves like a variable resistor, decreasing resistance to allow more flow when intraocular pressure is high. When pressure is low, the resistance to fluid outflow is high and the valve closes, thereby preventing hypotony. By means of the valve mechanism, the AGV-FP8 and the predicate AGV-FP7 maintain intraocular pressure within the appropriate physiological range.

    AI/ML Overview

    The provided text is a 510(k) Summary of Safety and Effectiveness for the Ahmed® Glaucoma Valve Model FP8. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a study with the common structure of an AI/ML device study.

    Therefore, many of the requested elements (e.g., sample size for test set, number of experts, adjudication method, MRMC comparative effectiveness study, standalone performance, training set size, ground truth for training set) are not applicable or cannot be extracted directly from this type of medical device submission.

    However, I can extract information related to performance testing in general and clinical study results that compare the device to others, which somewhat aligns with the spirit of the request.

    Here's an analysis based on the provided text, while acknowledging its limitations for an AI/ML context:

    1. Table of Acceptance Criteria and Reported Device Performance

    Strictly speaking, the document doesn't present "acceptance criteria" in the format one might expect for statistical performance metrics of an AI/ML model (e.g., "sensitivity >= X%"). Instead, it focuses on demonstrating that the performance of the new device (AGV-FP8) is equivalent to legally marketed predicate devices.

    The "performance" described is in terms of clinical outcomes, specifically success rates, IOP reduction, and reduction in glaucoma medications, adverse effects, and complications. The "acceptance criteria" are implied to be achieving outcomes that are not statistically significantly different from the predicate devices or showing effectiveness in specific patient populations.

    Acceptance Criteria (Implied)Reported Device Performance (Summary from Clinical Studies)
    Equivalence to Predicate Devices (FP7)No statistically significant differences between AGV-FP8 and AGV-FP7 in preservation of vision, IOP reduction, or decrease in the number of glaucoma medications. (Study 1)
    Effectiveness in Refractory Primary Congenital GlaucomaAGV-FP8 was found to be an effective treatment in this difficult-to-treat group. (Study 2)Similar long-term IOP reduction, decrease in number of glaucoma medications, and success rates compared to combined trabeculotomy-trabeculectomy for refractory primary congenital glaucoma. (Study 3)
    Effectiveness After Failed Glaucoma Drainage Device ImplantationAGV-FP8 implantation was described as a good choice for surgical treatment when the first glaucoma drainage device failed. (Study 4)
    Safety Profile (Complications/Adverse Events)In Study 3, a higher rate of hyphema was reported in the combined trabeculotomy-trabeculectomy group, with other complication rates similar between the two groups. In Study 4, corneal decompensation was found in some cases, but no other serious complications were reported. Absence of serious complications were not explicitly linked to specific thresholds.
    Physical and Chemical StabilityDevice maintains performance characteristics and structural integrity after aqueous exposure. Chemical analysis and aqueous aging chemical testing assess hazards and stability. (Non-clinical testing)
    BiocompatibilitySatisfactory biocompatibility. (Non-clinical testing)

    2. Sample Size Used for the Test Set and Data Provenance

    This information is typically for AI/ML model validation, which is not the primary focus here. Instead, the document refers to human clinical studies:

    • Study 1 (Koh et al., 2013): Retrospective review of records from adult refractory glaucoma patients. The exact number of patients undergoing AGV-FP8 or AGV-FP7 implantation is not explicitly stated in this summary but is implied to be a comparison between two groups. Provenance: Single center (details not provided, but the journal implies South Korea).
    • Study 2 (Dave et al., 2015): Retrospective noncomparative case series of eyes with refractory primary congenital glaucoma. The number of cases is not specified in the summary. Provenance: Not explicitly stated, but the journal implies India.
    • Study 3 (Helmy, 2016): Randomized, prospective, single-surgeon, comparative study including 66 eyes with refractory primary congenital glaucoma. Half underwent AGV-FP8 implantation. Provenance: Egyptian patients.
    • Study 4 (Ko et al., 2016): Retrospective review of clinical histories of 23 refractory glaucoma patients, 21 of whom underwent AGV-FP8 implantation. Provenance: Not explicitly stated, but the journal implies South Korea.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This concept of "ground truth" established by experts for a test set is not directly applicable to the clinical studies referenced in this 510(k) summary. These studies are clinical outcomes studies where the "ground truth" is typically the observed patient outcome (e.g., actual IOP, preserved vision, occurrence of complications) as assessed by the treating clinicians, not an independent panel establishing a "ground truth" for an AI model.

    • Study 1: "two surgeons at a single center" performed the implantations and likely contributed to follow-up data. Their specific qualifications (e.g., number of years of experience) are not stated.
    • Study 2: "a single surgeon" implanted the AGV-FP8. Qualifications not stated.
    • Study 3: "a randomized, prospective, single-surgeon, comparative study." Qualifications not stated.
    • Study 4: "Surgical outcomes of additional Ahmed glaucoma valve implantation in refractory glaucoma." Implies treating ophthalmic surgeons. Qualifications not stated.

    4. Adjudication Method for the Test Set

    As this is not an AI/ML diagnostic or prognostic study requiring ground truth consensus on images or other data, adjudication methods (like 2+1 or 3+1) are not typically used or reported in this context. Clinical outcomes are generally determined by treating physicians and medical records.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC study comparing human readers with and without AI assistance was done or is applicable here. The device is a physical implant (Ahmed Glaucoma Valve), not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The device is a physical medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in these clinical studies refers to the actual clinical outcomes observed in patients, such as:

    • Intraocular Pressure (IOP) measurements
    • Visual acuity
    • Number of glaucoma medications
    • Occurrence and type of adverse events/complications (e.g., hyphema, corneal decompensation)
    • Success rates (as defined by each study, e.g., defined IOP range without further surgical intervention for glaucoma).

    This is clinical outcomes data, as assessed and recorded by the surgical teams and clinicians in the respective studies.

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical implant, not an AI/ML model that requires a training set. The clinical "studies" are observational/comparative for the device's performance, not for training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an AI/ML model is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162060

    Validate with FDA (Live)

    Date Cleared
    2016-10-24

    (90 days)

    Product Code
    Regulation Number
    886.3920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ahmed™ Glaucoma Valve is indicated for the management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medications, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis

    Device Description

    The Ahmed™ Glaucoma Valve Model FP7 (AGV-FP7) is a valved aqueous drainage implant designed to regulate intraocular pressure in eyes suffering from intractable glaucoma. The Ahmed™ device is comprised of a silicone drainage tube that is connected to a valve mechanism. This valve mechanism is the same in the AGV-FP7 and the predicate AGV-S2. The valve mechanism consists of a silicone sheet folded and pressed between two complimentary polypropylene plates. The valve mechanism is securely positioned in a pocket inside of a silicone endplate that serves to distribute the aqueous humor from the anterior chamber of the eye over the surface of the endplate. The valve in the AGV-FP7 behaves like a variable resistor, decreasing resistance to allow more flow when intraocular pressure is high. When pressure is low, the resistance to fluid outflow is high and the valve closes, thereby preventing hypotony. By means of the valve mechanism, the AGV-FP7 maintains intraocular pressure within the appropriate physiological range.

    In both the AGV-S2 and AGV-FP7, the silicone sheet is folded and pressed between two polypropylene plates. In the AGV-S2, the bottom polypropylene plate is comprised of the polypropylene endplate body. In the AGV-FP7, the polypropylene bottom plate is separate from the silicone endplate material. The valve mechanism is inserted into a pocket in the silicone endplate to fixate the valve components to the endplate. Additional differences include stiffening ribs in the posterior half of the AGV-FP7 to add stiffness to the flexible endplate, The other predicate device, the Baerveldt Glaucoma Implant (BGI) also consists of a flexible silicone endplate which shares some features with the AGV-FP7, though the BGI endplate is larger in area. The AGV-FP7 endplate has the same curvature as the average human eye at its equator and also protects the valve from blockage by fibrous tissue. The endplate is made of flexible silicone. Inflammation and scarring around flexible silicone implants in animal ocular tissue was less pronounced than that found around rigid polypropylene.

    AI/ML Overview

    The provided text is a Premarket Notification [510(k)] Summary for the Ahmed™ Glaucoma Valve Model FP7. It describes the device, its intended use, and claims substantial equivalence to existing predicate devices. However, this document does not contain the detailed study information needed to fill out all the fields requested in your prompt.

    Here's what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document mentions that the AGV-FP7 has "similar IOP and complication rates to the predicate devices." Specifically, it states, "compared to the AGV-S2, the IOP of the AGV-FP7 groups was lower (within the acceptable physiological range) and fewer complications were reported."

    However, explicit acceptance criteria (e.g., "IOP must be reduced by X mmHg" or "complication rate must be < Y%") are NOT provided in this document. Therefore, a table with specific criteria and numerical performance cannot be fully constructed from the given text.

    Acceptance Criteria (Not explicitly stated in document)Reported Device Performance (as stated in document)
    Implicit: Similar IOP to predicate devices"similar IOP... to the predicate devices."
    Implicit: Comparable complication rates to predicate devices"similar... complication rates to the predicate devices."
    Implicit: IOP within acceptable physiological range"IOP of the AGV-FP7 groups was lower (within the acceptable physiological range)" compared to AGV-S2.
    Implicit: Fewer complications than predicate devices"fewer complications were reported" compared to AGV-S2.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document references several clinical studies:

    • The Ahmed Baerveldt Comparison Study:
      • Design: "multicenter, prospective, randomized, controlled clinical trial"
      • Sample Size: Not explicitly stated in this summary.
      • Provenance: "multicenter" implies multiple locations, but specific countries are not mentioned.
    • Comparison of Polypropylene and Silicone Ahmed Glaucoma Valves (Ishida K. et al.):
      • Design: "prospective, multicenter, comparative study"
      • Sample Size: Not explicitly stated in this summary.
      • Provenance: "multicenter," but specific countries are not mentioned.
    • Comparison of silicone and polypropylene Ahmed glaucoma valves: two-year follow-up (Mackenzie PJ et al.):
      • Design: "single surgeon, retrospective, consecutive case series."
      • Sample Size: Not explicitly stated in this summary.
      • Provenance: Not specified, but "single surgeon" suggests a single institution.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in the document. The studies listed are clinical trials, where patient outcomes (IOP, complications) are typically objectively measured or clinically observed, rather than adjudicated by experts in the context of establishing a 'ground truth' for an AI device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided. As these are clinical trials of a medical device (glaucoma valve) rather than, for example, an image analysis AI, the concept of "adjudication method for the test set" in the context of expert consensus on data interpretation doesn't directly apply here in the way it would for an AI performance study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a MRMC comparative effectiveness study was not done. The device in question is an implanted glaucoma valve, not an AI-assisted diagnostic tool for human readers. Therefore, there is no discussion of human readers or AI assistance in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, this is not an AI algorithm. This is a physical medical device (an aqueous shunt). The "testing" described is non-clinical (physical, chemical, sterilization, biocompatibility) and clinical (patient outcomes after surgical implantation).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the clinical studies, the "ground truth" (or outcome measures) were:

    • Intraocular Pressure (IOP): This is a direct, measurable physiological parameter, which can be considered objective outcomes data.
    • Complication Rates: These are clinically observed events, also considered outcomes data.

    8. The sample size for the training set:

    Not applicable. This document describes a physical medical device, not a machine learning model. Therefore, there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable. There is no "training set" or "ground truth" established for it, as this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060644

    Validate with FDA (Live)

    Date Cleared
    2006-09-18

    (192 days)

    Product Code
    Regulation Number
    886.3920
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ahmed™ Glaucoma Valve is indicated for the management of refractory qlaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medication, congenital or infantile dlaucoma, and refractory glaucomas resulting from aphakia or uveitis.

    Device Description

    The Ahmed™ Glaucoma Valve (AGV) S3 is a valved aqueous drainage implant designed to regulate intraocular pressure in eyes suffering from intractable glaucoma. The Ahmed™ device is comprised of a silicone drainage tube that is connected to a valve membrane. The valve membrane is sandwiched between a top plate made of polyprolylene and a complementary bottom plate made of polypropylene. The bottom plate is extended to provide for aqueous distribution and drainage. The valve body conforms to the shape of the globe at its equator and protects the valve from blockage by fibrous tissue growth.

    The Ahmed Valve Model M4 valve system is exactly the same as the Models S2 and S3. The M4 valve system is enclosed within a porous material (MEDPOR®) made of polyethylene (MEDPOR® is well established as a safe material and has been approved by the FDA for ocular use).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Ahmed Glaucoma Valve Model M4. The primary goal of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, rather than to prove specific performance against detailed acceptance criteria using a clinical study with quantifiable metrics like sensitivity or specificity for an AI device.

    Therefore, the following information is extracted based on the context of a 510(k) for a medical device that is not an AI algorithm. The device, an Ahmed Glaucoma Valve Model M4, is a physical implant, and the "study" referred to is biocompatibility and comparative physical/biological effects, not an AI performance study.

    Here's the breakdown based on the provided text, adapted to reflect that this is not an AI device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission for a non-AI medical device, the "acceptance criteria" are related to demonstrating substantial equivalence to a predicate device, particularly concerning safety and effectiveness through biocompatibility and functional comparison. Specific performance metrics like sensitivity/specificity for AI are not applicable here.

    Acceptance Criteria (Implied for 510(k))Reported Device Performance
    Material Biocompatibility: The new material (MEDPOR®) in Model M4 should be biocompatible.The M4 valve system is enclosed within MEDPOR® (polyethylene), which is "well established as a safe material and has been approved by the FDA for ocular use." An animal study specifically compared the AGV™-S3 and AGV™-M4 and was "successful in demonstrating the biocompatibility of the M4."
    Functional Equivalence: The M4 valve system should function similarly to predicate devices (S2 and S3) in terms of aqueous outflow regulation."The Ahmed Valve Model M4 valve system is exactly the same as the Models S2 and S3" in its core valving mechanism. The animal study showed that "resistance to outflow from the M4 to the surrounding tissue was significantly reduced when compared to the S3," suggesting an improvement in outflow characteristics related to the new porous material. The overall design for regulating intraocular pressure is considered equivalent.
    Reduced Fibrosis/Improved Vascularization (with new material): The porous material in M4 should ideally improve tissue integration compared to non-porous designs.The animal study showed that "the use of porous polyethylene in the M4 resulted in decreased collagen synthesis and increased vascularization in the tissue surrounding the implant when compared to the S3."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Animal Study): Not explicitly stated. The text only mentions "An animal study comparing the AGV™-S3 and AGV™-M4 was conducted."
    • Data Provenance: The study was an "animal study" conducted at "Duke University." It lasted "for over six months." It is a prospective study in an animal model.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This is not applicable to a non-AI medical device animal study. Ground truth, in the context of an animal study for a physical implant, would be derived from objective measurements (e.g., histology for collagen synthesis and vascularization, pressure measurements for outflow resistance, direct observation of biocompatibility).

    4. Adjudication Method for the Test Set

    Not applicable to a non-AI medical device animal study. Adjudication methods like 2+1 or 3+1 are typically used for human expert review in diagnostic studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (often AI-assisted) where human readers interpret cases. The study mentioned is an animal study evaluating the physical and biological characteristics of a glaucoma implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is a physical medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    For the animal study comparing the AGV™-M4 and AGV™-S3, the ground truth would be based on:

    • Histopathology/Biological Markers: For "decreased collagen synthesis and increased vascularization."
    • Physiological Measurements: For "resistance to outflow."
    • Clinical Observation/Pathology: For "biocompatibility."

    8. The Sample Size for the Training Set

    Not applicable. This is a physical implant, not an AI algorithm. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a physical implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991072

    Validate with FDA (Live)

    Date Cleared
    1999-07-13

    (104 days)

    Product Code
    Regulation Number
    886.3920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ahmed™ Glaucoma Valve is indicated for the management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medication, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.

    Device Description

    The Ahmed™ Glaucoma Valve Bi-plate is a modification of an already approved device called the Ahmed™ Glaucoma Valve (AGV™) Ref. 510 (k) 925636 dated November 12, 1993. The Ahmed™ Glaucoma Valve (AGV™) is an ophthalmic implant for use in intractable Glaucoma. The device features a specially engineered, one way silicone membrane valve system designed to prevent collapse of the anterior chamber (AC) due to hypotony (abnormally low intraocular pressure) and to reduce excessive intraocular pressure by venting aqueous out of the anterior chamber through this control one way valve. The AGV™ implant consists of a silicone drainage tube, a polypropylene valve body to house the valve membrane, and to protect it from occlusion due to fibrosis. All materials used in the manufacturing of this device are of medical grade quality. No metallic or toxic substances are used in the manufacturing of this device. The AGV™ is terminally sterilized by gamma radiation. The AGV™ has a surface area of 184mm².

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Ahmed Glaucoma Valve Bi-plate:

    Disclaimer: The provided text is a 510(k) summary (K991072) for a medical device seeking substantial equivalence, not a detailed clinical study report with explicit acceptance criteria. Therefore, the "acceptance criteria" discussed below are inferred from the safety and effectiveness claims and comparisons to the predicate device. The "reported device performance" is the performance of the predicate device (Ahmed Glaucoma Valve, single plate) and the general claims made about the new device (Ahmed Glaucoma Valve Bi-plate) based on literature about double-plate implants.


    Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission for substantial equivalence, the primary "acceptance criteria" are demonstrating that the Ahmed Glaucoma Valve Bi-plate is as safe and effective as a legally marketed predicate device (Molteno Double Plate Implant, K875099, and the original Ahmed Glaucoma Valve, K925636). The performance metrics revolve around:

    • Intraocular Pressure (IOP) Reduction: The device should effectively lower IOP in refractory glaucoma patients.
    • Maintenance of IOP within normal range: After surgery, IOP should be maintained at a healthy level.
    • Prevention of Hypotony: A key feature of the Ahmed Glaucoma Valve (and claimed for the Bi-plate) is its ability to prevent abnormally low IOP.
    • Safety Profile: The device should have an acceptable complication rate.
    • Biocompatibility: The materials used should be non-toxic and biocompatible.

    Since the submission specifically states, "AGV Bi-plate and Molteno Double Plate are substantially equivalent," and refers to the existing Ahmed Glaucoma Valve, the performance of these devices serves as the benchmark.

    Acceptance Criteria (Inferred from goals of glaucoma shunts)Reported Device Performance (From clinical papers on AGV and Molteno)
    IOP ReductionAhmed Glaucoma Valve (single plate, K925636 reference): Mean IOP reduced from 32.7 ± 0.8 mmHg before surgery to 15.9 ± 0.6 mmHg at most recent follow-up (mean 13.4 ± 0.7 months). (Huang et al., 1999) Molteno Double Plate (literature review): Mean pre-operative pressure of 28.3mmHg reduced to 12.75mmHg (Molteno, 1981).
    Reduction in Anti-glaucoma MedicationsAhmed Glaucoma Valve (single plate): Medications decreased from 2.7 ± 0.1 before surgery to 1.1 ± 0.1 after surgery. (Huang et al., 1999)
    Sustained Success RateAhmed Glaucoma Valve (single plate): Cumulative probability of success was 87% at 1 year and 75% at 2 years. (Huang et al., 1999)
    Visual Acuity Maintenance/ImprovementAhmed Glaucoma Valve (single plate): Visual acuity improved or within one Snellen line in 131 eyes (82%). (Huang et al., 1999)
    Prevention/Reduction of HypotonyAhmed Glaucoma Valve (single plate): Transient post-operative hypotony in 13 eyes (8%). Molteno Double Plate (literature review): Intraocular pressure was maintained at more than 5mm Hg in 11 out of 12 cases. Molteno also indicates that the use of a double plate helped in reducing post-operative hypotony.
    Acceptable Complication ProfileAhmed Glaucoma Valve (single plate): Most common complication: obstruction of the tube (11%). Molteno Double Plate (literature review): Implied acceptable, with focus on hypotony reduction.
    Biocompatibility & Physical IntegrityAhmed Glaucoma Valve (single plate, K925636 reference): Sensitive in vitro biocompatibility testing demonstrated non-toxicity, non-irritation, and biocompatibility. In vitro physical testing demonstrated valve efficacy and physical integrity.

    Study Details Proving Acceptance Criteria

    The document refers to several studies and prior 510(k) submissions rather than presenting a single, new clinical trial for the Ahmed Glaucoma Valve Bi-plate. The main approach is through demonstrating substantial equivalence to existing devices, supported by:

    1. Prior Approval of the Single-Plate Ahmed Glaucoma Valve (K925636): This forms the foundation of the device's safety and effectiveness.
    2. Literature Review of Double-Plate Implants (especially Molteno Double Plate): This supports the safety and effectiveness of increasing surface area.
    3. Comparisons of Technological Characteristics: Demonstrating similarity in materials, design, and function between the AGV Bi-plate and the predicate Molteno Double Plate.

    Here's a breakdown based on the provided text:

    1. Sample sizes used for the test set and the data provenance:

      • Ahmed Glaucoma Valve (single plate) - Huang et al. (1999):
        • Sample Size: 159 eyes (144 patients).
        • Data Provenance: Multi-center, retrospective clinical follow-up. Country of origin not specified, but the journal "Am J Ophthal." is a US-based publication.
      • Ahmed Glaucoma Valve (single plate) - K925636 Clinical Testing (50 patients):
        • Sample Size: 50 patients.
        • Data Provenance: Clinical study performed at five centers for six months. Retrospective or prospective not explicitly stated for this particular summary, but prior 510(k)s often involve prospective data collection. This data would be from the US, as it's a US FDA submission.
      • Molteno Double Plate - Molteno (1981) paper:
        • Sample Size: 12 patients for the double plate group.
        • Data Provenance: Clinical study. Country of origin not specified, but Molteno is associated with New Zealand.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided directly within the 510(k) summary. For the studies cited (Huang et al., Molteno 1981, and the initial 50-patient study), the "ground truth" (e.g., IOP measurements, complication assessments) would have been established by the clinicians participating in those studies (presumably ophthalmologists or glaucoma specialists), but their specific number or qualifications are not detailed here. For 510(k) submissions of this nature, the "ground truth" often comes from standard clinical practice and validated measurement techniques rather than an independent expert panel for the submission itself.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not provided in the 510(k) summary for any of the cited clinical studies. Clinical studies typically have internal review processes, but formal adjudication methods like "2+1" or "3+1" are not mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a physical implant, not an AI or imaging diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device. The "standalone performance" would relate to its physical function (e.g., fluid flow, pressure regulation), which was assessed via in vitro and in vivo animal studies.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Clinical Outcomes Data: This is the primary type of ground truth used. It includes:
        • Intraocular Pressure (IOP) measurements.
        • Number of anti-glaucoma medications.
        • Visual acuity.
        • Incidence and type of complications (e.g., hypotony, tube obstruction, iritis).
        • Success rates (based on predefined clinical criteria like target IOP without severe complications).
      • In vitro and In vivo (animal) studies: For biocompatibility and physical function.
    7. The sample size for the training set:

      • This term "training set" is typically used in the context of machine learning or AI models. It is not applicable here as the device is a physical implant. The development of the original Ahmed Glaucoma Valve would have involved iterative design and testing, but not in the framework of training sets.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the AI/ML sense. The "ground truth" for the device's performance claims comes from clinical data and scientific literature on similar devices, as described above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980657

    Validate with FDA (Live)

    Date Cleared
    1998-04-20

    (74 days)

    Product Code
    Regulation Number
    886.3920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ahmed Glaucoma Valve is indicated for the management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medication, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.

    Device Description

    The Ahmed™ Glaucoma Valve (AGV™) is a ophthalmic implant for use in intractable glaucoma. The device features a specially engineered, one-way silicone membrane valve system designed to prevent collapse of the anterior chamber (AC) due to hypotony, abnormally low intraocular pressure, and a build-up of excessive intraocular pressure, by venting aqueous humor from the eye. The AGV™ implant consists of a silicone drainage tube and polypropylene valve body to house the valve membrane and protect it from occlusion by fibrosis. All materials used in the manufacture of the device are of medical grade quality. No toxic substances are used in the manufacturing process. The AGV is terminally sterilized by gamma ray radiation.

    AI/ML Overview

    The Acceptance Criteria and Device Performance for the Ahmed™ Glaucoma Valve Model S3 are based on a clinical study of its predicate device, the Ahmed™ Glaucoma Valve Model S2. The S3 model was deemed substantially equivalent to the S2 model.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Success Definition)Reported Device Performance (at 6 months post-operatively)
    "Success": Post-operative Intraocular Pressure (IOP) of ≤ 21 mmHg and ≥ 5 mmHg, while maintaining or improving visual acuity, without the use of glaucoma medications.Visual acuity was preserved or improved, and IOPs were reduced to 21 mmHg or less and greater than 5 mmHg in 43 out of 50 subjects (86%). The mean post-operative IOP at six months was 15.33 ± 4.5 mmHg. The document implies without medication, as a "qualified success" would delineate medication use.
    "Qualified Success": Post-operative IOP of ≤ 21 mmHg and ≥ 5 mmHg, while maintaining or improving visual acuity, with the use of glaucoma medications.Not explicitly reported as a separate metric; however, the "Success" rate (86%) indicates a high proportion of patients met the primary endpoint, presumably without needing additional medication in many cases. Any patients needing medication while still meeting the IOP and visual acuity criteria would fall into this category.
    "Failure": Post-operative IOP of > 21 mmHg.7 out of 50 subjects (14%) were considered "failures" according to this definition.
    No valve failures observed in non-clinical experiments (in vitro and in vivo animal studies)."Aside from destructive testing, no valve failures were observed in any experiments."
    Biocompatibility and non-toxicity of materials."Sensitive in vitro biocompatibility testing, performed by several methods, demonstrated that the valve and its components are non-toxic, non-irritating, and biocompatible."
    Efficacy in controlling IOP in animal models."An in vivo, long-term and short-term animal studies using rabbits in which the AGV™ was implanted demonstrated it efficacy with regard to control of IOP, with the fellow eye used as a control, tolerance of the device, and offered further substantiation in this animal model of biocompatibility of the valve and its components."
    Mechanism to prevent anterior chamber collapse due to hypotony and vent aqueous humor from the eye."The device features a specially engineered, one-way silicone membrane valve system designed to prevent collapse of the anterior chamber (AC) due to hypotony, abnormally low intraocular pressure, and a build-up of excessive intraocular pressure, by venting aqueous humor from the eye." "No cases of collapsed chambers were reported."
    Immediate control of intraocular pressure upon implantation."In the immediate post-operative period (4-28 hours), the mean preoperative IOP decreased to 9.66 ± 7.06 mmHg."

    2. Sample Size for the Test Set and Data Provenance:

    • Sample Size: 50 subjects.
    • Data Provenance: Prospective clinical study conducted at six (6) institutions. The country of origin is not explicitly stated, but the submission is to the US FDA, implying it was likely conducted in the US or under US regulatory standards.

    3. Number of Experts and Qualifications for Ground Truth:

    • The document does not explicitly state the number of experts used to establish the ground truth for the test set or their specific qualifications. For medical device studies of this nature, clinical investigators (ophthalmologists, for this device) at the participating institutions would be responsible for patient assessments and data collection, which inherently forms the basis of the "ground truth." The study involved monitoring by medical professionals.

    4. Adjudication Method:

    • The document does not explicitly detail an adjudication method (e.g., 2+1, 3+1) for the test set. Clinical trials typically involve multiple investigators and a study monitor following a pre-defined protocol. The results presented are the aggregated outcomes from the participating institutions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a clinical trial evaluating the performance of a medical device (Ahmed™ Glaucoma Valve) in human subjects, not an assessment of human readers' performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study:

    • No, a standalone (algorithm only) performance study was not done. The device is a physical implant, not an AI algorithm. The studies focused on the performance of the physical device after implantation.

    7. Type of Ground Truth Used:

    • The ground truth was established through clinical outcomes data collected directly from patients after device implantation. This included measurements of intraocular pressure (IOP), assessment of visual acuity, and monitoring for post-operative complications, evaluated by medical professionals.

    8. Sample Size for the Training Set:

    • This information is not applicable. The device is a physical implant, not an AI algorithm that requires a "training set" in the computational sense. The "training" or development of the device would involve engineering, material science, and pre-clinical testing, not a data-based training set.

    9. How the Ground Truth for the Training Set was Established:

    • This information is not applicable for the same reason as above. The "ground truth" for the development of the physical device would be derived from physiological principles, material properties, and pre-clinical models (in vitro and animal studies) that guide its design and demonstrate its intended function.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1