510(k) Data Aggregation

    K Number
    K180239

    Validate with FDA (Live)

    Date Cleared
    2018-05-16

    (107 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Additive Orthopaedics 3D Printed Bone Segments are intended to be used for internal bone fixation for bone fractures or osteotomies in the ankle and foot, such as:

    *Cotton (opening wedge) osteotomies of the medial cuneiform *Evans lengthening osteotomies

    The Additive Orthopaedics 3D Printed Bone Segments are intended for use with ancillary plating fixation.

    The Additive Orthopaedics 3D Printed Bone Segments are not intended for use in the spine.

    It is a patient specific device.

    Device Description

    The Additive Orthopaedics Patient Specific 3D Printed Bone Segments is a simple one piece device constructed individually for each patient using CT image data. It is intended to be used for internal bone fixation for bone fractures or osteotomies in the foot and ankle. The segments are additively manufactured from medical grade titanium alloy (Ti-6AL-4V Eli). It is a patient specific device. The bone segments come in a variety of configurations that depend on the geometry of the application. The surgeon approves the design of the 3D Printed Bone Segments by comparing his/her design requirements to engineering drawings prior to the construction of the implant device.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, not an AI/ML device. Therefore, the requested information about acceptance criteria and studies (such as MRMC, standalone performance, ground truth, sample sizes for training/test sets, and expert qualifications) is not typically found in this type of document because it pertains to the evaluation of AI/ML algorithm performance.

    Here's what can be extracted from the document regarding the device's evaluation, rephrased to align with the spirit of the request, focusing on how the device meets the regulatory requirements for "substantial equivalence":

    Device Name: Additive Orthopaedics Patient Specific 3D Printed Bone Segments
    510(k) Number: K180239


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a non-AI/ML device submission, there are no "acceptance criteria" in the traditional sense of performance metrics like AUC, sensitivity, or specificity. Instead, the "acceptance criteria" are related to demonstrating substantial equivalence to a predicate device. The "performance" is shown through comparative testing against that predicate.

    Feature/TestAcceptance Criteria (for Substantial Equivalence to Predicate)Reported Device Performance
    Indications for UseNearly identical to predicate device.Verified to be nearly identical to predicate.
    MaterialIdentical to predicate device (medical grade titanium alloy (Ti-6AL-4V Eli)).Verified to be identical.
    Manufacturing ProcessIdentical to predicate device (additive manufacturing).Verified to be identical.
    Dimensions & Geometry (patient-specific)Within the range of sizes claimed for the predicate device, developed in a process equivalent to the reference device, and verified by the surgeon.Demonstrated to meet these conditions.
    Morphological CharacterizationComparable to predicate device.Results demonstrated identity to the predicate device.
    Mechanical Testing (friction, roughness, durability/abrasion, compressive fatigue)Comparable to predicate device.Results demonstrated identity to the predicate device.
    Biocompatibility TestingComparable to predicate device.Results demonstrated identity to the predicate device.

    Study Proving Device Meets Criteria (Substantial Equivalence Study):

    The submission highlights a substantial equivalence study based on non-clinical evidence.

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not applicable and not provided in the document. The evaluation is based on demonstrating equivalence in materials, manufacturing, indications, and non-clinical performance characteristics (morphological, mechanical, biocompatibility) rather than a "test set" of patient data for an algorithm. The "data" here refers to test results from the device itself and the predicate.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not applicable. "Ground truth" in the context of an AI/ML device (e.g., expert consensus on medical images) is not relevant for this type of device submission. The verification of the patient-specific design is done by the surgeon.

    4. Adjudication Method for the Test Set:

    Not applicable. There is no "test set" in the sense of evaluating diagnostic performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which doesn't apply to this 3D-printed bone segment device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical, patient-specific implant, not a standalone algorithm.

    7. The type of ground truth used:

    The concept of "ground truth" (e.g., pathology, outcomes data) as it applies to AI/ML diagnostic or prognostic devices is not relevant here. For device design, the "ground truth" for the patient-specific geometry is derived from the patient's CT image data and subsequently "verified by the surgeon."

    8. The Sample Size for the Training Set:

    Not applicable. There is no machine learning "training set" for this device. The device is custom-designed for each patient based on their CT scan data.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set. The patient-specific designs are generated using individual patient CT image data, and the final design is approved by the surgeon.