Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K211572

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-13

    (206 days)

    Product Code
    Regulation Number
    882.4175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The slim clip applier is used to open, close and apply permanent/temporary Aesculap YASARGIL titanium aneurysm clips.

    Device Description

    The Aesculap Slim Clip Appliers are manufactured from stainless steel and are available in various lengths, jaw angulations, with or without latches. This submission intends to add a coating to the previously cleared clip appliers. Each clip applier is individually laser marked with the type of clip designated for use. In addition, at least one identification plug is located within one of the handles to aid in identifying the aneurysm clip designation.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Aesculap Slim Clip Applier, formatted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestAcceptance Criteria (Implied)Reported Device PerformanceConclusion
    Clip Opening VerificationMinimum opening width of aneurysm clips is met.PassThe Aesculap Slim Clip Appliers meet the test acceptance criteria.
    Biocompatibility (Cytotoxicity)Reactivity grade of 0.Reactivity grade of 0.Non-cytotoxic
    Biocompatibility (Sensitization)No evidence of delayed dermal contact sensitization.No evidence of sensitization.Non-sensitive
    Biocompatibility (Pyrogenicity)Temperature rise below 0.5 °C.Below 0.5 °C.Non-pyrogenic
    Biocompatibility (Irritation)Difference between test article extract overall mean score and corresponding control extract overall mean score was 0.0.Difference of 0.0.Normal
    Biocompatibility (Systemic Toxicity)No mortality or evidence of systemic toxicity.No mortality or systemic toxicity.Non-Toxic
    Biocompatibility (Hemolysis)Hemolytic index of 0.0%.Hemolytic index of 0.0%.Non-Hemolytic

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the mechanical "Verification test for clip opening." It merely states "Pass" for the test.

    For biocompatibility testing, the sample sizes are implied by the test descriptions:

    • Cytotoxicity: "Test articles were extracted for 24 hours per ISO 10993-5." (Specific number of test articles not given).
    • Sensitization: "in the guinea pig." (Specific number of guinea pigs not given).
    • Pyrogenicity: "following intravenous injection in rabbits." (Specific number of rabbits not given).
    • Irritation: "All animals appeared normal throughout the study." (Specific number of animals not given).
    • Systemic Toxicity: "extracts injected into mice." (Specific number of mice not given).
    • Hemolysis: "The test article extract was tested in vitro." (Specific number of test articles or replicates not given).

    The data provenance is not explicitly stated regarding country of origin or whether it's retrospective/prospective. However, given that these are bench and biocompatibility tests conducted for a 510(k) submission, they are typically conducted prospectively in a controlled laboratory setting by the manufacturer or a contract research organization.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of device and testing. The ground truth for mechanical performance is based on engineering specifications and physical measurements. The ground truth for biocompatibility is based on established ISO standards and observed biological reactions in animal models or in vitro tests, rather than expert consensus on medical images or clinical outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. This document describes bench testing and biocompatibility testing against defined physical and biological parameters, not subjective assessments requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices where human readers interpret results, and the AI's impact on their performance is evaluated. The Aesculap Slim Clip Applier is a surgical instrument, and its evaluation focuses on mechanical performance and biocompatibility.

    6. Standalone Performance (i.e., algorithm only without human-in-the-loop performance)

    Yes, the studies described are essentially "standalone" in the sense that they evaluate the device's intrinsic mechanical properties and biocompatibility without a human operator's performance being the primary subject of the test (beyond the execution of the test itself). The "Verification test for clip opening" directly assesses the applier's mechanical function, and the biocompatibility tests evaluate the material's interaction with biological systems. There isn't an "algorithm" per se, but the device's performance is tested independently.

    7. Type of Ground Truth Used

    • For Clip Opening Verification: Engineering specifications and physical measurements of the clip's opening width.
    • For Biocompatibility Tests: Established biological endpoints and standards (e.g., ISO 10993 series) for cytotoxicity, sensitization, pyrogenicity, irritation, systemic toxicity, and hemolysis.

    8. Sample Size for the Training Set

    Not applicable. The Aesculap Slim Clip Applier is a mechanical surgical instrument, not an AI/machine learning device that requires a training set of data. The "training" in manufacturing would relate to process validation and quality control, not a data-driven model.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180914

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-17

    (8 days)

    Product Code
    Regulation Number
    882.4175
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Slip Clip Applier forceps are used for opening, closing, and applying Aesculap Y ASARGIL aneurysm clips.

    Device Description

    The Aesculap Slim Clip Appliers are manufactured from stainless steel (body & jaw). The previously cleared appliers are available in lengths ranging from 50 mm to 110 mm in 10 mm increments. In addition to the straight jaw configuration (0%), the Clip Appliers are available with jaws angled up & down (5° to 15°), left & right (5° to 15°), in increments of 5° in all lengths. This submission intends to add additional lengths of 120 mm and 130 mm in these configurations. This submission intends to add additional jaw angulations of 20° to 45° up and down and 20° to 30° left and right in 5° increments and lengths ranging from 50-130 mm in 10 mm increments. In addition to these configurations, clip appliers with latches will be available in lengths ranging from 70 mm to 110 mm increments with jaw angulations ranging from straight, 5° to 45° up and down and 5° to 30° left and right in 5° increments. Each Clip Applier is individually laser marked with the type of clip designated for use. In addition, at least one identification plug is located within one of the handles to aid in identifying the aneurysm clip designation.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Aesculap Slim Clip Applier, a device used for opening, closing, and applying Aesculap YASARGIL aneurysm clips. The submission aims to add new lengths and jaw angulations, as well as an optional latch feature, to existing cleared appliers.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestAcceptance Criteria (Methodology)Reported Device Performance (Results)Conclusion
    Benchmark Functional Test(1) Meet required interface and opening width of the aneurysm clip. (2) Successful release of designated aneurysm clip in the craniotomy. (3) Latch functionality.PassThe devices interfaced successfully with the aneurysm clips while providing the correct opening width. In addition, the devices successfully released the aneurysm clips within the craniotomy.
    Predicate/Subject Device Comparison TestComparison between predicate and subject device ensuring: (1) Required interface and opening width of designated aneurysm clip. (2) Successful performance of the clip applier with the aneurysm clip. (3) Latch functionality.PassThe performance testing demonstrated that the subject devices are substantially equivalent to the predicate devices.
    Usability Test(1) Ensuring identification of designated clip type for the specific clip applier. (2) Legibility of additional Clip Applier identification features, i.e., company name, article number, and UDI information. (3) Ensuring visualization of surgical site. (4) Ability to grasp, hold, and apply YASARGIL Aneurysm Clip. (5) Latch usability test.PassThe usability test demonstrated that the subject devices were properly identified and met the usability requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test (e.g., number of appliers, number of clips, number of simulated craniotomies). It generally refers to "all required testing" being completed. The tests are bench tests, meaning they were conducted in a laboratory setting, not with human subjects. Thus, data provenance in terms of country of origin or retrospective/prospective medical data is not applicable here as the test data is generated in a controlled environment.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not provided in the document. The tests performed are primarily functional and mechanical, suggesting that the "ground truth" is defined by engineering specifications and performance standards rather than expert clinical consensus.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the nature of the bench testing (Pass/Fail results against predefined criteria), formal adjudication by multiple parties for each test outcome is unlikely to have been a primary method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The device is a surgical instrument (aneurysm clip applier), and the studies described are bench tests to confirm its mechanical and functional equivalence to predicate devices, not studies involving human readers interpreting imaging data or clinical outcomes. Therefore, an effect size of human readers improving with or without AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the Aesculap Slim Clip Applier. This device is a manual surgical instrument, not an algorithm or AI-powered system.

    7. The Type of Ground Truth Used

    The ground truth for the acceptance criteria and testing appears to be based on:

    • Engineering Specifications: The required interface and opening width of the aneurysm clip, and the ability to release the clip successfully.
    • Functional Performance Standards: The successful performance of the clip applier with the aneurysm clip, and latch functionality.
    • Usability Requirements: Ensuring proper identification features, legibility, visualization of the surgical site, and the ability to grasp, hold, and apply the clip.

    These are primarily objective, measurable criteria related to the device's design and intended function.

    8. The Sample Size for the Training Set

    This is not applicable. The Aesculap Slim Clip Applier is a mechanical medical device, not a machine learning or AI algorithm that requires training data.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for a mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173271

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-10

    (29 days)

    Product Code
    Regulation Number
    882.4175
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Slim Clip Applier forceps are used for opening, closing, and applying Aesculap YASARGIL aneurysm clips.

    Device Description

    The Aesculap Slim Clip Applier forceps are manufactured from stainless steel (body & jaw). The appliers are available in lengths ranging from 50 mm to 110 mm. In addition to the straight jaw configuration, the Clip Appliers are available with jaws angled up & down, left & right.

    Each Clip Applier is individually laser marked with the type of clip designated for use. In addition, identification plugs to easily identify each clip designation are placed in the handles.

    AI/ML Overview

    The Aesculap Slim Clip Applier is a medical device for opening, closing, and applying Aesculap YASARGIL aneurysm clips.

    Here's an analysis of its acceptance criteria and the study proving it meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from "Performance Test completed" column in source)Reported Device Performance (from "Results" and "Conclusions" columns in source)
    Benchmark Functional Test:
    1. Meet required interface and opening width of the aneurysm clip.Pass. The devices interfaced successfully with the aneurysm clips while providing the correct opening width.
    2. Successful release of designated aneurysm clip in the craniotomy.Pass. In addition, the devices successfully released the aneurysm clips within the craniotomy.
    Predicate/Subject Device Comparison Test:
    Comparison between predicate and subject device ensuring:Pass. The performance testing demonstrated that the subject devices are substantially equivalent to the predicate devices.
    1. Required interface and opening width of designated aneurysm clip.(Implied by "Pass" and "substantially equivalent" conclusion).
    2. Successful release of aneurysm clip in the craniotomy.(Implied by "Pass" and "substantially equivalent" conclusion).
    Usability Test:
    1. Ensuring identification of designated clip type for the specific clip applier.Pass. The usability test demonstrated that the subject devices were properly identified and met the usability requirements.
    2. Legibility of additional Clip Applier identification features, i.e., company name, article number, and UDI information.Pass. (Implied by "Pass" and "met the usability requirements" conclusion).
    3. Ensuring visualization of surgical site.Pass. (Implied by "Pass" and "met the usability requirements" conclusion).
    4. Ability to grasp, hold, and apply YASARGIL Aneurysm Clip.Pass. (Implied by "Pass" and "met the usability requirements" conclusion).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the test set. It mentions "Bench testing results" and describes the types of tests performed. The data provenance is not explicitly mentioned (e.g., country of origin). The testing described is prospective bench testing focused on the mechanical and functional aspects of the device, rather than patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish ground truth for the test set. The tests described are objective performance and usability assessments.

    4. Adjudication Method for the Test Set

    No adjudication method is mentioned, as the tests appear to be objective performance evaluations rather than subjective assessments requiring expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not conducted as this device is a mechanical instrument (aneurysm clip applier) rather than an AI-powered diagnostic tool or image interpretation system. Therefore, there is no effect size reported for human readers with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A "standalone" study in the context of an algorithm is not applicable here, as this is a physical medical device. The "standalone" performance would be analogous to the "Benchmark Functional Test" and "Predicate/Subject Device Comparison Test" which assess the device's inherent functional capabilities without direct human interaction beyond its intended use.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on engineering specifications and functional requirements for aneurysm clip appliers, ensuring they correctly interface, open, close, and release aneurysm clips as intended, and meet usability standards.

    8. The Sample Size for the Training Set

    A "training set" is not applicable as this is a mechanical device, not an AI or machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML model, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1