Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153023

    Validate with FDA (Live)

    Date Cleared
    2015-11-06

    (22 days)

    Product Code
    Regulation Number
    876.4340
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ablatherm® Integrated Imaging device is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue.

    Device Description

    The Ablatherm Integrated Imaging is a computer controlled medical device intended to provide High Intensity Focused Ultrasound (also referred as HIFU) to ablate prostate tissue. The system consists of the following main sub-assemblies: Treatment Module, Control Module, Endorectal Probe, and Ablapak. HIFU is a unique process of delivering a large amount of heat energy to a confined space in a highly controlled manner. This energy heats the tissue to ablation levels while minimizing the effect on surrounding structures.

    The ultrasound energy is delivered via an endorectal probe, which includes an imaging system. The ultrasound waves propagate through the rectal wall and are focused on a portion of the prostate, generating intense heat and causing the ablation of tissue within the targeted area. The process is then repeated in a stepwise fashion to destroy the targeted tissues within the prostate. The apex, sphincter and rectum are preserved while prostate tissues are ablated.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Ablatherm Integrated Imaging

    The provided document describes the Ablatherm Integrated Imaging device, indicated for transrectal high-intensity focused ultrasound (HIFU) ablation of prostate tissue. The study presented is a clinical study designed to demonstrate the device's effectiveness through measures of prostate tissue ablation and patient outcomes.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a submission for a 510(k) premarket notification, formal "acceptance criteria" for performance are not explicitly stated in the same way they might be for a PMA. Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Sonablate 450) and providing clinical evidence of the device's ability to achieve its intended purpose (prostate tissue ablation) and its safety profile.

    Therefore, the table below consolidates the performance metrics presented as evidence of the device meeting its intended use for prostate tissue ablation and the associated clinical outcomes.

    Performance Metric (Indicative of Acceptance)Reported Device Performance (Ablatherm Integrated Imaging)Notes/Context
    Prostate Tissue Ablation
    Mean Prostate Volume Reduction (post-HIFU vs. baseline)13.6 cc (95% CI: 10.9, 16.3)Measured at first post-HIFU ablation biopsy. Mean post-HIFU volume: 9.0 cc; Baseline mean volume: 22.7 cc.
    Mean PSA Nadir (first 6 months post-HIFU)0.53 ng/mL (95% CI: 0.37, 0.69)
    Mean % PSA Reduction from Baseline to Nadir87.97% (95% CI: 84.80, 91.15)
    PSA Reduction from Baseline (1-month post-HIFU)78.14% (95% CI: 72.95, 83.34%)Mean PSA: 0.89 ng/mL
    PSA Reduction from Baseline (1-year post-HIFU)80.34% (95% CI: 75.77, 84.91)Mean PSA: 0.91 ng/mL
    PSA Reduction from Baseline (2-year post-HIFU)82.25% (95% CI: 77.61, 86.89%)Mean PSA: 0.72 ng/mL
    Phoenix Biochemical Survival Rate90.5% (95% CI: 85.2, 95.8%) (116 evaluable subjects)Defined as: No PSA between 6 and 24 months post-HIFU procedure ≥ PSA Nadir + 2 ng/ml.
    Negative Post-Ablation Biopsy Rate59% (95% CI: 50, 68) (Total cohort of 135 patients)For ITT analysis, 17 patients without biopsy were considered "positive".
    Safety Profile (Adverse Events)(All adverse events collected regardless of relationship to device/procedure, n=135)
    Any Adverse Event97.0%
    Erectile Dysfunction67.4% (61/135 unresolved at 2 years)12/61 subjects had pre-HIFU ED; 9/61 unresolved ED were mild. Expected due to targeting lateral margins.
    Urinary Incontinence38.5% (21/135 unresolved at 2 years)16/21 unresolved UI were mild.
    Urinary Obstruction (unresolved at 2 years)0.7% (1 subject)
    Stricture (unresolved at 2 years)2.2% (3 subjects)
    Bladder Neck Contracture (unresolved at 2 years)1.5% (2 subjects)
    Sepsis (unresolved at 2 years)0.7% (1 subject)Two cases of sepsis reported, both not considered related to device/procedure.
    Urinary Tract Infections (resolved within 2 years)34.1% (46 subjects)All resolved within 2 years.
    Bowel Injury (resolved within 2 years)3.0% (4 subjects)All resolved within 2 years. Includes "anal tears".

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 136 subjects were enrolled, of which 135 subjects underwent the HIFU procedure. One subject was excluded due to calcifications.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study seems to be a prospective clinical study (IDE study) where subjects were enrolled and followed over time. The first subject was enrolled on May 4, 2006, and the last on June 30, 2010.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts used to establish the ground truth or their qualifications. The clinical outcomes (prostate volume reduction, PSA levels, biopsy results, adverse events) were likely assessed by the treating physicians and clinical staff involved in the study, consistent with standard medical practice.

    4. Adjudication Method for the Test Set

    The document does not describe any formal adjudication method (e.g., 2+1, 3+1). Clinical data such as PSA levels, prostate volume, and biopsy results are generally based on objective measurements and established clinical diagnostic criteria, rather than subjective expert interpretation requiring adjudication in this context. Adverse events were collected and summarized.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is a treatment device (HIFU for prostate ablation), not an AI-assisted diagnostic or imaging interpretation device. Therefore, the concept of "human readers improve with AI" is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No standalone algorithm performance study was reported. This device is a medical instrument used by trained physicians, not an autonomous AI system.

    7. The Type of Ground Truth Used

    The ground truth used in the clinical study is a combination of:

    • Objective Medical Measurements:
      • Prostate Volume Reduction: Measured clinically.
      • PSA Levels: Standard laboratory test results.
      • Biopsy Results: Histopathological analysis (determines presence/absence of cancer cells post-ablation or tissue characteristics).
    • Clinical Outcomes/Patient Reported Outcomes:
      • Adverse Events: Clinically observed and reported events during follow-up.

    These are considered relevant clinical endpoints for evaluating the effectiveness and safety of a prostate ablation device.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of an AI/machine learning model. The clinical study described involved 135 subjects who underwent the HIFU procedure, which served as the primary clinical evidence for the device's performance in this 510(k) submission. Therefore, it's not applicable to the concept of a training set for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a training set for an AI/machine learning algorithm, this question is not applicable. The clinical data used to support the device's effectiveness and safety were gathered prospectively from the 135 subjects undergoing the procedure and evaluated against established clinical benchmarks and adverse event reporting.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1