510(k) Data Aggregation

    K Number
    K242022

    Validate with FDA (Live)

    Device Name
    Access Toxo IgG
    Date Cleared
    2025-03-28

    (260 days)

    Product Code
    Regulation Number
    866.3780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.

    This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.

    Device Description

    The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative detection of Toxoplasma gondii-specific IgG antibody in adult human serum using the Access Immunoassay Systems.

    The Access Toxo IgG assay consists of the reagent pack, calibrators, and quality controls (OCs), packaged separately. Other items needed to run the assay include substrate and wash buffer.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Beckman Coulter Access Toxo IgG assay, a chemiluminescent immunoassay for detecting IgG antibodies to Toxoplasma gondii in human serum. This product is intended to aid in the diagnosis of Toxoplasma gondii infection and assess the immune status of pregnant women.

    The submission claims substantial equivalence to a legally marketed predicate device, the Access Toxo IgG assay (K080869). The primary difference highlighted is the instrument used: the new device runs on the DxI 9000 Access Immunoassay Analyzer, while the predicate runs on the Access 2 Immunoassay System.

    Here's an analysis of the provided information, focusing on the study that proves the device meets the acceptance criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Strictly speaking, the document does not present "acceptance criteria" in a separate table with yes/no compliance. Instead, it details specific performance metrics and their measured values. The implicit acceptance criterion for most analytical performance studies (like imprecision and method comparison) is that the new device's performance is acceptable for its intended use and comparable to or better than the predicate. For Linearity, LoB, LoD, and LoQ, the acceptance criterion is that the study supports the claimed values.

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design/Claims)Reported Device Performance (Access Toxo IgG on DxI 9000)
    Method Comparison (vs. Access 2 Immunoassay System)High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) to demonstrate interchangeability between instruments.PPA: 100.00% (40/40) with 95% CI = 91.24% to 100% (for Reactive samples) NPA: 100.00% (99/99) with 95% CI = 96.26% to 100.00% (for Non-Reactive samples)
    Imprecision (Within-Laboratory)SD < 0.64 IU/mL at concentrations < 3.2 IU/mL; CV < 20.0% at concentrations > 3.2 IU/mL. (These are the design criteria mentioned, implying they are the acceptance threshold.)Sample 1 (2.7 IU/mL): Overall Precision SD 0.38 (13.9% CV) - Meets SD criterion (0.38 < 0.64) Samples 2-6 (8.3 - 361.0 IU/mL): Overall Precision CVs range from 9.6% to 12.1% - Meets CV criterion (all < 20.0%)
    Imprecision (Reproducibility / Between-Instrument)Not explicitly stated as a separate numerical criterion beyond the general imprecision goal, but implicitly to show consistent performance across instruments.Sample 1 (2.5 IU/mL): Reproducibility SD 0.29 (11.9% CV) Samples 2-6 (7.9 - 377.5 IU/mL): Reproducibility CVs range from 8.9% to 13.4%
    LinearityDemonstrated linearity across the claimed measuring interval.Demonstrated linearity across the measuring interval of 3.2 - 450 IU/mL.
    Limit of Blank (LoB)Claimed value supported by study.0.8 IU/mL (claimed)
    Limit of Detection (LoD)Claimed value supported by study.1.5 IU/mL (claimed)
    Limit of Quantitation (LoQ)Claimed value supported by study.3.2 IU/mL (claimed)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison:

      • Test Sample Size: 140 native serum samples.
      • Data Provenance: The document does not explicitly state the country of origin or whether the samples were retrospective or prospective. It refers to "native serum samples," which typically implies real-world patient samples.
    • Imprecision (Within-Laboratory):

      • Test Sample Size: 6 serum samples (analyzed across 240 replicates each).
      • Data Provenance: Not specified for origin or prospective/retrospective.
    • Imprecision (Reproducibility):

      • Test Sample Size: 6 serum samples (analyzed across 225 replicates each).
      • Data Provenance: Performed at an "internal site." Not specified for origin or prospective/retrospective.
    • Linearity, LoB, LoD, LoQ: The type and number of samples used for these studies are not specified, only that "a study" was performed and values were "claimed."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (experts, qualifications) is primarily relevant for studies involving human interpretation or clinical diagnosis, such as imaging studies where radiologists establish ground truth.

    For this in vitro diagnostic (IVD) device (an immunoassay), the "ground truth" for the test results is established by the measurement itself and comparison to a reference method or validated system (the predicate device). There are no human experts involved in interpreting the raw assay output to establish ground truth for method comparison or analytical performance studies like imprecision. The "truth" is the analytical measurement obtained by the reference method (the predicate device or a highly characterized sample).

    4. Adjudication Method for the Test Set

    Not applicable. As this is an IVD immunoassay, not a human reader study, adjudication methods (like 2+1 or 3+1 for imaging reads) are not relevant here. The comparison is between two analytical instruments/systems.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is an in vitro diagnostic immunoassay. It does not involve human readers interpreting images assisted by AI, nor is it an AI-driven device in the sense of image analysis. Its function is to quantitatively and qualitatively measure an analyte in a biological sample.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the primary studies listed (Method Comparison, Imprecision, Linearity, LoB, LoD, LoQ) are all "standalone" in the sense that they evaluate the analytical performance of the device itself (Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer) without human interpretation affecting the measurement outcome. The device's output is a quantitative (IU/mL) or qualitative (Reactive/Non-Reactive/Equivocal) result.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the method comparison study, the "ground truth" was essentially the results obtained from the predicate device (Access Toxo IgG assay on the Access 2 Immunoassay System). The goal was to show that the new device on the new instrument produced comparable results to the already cleared predicate.

    For the analytical performance studies (Imprecision, Linearity, LoB, LoD, LoQ), the "ground truth" refers to the inherent analytical properties of the assay and instrument, which are determined by testing characterized samples (e.g., known concentrations for linearity, low-concentration samples for LoD/LoB).

    8. The Sample Size for the Training Set

    This document does not provide information about a "training set" in the context of machine learning or AI models. For IVD assays, "training" typically refers to the development and optimization process that precedes validation studies. The analytical performance studies presented here are primarily validation studies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a "training set" in the context of AI/ML ground truth establishment is not relevant to this type of IVD device submission. Assay development and optimization for IVDs involve extensive R&D, but the concept of "ground truth for training" as seen in AI is not directly applicable to a chemical immunoassay.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080869

    Validate with FDA (Live)

    Date Cleared
    2008-05-23

    (53 days)

    Product Code
    Regulation Number
    866.3780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.

    Note: In the United States, this product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.

    Device Description

    The Access Toxo IgG Assay, Toxo IgG Calibrators, Toxo IgG QC, and the Access Immunoassay Analyzers (Access, Access 2, Synchron LXi 725, UniCel DxC 600i, UniCel Dxl 600 and UniCel Dxl 800) comprise the Access Immunoassay Systems for the quantitative and qualitative determination of anti- Toxoplasma gondii IgG in human serum.

    AI/ML Overview

    The Access Toxo IgG Assay is intended for the quantitative and qualitative determination of IgG antibodies to Toxoplasma gondii in human serum. It aids in the diagnosis of Toxoplasma gondii infection and can assess the immune status of pregnant women.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than defining explicit acceptance criteria. However, performance metrics were evaluated in comparison to a predicate device and a CDC panel. The key performance indicators and results are summarized below:

    Performance MetricAcceptance Criteria (Implied/Predicate-based)Reported Device Performance
    Analytical Sensitivity (LoQ)Not explicitly stated as an acceptance criterion, but low detection limit is desirable.Mean Limit of Quantitation (LoQ) was 3.2 IU/mL.
    Dilution Recovery (Linearity)Recovery close to 100% for WHO standards and high patient samples across dilutions.WHO Standard Recovery: Ranged from 78.1% to 108.5% across five dilutions of the Third International Standard for Anti-Toxoplasma Serum (TOXM). Linearity with High Patient Samples: Mean recovery was 100%, ranging from 93.8% to 109.1%.
    Analytical Specificity/InterferenceLow cross-reactivity with common interfering substances and other infectious agents. Ideally, no equivocal or reactive results in non-reactive samples.Out of 311 samples tested for cross-reactivity and interference, 9 samples (2.9%) that were non-reactive by another commercial assay were equivocal or reactive in the Access Toxo IgG Assay. Most categories showed 0 equivocal/reactive results. No significant effect from bilirubin, triolein, albumin, or hemoglobin.
    Reproducibility (Total %CV)Consistent results across different sites, calibration methods, and runs. While specific %CV thresholds are not stated as acceptance criteria, industry standards for immunoassays typically define acceptable variability.Combined Results (Daily Calibration): Average total %CV was 12.6% (ranging from 10.3% to 20.0% across samples A001-A007). Combined Results (Stored Calibration): Average total %CV was 15.3% (ranging from 13.7% to 17.3% across samples A001-A007).
    Method Agreement (Predicate Comparison)High positive and negative agreement with the predicate device (Abbott AxSYM Toxo IgG method).Positive Agreement: Ranged from 80.0% to 99.4% across different sites and sample types. Negative Agreement: Ranged from 94.7% to 100% across different sites and sample types.
    CDC Toxoplasma 1998 Human Serum Panel100% sensitivity and specificity (as demonstrated by predicate devices or general expectations for such panels).Exhibited 100% sensitivity and 100% specificity with the 100-member CDC Toxoplasma 1998 Human Serum Panel.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Analytical Specificity/Interference: 311 samples
    • Reproducibility: 9 serum samples, tested over 7 days with 5 replicates per run per day, across 3 sites (total N=35 replicates per sample per site category). Total 105 replicates for combined results.
    • Method Comparison (Predicate):
      • Site 1: 406 samples (frozen, predominantly prenatal screening, also males/non-pregnant females). Provenance: South-central France.
      • Site 2: 28 samples (fresh) and 433 samples (frozen) (prenatal screening, also males/non-pregnant females). Provenance: Northeastern United States.
      • Site 3 (Manufacturer's site): 558 prenatal specimens (frozen). Provenance: North-central France.
    • CDC Toxoplasma 1998 Human Serum Panel: 100-member panel. Provenance: CDC (United States).
    • Data Provenance: The data used for method comparison was from various locations including south-central France, north-central France, and the northeastern United States. Sample types included prenatal screening populations, males, and non-pregnant females. The data included both prospective (fresh) and retrospective (frozen) collections.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets.

    • For the analytical studies (sensitivity, linearity, specificity, reproducibility), the "ground truth" is based on the inherent properties of the samples (e.g., known concentrations of WHO standards, defined interfering substances).
    • For the method comparison studies, the "ground truth" was established by the predicate device (Abbott AxSYM Toxo IgG method). As such, no human experts were involved in establishing the ground truth for these comparison studies.
    • For the CDC Toxoplasma 1998 Human Serum Panel, the panel itself is a "masked, characterized serum panel." This implies the CDC (a reputable organization) established the true positive/negative status of these samples through their own established methods, which may involve expert consensus or other reference methods, but the specifics are not detailed here.

    4. Adjudication Method for the Test Set:

    • None specifically described for expert adjudication. The "ground truth" for method comparison was the result from the predicate device. If there were discrepancies between the new device and the predicate, the document does not describe a process of expert adjudication to resolve these. Discrepancies are simply presented as part of the agreement tables.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an automated immunoassay for in vitro diagnostic testing, not an imaging device or a system requiring human interpretation with AI assistance. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone study. The Access Toxo IgG Assay is an automated immunoassay system. All the performance data (analytical sensitivity, linearity, specificity, reproducibility, and method comparison) represents the performance of the algorithm/device alone, without human interpretation of the assay results during the testing process. Human intervention would be limited to standard laboratory procedures (sample loading, maintenance, result interpretation post-analysis).

    7. The Type of Ground Truth Used:

    • Predicate device results: For the method comparison studies, the results from the Abbott AxSYM Toxo IgG method served as the comparative ground truth.
    • Internationally recognized standards: For dilution recovery/linearity, the Third International Standard for Anti-Toxoplasma Serum (TOXM) from WHO was used.
    • Characterized serum panels: The CDC Toxoplasma 1998 Human Serum Panel, a masked and characterized panel, was used to assess sensitivity and specificity.
    • Known properties of samples/substances: For analytical sensitivity, specificity, and reproducibility, the ground truth was based on the inherent characteristics of the samples (e.g., low-dose samples, samples with known interferents).

    8. The Sample Size for the Training Set:

    The document does not provide details on a separate "training set" as it is typically understood in machine learning. This is an immunoassay device, and its development likely involved internal optimization and validation rather than training a machine learning model on a distinct dataset. If any "training" refers to internal development and optimization, the size and nature of such internal datasets are not disclosed. The studies described are primarily for validation and verification.

    9. How the Ground Truth for the Training Set Was Established:

    • As noted above, the concept of a "training set" as in machine learning is not directly applicable here. The document describes validation studies rather than a training process. Therefore, how the "ground truth for the training set" was established is not provided.