Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223878

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-17

    (80 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aevumed RAPID™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

    Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Hip: Capsular repair, Acetabular Labral Repair

    Device Description

    The Aevumed RAPIDTM Suture Anchor with HS FiberTM suture is a suture anchor manufactured from polyetheretherketone (PEEK) material preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed RAPIDTM Suture Anchors are available in diameter sizes: 2.4mm and 3.0mm. They are offered sterile and are for single use only

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (Aevumed RAPID™ Suture Anchors). It details the device's characteristics, indications for use, and a comparison to predicate devices, but does not describe a study involving AI or software, nor does it provide acceptance criteria and performance data in the context of an AI/software device evaluation.

    Therefore, I cannot extract the information required in your request (acceptance criteria, study details, sample sizes, expert qualifications, etc.) because the document describes a hardware medical device (suture anchors) and its mechanical testing, not a diagnostic or AI-driven system.

    The "Non-clinical Tests" section mentions mechanical testing for pullout strength and states:

    • Study: "Both the Aevumed RAPID™ Suture Anchor and predicate SutureTak™ Anchor (K140855) were mechanically tested for maximum pullout strength."
    • Result: "The Aevumed RAPID™ Suture Anchor demonstrated significantly higher pullout strength in comparison to the predicate SutureTak™ Anchor (K140855)."

    This is the only "performance" information related to a "study" in the document, and it pertains to the physical device's mechanical properties, not an algorithmic performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1