510(k) Data Aggregation

    K Number
    K171098

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-02

    (19 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Storage, management, and display of patient data, diagnostic data, videos, and images from computerized ophthalmic diagnostic imaging devices.

    Device Description

    AXIS Image Management System is a browser-based ophthalmic image management system designed for storage, retrieval, and review of images, videos, and reports originating from ophthalmic imaging devices.

    AXIS Image Management System and the predicate device are software systems that work on standard network infrastructure provided by the user. Such infrastructure may include a server, network, and computing devices.

    AXIS Image Management System provides a central database of patient information and exam image history. The AXIS software resides on a server provided by the user and interfaces with networked imaging instruments. Users utilize their existing browser to access the AXIS software to review images, videos, and reports.

    Images can be obtained from diagnostic devices in a variety of industry-standard formats including PDF, JPEG, DCM, and AVI.

    AI/ML Overview

    This document is a 510(k) premarket notification for the AXIS Image Management System. It primarily outlines the substantial equivalence to a predicate device and does not contain detailed acceptance criteria and study results in the typical format for performance studies of AI-powered medical devices.

    Here's an analysis based on the provided text, addressing the points where information is available or noting its absence:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of acceptance criteria for specific performance metrics (e.g., sensitivity, specificity, accuracy) because the device is an Image Management System, not a diagnostic or AI-powered analytical tool. Its performance is assessed functionally and through its compliance with standards.

    The closest equivalent to "acceptance criteria" are the functional characteristics and compliance with industry standards, which are met by the device.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Software-only systemMust be software-onlyYes
    Patient databaseMust have a patient databaseYes
    Imaging review capabilityMust allow imaging reviewYes
    Image annotation and measurement capabilityMust allow image annotation and measurementYes
    Application access methodMust provide user access to the systemBrowser on reviewing device accesses server (no special software installed on reviewing device). This is a difference from the predicate but is presented as an advantage (simpler to maintain/use and device agnostic), not a failure of a criterion.
    Secure LoginMust have secure loginYes
    Interface with electronic medical records (EMR)Must interface with EMRYes
    Connects to imaging instruments via DICOM and non-DICOM methodsMust connect to imaging instruments (DICOM and non-DICOM)Yes
    DICOM ComplianceMust be DICOM-compliantAXIS is DICOM-compliant as stipulated in its DICOM Conformity Statement.
    Functional Performance (Software Verification & Validation)Must perform as intendedPerformance testing was conducted on AXIS Image Management System as part of the software verification and validation and was found to perform as intended.
    Safety and Effectiveness Equivalence to PredicateMust be as safe, as effective, and perform as well as predicate.Based upon the results of the data as summarized, AXIS Image Management System has demonstrated that it is as safe, as effective, and performs as well as or better than the predicate device. Furthermore, based on the comparison with the predicate device...AXIS Image Management System is deemed to be substantially equivalent to the predicate device.

    2. Sample size used for the test set and the data provenance

    Not applicable. This device is an image management system, not one that performs analysis on a dataset of patient images requiring performance metrics like sensitivity/specificity. The "test set" would refer to the testing done during software verification and validation, which is typically a functional and system-level test, not a clinical data test set. The document does not specify details of the samples used in such testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As this device is not performing a diagnostic task, there is no "ground truth" to be established by experts in the context of clinical performance like disease detection. The "ground truth" for its functionality would be whether it performs its stated functions (storage, retrieval, display, etc.) correctly according to its design specifications.

    4. Adjudication method

    Not applicable, for the same reasons as points 2 and 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Performance Data: None required or submitted." As this is an image management system, its primary function is to store, manage, and display data, not to interpret or diagnose, and therefore, it does not involve human readers interpreting cases with or without AI assistance in the way a diagnostic AI would.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This refers to the performance of a diagnostic or analytical algorithm. The AXIS Image Management System is not such an algorithm; it is an infrastructure system. Its "standalone" performance refers to its ability to perform its specified functions (e.g., DICOM compliance, data storage, retrieval) without human intervention in the data processing flow, but this is a functional test, not a diagnostic performance test. The document states that "Performance testing was conducted on AXIS Image Management System as part of the software verification and validation and was found to perform as intended," which implies standalone functional testing.

    7. The type of ground truth used

    For an image management system, the "ground truth" is typically its functional correctness and adherence to standards. For example, a successful ground truth for a storage function would be that data stored can be retrieved accurately and completely. For DICOM compliance, the ground truth is successful interoperability as per the standard. The document states it is "DICOM-compliant as stipulated in its DICOM Conformity Statement" and that it "was found to perform as intended" through software verification and validation.

    8. The sample size for the training set

    Not applicable. The AXIS Image Management System is not an AI/ML device that requires training data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.