Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K103211

    Validate with FDA (Live)

    Device Name
    AVEA VENTILATOR
    Manufacturer
    Date Cleared
    2011-05-12

    (192 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on neonatal through adult patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.

    Device Description

    The AVEA is a servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for neonatal through adult patients. Its graphical user interface module (UIM) has a flat panel color LCD with real time charting and digital monitoring capabilities, a touch screen for interaction, membrane kevs and a dial for changing settings and operating parameters. It also has an internal gas delivery system with servo controlled active inhalation and exhalation functions. Using internal batteries this provides inter-hospital transport as well as back up capability due to loss of AC power. The AVEA may be configured as a conventional ventilator or non-invasive positive pressure ventilator (NPPV). It has been designed to function using commonly available accessories.

    AI/ML Overview

    The provided document is a 510(k) summary for the AVEA Ventilator, focusing on modifications for Volume Guarantee and Nasal Intermittent Positive Pressure Ventilation. It states that "Performance testing verified that the AVEA Ventilator meets its performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States." However, it does not provide specific details about the acceptance criteria or the study that proves the device meets those criteria, as requested in the prompt.

    Therefore, I cannot provide a complete answer to your request based on the provided text. The document confirms that performance testing was done, but it omits the actual results and methodology of that testing.

    Here's what I can extract and what is missing based on your request:

    1. Table of acceptance criteria and the reported device performance:

    • Missing from the document. The document states that performance testing "verified that the AVEA Ventilator meets its performance requirements," but it does not list these requirements (i.e., acceptance criteria) or the specific reported device performance against them.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Missing from the document. The document mentions "Performance testing" but does not specify the sample size, data provenance, or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable / Missing from the document. This type of information is typically relevant for AI/ML-based diagnostic devices where a "ground truth" needs to be established, often by human experts reviewing medical images or patient data. The AVEA Ventilator is a mechanical ventilator, and its performance testing would likely involve engineering and physiological measurements rather than expert review for "ground truth." Therefore, this information is not relevant in the context of this device or is completely absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable / Missing from the document. Similar to point 3, adjudication methods are usually for resolving discrepancies in expert interpretations, which is not relevant for a mechanical ventilator's performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable / Missing from the document. An MRMC study is relevant for diagnostic AI tools involving human readers. This is not applicable to a mechanical ventilator like the AVEA Ventilator.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable / Missing from the document. This question is also focused on AI/ML algorithms. The AVEA Ventilator has software that implements Volume Guarantee and Nasal Intermittent Mandatory Ventilation, which are automated functions. The document states "software implement Volume Guarantee which is the automated requlation of inspiratory pressure," implying standalone algorithmic function within the device. However, the performance study details for these automated functions are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable / Missing from the document. As mentioned in point 3, the concept of "ground truth" in the context of expert review is unlikely to apply to the performance testing of a mechanical ventilator. The ground truth for such a device would typically be derived from engineering specifications, established physiological parameters, and direct measurement against those standards. The specific methods are not detailed.

    8. The sample size for the training set:

    • Not applicable / Missing from the document. A "training set" typically refers to data used to train an AI/ML model. While the ventilator has "software-driven" functions, the document does not suggest an AI/ML model that would require a distinct training set in the conventional sense. The "training" would be more akin to software development and verification/validation against specifications.

    9. How the ground truth for the training set was established:

    • Not applicable / Missing from the document. For the reasons stated in points 7 and 8, this information is not provided and likely not relevant in the AI/ML context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K081837

    Validate with FDA (Live)

    Device Name
    AVEA VENTILATOR
    Date Cleared
    2008-08-13

    (44 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVEA Ventilator is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on adult, pediatric, and neonatal patients. It should be operated by properly trained clinical personnel, under the direction of a physician.

    Device Description

    The AVEA Ventilator is a fourth generation servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for neonatal through adult patients. Its revolutionary user interface module (UIM) provides maximum flexibility wit6h simple operator interaction. It has a flat panel color LCD with real time graphic displays and digital monitoring capabilities, a touch screen for easy interaction, membrane keys and a dial for changing settings and operating parameters. A precision gas delivery engine with servo controlled active inhalation and exhalation improves performance over previous generations. The only design change is an optional Disposable Expiratory Filter / Water Trap to be used exclusively with the AVEA Ventilator.

    AI/ML Overview

    The provided text is a 510(k) summary for the AVEA Ventilator's optional Disposable Expiratory Filter / Water Trap. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device performance claim through a clinical study with specific acceptance criteria.

    The key points from the document regarding "acceptance criteria" and "study" are:

    • Acceptance Criteria: The device is considered to meet acceptance criteria if it is "substantially equivalent in terms of BFE / VFE and resistance performance" to the predicate device (Clear-Guard II, K990949) and "has the same indicated use as the reusable existing filter / water trap" (integral to the AVEA Ventilator, K013642).
    • Study: "Verification and Validation Testing demonstrated that the AVEA Disposable Expiratory Filter / Water Trap meets its performance requirements at both: Component Level and System Level, and, that this device is substantially equivalent to medical devices currently legally marketed in the United States."

    This 510(k) summary does not provide the kind of detailed clinical study information (e.g., sample sizes, ground truth establishment, expert qualifications, MRMC studies) that would typically be associated with a device making a new performance claim or AI algorithm. It is a "substantial equivalence" review for a modification (an optional disposable filter) to an already cleared ventilator.

    Therefore, the requested table and detailed information mostly cannot be extracted because the provided document does not contain that level of detail for this type of submission.

    Here's a breakdown of what can be inferred or stated from the text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Substantially equivalent in BFE (Bacterial Filtration Efficiency) / VFE (Viral Filtration Efficiency) performance to predicate device (Clear-Guard II - K990949).Not explicitly quantified in this document, but the submission concludes that it "is substantially equivalent in terms of BFE / VFE" to the predicate.
    Substantially equivalent in resistance performance to predicate device (Clear-Guard II - K990949).Not explicitly quantified in this document, but the submission concludes that it "is substantially equivalent in terms of ... resistance performance" to the predicate.
    Has the same indicated use as the reusable existing filter / water trap (integral to AVEA Ventilator, K013642).Stated: "Has the same indicated use as the reusable existing filter / water trap which is integral to the AVEA Ventilator approved under K013642." The intended use remains unchanged.
    Meets performance requirements at both Component Level and System Level.Stated: "Verification and Validation Testing demonstrated that the AVEA Disposable Expiratory Filter / Water Trap meets its performance requirements at both: Component Level and System Level."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document refers to "Verification and Validation Testing" but does not break down the specific sample sizes used for BFE/VFE or resistance testing.
    • Data Provenance: Not specified. It's likely laboratory testing rather than patient data, given the nature of the filter.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable as this is not a diagnostic device involving expert interpretation or ground truthing of clinical data in the traditional sense. Performance is based on physical and biological filtration efficiency measurements.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable for the type of testing described (performance testing of a filter).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This device is a ventilator filter, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, this is a hardware device (filter), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" here is the measured performance metrics (BFE, VFE, resistance) against established engineering standards and the performance of the predicate device. It's not clinical "ground truth" derived from patient outcomes or expert consensus.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062093

    Validate with FDA (Live)

    Device Name
    AVEA VENTILATOR
    Date Cleared
    2006-09-20

    (58 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on adult, pediatric, and neonatal patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.

    Device Description

    The AVEA is a servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for neonatal through adult patients. Its graphical user interface module (UIM) has a flat panel color LCD with real time charting and digital monitoring capabilities, a touch screen for interaction, membrane keys and a dial for changing settings and operating parameters. It also has an internal gas delivery system with servo controlled active inhalation and exhalation functions. Using internal batteries this provides inter-hospital transport as well as back up capability due to loss of AC power. The AVEA may be configured as a conventional ventilator or non-invasive positive pressure ventilator (NPPV). It has been designed to function using commonly available accessories.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (AVEA Ventilator), not a study report. Therefore, much of the requested information regarding acceptance criteria and study details for device performance, ground truth, and expert evaluation is not available in the provided text.

    However, based on the submission, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or detailed reported device performance in a table format. It states generally that: "Performance testing verified that the AVEA Ventilator meets it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States."

    The focus of this 510(k) is on modifications to an already cleared device, asserting substantial equivalence. The specific modifications are:

    • "Software update encompasses a modification to the current . Non-Invasive ventilation previously cleared under K013642 of which allows for a particular mode on the Infant Flow Plus, Nasal CPAP."
    • "This Nasal CPAP mode that is emulated is accomplished by a software modification only, utilizing existing AVEA hardware."
    • "This Nasal CPAP mode is only for single level continuous positive airway pressure to nasal pronos."

    Therefore, the "acceptance criteria" for this specific submission would implicitly be:

    • The modified AVEA Ventilator, with the new Nasal CPAP software, performs equivalently to the predicate devices for continuous ventilation and specifically for the Nasal CPAP mode.
    • The software modification does not adversely affect existing validated functionalities.

    Given the nature of a ventilator, performance metrics would likely revolve around aspects such as:

    • Accuracy of delivered pressure/volume/flow.
    • Response time to patient effort.
    • Safety alarms and limits.
    • Biocompatibility (not directly covered by the software modification but generally a part of medical device approval).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on the sample size for any specific test set, nor the data provenance (country of origin, retrospective/prospective). The submission refers to "Performance testing" in a general sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. Since this is an engineering performance assertion for a ventilator and not a diagnostic device relying on expert interpretation of images or signals, the concept of "ground truth" derived from expert consensus in that context is unlikely to apply directly in the same way. Performance testing for a ventilator typically involves engineering and physiological measurements rather than expert clinical consensus on data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable/not provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving expert interpretation where disagreements need to be resolved for ground truth establishment. This is not the type of study described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a ventilator device, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept doesn't directly apply as the AVEA Ventilator is a device that interacts with a patient and is operated by trained clinical personnel. The "performance testing" mentioned would be for the device's functionality, which by its nature is "standalone" in terms of its mechanical and software operation, but in a real-world scenario, it's always "with human-in-the-loop" for patient management. The document states it is "servo-controlled, software-driven."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Given the device type, "ground truth" would likely be derived from:

    • Engineering specifications and validated measurement equipment (e.g., flow sensors, pressure transducers) for evaluating the ventilator's physical outputs.
    • Physiological models or test lung systems to simulate patient conditions.
    • Compliance with recognized standards (e.g., ISO standards for ventilators).
      However, the specific methods are not detailed in the provided text.

    8. The sample size for the training set

    Not applicable. This device is not described as utilizing a machine learning model that requires a distinct "training set" in the context of AI/ML. The software modification is likely a rule-based or control-algorithm update, not an autonomously learning system.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a machine learning training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1