510(k) Data Aggregation

    K Number
    DEN150055

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-21

    (380 days)

    Product Code
    Regulation Number
    878.3510
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AeroForm Tissue Expander System is used for soft tissue expansion in breast reconstruction following mastectomy, for the treatment of underdeveloped breasts, and for the treatment of soft tissue deformities in the breast.

    The AeroForm Expander is intended for temporary subcutaneous or submuscular implantation and is not intended for use beyond six months.

    Device Description

    The AeroForm® Tissue Expander System is comprised of the AeroForm Tissue Expander, the Dosage Controller, and the Physician Master Key. The AeroForm Tissue Expander is a sterile implant with an outer textured silicone shell and a non-distensible inner bag which are anatomically shaped to allow for directional expansion in the lower, anterior pole. The AeroForm Tissue Expander contains a reservoir of compressed carbon dioxide (CO2), which is released within the AeroForm Tissue Expander by using the Dosage Controller. The outer shell has suture tabs to allow fixation of the device to the surrounding tissues to prevent rotation. A receiving antenna and electronics within the AeroForm Tissue Expander enable communication with the Dosage Controller. The AeroForm Tissue Expander has no intrinsic electrical power, batteries, or software, and can only be activated by the Dosage Controller. There are 3 different sizes of the expander, including small (400cc), medium (600cc), and large (850cc). The Dosage Controller is a small, hand held battery-powered, non-sterile device. It activates the AeroForm Tissue Expander to release the programmed amount of CO2 gas (10cc). The Dosage Controller is configured to provide coded instructions to its bonded AeroForm Tissue Expander. It has a single push button for ease of use and a bank of indicator lights (LEDs) and tones. The Physician Master Key is used only by the physician in the operating room or in the physician office. With the Physician Master Kev inserted into the Dosage Controller receptacle, the Dosage Controller can activate the Expander with no daily limit on the expansion volume.

    AI/ML Overview

    The AeroForm® Tissue Expander System underwent extensive testing, encompassing biocompatibility, shelf-life, electromagnetic compatibility, software verification, and performance bench testing. A pivotal clinical study, XPAND trial (CPT-0003), and a Continued Access Study (CAS) provided critical insights into its performance and safety in a real-world setting.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for several non-clinical tests and the corresponding reported device performance are summarized below:

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility
    Cytotoxicity (ISO MEM Elution)The biological response of the test samples extract must be grade 2 (mild or less)Pass
    Sensitization (Maximization)The test samples must show a grade of 1 or less when compared to the controlPass
    Irritation (ISO Intracutaneous)The difference between the mean score of the test samples and control groups, in terms of erythema and edema, must be less than 1.0Pass
    Acute Systemic Toxicity (USP Systemic Injection)The test mice must not show a significantly greater reaction than the control micePass
    Acute Systemic Toxicity (USP Mediated Pyrogen)The test rabbits must not show a temperature difference from the control of greater than or equal to 0.5°CPass
    Sub-chronic toxicityEvidence of irritation in the test animals will be scored and compared to the control animalsPass
    Chronic ToxicityEvidence of irritation in the test animals will be scored and compared to the control animalsPass
    Genotoxicity (Bacterial Reverse Mutation)The mean number of test revertants must be less than 2x the mean number of negative control revertantsPass
    Genotoxicity (Mouse Micronucleus)The average % MN-RET for the test group must be less than 1.0%Pass
    Genotoxicity (Mouse Lymphoma)The mutant frequency of the test sample must be 2X or less than that of the controlPass
    Implantation (2 week and 9 week)The average macroscopic scores for test sample sites compared to control sites are correlated to a grade. The microscopic findings showing cellular changes will be graded according to severity (0-4) and any resulting irritant response indicated by a difference between test and control values will be graded as nonirritant, slight, moderate or severe.Pass
    Particulate TestingParticle size [Redacted] (b) (4) particles per mLParticle size [Redacted] (b) (4) particles per mL(per EN45502-1)Pass
    LAL (Limulous Amoebocyte Lysate) TestingDetected EU/device level is below (b) (4) EU/devicePass
    CarcinogenicityAnalysis of the biological test results, chemical characterization data and literature reviewPass by analysis and risk assessment
    Performance - Bench Testing
    Cycle Rub TestThere are no major leaks or tears in the film or the seals, after 25,000 cycles.The Expander successfully met the no leak and tear requirements after 25,000 cycles.
    Impact TestThere are no ruptures after the impact test.No ruptures were reported after the impact test.
    Impact Force TestThe device shall not fail or leak when experiencing forces less than 1700 lbs.All acceptance criteria were met. The reservoir and valve assembly withstood an impact load of ~4X the human tolerance levels of 1700lbs without leakage. Impact loads to ~7X the human tolerance level resulted in slow leaks.
    Leak TestThe CO2 concentration shall not increase more than 100 PPM.The Expander successfully met the leak test requirements following functional testing.
    Permeability TestSmall & Medium Size: Each test article shall have a daily average permeation rate of 0-4.1 mL CO2 per day. Large Size: Each test article shall have a daily average permeation rate of 0-6.0 mL CO2 per day.Small and medium test articles met the daily average permeation rate of < 4.1 mL CO2 per day. The large sample met the requirement for < 6.0 mL CO2 per day.
    Shell elongationMaximum set shall be less than 10%.All test articles passed the acceptance criteria.
    Shell breaking forceUltimate breaking force in tension shall be no less than 2.5lbs.All test articles passed the acceptance criteria.
    Phone dial / shell joint testAdhered or fused joints or seams that are critical to the integrity of the device envelope shall not fail when the shell adjacent to the joint is stressed at 200% elongation for 10 seconds. Zero failures results in a minimum 95/90 Confidence/Reliability interval.All samples passed the phone dial / shell joint seam test with no failures.
    Suture tab joint testThe joint shall not fail after being exposed to 200% elongation. Zero failures results in a minimum 95/90 Confidence/Reliability interval.All samples remained fully intact, and achieved at least 200% elongation without any failures.
    Suture tab tear strengthThe Suture Tab fixation hole shall withstand a tensile tear-out force of at least 2.5lb. Zero failures results in a minimum 95/90 Confidence/Reliability interval.All suture tab fixation holes withstood a tensile tear-out force of at least 2.5lbs.

    Clinical Performance:

    Clinical MetricAcceptance CriteriaReported Device Performance (AeroForm)
    Primary Endpoint: Treatment Success Rate (per breast, excluding non-device related failures)Expected ≥ 95% with non-inferiority margin > -10% compared to saline group.96.1% (149/155 breasts). Difference to Saline: -2.7% (lower CL -7.3%), meeting non-inferiority.
    Median Days to Complete ExpansionNot explicitly stated as acceptance criteria, but a key secondary effectiveness endpoint.21.0 days (95% CI 15.0, 24.0)
    Median Days to Complete ReconstructionNot explicitly stated as acceptance criteria, but a key secondary effectiveness endpoint.108.5 days (95% CI 99.0, 117.0)
    Subject Satisfaction (Overall)Not explicitly stated as acceptance criteria, but a key secondary effectiveness endpoint.78% (AeroForm overall), 84% (V2.5 only)
    Physician Satisfaction (Overall)Not explicitly stated as acceptance criteria, but a key secondary effectiveness endpoint.68% (AeroForm overall), 83% (V2.5 only)
    Adverse Event Rates (Any)Not explicitly stated as acceptance criteria, but a key safety endpoint for comparison.63.3% (AeroForm arm) vs 63.5% (Saline arm)
    Device-Related AE RatesNot explicitly stated as acceptance criteria, but a key safety endpoint for comparison.21.2% (AeroForm arm) vs 19.2% (Saline arm)
    Device Malfunctions (V2.5 only)Reduction in malfunction rates, particularly under expansion/deflation, after design modification.Under expansion/deflation reduced from 46.7% (V2.0) to 15.4% (V2.5).
    XPAND II CAS Treatment Success Rate100% (n=18 breasts evaluated)
    XPAND II CAS Subject Satisfaction89% (8/9 subjects)
    XPAND II CAS Physician Satisfaction90% (9/10 physicians) for device performance; 80% (8/10) for expansion results.

    2. Sample Size and Data Provenance

    Pivotal Study (XPAND Trial - CPT-0003):

    • Sample Size (Test Set - Effectiveness Analysis): 151 patients (99 AeroForm patients and 52 saline patients for effectiveness analysis). The analysis was per breast: 155 AeroForm breasts and 83 Saline breasts.
    • Sample Size (Test Set - Safety Analysis): 158 subjects enrolled (106 randomized to AeroForm, 52 to Saline). One AeroForm subject was randomized but no device implanted, so 157 subjects for safety analysis.
    • Data Provenance: Prospective, multi-center, randomized, controlled, open-label study conducted at 17 investigational sites in the U.S.

    Continued Access Study (XPAND II CAS):

    • Sample Size (Test Set): 28 subjects (49 breasts) enrolled as of June 8, 2016. Data available for 11 subjects (21 breasts) who completed the final study visit.
    • Data Provenance: Prospective, multi-center, single-arm study of the V2.5 device.

    Prior Investigations (Australia):

    • Sample Size: 40 subjects (71 AeroForm Tissue Expanders) in the PACE 1 and PACE 2 studies. 21 subjects (34 AeroForm Expanders) in a second Australian study.
    • Data Provenance: Two-phase single-center, prospective, open-label clinical study (PACE) in Perth, Australia. Second study also conducted in Australia. These were prior investigations, likely used for early feasibility and design refinement.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologist with 10 years of experience) used to establish ground truth for image-based assessments. However, the ground truth for the clinical study was based on:

    • Treatment Success: Defined by successful tissue expansion and exchange to a permanent breast implant, unless precluded by a non-device-related event. This assessment would have been made by the investigating physician (surgeon) at each site.
    • Adverse Events: Clinical observations and diagnoses made by the investigator (physician).
    • Patient-reported outcomes: Patient satisfaction and pain ratings, which are self-reported.
    • Physician satisfaction: Reported by the investigating physicians.

    These are clinical outcomes determined by the treating physicians and patient self-assessment, rather than independent expert consensus on specific objective findings like image interpretation.

    4. Adjudication Method for the Test Set

    The document does not specify a formal adjudication method (e.g., 2+1, 3+1) for the clinical outcomes in the pivotal study or the CAS. The clinical outcomes (treatment success, adverse events) were reported by the investigators at each site. Given the nature of breast reconstruction outcomes, these are typically determined by the operating surgeon and clinical staff based on direct patient assessment and surgical findings, rather than independent adjudication of ambiguous findings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The pivotal study was a direct comparison between the AeroForm Tissue Expander System and standard saline tissue expanders, both used by human clinicians. The comparison was of device performance and patient outcomes, not of human reader improvement with or without AI assistance, as the device itself is not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance

    The AeroForm Tissue Expander System is a physical medical device (implant, controller, master key), not a diagnostic algorithm or AI system. Therefore, a standalone (algorithm-only) performance study is not applicable and was not conducted. The software component manages the device's function (e.g., CO2 dosing, communication) and its validation focused on software reliability, safety, and hazard analysis, as outlined in the "SOFTWARE" section.

    7. Type of Ground Truth Used

    • Clinical Studies (XPAND Trial & CAS): The ground truth for effectiveness was primarily based on clinical outcomes data, specifically whether tissue expansion was successful and led to exchange with a permanent breast implant. Safety ground truth was established through recorded adverse events and device malfunctions observed during the clinical course. Patient and physician satisfaction ratings were also used.
    • Non-Clinical (Bench, Biocompatibility, EMC) Studies: Ground truth for these studies was established by pre-defined quantitative and qualitative acceptance criteria for specific physical, chemical, and electrical properties, as measured by standardized testing methodologies.

    8. Sample Size for the Training Set

    The document does not mention a separate "training set" in the context of machine learning or AI models, as the AeroForm Tissue Expander System is a physical device, not an AI/ML product.

    However, if "training set" is meant to refer to early clinical data used to refine the device or inform the pivotal study design:

    • PACE Studies (Australia): 40 subjects (71 AeroForm Expanders V1.0 and V2.0).
    • Second Australian Study: 21 subjects (34 AeroForm Expanders).
    • Sheep Implant Study: 2 male ovine subjects, 12 implants. This was a preliminary safety evaluation.

    These earlier studies and animal trials served as an evolutionary process, leading to modifications (V2.0 to V2.5) that reduced malfunctions, similar to how a training set might inform model refinements.

    9. How the Ground Truth for the Training Set was Established

    As clarified in point 8, there isn't a "training set" in the AI sense for this physical device. However, the "ground truth" for the early studies (PACE, second Australian study, sheep study) that informed device development and modification was established through:

    • Clinical Observation: Direct assessment by surgeons/investigators of device function, tissue expansion, patient comfort, and any adverse events or complications.
    • Surgical Outcomes: Success of expansion, ability to proceed to permanent implant exchange, and any device-related failures requiring intervention.
    • Animal Study Observations: Monitoring of animal subjects for pain, discomfort, and general health, along with assessment of device function in vivo.
    • Device Performance Monitoring: Tracking of CO2 release, communication, and instances of under/over expansion or deflation.

    These observations and outcomes from early non-clinical and clinical experience directly led to the design modifications (e.g., V2.5 with enhanced inner liner and software reprogramming) aimed at improving device performance and reducing malfunctions observed in the V2.0 version.