510(k) Data Aggregation

    K Number
    K250469

    Validate with FDA (Live)

    Date Cleared
    2025-06-13

    (115 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system.

    The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide.

    The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants.

    Device Description

    The Accuro® 3S Needle Guide Kit consists of the following single use sterile disposable components: Accuro® 3S Needle Guide with integrated probe cover, conductive ultrasound gel, and a patient drape with two integrated probe elastic bands.

    The kit components are assembled into a sterilized CSR Wrap then packaged in a sterile tray.

    The final kit is sterilized under ethylene oxide.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the Accuro® 3S Needle Guide Kit. The clearance is based on a determination of substantial equivalence to a legally marketed predicate device (K171348 Pinpoint™ GT Needle Guide Kits), supported by non-clinical testing.

    Here's an analysis of the provided text in relation to acceptance criteria and study details:

    Key Takeaway Points from the provided text:

    • Device: Accuro® 3S Needle Guide Kit (needle guide for ultrasound-guided procedures).
    • Clearance Type: FDA 510(k) (Premarket Notification).
    • Basis for Clearance: Substantial Equivalence to a predicate device (K171348 Pinpoint™ GT Needle Guide Kits).
    • Key Difference from Predicate: Inclusion of a patient drape with elastic bands to stabilize the needle guide.
    • Study Type: Primarily non-clinical benchtop testing and simulated use evaluations. No clinical studies were required or performed to support this 510(k) submission. This means the acceptance criteria and performance data are entirely based on non-clinical methods.

    Given that no clinical studies were performed, and the evaluation focused on non-clinical testing to demonstrate substantial equivalence, several of the requested points related to clinical studies, human-in-the-loop performance, and expert ground truth cannot be answered from this document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the types of non-clinical tests performed but does not explicitly list quantitative acceptance criteria with corresponding performance results in a structured table. It states that "Design functionality of the subject device was tested to ensure that it meets the requirements of the intended end users" and lists various tests without specifying the target values or the measured outcomes.

    However, based on the non-clinical tests listed, we can infer the categories of acceptance criteria and the nature of the "reported device performance" (which is implied to be successful since clearance was granted).

    Acceptance Criteria Category (Inferred)Reported Device Performance (Implied)
    Mechanical Performance
    Needle guide dimensions accuracyMet specified dimensional requirements
    Needle insertion force & damage during insertionAcceptable force for insertion, no damage to needle or guide
    Needle angulation accuracyMaintained accurate angulation relative to ultrasound plane
    Needle tip depth accuracyMaintained accurate needle tip depth
    Needle guide attachment/removal forceAcceptable force for attachment and removal, secure connection
    Needle gate pinch forceAcceptable pinch force (e.g., to hold the needle securely but allow movement)
    Elastic bands holding forceSufficient force to stabilize positioning
    Elastic bands attachment durabilityDurable attachment, withstood simulated use conditions
    Integrity & Sterility
    Sterile barrier integrityMaintained sterility, no breaches
    Packaging integrityPackaging remained intact and protected sterility
    Ethylene Oxide Sterilization ResidualsResidues within acceptable limits (per ISO 10993-7)
    Biocompatibility
    In vitro CytotoxicityNo cytotoxic effects (per ISO 10993-5)
    Bacterial EndotoxinsEndotoxin levels within acceptable limits (per AAMI/ANSI ST72)
    Assessment of Hemolytic PropertiesNo significant hemolytic properties (per ASTM F756)
    Irritation and Skin SensitizationNo significant irritation or sensitization (per ISO 10993-10)
    Systemic ToxicityNo systemic toxic effects (per ISO 10993-11)
    Imaging Quality
    Imaging quality following administration of needle guide and probe coverUncompromised ultrasound imaging quality
    Probe midline needle insertionNeedle path aligned with probe midline
    General
    Material dimensionsMet specified material dimensions
    Packaging configurationPackaging configuration was appropriate and consistent
    Risk analysisRisks identified and mitigated as per ISO 14971
    Quality System ComplianceMet 21 CFR Part 820 requirements

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the sample sizes used for each non-clinical test. It generally states that "Design functionality...was tested" and lists the types of tests (e.g., "Needle guide dimensions," "Needle angulation accuracy").
    • Data Provenance: The studies were non-clinical benchtop testing and simulated use evaluations. The document does not explicitly state the country of origin for the data collection, but it is implied to be internal testing by Rivanna Medical, Inc. (located in Charlottesville, Virginia, USA). The studies were prospective in nature, designed specifically to support this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: For the "simulated use evaluations," the document mentions evaluations "performed by representative end users." It does not specify the number or detailed qualifications of these representative end users.
    • Qualifications of Experts: The term "representative end users" implies healthcare professionals who would typically use such a device (e.g., radiologists, interventionalists, nurses who perform ultrasound-guided procedures). However, specific qualifications (e.g., "radiologist with 10 years of experience") are not provided. Given the non-clinical nature, this would likely be an engineering/usability assessment by medical professionals rather than a clinical "ground truth" derived from patient data.

    4. Adjudication Method for the Test Set

    Since the studies were non-clinical benchtop and simulated use evaluations, the concept of an "adjudication method" as typically applied to human readers or clinical image interpretation (e.g., 2+1, 3+1 consensus) is not applicable. The "ground truth" for these tests would be established by engineering specifications, physical measurements, and direct observation of functionality and usability by testers.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, Accuro® 3S Needle Guide Kit, did not require clinical studies to support the determination of substantial equivalence." An MRMC study is a type of clinical study involving human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Not applicable. The Accuro® 3S Needle Guide Kit is a physical medical device (a needle guide and cover), not an algorithm or AI software. Therefore, there is no "standalone algorithm" performance to evaluate.

    7. The Type of Ground Truth Used

    • Given that no clinical studies were performed, the "ground truth" was established through non-clinical methods:
      • Engineering Specifications/Benchtop Measurements: For tests like dimensions, force, angulation, and depth accuracy.
      • Simulated Use Observations: For usability, attachment durability, and general functionality assessments by "representative end users."
      • Adherence to Standards: For sterility, biocompatibility, and packaging integrity, the “ground truth” is compliance with, or passing, specified international standards (e.g., ISO 10993 series, ASTM standards).

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical kit, not an AI/ML algorithm or software that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243937

    Validate with FDA (Live)

    Device Name
    Accuro 3S
    Date Cleared
    2025-05-23

    (154 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accuro 3S is a diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. Accuro 3S is intended to be used in a hospital or medical clinic environment at the point of care.

    Accuro 3S supports B-mode imaging and a SpineNav-AI™ image processing software. Accuro 3S clinical applications include: musculoskeletal conventional and superficial, and guidance for needle or catheter placement. A typical examination using Accuro 3S is guidance of neuraxial anesthesia.

    Device Description

    The Accuro® 3S is an ultrasound imaging device intended for use by qualified and trained healthcare professionals in hospital or medical clinic environments. The device offers B-mode imaging and a SpineNav-AI™ image processing software.

    The Accuro 3S is a portable system with a small footprint that can be easily maneuvered within the intended use environment and at the point of care. The device features a touchscreen interface and articulated monitor arm to optimize viewing angle. An integrated battery pack allows the system to operate without wall power. Accuro 3S interfaces with healthcare IT networks to implement DICOM-based patient and image archival workflows, with image storage in an external PACS. The device utilizes a Dual-Array™ convex probe.

    SpineNav-AI is an automated software tool that utilizes machine learning technologies to facilitate workflows associated with musculoskeletal imaging assessments of the lumbar spine.

    The Accuro 3S Dual-Array probe comprises a side-by-side convex transducer array design. Each convex array has identical specifications: 64 elements, 3.5 – 4.0 MHz nominal center frequency, 480-micron pitch, and 50 mm radius of curvature.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the Accuro 3S device meets them, based on the provided FDA 510(k) clearance letter.

    Important Note: The provided document is an FDA 510(k) clearance letter, not a detailed study report. Therefore, some information, especially regarding the specifics of the AI study design (e.g., adjudication method beyond concensus, full statistical analysis of MRMC), is inferred or not explicitly stated in the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter/MetricAcceptance CriteriaReported Device Performance
    SpineNav-AI™ Anatomical Annotation
    Per-frame accuracy (Individual anatomical features)$\ge$ 70%82.1% - 99.3% (Range across all anatomical structures)
    Per-sequence detection success rate (Individual anatomical features)$\ge$ 80%91.7% - 100% (Range across all anatomical structures)
    DICE score (for all anatomical structures)Not explicitly stated as an "acceptance criteria" but included in results0.64 - 0.87 (Range across all anatomical structures)
    Epidural Region Indicator
    Accuracy (Lateral dimension)Not explicitly stated as a numerical "acceptance criteria"1.61 (± 2.57) mm compared to radiologist panel ground truth
    Accuracy (Depth dimension)Not explicitly stated as a numerical "acceptance criteria"2.42 (± 3.41) mm compared to radiologist panel ground truth
    Per-frame detection success rate95.5% (This appears to be the result that also functioned as the specific acceptance criteria for this metric)95.5%

    Note on Acceptance Criteria: The document states that "success criteria for each anatomical annotation produced with SpineNav-AI were derived via a preliminary survey with intended users." For the Epidural Region Indicator's accuracy, specific numerical acceptance criteria are not explicitly stated in the same manner as the per-frame and per-sequence detection rates. The reported values are presented as the results of the evaluation against the radiologist panel ground truth.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): 120 sequences and 10,080 image frames obtained from a total of 81 individuals.
    • Data Provenance:
      • Country of origin: Not explicitly stated, but collected at "seven (7) geographically diverse sites," implying multi-site collection.
      • Retrospective or Prospective: Not explicitly stated, but the description of "test dataset comprising a diverse range of demographic variables" and being "collected at seven (7) geographically diverse sites" suggests data collected specifically for this validation, making it likely prospective or a carefully curated retrospective collection. The phrase "completely distinct from the test dataset used for validation" for the training data further supports the idea of a dedicated test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three board-certified radiologists.
    • Qualifications: "Board certified radiologists." No specific years of experience or subspecialty are detailed beyond board certification.

    4. Adjudication Method for the Test Set

    • Adjudication Method: The ground truth for anatomical labels was "established from a panel of three board certified radiologists." This strongly implies a consensus method where the three radiologists collectively determined the ground truth. The specific nature of how disagreements were resolved (e.g., majority vote, discussion-based consensus, or presence of a tie-breaker) is not explicitly detailed but falls under a consensus approach. It is not a 2+1 or 3+1 setup, but rather a direct panel consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, an MRMC comparative effectiveness study was not explicitly stated as having been done. The study focuses on the standalone performance of the AI against expert ground truth. Assistance to human readers is not mentioned as part of the validation design.
    • Effect size of improvement: Not applicable, as an MRMC study comparing human readers with and without AI assistance was not described.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes. The "AI Summary of Tests" section exclusively describes the performance of the SpineNav-AI™ software tool against ground truth established by experts, without human intervention or interaction. Metrics like per-frame accuracy, per-sequence detection success rate, and DICE scores are typical for standalone algorithm performance evaluation.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert Concensus. The ground truth for anatomical labels was "established from a panel of three board certified radiologists." This means the truth was defined by the agreement of multiple human experts.

    8. Sample Size for the Training Set

    • Sample Size (Training Set): 25,536 images from 147 subjects.

    9. How Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established: The document states that the training data and test data are "completely distinct." While it explicitly mentions how ground truth was established for the test set (panel of three board-certified radiologists), it does not explicitly state how the ground truth for the training set was established. It is a common practice for training data ground truth to also be established by experts, potentially with varying levels of expert involvement or adjudication compared to a highly scrutinized test set, but this specific detail is not provided in the submitted text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171594

    Validate with FDA (Live)

    Device Name
    Accuro
    Date Cleared
    2017-10-20

    (142 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accuro is intended for diagnostic ultrasound imaging of the human body in the following clinical applications: Abdominal, Musculoskeletal (conventional and superficial), Cardiac, Peripheral Vascular, Small Organs (e.g. thyroid, breast), and Pediatrics. A typical examination using the Accuro is guidance of neuraxial anesthesia.

    Device Description

    The Accuro is a hand-held device that features real-time B-mode ultrasound imaging only. Additional features include a compact size and a simple user interface.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Accuro" ultrasonic pulsed echo imaging system. The submission expands the indications for use of a previously cleared device (Accuro 3000 Ultrasound Scanner, K132736). As such, the information provided focuses on the safety and performance validation against standards rather than a clinical study evaluating diagnostic accuracy with acceptance criteria.

    Key takeaway: This 510(k) submission primarily focuses on expanding the indications for use of an existing device and demonstrates substantial equivalence through adherence to recognized standards and safety testing, rather than presenting a clinical study with specific diagnostic performance acceptance criteria. Therefore, many of the requested data points related to clinical study design (sample size for test set, expert ground truth, MRMC, etc.) are not explicitly available in this submission.

    Here's a breakdown of the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the acceptance criteria are not in the form of diagnostic accuracy metrics (e.g., sensitivity, specificity) for a clinical algorithm. Instead, the device's performance is demonstrated through adherence to various international and national safety and performance standards for medical electrical equipment and diagnostic ultrasound.

    Acceptance Criteria (Standards Adhered To)Reported Device Performance
    IEC60601-1:2005 (Medical Electrical Equipment - General Requirements for Safety and Essential Performance)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    IEC60601-1-2:2007 (Medical Electrical Equipment - General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    IEC60601-2-37:2007 (Medical Electrical Equipment - Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    ISO14971:2007 (Application of risk management to medical devices)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    NEMA UD-2:2004 (Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    NEMA UD-3:2004 (Standard for Real-Time Display of Thermal and Mechanical Acoustic Output indices on Diagnostic Ultrasound Equipment)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)
    ISO 10993-1: (Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing Within a Risk Management Process)"Verification and Validation activities were designed and performed to demonstrate that the Accuro met pre-determined performance specifications." (Implies compliance)

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the given text. The filing refers to "Verification and Validation activities" and "in-house protocols" but does not detail a clinical test set size, nor the provenance (country of origin, retrospective/prospective) of any data used for such testing.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the given text. As the submission focuses on standard compliance and expanded indications for an existing device rather than a new diagnostic accuracy study, the establishment of ground truth by clinical experts is not mentioned.


    4. Adjudication Method (for the test set)

    This information is not provided in the given text.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study is not mentioned in the provided text. The submission focuses on the device's technical characteristics, safety, and performance standards adherence for expanding its indications of use.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The "Accuro" is described as a "hand-held device that features real-time B-mode ultrasound imaging only." It is a diagnostic ultrasound system intended for use by a human operator for image guidance and diagnosis. It does not appear to be an AI algorithm with standalone performance, but rather a tool to be used by a clinician. Therefore, a standalone algorithm performance study as typically understood for AI/ML devices is not applicable and not mentioned.


    7. The Type of Ground Truth Used

    The text does not specify the type of ground truth used, as it does not detail a clinical study with diagnostic accuracy endpoints. The "ground truth" in the context of this submission would likely relate to the accurate measurement of physical parameters for the engineering validation required by the listed standards (e.g., acoustic output, electromagnetic compatibility, image quality phantoms).


    8. The Sample Size for the Training Set

    This information is not applicable/provided. The Accuro is an ultrasound imaging device, not an AI/ML algorithm that typically requires a large training set of clinical data. The "training" for such a device would involve its design and calibration against engineering specifications, not statistical learning from a dataset.


    9. How the Ground Truth for the Training Set was Established

    This information is not applicable/provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132736

    Validate with FDA (Live)

    Date Cleared
    2014-03-11

    (189 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accuro 3000 ultrasound scanner is intended for diagnostic ultrasound imaging of the human body in the following clinical applications: Abdominal, Musculoskeletal (Conventional and superficial), Cardiac, Peripheral vascular. A typical examination using the Accuro 3000 is guidance of neuraxial anesthesia.

    Device Description

    The Accuro 3000 Ultrasound Imaging Device is a hand-held device that features real-time B-mode ultrasound imaging only. Additional features include a compact size and a simple user interface.

    AI/ML Overview

    The provided text describes specific details regarding the Accuro 3000 Ultrasound System, but it does not contain information about acceptance criteria or a study that specifically proves the device meets acceptance criteria.

    The document does state: "Verification and Validation activities were designed and performed to demonstrate that the Accuro 3000 met pre-determined performance specifications." However, it does not provide details of these specifications, the study design, or the results.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance or information about sample sizes, ground truth establishment, or multi-reader multi-case studies based solely on the provided text.

    The document primarily focuses on:

    • Administrative details (submitter, contact, dates, trade names).
    • Device classification and comparison to predicate devices.
    • Device description (hand-held, real-time B-mode ultrasound).
    • Indications for use (Abdominal, Musculoskeletal, Cardiac, Peripheral vascular, with typical use being guidance of neuraxial anesthesia).
    • Technological characteristics (equivalent to predicate devices).
    • Safety & Performance Tests: A list of IEC, UL, and NEMA standards that were used, implying the device met these standards for safety and acoustic output, but no specific performance metrics are given.

    In summary, the provided text confirms that verification and validation activities were performed using specific standards, but it does not detail the acceptance criteria or report performance data against those criteria.