Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K191318

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-06-14

    (30 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IN ANATOMIC:
    The proximal body, stem, assembly screw, locking cap, optional spacer(s), and humeral head may be used together, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.

    The AEQUALIS™ FLEX REVIVE™ Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The AEQUALIS™ FLEX REVIVE™ Shoulder System is indicated for use as a replacement of shoulder joints disabled by:

    • Rheumatoid arthritis with pain
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Revision of other devices if sufficient bone stock remains

    IN REVERSE:
    The AEQUALIS™ FLEX REVIVE™ Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle with pain disabled by:

    • Rheumatoid arthritis
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Massive and non-repairable rotator cuff tear
    • · Revision of the devices if sufficient bone stock remains

    The reversed tray and polyethylene insert are in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of the humeral assembly during revision with a functional deltoid muscle.

    Device Description

    AEQUALIS™ FLEX REVIVE™ Shoulder System (AFR) is a fully convertible anatomic and reversed shoulder arthroplasty system that is designed to be used with existing Tornier implant systems. It is a non-constrained system intended for total or partial replacement of the glenohumeral articulation. AFR includes a proximal body (metaphysis), a stem, assembly screw, and locking cap. The proximal body has a female taper that is compatible with Ascend Flex Humeral Head and Ascend Flex reversed trays and poly inserts.

    The AFR Shoulder System is implanted by a surgeon and is designed to allow the surgeon to select the components to size the shoulder system for the patient. It allows the shoulder to be constructed in an anatomical or reversed configuration using cleared Flex humeral heads or Ascend Flex reversed trays and inserts. In addition, the AFR Shoulder System can be transformed from anatomic to reverse shoulder prosthesis without the removal of the humeral implant assembly during revision surgery.

    The humeral length is measured to determine the overall humeral implant construct length. The length is assembled from 120 mm (using the short proximal body and stem) to 300 mm (using the standard proximal body, spacers, and stem) in 10 mm increments with the spacers as needed and either of the two available lengths of the proximal body for patient specificity.

    The proximal body, stem, and spacers are made from Ti6A14V per ASTM F-136. The proximal body and stem have a Titanium plasma spray coating. The assembly screw and locking cap are made of CoCr per ISO 5832-12. All implant parts are single use and packaged sterile, using gamma radiation at a minimum dose of 25 kGy to an SAL of 1x10-6.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the AEQUALIS™ FLEX REVIVE™ Shoulder System. This document focuses on demonstrating substantial equivalence to a predicate device rather than conducting a de novo study or a clinical trial to establish the device's performance against specific acceptance criteria for diagnostic output.

    Therefore, the information requested in your prompt regarding acceptance criteria, performance studies, sample sizes, expert involvement, and ground truth establishment, which are typical for studies evaluating the performance of an AI/diagnostic device, is not applicable to this document.

    Here's why and what information can be extracted from the document:

    • Type of Device: The AEQUALIS™ FLEX REVIVE™ Shoulder System is a shoulder joint prosthesis, not a diagnostic device, AI software, or a device that produces a measurable output requiring a performance study against specific acceptance criteria like sensitivity, specificity, or accuracy.
    • Regulatory Pathway: The document is a 510(k) submission, which aims to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This typically involves performance bench testing, biocompatibility, and sometimes design verification, but rarely requires large-scale clinical studies proving diagnostic or AI performance with clearly defined acceptance criteria in the way you've described.
    • "Non-clinical Performance Testing" Section: This section explicitly states the types of tests performed:
      • Packaging Validation Testing
      • Biocompatibility Evaluation
      • Press-Fit Analysis
      • Simulated Use Test
        These are engineering, material, and mechanical tests, not diagnostic performance evaluations.
    • "Clinical Testing" Section: This section explicitly states: "No clinical studies were performed." This is crucial as it directly answers point 5 and 6 of your prompt, indicating that no studies were done to assess human reader improvement with AI or standalone algorithm performance.

    Based on the provided text, none of the specific details about acceptance criteria for AI/diagnostic performance studies, sample sizes for those studies, expert involvement, adjudication methods, MRMC studies, or specific ground truth for such studies can be extracted because they are not relevant to this type of device submission.

    However, if we interpret "acceptance criteria" broadly in the context of this 510(k) submission, it refers to passing the non-clinical performance tests and demonstrating substantial equivalence to the predicate device.

    Information that can be extracted or inferred from the document related to its regulatory acceptance is presented below, though it does not align with your specific questions about AI/diagnostic performance studies.


    Acceptance Criteria and Device Performance (in the context of a 510(k) for a prosthesis):

    Acceptance Criteria Category (as per 510(k) for a prosthesis)Reported Device Performance / Assessment
    Substantial Equivalence:Demonstrated to be substantially equivalent to the predicate device (K181420).
    Intended Use:Identical to the predicate device.
    Principles of Operation:Identical to the predicate device.
    Conditions for Use:Identical to the predicate device.
    Design Differences:Differences (new distal stem components in additional lengths and a partially coated format) do not raise new issues of safety or effectiveness.
    Material Composition:Proximal body, stem, spacers: Ti6A14V per ASTM F-136. Assembly screw, locking cap: CoCr per ISO 5832-12. (Compared to predicate, implied to be acceptable).
    Sterilization:Gamma radiation at a minimum dose of 25 kGy to an SAL of 1x10^-6. (Standard acceptance, implied).
    Packaging Validation Testing:Device passed all testing.
    Biocompatibility Evaluation:Device passed all testing.
    Press-Fit Analysis:Device passed all testing.
    Simulated Use Test:Device passed all testing.

    Study Details (as they pertain to this 510(k) submission, not an AI/diagnostic study):

    1. Sample sizes used for the test set and the data provenance: Not applicable in the context of AI/diagnostic performance. The document describes "non-clinical bench testing" and "process validations" for which specific sample sizes (e.g., number of components tested for strength, wear, etc.) are performed but not detailed in this summary document. There is no "test set" of patient data for diagnostic performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device is an implant, not a diagnostic tool requiring expert interpretation of images or data for ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. There is no "test set" of patient cases or diagnostic interpretations for adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was done. The document explicitly states: "No clinical studies were performed." This is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No standalone algorithmic performance study was done. This is not an AI/algorithmic device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for an AI/diagnostic context. For the device itself, the "ground truth" for its performance would be engineering specifications, material standards, and mechanical test results (e.g., passing pre-defined load limits, wear rates, etc.).
    7. The sample size for the training set: Not applicable. This device does not use a training set for an AI model.
    8. How the ground truth for the training set was established: Not applicable. This device does not have a training set.

    In summary, the provided document is a regulatory communication for a medical device (a shoulder prosthesis) seeking 510(k) clearance based on substantial equivalence, not a study evaluating the performance of an AI or diagnostic device against specific clinical or imaging acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181420

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-20

    (112 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IN ANATOMIC:

    The proximal body, stem, assembly screw, locking cap, optional spacer(s), and humeral head may be used together, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.

    The AEQUALIS FLEX REVIVE Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The AEQUALIS FLEX REVIVE Shoulder System is indicated for use as a replacement of shoulder joints disabled by:

    • Rheumatoid arthritis with pain
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Revision of other devices if sufficient bone stock remains

    IN REVERSE:

    The AEQUALIS FLEX REVIVE Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle with pain disabled by:

    • Rheumatoid arthritis
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Massive and non-repairable rotator cuff tear
    • · Revision of the devices if sufficient bone stock remains

    The reversed tray and polyethylene insert are in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of the humeral assembly during revision surgery for patients with a functional deltoid muscle.

    Notes:

    • All components are single use.
    • · The coated humeral stem is intended for cemented or cementless use.
    • · The all-poly glenoid components are intended for cemented use only
    • · The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
    • Titanium humeral heads are intents with suspected cobalt alloy material sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt alloy.
    Device Description

    AEQUALIS FLEX REVIVE Shoulder System (AFR) is a fully convertible anatomic and reversed shoulder arthroplasty system that is designed to be used with existing Tornier implant systems. It is a non-constrained system intended for total or partial replacement of the glenohumeral articulation. AFR includes a proximal body (metaphysis), spacer(s), a stem, assembly screw, and locking cap. The proximal body has a female taper that is compatible with Ascend Flex Humeral Head and Ascend Flex reversed trays and poly inserts.

    The AFR Shoulder System is implanted by a surgeon and is designed to allow the surgeon to select the components to size the shoulder system for the patient. It allows the shoulder to be constructed in an anatomical or reversed configuration using cleared Ascend Flex humeral heads or Ascend Flex reversed trays and inserts. In addition, the AFR Shoulder System can be transformed from anatomic to reverse shoulder prosthesis without the removal of the humeral implant assembly during revision surgery.

    The humeral length is measured to determine the overall humeral implant construct length. The length is assembled from 120 mm (using the short proximal body and stem) to 180 mm (using the standard proximal body, spacers, and stem) in 10mm increments with the spacers as needed and either of the two available lengths of the proximal body for patient specificity.

    The proximal body, stem, and spacers are made from Ti6Al4V per ASTM F-136. The proximal body and stem have a Titanium plasma coating. The assembly screw and locking cap are made from of CoCr per ISO 5832-121. They are single use and packaged sterile, using gamma radiation at a minimum dose of 25 KGy to an SAL of 1x10-6.

    AI/ML Overview

    The provided text describes the AEQUALIS FLEX REVIVE Shoulder System, a medical device, and summarizes the non-clinical performance testing conducted to support its substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test PerformedAcceptance CriteriaReported Device Performance and Conclusion
    Fatigue TestingThe humeral stems were tested to 10 million cycles (5 million in each configuration) without failure, fracture, or dislocation when inspected after each tested configuration at a magnification of 50X.Met acceptance criteria. Tested without failure, fracture, or dislocation.
    Range of MotionThe Aequalis Flex Revive Shoulder System met the range of motion requirements in ASTM F1378-12.Met acceptance criteria.
    Taper TensileTaper resultant force were not significantly lower than those of the Ascend Flex taper connections when analyzed with a 2-sample T-test.Met acceptance criteria. Taper resultant force was not significantly lower than predicate.
    Tensile Static TorqueThe disassociation torque of the male taper from the female taper was not statistically lower than 1.72Nm, or two times the anticipated frictional torque between glenoid sphere and the reversed insert.Met acceptance criteria. Disassociation torque was not statistically lower than specified value.
    Press Fit PreparationThe AFR Shoulder System achieved a press-fit for the proximal body and distal stem allowing the implant to attain proper seating and acceptable fixation.Met acceptance criteria. Achieved proper seating and acceptable fixation.
    Cemented PreparationThe AFR Humeral Stem cement mantle thickness met or exceeded the Ascend Flex cement mantle thickness.Met acceptance criteria. Cement mantle thickness met or exceeded predicate.
    BiocompatibilityBiocompatibility requirements per ISO 10993-1 were met.Met acceptance criteria.
    EndotoxinEndotoxin results per ST72 demonstrated <20EU/device.Met acceptance criteria.
    PackagingDistribution testing met requirements per ASTM D4169-16.Met acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document summarizes non-clinical performance testing which typically involves bench testing of mechanical properties and material characteristics, not clinical studies with patient data. Therefore, the concepts of "test set sample size" and "data provenance (e.g., country of origin, retrospective/prospective)" as they relate to clinical data are not applicable here. The tests are performed on manufactured device components.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. Ground truth for mechanical and material tests is established by engineering specifications, recognized standards (ASTM, ISO), and analytical methods, not by expert consensus from clinical professionals for individual test cases.

    4. Adjudication Method for the Test Set:

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies for reconciling discrepancies in expert opinions on patient data. For non-clinical performance testing, compliance with predetermined acceptance criteria against recognized standards is the primary method of evaluation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for a shoulder implant system, not an AI-powered diagnostic or assistive device for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a physical medical device (shoulder implant), not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The ground truth for the non-clinical performance testing and acceptance criteria is based on:

    • Recognized Industry Standards: ASTM (e.g., F1378-12, D4169-16), ISO (e.g., 10993-1, 5832-121).
    • Internal Engineering Specifications: Derived from device design, intended use, and comparison to predicate devices (e.g., specific torque values, cement mantle thickness targets).
    • Comparative Analysis to Predicate Devices: Demonstrating that the new device's performance is not significantly inferior to that of a legally marketed predicate device.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not a machine learning model, so there is no training set in the AI sense. The "training" for the device's design and manufacturing would involve engineering principles, materials science, and previous product development experience.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. See point 8.