Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171150

    Validate with FDA (Live)

    Device Name
    AW-100
    Manufacturer
    Date Cleared
    2018-03-08

    (323 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used for endodontics (canal preparation and canal cleaning).

    Device Description

    The prosed device is used by dental professionals for removal of dentine debris in root canals in the field of endodontics. A turbine hose from a dental unit that supplies the device with compressed air and water is attached of the device in order to use the device. The compressed air is the energy source of the device to generate mechanical vibration. Water from the dental unit is used for cooling and cleaning the treatment site. The device consists of the handpiece including the connection to a hose and file holder.

    The proposed device has two model 4H connects to the Midwest 4-hole turbine hose and the model K A connects to the KaVo MULTIflex® quick-disconnect coupling.

    The associated accessories include:

    • file holder to attach endodontics files
    • replacement wrench for the holder

    The proposed device and accessories are supplied non-sterile and must be sterilized in an autoclave prior to the first use and after every patient by the end-user.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the AW-100 ultrasonic scaler. This document focuses on demonstrating substantial equivalence to a predicate device, not on proving the device meets specific acceptance criteria through a clinical study in the way typically required for AI/ML devices or new medical technologies.

    Therefore, many of the requested points regarding sample size, expert ground truth, adjudication methods, MRMC studies, and training set details are not applicable to this type of submission as they were not required or performed.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance was evaluated against the applicable international standard ISO 18397:2016 for powered scalers. The document states that the device passed all performance criteria outlined in this standard.

    Acceptance Criteria (from ISO 18397:2016 - details not fully described in the document)Reported Device Performance (AW-100)
    Output frequency (as per ISO 18397:2016)2000 to 3000 Hz (measured at 0.28 MPa air pressure) – Passed
    Amplitude (as per ISO 18397:2016)Tested at 0.28 MPa air pressure – Passed
    Resistance to autoclave sterilizationTested as per relevant standards – Passed
    Corrosion resistanceTested as per relevant standards – Passed
    Basic safety and performance requirements (as per relevant standards)Evaluated in accordance with elements of applicable international standards for medical devices – Passed
    Risk managementEvaluated in accordance with elements of applicable international standards for medical devices – Passed
    Biocompatibility testingEvaluated in accordance with elements of applicable international standards for medical devices – Passed
    Reprocessing validationEvaluated in accordance with elements of applicable international standards for medical devices – Passed

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated as a number of devices or cases. Non-clinical bench testing was performed on the AW-100 device.
    • Data Provenance: Not applicable. This was a non-clinical bench performance study, likely conducted by the manufacturer, Micron Corporation, in Japan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth for non-clinical bench testing is typically established by engineering measurements and compliance with standard specifications, not by expert medical opinion.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was non-clinical bench testing against industry standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an ultrasonic scaler, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance was compliance with the technical specifications and safety requirements outlined in international standards, specifically ISO 18397:2016 (Dentistry – Powered scaler) and ISO 9168:2009 (Dentistry -- Hose connectors for air driven dental handpieces). This involves objective measurements of physical parameters such as frequency, amplitude, and material properties.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    Not applicable.